The DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Vibro- and Electrotactile User Feedback on Hand Opening for Myoelectric Forearm Prostheses
Open accessAbstract
Many of the currently available myoelectric forearm prostheses stay unused because of the lack of sensory feedback. Vibrotactile and electrotactile stimulation have high potential to provide this feedback. In this study, performance of a grasping task is investigated for different hand opening feedback conditions on 15 healthy subjects and validated on three patients. The opening of a virtual hand was controlled by a scroll wheel. Feedback about hand opening was given via an array of eight vibrotactile or electrotactile stimulators placed on the forearm, relating to eight hand opening positions. A longitudinal and transversal orientation of the array and four feedback conditions were investigated: no feedback, visual feedback, feedback through vibrotactile or electrotactile stimulation, and addition of an extra stimulator for touch feedback. No influence of array orientation was shown for all outcome parameters (duration of the task, the percentage of correct hand openings, the mean position error, and the percentage deviations up to one position). Vibrotactile stimulation enhances the performance compared to the nonfeedback conditions. The addition of touch feedback further increases the performance, but at the cost of an increased duration. The same effects were found for the patient group, but the task duration was around 25% larger.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The text we found for this paper looks like another document, so the self-check wasn’t run.
Citations
Cites
Electrotactile and vibrotactile displays for sensory substitution systems
Kaczmarek, K. A., Webster, J. G., Bach-y-Rita, P., & Tompkins, W. J. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.68204 · Cited by 827
Upper-Limb Prosthetics: Critical Factors in Device Abandonment
Biddiss, E., & Chau, T. (2007) · American Journal of Physical Medicine & Rehabilitation · doi:10.1097/phm.0b013e3181587f6c · Cited by 645
Direct neural sensory feedback and control of a prosthetic arm
Dhillon, G. S., & Horch, K. W. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.856072 · Cited by 573
Tactile Displays: Guidance for Their Design and Application
Jones, L. A., & Sarter, N. B. (2008) · Human Factors: The Journal of the Human Factors and Ergonomics Society · doi:10.1518/001872008x250638 · Cited by 561
Epidemiologic Overview of Individuals with Upper-Limb Loss and Their Reported Research Priorities
Atkins, D. J., Heard, D. C. Y., & Donovan, W. H. (1996) · JPO Journal of Prosthetics and Orthotics · doi:10.1097/00008526-199600810-00003 · Cited by 508
Myoelectric forearm prostheses: State of the art from a user-centered perspective
Peerdeman, B., Boere, D., Witteveen, H., Huis in `tVeld, R., Hermens, H., Stramigioli, S., Rietman, H., Veltink, P., & Misra, S. (2011) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.2010.08.0161 · Cited by 469
Self-citationVibrotactile localization on the arm: Effects of place, space, and age
Cholewiak, R. W., & Collins, A. A. (2003) · Perception & Psychophysics · doi:10.3758/bf03194834 · Cited by 325
Vibrotactile localization on the abdomen: Effects of place and space
Cholewiak, R. W., Brill, J. C., & Schwab, A. (2004) · Perception & Psychophysics · doi:10.3758/bf03194989 · Cited by 300
Comparison of Skin Stretch and Vibrotactile Stimulation for Feedback of Proprioceptive Information
Bark, K., Wheeler, J. W., Premakumar, S., & Cutkosky, M. R. (2008) · Symposium on Haptic Interfaces for Virtual Environment and Teleoperator Systems · doi:10.1109/haptics.2008.4479916 · Cited by 169
Wheeler, J. W., Bark, K., Savall, J., & Cutkosky, M. R. (2010) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2009.2039602 · Cited by 168
Research works printed but not linked: 2
- cheung, 2008, Tactile Displays for Orientation Navigation and Communication in Air Sea and Land Environments
- cholewiak, 2005, The representation of space through static and dynamic tactile displays, presented at the Proc Virtual Reality Int Conf Conjunction Human Comput Interaction
Cited by
A neural interface provides long-term stable natural touch perception
Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3008669 · Cited by 751
“Sensory substitution, such as vibration on a residual limb with intensity proportional to pressure at the prosthetic fingertip, improves prosthetic control in limited situations (8, 9), but has not been widely adopted as the vibration is often described as distracting (8).”
BackgroundGeorge, J. A., Kluger, D. T., Davis, T. S., Wendelken, S. M., Okorokova, E. V., He, Q., Duncan, C. C., Hutchinson, D. T., Thumser, Z. C., Beckler, D. T., et al. (2019) · Science Robotics · doi:10.1126/scirobotics.aax2352 · Cited by 388
Parajuli, N., Sreenivasan, N., Bifulco, P., Cesarelli, M., Savino, S., Niola, V., Esposito, D., Hamilton, T. J., Naik, G. R., Gunawardana, U., et al. (2019) · Sensors · doi:10.3390/s19204596 · Cited by 332
“However, the basic control of the majority of those myo-activated limbs has followed the same operating principles (muscle contractions) for more than half of a century [9,10].”
BackgroundXu, B., Akhtar, A., Liu, Y., Chen, H., Yeo, W., Park, S. I., Boyce, B., Kim, H., Yu, J., Lai, H., et al. (2016) · Advanced Materials · doi:10.1002/adma.201504155 · Cited by 304
Restoration of sensory information via bionic hands
Bensmaia, S. J., Tyler, D. J., & Micera, S. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-00630-8 · Cited by 257
Metadata
- Authors
- Witteveen, H. J. B., Droog, E. A., Rietman, J. S., & Veltink, P. H.
- Affiliations
- University of Twente (NL)
- Roessingh Research and Development (NL)
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2012-08
- Volume
- 59
- Issue
- 8
- Pages
- 2219-2226
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- OpenAlex
- W2000332758
- Cited by
- 150
- References
- 23
- Citations by year
- 2026: 2 · 2025: 3 · 2024: 12 · 2023: 13 · 2022: 15 · 2021: 13 · 2020: 13 · 2019: 13 · 2018: 15 · 2017: 8 · 2016: 13 · 2015: 12 · 2014: 11 · 2013: 6 · 2012: 1
- FWCI
- 5.46
- Type
- journal article
- Language
- en
- Open access
- green
- research.utwente.nl
- Licence
- other-oa
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- University of Twente Research Information (publishedVersion, other-oa)
- University of Twente Research Information (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Physical Sciences › Engineering › Biomedical Engineering › Prosthetics and Rehabilitation Robotics
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Keywords
- vibrotactile feedback, electrotactile feedback, sensory feedback, hand opening, grasping tasks, prosthetic hand control, tactile feedback, upper limb amputees
- MeSH terms
- Adult, Artificial Limbs, Electric Stimulation, Feedback, Hand, Humans, Middle Aged, Movement, Physical Stimulation, Psychomotor Performance, Vibration, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Vibro- and Electrotactile User Feedback on Hand Opening for Myoelectric Forearm Prostheses. https://app.cochat.ai/papers/10.1109/tbme.2012.2200678
- The paper
- Witteveen, H. J. B., Droog, E. A., Rietman, J. S., & Veltink, P. H. (2012). Vibro- and Electrotactile User Feedback on Hand Opening for Myoelectric Forearm Prostheses. IEEE Transactions on Biomedical Engineering. https://doi.org/10.1109/tbme.2012.2200678
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.