Four Types of Pulse Oximeters Accurately Detect Hypoxia during Low Perfusion and Motion
Abstract
Background Pulse oximeter performance is degraded by motion artifacts and low perfusion. Manufacturers developed algorithms to improve instrument performance during these challenges. There have been no independent comparisons of these devices. Methods We evaluated the performance of four pulse oximeters (Masimo Radical-7, USA; Nihon Kohden OxyPal Neo, Japan; Nellcor N-600, USA; and Philips Intellivue MP5, USA) in 10 healthy adult volunteers. Three motions were evaluated: tapping, pseudorandom, and volunteer-generated rubbing, adjusted to produce photoplethsmogram disturbance similar to arterial pulsation amplitude. During motion, inspired gases were adjusted to achieve stable target plateaus of arterial oxygen saturation (SaO2) at 75%, 88%, and 100%. Pulse oximeter readings were compared with simultaneous arterial blood samples to calculate bias (oxygen saturation measured by pulse oximetry [SpO2] - SaO2), mean, SD, 95% limits of agreement, and root mean square error. Receiver operating characteristic curves were determined to detect mild (SaO2 Results Pulse oximeter readings corresponding to 190 blood samples were analyzed. All oximeters detected hypoxia but motion and low perfusion degraded performance. Three of four oximeters (Masimo, Nellcor, and Philips) had root mean square error greater than 3% for SaO2 70 to 100% during any motion, compared to a root mean square error of 1.8% for the stationary control. A low perfusion index increased error. Conclusions All oximeters…
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“it is known that pulse oximeters underperform under circumstances of low perfusion [33]”
From Investigating the accuracy of blood oxygen saturation measurements in common consumer smartwatches (2023) · cites this paper as [33]
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Beckett, A. H., Rowland, M., & TURNER, P. (1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,917
Statistical methods for assessing agreement between two methods of clinical measurement
Bland, J. M., & Altman, D. G. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2009.10.001 · Cited by 5,220
Agreement Between Methods of Measurement with Multiple Observations Per Individual
Bland, J. M., & Altman, D. G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329422 · Cited by 1,864
Simple, accurate equations for human blood O2 dissociation computations
Severinghaus, J. W. (1979) · Journal of Applied Physiology · doi:10.1152/jappl.1979.46.3.599 · Cited by 751
Jubran, A. (2015) · Critical Care · doi:10.1186/s13054-015-0984-8 · Cited by 511
Pulse oximetry: fundamentals and technology update
Nitzan, M., Romem, A., & Koppel, R. (2014) · Medical Devices: Evidence and Research · doi:10.2147/mder.s47319 · Cited by 438
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427
Self-citationJubran, A. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2399-x · Cited by 400
The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)
Barker, S. J., & Shah, N. K. (1997) · Anesthesiology · doi:10.1097/00000542-199701000-00014 · Cited by 197
Cited by
Sorted by citations
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114
Current progress of photoplethysmography and SPO2 for health monitoring
TAMURA, T. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00097-w · Cited by 274
Valbuena, V. S. M., Seelye, S., Sjoding, M. W., Valley, T. S., Dickson, R. P., Gay, S. E., Claar, D., Prescott, H. C., & Iwashyna, T. J. (2022) · BMJ · doi:10.1136/bmj-2021-069775 · Cited by 127
Valbuena, V. S. M., Barbaro, R. P., Claar, D. D., Valley, T. S., Dickson, R. P., Gay, S. E., Sjoding, M. W., & Iwashyna, T. J. (2021) · CHEST Journal · doi:10.1016/j.chest.2021.09.025 · Cited by 118
A review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119
