A neural interface provides long-term stable natural touch perception
Open accessAbstract
Touch perception on the fingers and hand is essential for fine motor control, contributes to our sense of self, allows for effective communication, and aids in our fundamental perception of the world. Despite increasingly sophisticated mechatronics, prosthetic devices still do not directly convey sensation back to their wearers. We show that implanted peripheral nerve interfaces in two human subjects with upper limb amputation provided stable, natural touch sensation in their hands for more than 1 year. Electrical stimulation using implanted peripheral nerve cuff electrodes that did not penetrate the nerve produced touch perceptions at many locations on the phantom hand with repeatable, stable responses in the two subjects for 16 and 24 months. Patterned stimulation intensity produced a sensation that the subjects described as natural and without "tingling," or paresthesia. Different patterns produced different types of sensory perception at the same location on the phantom hand. The two subjects reported tactile perceptions they described as natural tapping, constant pressure, light moving touch, and vibration. Changing average stimulation intensity controlled the size of the percept area; changing stimulation frequency controlled sensation strength. Artificial touch sensation improved the subjects' ability to control grasping strength of the prosthesis and enabled them to better manipulate delicate objects. Thus, electrical stimulation through peripheral nerve electrodes produced long-term sensory restoration after limb loss.
Overview
CoChat Summary
TL;DR
Implanted nonpenetrating peripheral nerve cuff electrodes provided stable, natural touch sensations in the phantom hands of two upper-limb amputees for more than 1 year. Patterned stimulation intensity produced natural pressure, tapping, vibration, and moving touch, and sensory feedback improved grasping control and confidence.
Key findings
- Two subjects with upper limb amputation received implanted nerve cuff electrodes and reported stable touch percepts for 16 and 24 months; stimulation produced sensation from 19 of 20 channels at 15 unique locations in subject 1 and 14 of 16 channels at 9 unique locations in subject 2.
- Constant stimulation intensity typically produced paresthesia, described as “electrical,” in 96% of 151 trials over 10 months; time-variant pulse-width modulation instead produced perceptions described as natural, including pulsing pressure, constant pressure, light moving touch, tapping, and vibration.
- Perceived pressure intensity correlated with maximum pulse width (P < 0.05, R² = 0.85, n = 22), and matched pressure sensation correlated with log stimulation frequency in subject 1 (ANOVA, P < 0.05, R² = 0.70, n = 25).
- In a cherry-stem pulling task, success improved from 43% without sensory or audiovisual feedback to 93% with sensory feedback alone (P < 0.001, n = 15 per condition), and from 77% with audiovisual feedback alone to 100% with both (P < 0.005); self-reported confidence was higher with sensory feedback (one-tailed t test, P = 0.0305).
Methods
- Two male subjects with upper limb amputation received nonpenetrating peripheral nerve cuff electrodes: subject 1 had cuffs on median, radial, and ulnar nerves in the mid-forearm (20 channels total), and subject 2 had cuffs on median and radial nerves in the mid-upper arm (16 channels total).
- During monthly 4- to 6-hour visits, experimenters delivered trains of electrical pulses, identified sensory thresholds, and had subjects describe and trace perceived sensations on a hand schematic; stimulation parameters included pulse amplitude, pulse width, frequency, and time-variant pulse-width modulation.
- Functional testing included a cherry-stem pulling task under four sensory/audiovisual feedback conditions, force matching with a sensor, and pain surveys using the Trinity Amputation and Prosthesis Experience Scales (TAPES).
Limitations
- The study reports only two human subjects, so findings are based on a very small sample and no controlled comparison group.
- The paper states that the phantom pain findings “warrant more rigorous investigation” and that direct comparison with prior natural tactile sensation work is difficult because natural sensation had not been explicitly quantified or detailed.
- The paper notes that higher-density contact arrays may further improve hand coverage; the text also makes evident that many outcomes rely on subjective verbal reports, although some tests blinded subjects to stimulation level.
Model-written summary of a shortened version of the paper’s full text (it is longer than we read). Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
4 Valid · 1 Not valid · 12 Not checked · 82 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“Nerve electrodes activate populations of axons on the basis of size and location relative to the stimulating contact (23).”
