A method to determine the uncertainty in reading the location of a concave meniscus in conical graduated volumetric instruments (CGVI) is presented.

Overview

CoChat Summary

TL;DR

The paper presents a method to estimate the uncertainty contribution from reading a concave meniscus in conical graduated volumetric instruments (CGVI), where internal diameter varies with each graduation line. It measures selected internal diameters with an optical comparator, interpolates the rest using cubic splines, and computes meniscus reading uncertainty from the local cross-section area.

Key findings

  • Optical comparator relative measurement uncertainties ranged between 0.004% and 0.05%, while the internal diameter interpolation error was one percent at most.
  • When interpolated internal diameters were used instead of measured ones, the meniscus reading uncertainty contribution had absolute errors of two percent at most.
  • For a 100 mL centrifuge tube, the meniscus reading uncertainty contribution, \(u_{\text{meniscus}}\), ranged from 0.0136 mL at 0.05 mL nominal volume to 0.4285 mL at 30 mL nominal volume, with relative errors up to 1.99% when using interpolated diameters.
  • Internal diameter measurements are reported for a 1 L Imhoff cone, e.g., 5.530 mm at 0.1 mL and 96.184 mm at 1000 mL, and for a 100 mL centrifuge tube, e.g., 5.892 mm at 0.05 mL and 33.039 mm at 30 mL.

Methods

  • Internal diameters of CGVI were measured with an optical comparator having X-axis range 0–250 mm, Y-axis range 0–150 mm, 0.001 mm display resolution, and a 20X lens; six repetitions were used for every internal diameter measurement.
  • Cubic splines were fitted in R using `splines::interpSpline(V,D)$coef` to interpolate internal diameters at unmeasured graduation lines; the paper suggests less than half of the total required internal diameter values may be enough.
  • Meniscus reading uncertainty was computed as \(u_{\text{meniscus}} = u_{\text{position}} \cdot A\), with \(u_{\text{position}} = 0.05\) mm assumed and \(A\) the cross-section area at the meniscus location; the measurement method was assessed against international standards for cylindrical glassware.

Limitations

  • The paper does not include an explicit limitations section.
  • Evident from the text: the internal diameter measurement method was validated against international standards for cylindrical glassware, not against CGVI reference standards; interpolation performance was assessed with eleven additional internal diameters for one centrifuge tube, while the Imhoff cone interpolation was not similarly evaluated.
  • The paper states that the optical comparator is more expensive and more complex equipment than a Vernier caliper, and that measuring all graduation lines of CGVI would be impractical, time-consuming, and expensive.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1109/access.2021.3123961

Cites

11 of 11 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

FIGURE 1 Close-up on the tip of a CGVI (an Imhoff cone). Notice how uneven the wall thickness is towards the tip.
FIGURE 1 Close-up on the tip of a CGVI (an Imhoff cone). Notice how uneven the wall thickness is towards the tip.CC-BY · From the PDF
FIGURE 2 Leveling and Imhoff cone on its stand with aid of a spirit level.
FIGURE 2 Leveling and Imhoff cone on its stand with aid of a spirit level.CC-BY · From the PDF
FIGURE 3 Internal diameter measurement of a centrifuge tube with an optical comparator.
FIGURE 3 Internal diameter measurement of a centrifuge tube with an optical comparator.CC-BY · From the PDF
(d) One-mark volumetric flask, 100 mL
(d) One-mark volumetric flask, 100 mLCC-BY · From the PDF
Image 5
CC-BY · From the PDF
Image 6
CC-BY · From the PDF
Image 7
CC-BY · From the PDF
Image 8
CC-BY · From the PDF
Image 9
CC-BY · From the PDF
Image 10
CC-BY · From the PDF
Image 11
CC-BY · From the PDF

Verification

Claims the paper makes

5 Valid · 2 Not valid · 6 Not checked · 22 Own results and interpretations

18 more claims

Create a free account to see all 35 claims and every source verdict.

See all claims — free account
  • Probably not a substantial claim. 4
  • It may state two separate claims as one. 6
  • The quote only partly states this claim. 2

Create a free account to see these claims.

