Validation of brachial artery pressure reconstruction from finger arterial pressure

Guelen, Ilja, Westerhof, Berend E, van der Sar, Gertrude L, van Montfrans, Gert A, Kiemeneij, Ferdinand, Wesseling, Karel H, Bos, Willem Jan W · 2008 · Journal of Hypertension
Cited by 186

The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: Measurement of finger artery pressure with Finapres offers noninvasive continuous blood pressure, which, however, differs from brachial artery pressure. Generalized waveform filtering and level correction may convert the finger artery pressure waveform to a brachial waveform. An upper-arm cuff return-to-flow measurement may be used to calibrate the blood pressure on an individual basis. We tested these corrective methods as implemented in the Finometer device. METHODS: Intrabrachial artery pressure (BAP) and finger artery pressures were recorded simultaneously in 37 cardiac patients, aged 41-83 years, who underwent a cardiac catheterization procedure. Finger artery pressures were compared after waveform filtering and level correction and after an additional return-to-flow calibration. Measurements were performed in supine and sitting positions. Accuracy and precision were considered clinically acceptable if the mean and standard deviation of the return-to-flow intrabrachial artery pressure (reBAP)-BAP differences were smaller than 5 +/- 8 mmHg (Association for the Advancement of Medical Instrumentation requirements). RESULTS: Finger artery systolic, diastolic and mean pressures for the group differed from that of intrabrachial artery pressure by -10 +/- 13, -12 +/- 8 and -16 +/- 8 mmHg, respectively. After waveform filtering and level correction the filtered level corrected arterial pressure differed by -1 +/- 11, -0 +/- 7 and -2 +/- 7 mmHg. After individual calibration, reBAP differed by 3 +/- 8, 4 +/- 6 and 3 +/- 5 mmHg. Comparable results were found in the sitting position but only when the supine return-to-flow calibration was used. CONCLUSION: Reconstruction of intrabrachial artery pressure from finger artery pressure with waveform filtering and level correction reduces the pressure differences substantially, with diastolic and mean within Association for the Advancement of Medical Instrumentation requirements. After one supine return-to-flow calibration, all pressure differences meet the requirements. Return-to-flow calibration should not be repeated in sitting position.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/hjh.0b013e3282fe1d28

Cites

4 of 4 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 186

Metadata

Authors
Guelen, I., Westerhof, B. E., van der Sar, G. L., van Montfrans, G. A., Kiemeneij, F., Wesseling, K. H., & Bos, W. J. W.
Venue
Journal of Hypertension
Published
2008-07
Volume
26
Issue
7
Pages
1321-1327
ISSN
0263-6352 (ISSN-L)
OpenAlex
W1985227030
Cited by
186
References
4
FWCI
7.03
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Validation of brachial artery pressure reconstruction from finger arterial pressure. https://app.cochat.ai/papers/10.1097/hjh.0b013e3282fe1d28
The paper
Guelen, I., Westerhof, B. E., van der Sar, G. L., van Montfrans, G. A., Kiemeneij, F., Wesseling, K. H., & Bos, W. J. W. (2008). Validation of brachial artery pressure reconstruction from finger arterial pressure. Journal of Hypertension. https://doi.org/10.1097/hjh.0b013e3282fe1d28

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.