Pulse Oximeter Performance, Racial Inequity, and the Work Ahead
Okunlola, O. E., Lipnick, M. S., Batchelder, P. B., Bernstein, M., Feiner, J. R., & Bickler, P. E. (2022) · Respiratory Care · doi:10.4187/respcare.09795 · Cited by 96
Self-citationThe Use of Pulse Oximetry in the Assessment of Acclimatization to High Altitude
Dünnwald, T., Kienast, R., Niederseer, D., & Burtscher, M. (2021) · Sensors · doi:10.3390/s21041263 · Cited by 82
Gudelunas, M. K., Lipnick, M. S., Hendrickson, C. M., Vanderburg, S. B., Okunlola, B., Auchus, I., Feiner, J., & Bickler, P. E. (2023) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000006755 · Cited by 68
Self-citationContinuous vital sign monitoring after major abdominal surgery—Quantification of micro events
Duus, C. L., Aasvang, E. K., Olsen, R. M., Sørensen, H. B. D., Jørgensen, L. N., Achiam, M. P., & Meyhoff, C. S. (2018) · Acta Anaesthesiologica Scandinavica · doi:10.1111/aas.13173 · Cited by 59
Peripheral perfusion index of pulse oximetry in adult patients: a narrative review
Sun, X., He, H., Xu, M., & Long, Y. (2024) · European journal of medical research · doi:10.1186/s40001-024-02048-3 · Cited by 55
Predicting patient decompensation from continuous physiologic monitoring in the emergency department
Sundrani, S. C., Chen, J., Jin, B. T., Abad, Z. S. H., Rajpurkar, P., & Kim, D. A. (2023) · npj Digital Medicine · doi:10.1038/s41746-023-00803-0 · Cited by 55
Poorzargar, K., Pham, C., Ariaratnam, J., Lee, K., Parotto, M., Englesakis, M. F., Chung, F. F., & Nagappa, M. (2022) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-021-00797-8 · Cited by 51
Benefits and risks of oxygen therapy during acute medical illness: Just a matter of dose!
Allardet-Servent, J., Sicard, G., Metz, V., & Chiche, L. (2019) · La Revue de Médecine Interne · doi:10.1016/j.revmed.2019.04.003 · Cited by 51
Lauterbach, C. J., Romano, P. A., Greisler, L. A., Brindle, R. A., Ford, K. R., & Kuennen, M. R. (2020) · Research Quarterly for Exercise and Sport · doi:10.1080/02701367.2020.1759768 · Cited by 43
Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients
Blanchet, M.-A., Mercier, G., Delobel, A., Nayet, E., Bouchard, P.-A., Simard, S., L’Her, E., Branson, R. D., & Lellouche, F. (2023) · Respiratory Care · doi:10.4187/respcare.10582 · Cited by 42
The Importance of Data Quality Control in Using Fitbit Device Data From the Research Program
Lederer, L. E., Breton, A., Jeong, H., Master, H., Roghanizad, A. R., & Dunn, J. (2023) · JMIR mhealth and uhealth · doi:10.2196/45103 · Cited by 42
Ráfl, J., Bachman, T. E., Rafl-Huttova, V., Walzel, Š., & Rožánek, M. (2022) · Digital Health · doi:10.1177/20552076221132127 · Cited by 39
Investigating the accuracy of blood oxygen saturation measurements in common consumer smartwatches
Jiang, Y., Spies, C., Magin, J., Bhosai, S. J., Snyder, L., & Dunn, J. (2023) · PLOS Digital Health · doi:10.1371/journal.pdig.0000296 · Cited by 36
“For example, it is known that pulse oximeters underperform under circumstances of low perfusion [33].”
BackgroundChesley, C. F., Lane-Fall, M. B., Panchanadam, V., Harhay, M. O., Wani, A. A., Mikkelsen, M. E., & Fuchs, B. D. (2022) · Respiratory Care · doi:10.4187/respcare.09769 · Cited by 38
Marillier, M., Bernard, A., Vergès, S., Moran‐Mendoza, O., O’Donnell, D. E., & Neder, J. A. (2021) · Thorax · doi:10.1136/thoraxjnl-2020-215135 · Cited by 34
Yu, Y. S., Wang, J., Wang, Q., Wang, J., Min, J., Wang, S., Wang, P., Huang, R., Xiao, J. Z., Zhang, Y., et al. (2020) · Annals of Translational Medicine · doi:10.21037/atm-20-2614 · Cited by 32
Santos, M., Vollam, S., Pimentel, M. A. F., Areia, C. M., Young, L., Román, C., Ede, J., Piper, P., King, E., Harford, M., et al. (2021) · Journal of Medical Internet Research · doi:10.2196/28890 · Cited by 31
Facing SpO2 and SaO2 discrepancies in ICU patients: is the perfusion index helpful?