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

4 of 4

Citations

Cites

Show all 11 references

Research works printed but not linked: 19

  1. 10. Standard Test Method for Water and Sediment in Fuel Oils by the Centrifuge Method (Laboratory Procedure), ASTM Standard, D1796-11, 2011.
  2. 11. Method for Imhoff Cone Test, Bureau of Indian Standards, IS 7232, 2007.
  3. 12. Quality management system -Requirements, ISO Standard, ISO 9001, 2015.
  4. 13. Environmental management system -Requirements with guidance for use, ISO Standard, ISO 14001, 2015.
  5. 16. Determination of uncertainty for volume measurements made using the gravimetric method, ISO Standard, ISO 20461, 2000.
  6. 17. Laboratory glassware -Volumetric instruments -Methods for testing of capacity and for use, ISO Standard, ISO 4787, 2010.
  7. 6. Laboratory glassware -Single-volume pipettes, ISO Standard, ISO 648, 2008.
  8. 5. Laboratory glassware -Burettes, ISO Standard, ISO 385, 2005.
  9. 7. Laboratory glassware -One-mark volumetric flasks, ISO Standard, ISO 1042, 1998.
  10. 2007, Laboratory Glassware—Graduated Pipettes
  11. 2005, Laboratory Glassware—Graduated Measuring Cylinders
  12. 18. Guidelines on the determination of uncertainty in gravimetric volume calibration, EURAMET cg-19, 2018.
  13. 19. Standard Practice for Calibration of Laboratory Volumetric Apparatus, ASTM Standard, E542-01, 2012.
  14. 22. Laboratory glass or plastics ware -Imhoff sedimentation cones, DIN 12672, 2001.
  15. 21. S. S. Zumdahl, S. A. Zumdahl and J. Hall, "Lab Manual Experimental Chemistry, 9 th ed., "USA: Cengage Learning, 2014, pp. ix.
  16. 25. OMAR-JAIR PURATA-SIFUENTES received the B.S. degree in chemical engineering from the Instituto Tecnológico de Ciudad Madero, Tamauli- pas, México, in 1994 and the M.S. and Ph.D. degrees in chemical engineering from the Instituto Tecnológico de Celaya, Guanajuato, México, in 2004. He also received a M.Eng. degree in quality and productivity systems from the Instituto Tec- nológico y de Estudios Superiores de Monterrey, in 2013. From 2004 to 2013 he was working as a Laboratory Manager or Quality Manager in various companies. Since 2013, he is a full-time Professor with the Organizational Studies Department, Universidad de Guanajuato. He has authored several journals and proceedings papers. His research lines covers applied metrology, integration of management systems and lean six-sigma.
  17. 27. HAIDEÉ-LUCÍA GONZÁLEZ-GÓMEZ re- ceived the B.S. in industrial engineering from the Instituto Tecnológico de León, Guanajuato, México, in 2003. Since 2004 she has been working in the accred- ited metrology laboratories, in CIATEC, León, Guanajuato, México, where she is currently a Lead Specialist in Volume, Density and Viscosity. She has several technical stays at the Centro Nacional de Metrología (CENAM) in volume, density, hard- ness, mass and viscosity. She has co-authored several proceedings papers.
  18. 28. Ms. González-Gómez is currently an Authorized Signatory by the Mex- ican Accreditation Entity, a.c. (ema) in the volume, density and viscosity quantities. JOSÉ-ALBERTO ECHAVARRI-RODRÍGUEZ received the B.S. degree in industrial engineering from the Universidad Tecnológica de León, Gua- najuato, México, in 2012. Since 2006 he has been working in the accred- ited metrology laboratories, in CIATEC, León, Guanajuato, México. He has several technical stays at the Centro Nacional de Metrología (CE- NAM) in volume, density, hardness, mass and viscosity. He has co-authored several proceedings papers.
  19. 29. Mr. Echavarri-Rodríguez is currently an Authorized Signatory by the Mexican Accreditation Entity, a.c. (ema) in the volume and dimensional quantities, and holds an AUKON certification from Zeiss. ELVIA FUNES-RODRÍGUEZ received the B.S. degree in chemical engineering from the Univer- sidad de Guanajuato, México, in 1992 and the M.S. in chemical engineering from the Instituto Tecnológico de Celaya, Guanajuato, México, in 1995. She also has graduated studies in statistics from the Centro de Investigación en Matemáticas, A.C. (CIMAT), Guanajuato, México, in 2007. From 2003 to 2018 he was working as a Metrol- ogy Manager in the accredited metrology labora- tories, in CIATEC, León, Guanajuato, México. Since 2019 she becomes an Independent Metrologist in León, Guanajuato. She has several technical stays at the Centro Nacional de Metrología (CENAM) in mass and density. She has led more than 50 technological development projects related to metrology. She has co-authored several journals and proceedings papers.

Cited by

See all 3

Metadata

Authors
Purata-Sifuentes, O.-J., Gonzalez-Gomez, H.-L., Echavarri-Rodriguez, J.-A., & Funes-Rodriguez, E.
Affiliations
  • Universidad de Guanajuato (MX)
  • Centro de Innovación Aplicada en Tecnologías Competitivas (MX)
Venue
IEEE Access
Published
2021
Volume
9
Pages
147801-147808
ISSN
2169-3536 (ISSN-L)
OpenAlex
W3209340290
Cited by
3
References
11
Citations by year
2024: 2 · 2023: 1
FWCI
0.44
Type
journal article
Language
en
Open access
Locations
  • IEEE Access (publishedVersion, cc-by)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
Universidad de Guanajuato
Topics
  • Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
  • Physical Sciences › Engineering › Mechanical Engineering › Advanced Measurement and Metrology Techniques
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Advanced Electrical Measurement Techniques
Keywords
volume measurement, internal diameter measurement, optical comparator, cubic spline interpolation, measurement uncertainty

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-11, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-11 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.