What the sources say
- Supportedcites [23] source
In Methods §2.2 (page unavailable), the paper models axons with varying diameters and positions relative to a stimulating contact, then determines which propagate an action potential. Results §3.1 also describes the assumption that larger-diameter axons are recruited before smaller ones. Together, these passages support the paraphrase for the nerve electrode studied. (LLM, conf=0.88)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“When the peak pulse width was set to a first noticeable level, B = Bth, the sensation was described as the back of a pen repeatedly pushing “very lightly” on a small, localized area of the skin (Fig. 3A).”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“There was a psychometric correlation between perceived intensity of the pressure pulse and the maximum pulse width.”
plainstates an associationn = 22 ·- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Of the 16 available channels implanted in subject 2, stimulation produced sensation from 14 channels at 9 unique locations (fig. S1).”
plainstates a causeSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The tapping matched the frequency of the intensity modulation, fmod, as fmod varied between 0.5 and 4 Hz.”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
61 more claims
Create a free account to see all 99 claims and every source verdict.
- Probably not a substantial claim. 10
- It may state two separate claims as one. 24
- The quote only partly states this claim. 1
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“It is proven that FINE and the spiral cuff electrode, when used for stimulation of the afferent axons, produce stable and somatotopically selective results in human subjects in long-term applications, providing appropriate sensory information to the prosthesis wearer [73].”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [73]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Complex qualities of touch such as vibration on the skin, tingling, tapping, stinging, brushing, and itch have been observed during experimentation (Tan et al., 2014).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“and that long-term stability and selectivity has demonstrated using FINE (Ortiz-Catalan et al., 2014; Tan et al., 2014).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Stimulation of the nerve has been shown to induce a variety of sensations with complex qualities, such as tapping on the skin, vibration, tingling, stinging, brushing, and itch (Clark et al., 2014;Tan et al., 2015;2014).”
From Biomimetic approaches to bionic touch through a peripheral nerve interface (2015) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Differences in evoked sensations across electrodes likely reflect differences in the submodality distribution of afferents near the electrode, as evidenced by the fact that stimulation through electrodes with glabrous and non-glabrous receptive fields elicits distinct sensations, reflecting differences in the submodalities that innervate these skin regions (Tan et al., 2014).”
From Biomimetic approaches to bionic touch through a peripheral nerve interface (2015) · cites this paper as Tan et al., 2014
Experimental — accuracy not yet measured
14 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ 151 trials over a 10-month period. We systematically mapped the sensory perceptions
“When sensory feedback was not provided, the subject successfully plucked 43 and 77% of the cherries without (S−/AV−) or with (S−/AV+) audiovisual feedback, respectively.”
§ 151 trials over a 10-month period. We systematically mapped the sensory perceptions
“n = 15 per condition”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ RESULTS
“stimulation produced sensation from 19 channels at 15 unique locations on the perceived hand”
§ DISCUSSION
“There was independent control of 19 different locations on the hand with only three implanted cuffs”
Citations
Cites
Synaesthesia in phantom limbs induced with mirrors
(1996) · Proceedings of the Royal Society of London. Series B: Biological Sciences · doi:10.1098/rspb.1996.0058 · Cited by 1,246
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415
Double nerve intraneural interface implant on a human amputee for robotic hand control
Rossini, P. M., Micera, S., Benvenuto, A., Carpaneto, J., Cavallo, G., Citi, L., Cipriani, C., Denaro, L., Denaro, V., Di Pino, G., et al. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.01.001 · Cited by 396
Dhillon, G. S., Lawrence, S. M., Hutchinson, D. T., & Horch, K. W. (2004) · The Journal of Hand Surgery · doi:10.1016/j.jhsa.2004.02.006 · Cited by 354 · cited 2×
“There is mounting evidence that peripheral nerves are somatotopically organized (24–26), and hence, multi-channel cuff electrodes are able to produce somatotopically selective results similar to those reported in acute trials with intrafascicular electrodes (5, 12, 16).”