Thijssen, M. G., Janssen, L., le Noble, J. L. M. L., & Foudraine, N. A. (2019) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-019-00371-3 · Cited by 30
Pham, L. V., Jun, J. C., & Polotsky, V. Y. (2022) · Handbook of clinical neurology · doi:10.1016/b978-0-323-91532-8.00017-3 · Cited by 28
Elvekjær, M., Carlsson, C. J., Rasmussen, S. M., Porsbjerg, C. M., Grønbæk, K. K., Haahr-Raunkjær, C., Sørensen, H. B. D., Aasvang, E. K., Meyhoff, C. S., & WARD-Project Group (2021) · Physiological Measurement · doi:10.1088/1361-6579/ac010c · Cited by 22
Baker, K. S., Petzold, M., Mucunguzi, A., Wharton–Smith, A., Dantzer, E., Habte, T., Matata, L., Nanyumba, D., Okwir, M., Posada, M., et al. (2021) · EClinicalMedicine · doi:10.1016/j.eclinm.2021.101040 · Cited by 23
Areia, C. M., Vollam, S., Piper, P., King, E., Ede, J., Young, L., Santos, M., Pimentel, M. A. F., Román, C., Harford, M., et al. (2020) · BMJ Open · doi:10.1136/bmjopen-2019-034404 · Cited by 22
Schrading, W. A., McCafferty, B., Grove, J., & Page, D. B. (2020) · Journal of the American College of Emergency Physicians Open · doi:10.1002/emp2.12292 · Cited by 21
Proença, M., Ambuehl, J., Bonnier, G., Meister, T. A., Valentin, J., Soria, R., Ferrario, D., Lemay, M., & Rexhaj, E. (2023) · Scientific Reports · doi:10.1038/s41598-023-33205-z · Cited by 20
Pulse oximetry in paediatric primary care in low-income and middle-income countries
McCollum, E. D., King, C., Colbourn, T., Graham, H., Bernstein, M., Wilson, I. H., & Checkley, W. (2019) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(19)30358-3 · Cited by 18
Self-citationRasmussen, P. S., Aasvang, E. K., Olsen, R. M., Haahr-Raunkjær, C., Elvekjær, M., Sørensen, H. B. D., & Meyhoff, C. S. (2020) · Acta Anaesthesiologica Scandinavica · doi:10.1111/aas.13711 · Cited by 17
Giuliano, K. K., Bilkovski, R. N., Beard, J., & Lamminmäki, S. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-023-01029-x · Cited by 19
Heart-rate tuned comb filters for processing photoplethysmogram (PPG) signals in pulse oximetry
Alkhoury, L., Choi, J.-W., Wang, C., Rajasekar, A., Acharya, S., Mahoney, S., Shender, B. S., Hrebien, L., & Kam, M. (2021) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-020-00539-2 · Cited by 15
Photoplethysmography in oxygenation and blood volume measurements
Abay, T. Y., & Kyriacou, P. A. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-823374-0.00003-7 · Cited by 16
Nguyen, L. S., Helias, M., Raia, L., Nicolas, E., Jaubert, P., Benghanem, S., Hamou, Z. A., Dupland, P., Charpentier, J., Pène, F., et al. (2022) · Scientific Reports · doi:10.1038/s41598-021-02634-z · Cited by 15
de Carvalho, E. B., Leite, T. R. S., Sacramento, R. F. D. M., do Nascimento, P. R. L., Samary, C. S., Rocco, P. R. M., & Silva, P. L. (2022) · Revista Brasileira de Terapia Intensiva · doi:10.5935/0103-507x.20220013-pt · Cited by 14
McCollum, E. D., King, C., Ahmed, S., Hanif, A. A. M., Roy, A. D., Islam, M. A., Colbourn, T., Schuh, H. B., Ginsburg, A. S., Hooli, S., et al. (2021) · BMJ Open Respiratory Research · doi:10.1136/bmjresp-2021-001023 · Cited by 13
Areia, C. M., Vollam, S., Young, L., Biggs, C. A., Pimentel, M. A. F., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., & Watkinson, P. (2021) · BMJ Open · doi:10.1136/bmjopen-2020-047715 · Cited by 12
Zestic, J., Brecknell, B., Liley, H. G., & Sanderson, P. M. (2018) · Human Factors The Journal of the Human Factors and Ergonomics Society · doi:10.1177/0018720818793769 · Cited by 12
Wu, W., Stapelfeldt, F.-N., Kroker, S., Wasisto, H. S., & Waag, A. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3048118 · Cited by 12
Helmer, P., Rodemers, P., Hottenrott, S., Leppich, R., Helwich, M., Pryss, R. C., Kranke, P., Meybohm, P., Winkler, B. E., & Sammeth, M. (2023) · iScience · doi:10.1016/j.isci.2023.108155 · Cited by 11
A review on remote health monitoring sensors and their filtering techniques
Rajeswari, D. S., Hitha Shree, J., Simha, A. L. N., Subhani, M. N., Shivaleelavathi, B. G., & Yatnalli, V. (2021) · Global Transitions Proceedings · doi:10.1016/j.gltp.2021.08.022 · Cited by 12
Assessment of skin pigmentation-related bias in pulse oximetry readings among adults
Khanna, A. K., Beard, J., Lamminmäki, S., Närväinen, J., Antaki, N., & Yapici, H. O. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-023-01095-1 · Cited by 12
Usual and Advanced Monitoring in Patients Receiving Oxygen Therapy
Lellouche, F., & L’Her, E. (2020) · Respiratory Care · doi:10.4187/respcare.07623 · Cited by 12
Mulder, E., Schagatay, E. K. A., & Sieber, A. (2021) · Frontiers in Physiology · doi:10.3389/fphys.2021.649674 · Cited by 9
Gylys, R., Feiner, J., Pologe, J. A., Delianides, T. P., Sutter, S. A., Bickler, P. E., & Lipnick, M. S. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-023-01031-3 · Cited by 9
Self-citationBubb, C. A. B., Weber, M., Kretsch, N., Heim, R. R., Zellhuber, I., Schmid, S., Kagerbauer, S. M., Kreuzer, J., Schaller, S. J., Blobner, M., et al. (2023) · Digital Health · doi:10.1177/20552076231211169 · Cited by 9
Bradke, B. S., Miller, T. A., & Everman, B. (2021) · Sensors · doi:10.3390/s21134543 · Cited by 9
Single-wavelength, single-shot pulse oximetry using an LED-generated vector beam
Jakachira, R., Diouf, M., Lin, Z., Burrow, J. A., Howes, A. M., Oguntolu, T. O., Carter, R., III, DUNSIGER, S. I., & Toussaint, K. C. (2022) · Optics Express · doi:10.1364/oe.461871 · Cited by 8
Wong, M. D., Chung, H., & Chawla, J. (2019) · Journal of Paediatrics and Child Health · doi:10.1111/jpc.14606 · Cited by 8
Pulse Oximetry Discrepancies and Occult Hypoxemia in ICU Patients: Predictors and Clinical Outcomes
Saidy, S. O., Iqbal, A., & Baig, S. H. (2025) · Journal of Intensive Care Medicine · doi:10.1177/08850666251351594 · Cited by 10
Some effects of putative G-protein biased mu-opioid receptor agonists in male rhesus monkeys
Cornelissen, J. C., Blough, B. E., Bohn, L., Negus, S. S., & Banks, M. L. (2021) · Behavioural Pharmacology · doi:10.1097/fbp.0000000000000634 · Cited by 6
Hunasikatti, M. (2021) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.9178 · Cited by 5
Basing intubation of acutely hypoxemic patients on physiologic principles
Laghi, F. A., Shaikh, H., & Caccani, N. (2024) · Annals of Intensive Care · doi:10.1186/s13613-024-01327-w · Cited by 5
Hirota, K., Mayahara, T., Fujii, Y., & Nishi, K. (2022) · COVID · doi:10.3390/covid2010004 · Cited by 4