ResultRobotic touch shifts perception of embodiment to a prosthesis in targeted reinnervation amputees
Marasco, P. D., Kim, K., Colgate, J. E., Peshkin, M. A., & Kuiken, T. A. (2011) · Brain · doi:10.1093/brain/awq361 · Cited by 335
Augurelle, A.-S., Smith, A. M., Lejeune, T., & Thonnard, J.-L. (2003) · Journal of Neurophysiology · doi:10.1152/jn.00249.2002 · Cited by 330
The Importance of the Sense of Touch in Virtual and Real Environments
Robles-De-La-Torre, G. (2006) · IEEE Multimedia · doi:10.1109/mmul.2006.69 · Cited by 311
PARÆSTHESIÆ FROM ECTOPIC IMPULSE GENERATION IN HUMAN SENSORY NERVES
OCHOA, J. L., & TOREJÖK, H. E. (1980) · Brain · doi:10.1093/brain/103.4.835 · Cited by 213
Horch, K., Meek, S., Taylor, T. G., & Hutchinson, D. T. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2162635 · Cited by 194
Research works printed but not linked: 1
- C. D. Murray in Psychoprosthetics P. Gallagher D. Desmond M. MacLachlan Eds. (Springer London 2008) pp. 119–129.
Cited by
Pursuing prosthetic electronic skin
Chortos, A., Liu, J., & Bao, Z. (2016) · Nature Materials · doi:10.1038/nmat4671 · Cited by 2,396
A skin-inspired organic digital mechanoreceptor
Tee, B. C.-K., Chortos, A., Berndt, A., Nguyen, A. K., Tom, A., McGuire, A., Lin, Z. C., Tien, K., Bae, W.-G., Wang, H., et al. (2015) · Science · doi:10.1126/science.aaa9306 · Cited by 882
Literature Review on Needs of Upper Limb Prosthesis Users
Cordella, F., Ciancio, A. L., Sacchetti, R., Davalli, A., Cutti, A. G., Guglielmelli, E., & Zollo, L. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00209 · Cited by 752
Intracortical microstimulation of human somatosensory cortex
Flesher, S. N., Collinger, J. L., Foldes, S. T., Weiss, J. M., Downey, J. E., Tyler-Kabara, E. C., Bensmaia, S. J., Schwartz, A. B., Boninger, M. L., & Gaunt, R. A. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf8083 · Cited by 704
Brain-Machine Interfaces: From Basic Science to Neuroprostheses and Neurorehabilitation
Lebedev, M. A., & Nicolelis, M. A. L. (2017) · Physiological Reviews · doi:10.1152/physrev.00027.2016 · Cited by 653
“One study (791) employed peripheral nerve cuff electrodes implanted in two patients with arm amputation for more than 1 yr.”
Background
Metadata
- Authors
- Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J.
- Affiliations
- Louis Stokes Cleveland VA Medical Center (US)
- MetroHealth Medical Center (US)
- Corresponding author
- Tan, Daniel W.
- Schiefer, Matthew A.
- Tyler, Dustin J.
- Venue
- Science Translational Medicine
- Published
- 2014-10-08
- Volume
- 6
- Issue
- 257
- Pages
- 257ra138-257ra138
- ISSN
- 1946-6234 (ISSN-L)
- 1946-6242
- PMC
- PMC5517305
- OpenAlex
- W2157094999
- Semantic Scholar
- 308d067b912e90f2bd3958d5248e2d960435ea5c
- Cited by
- 751
- References
- 29
- Citations by year
- 2026: 24 · 2025: 41 · 2024: 57 · 2023: 59 · 2022: 81 · 2021: 75 · 2020: 74 · 2019: 87 · 2018: 82 · 2017: 64 · 2016: 64 · 2015: 37 · 2014: 5
- FWCI
- 25.89
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Science Translational Medicine (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- U.S. Department of Veterans Affairs (I01 RX001334, IK1 RX000724)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
- Keywords
- peripheral nerve interfaces, electrical stimulation, tactile perception, upper limb amputation, prosthetic hand, sensory feedback, phantom hand, long-term stability, paresthesia, grasp control, fine motor control, sensory recovery
- MeSH terms
- Amputees, Artificial Limbs, Electric Stimulation, Electrodes, Implanted, Hand, Humans, Male, Middle Aged, Motor Skills, Paresthesia, Peripheral Nerves, Phantom Limb, Pressure, Sensation, Touch Perception, Feedback, Sensory, Translational Research, Biomedical
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-07, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-07 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A neural interface provides long-term stable natural touch perception. https://app.cochat.ai/papers/10.1126/scitranslmed.3008669
- The paper
- Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J. (2014). A neural interface provides long-term stable natural touch perception. Science Translational Medicine. https://doi.org/10.1126/scitranslmed.3008669
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.