Mark, N., Lyubin, A., Gerasi, R., Ofir, D., Tsur, A. M., Chen, J., & Bader, T. (2021) · Military Medicine · doi:10.1093/milmed/usaa314 · Cited by 4
Katayama, S., Shima, J., Tonai, K., Koyama, K., & Nunomiya, S. (2021) · Scientific Reports · doi:10.1038/s41598-021-88608-7 · Cited by 4
Kumar, M. G., Kaur, S., & Kumar, R. (2021) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-021-00747-4 · Cited by 4
Monitoring free flaps and replanted digits via perfusion index – A proof of concept study
Strauss, C. J., Anker, A., Klein, S., Kemper, R., Brébant, V., Prantl, L., & Geis, S. (2022) · University of Regensburg Publication Server (University of Regensburg) · Cited by 4
Maso, I., Luvizutto, G. J., Miranda, J. M. D. A., do Nascimento, C. F., Miranda, L. A., Pinto, E. B., dos Santos, F. M. K., Souza, R. M., Moro, C. H. C., Bazán, R., et al. (2025) · Arquivos de Neuro-Psiquiatria · doi:10.1055/s-0045-1806924 · Cited by 3
Evaluating the Accuracy of Low-Cost Wearable Sensors for Healthcare Monitoring
Filgueiras, T. P., Bertemes-Filho, P., & Noveletto, F. (2025) · Micromachines · doi:10.3390/mi16070791 · Cited by 3
Monitoring free flaps and replanted digits via perfusion index – A proof of concept study
Strauss, C., Anker, A., Klein, S., Kemper, R., Brebant, V., Prantl, L., & Geis, S. (2021) · Clinical Hemorheology and Microcirculation · doi:10.3233/ch-211295 · Cited by 3
Accuracy of Oxygen Saturation Measurements in Patients with Obesity Undergoing Bariatric Surgery
Xiong, Y., Pan, G., Huang, W., Yang, W., Hu, R., Mai, Y., Chen, L., Miao, J., & Peng, X. (2022) · Obesity Surgery · doi:10.1007/s11695-022-06221-7 · Cited by 3
Marlar, A. I., Knabe, B. K., Taghikhan, Y., Applegate, R. L., & Fleming, N. (2024) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-024-01211-9 · Cited by 3
Allen, P. E., Akinwumi, A. D., Kroeze, E. G., Leigh, P. Y., Ramirez, S. N., Smart, G. L., Thomas, T. M., & Paris, H. L. R. (2024) · Obesities · doi:10.3390/obesities4020016 · Cited by 2
Vikne, H., Kjeserud, J. A., Westgaard, W., Westlie, R. B., Kåsin, J. I., Medbø, J. I., Gjøvaag, T. F., & Holmedahl, N. H. (2025) · PLoS ONE · doi:10.1371/journal.pone.0326674 · Cited by 3
Hegarty‐Craver, M., Womack, D. S., Thornburg, J. W., Boe, T., Archer, M. J., & Calfee, M. W. (2025) · JMIR Formative Research · doi:10.2196/72324 · Cited by 2
Elia, A., & Keramidas, M. E. (2026) · Respiratory Physiology & Neurobiology · doi:10.1016/j.resp.2026.104603 · Cited by 2
Conventional Oxygen Therapy: Technical and Physiological Issues
Lellouche, F., & L’Her, E. (2021) · Springer eBooks · doi:10.1007/978-3-030-42454-1_1 · Cited by 3
Design of a Physiological Monitoring Device for Holistic Telemedicine Applications
Kullayappa, G. R., Kumar, C. M., & Alam, K. T. (2022) · MAPAN · doi:10.1007/s12647-022-00577-y · Cited by 2
Kaçmaz, Ö., Gökdemir, G. Ş., Özel, M., & Gökdemir, M. T. (2022) · Harran Üniversitesi Tıp Fakültesi Dergisi · doi:10.35440/hutfd.1067974 · Cited by 2
Huang, L., Li, M., Zhao, Y., & He, D. K. (2018) · IEEE International Conference on Mechatronics and Automation (ICMA) · doi:10.1109/icma.2018.8484369 · Cited by 2
Acil Serviste Çalışan Hemşirelerin Pulse Oksimetre Bilgisine Eğitimin Etkisi
ALTUN, N., Uğraş, G. A., & Yüksel, S. (2021) · Sağlık Akademisi Kastamonu · doi:10.25279/sak.692942 · Cited by 1
Clinical evaluation for the pharyngeal oxygen saturation measurements in shocked patients
Hassan, E. A., Mohamed, S. N., Hamouda, E. H., & Ahmed, N. T. (2022) · BMC Nursing · doi:10.1186/s12912-022-01073-z · Cited by 1
Khan, A. R., Rabbani, B., Jameel, A., Khan, B., Husain, T., & Ahmed, J. (2024) · ANNALS OF ABBASI SHAHEED HOSPITAL AND KARACHI MEDICAL & DENTAL COLLEGE · doi:10.58397/ashkmdc.v29i1.794 · Cited by 1
Anaesthetic Management of a Patient with Heterotaxy Syndrome for Emergency Mechanical Thrombectomy
Madhusudhan, S., Sethuraman, M., & Vimala, S. (2025) · Neurology India · doi:10.4103/neurol-india.neurol-india-d-24-00401 · Cited by 1
Inter-Finger Variability of SpO2 During Hypoxemia and Step Resaturation
Walzel, Š., Rafl-Huttova, V., Rožánek, M., Kudrna, P., Rybář, M., & Ráfl, J. (2025) · Healthcare · doi:10.3390/healthcare13202648 · Cited by 1
Karlsson, N., Lindstrand, F. O., Johansson, L., & Sjödin, C. (2026) · Journal of Intensive Care · doi:10.1186/s40560-026-00847-w · Cited by 2
Wu, X., Qin, H., Huang, H., Jiang, J., Wu, X., Liu, Z., Li, M., & Li, S. (2026) · The Clinical Respiratory Journal · doi:10.1111/crj.70176 · Cited by 1
Analysis of the difference in pulse oxygen saturation between the ventral and dorsal fingers
Zhen, Y., Zhu, J., Liu, L., Wang, B., Wang, L., Cui, W., & Liu, D. (2021) · International Journal of Nursing Practice · doi:10.1111/ijn.12916 · Cited by 2
Development of Personalized Health Detection Device for Covid-19
Yatnalli, V., Bhusare, S. S., Shivaleelavathi, B. G., Rajeswari, D. S., HithaShree, J., Simha, A. L. N., & Subhani, M. N. (2024) · International Conference on Intelligent Data Communication Technologies and Internet of Things (IDCIoT) · doi:10.1109/idciot59759.2024.10468030 · Cited by 1
Transforming in-clinic post-operative and intermediate care with cosinuss°
Zellhuber, I., Schade, M., Adams, T., Blobner, M., Weber, M., & Bubb, C. A. B. (2024) · Computational and Structural Biotechnology Journal · doi:10.1016/j.csbj.2024.10.002 · Cited by 1
Interference of pulse oximeter reading in an undiagnosed case of Hansen's disease
Dwivedi, D., Sukhadeve, A., & Jahan, N. (2019) · International Journal of Mycobacteriology · doi:10.4103/ijmy.ijmy_118_19 · Cited by 1
Sensorex: Review on Saturation Extraction Algorithms on Demand of Low-cost Optical Pulse Oximetry
Lipchak, D., Chupov, A., & Soldatchenkov, A. (2021) · Ural Symposium on Biomedical Engineering, Radioelectronics and Information Technology (USBEREIT) · doi:10.1109/usbereit51232.2021.9455023 · Cited by 1
Mitigation of Motion Artifacts in Pulse Oximetry through Redundant Sensors
Alkhoury, L., Choi, J., Wang, C., Mahoney, S., Shender, B. S., Hrebien, L., & Kam, M. (2022) · Signal Processing: Algorithms, Architectures, Arrangements, and Applications · doi:10.23919/spa53010.2022.9927970 · Cited by 1
Considerations while using Fitbit Data in the All of Us Research Program: Tutorial (Preprint)Preprint
Lederer, L. E., Breton, A., Jeong, H., Master, H., Roghanizad, A. R., & Dunn, J. (2022) · doi:10.2196/preprints.45103 · Cited by 1
Performance of Wearable Pulse Oximetry During Controlled Hypoxia InductionPreprint
Jiang, Y., Spies, C. S., Wang, W. K., Roghanizad, A. R., Bhosai, S. J., Snyder, L. D., Burke, A., MacLeod, D., & Dunn, J. (2024) · medRxiv · doi:10.1101/2024.07.16.24310506 · Cited by 1
Vargas, L. C., & Mendoza, J. G. (2018) · Revista Científica Odontológica · doi:10.21142/2523-2754-0601-2018-39-50
INTENSIFIED HOME-MANAGEMENT FOR WORRISOME COVID-19 ADULT PATIENTS
Jonckheere, P., Belche, J. L., Castanares‐Zapatero, D., Henrard, G., Jespers, V., Laet, C. E. D. H. D., Mistiaen, P., Mouillet, M., Orban, T., Ricour, C., et al. (2021) · Open Repository and Bibliography (University of Liège)
Bourgain, J.-L. (2021) · EMC - Anestesia-Rianimazione · doi:10.1016/s1283-0771(21)44996-0
Bourgain, J.-L. (2021) · EMC - Anestesia-Reanimación · doi:10.1016/s1280-4703(21)45024-3
Response to letter to the editor
Eisenberg, M. E., & Levin, R. (2019) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-019-00407-8
The Search for Optimal Oxygen Saturation Targets in Sepsis Patients: a retrospective studyPreprint
Wang, K., Hao, Q., Hu, Z., Wang, J., Juan, Z., & wang, Y. (2022) · Research Square · doi:10.21203/rs.3.rs-2190892/v1
Filippou, S., Iakovidis, P., Lytras, D., Kasimis, K., Solomonidou, F., & Kopsidas, C. (2023) · Physiologia · doi:10.3390/physiologia3020018
Marforio, S. A., Hansson, E. E., & Josenby, A. L. (2024) · Physical & Occupational Therapy In Pediatrics · doi:10.1080/01942638.2024.2386475
Walzel, Š. (2024) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-62502-2_3
Adylov, T. S., Shestak, E. V., & Starkov, V. Y. (2024) · General Reanimatology · doi:10.15360/1813-9779-2024-4-48-54
Hegarty‐Craver, M., Womack, D. S., Thornburg, J., Boe, T., Archer, M. J., & Calfee, W. (2025) · doi:10.2196/preprints.72324
Hemkumar, A., Hermann, J., Wilson, B., & Rajbhandary, P. L. (2025) · Design and Quality for Biomedical Technologies · doi:10.1117/12.3042294
Remote Assessment of the Respiratory and Cardiovascular Systems
Tonelli, R., Bruzzi, G., Marchioni, A., & Clini, E. M. (2025) · TELe-Health · doi:10.1007/978-3-031-89661-3_13
Metadata
- Authors
- Louie, A., Feiner, J. R., Bickler, P. E., Rhodes, L., Bernstein, M., & Lucero, J.
- Affiliations
- University of California, San Francisco (US)
- Corresponding author
- Lucero, Jennifer
- Venue
- Anesthesiology
- Published
- 2018-03
- Volume
- 128
- Issue
- 3
- Pages
- 520-530
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2773802879
- Cited by
- 106
- References
- 24
- Citations by year
- 2026: 6 · 2025: 13 · 2024: 11 · 2023: 13 · 2022: 17 · 2021: 26 · 2020: 9 · 2019: 7 · 2018: 4
- FWCI
- 2.46
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Anesthesiology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Keywords
- pulse oximeter, hypoxemia detection, motion artifacts, low perfusion, arterial oxygen saturation, photoplethysmography, pulse oximetry accuracy, device comparison, hypoxia sensitivity
- MeSH terms
- Adult, Algorithms, Hypoxia, Female, Humans, Male, Motion, Oximetry, Oxygen, Reference Values, Reproducibility of Results, Artifacts, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review