Automatic Annotation of Hip Anatomy in Fluoroscopy for Robust and Efficient 2D/3D Registration

Open access
Grupp, Robert B., Unberath, Mathias, Gao, Cong, Hegeman, Rachel A., Murphy, Ryan J., Alexander, Clayton P., Otake, Yoshito, McArthur, Benjamin A. +2 more·2020·International Journal of Computer Assisted Radiology and Surgery

Abstract

Fluoroscopy is the standard imaging modality used to guide hip surgery and is therefore a natural sensor for computer-assisted navigation. In order to efficiently solve the complex registration problems presented during navigation, human-assisted annotations of the intraoperative image are typically required. This manual initialization interferes with the surgical workflow and diminishes any advantages gained from navigation. We propose a method for fully automatic registration using annotations produced by a neural network. Neural networks are trained to simultaneously segment anatomy and identify landmarks in fluoroscopy. Training data is obtained using an intraoperatively incompatible 2D/3D registration of hip anatomy. Ground truth 2D labels are established using projected 3D annotations. Intraoperative registration couples an intensity-based strategy with annotations inferred by the network and requires no human assistance. Ground truth labels were obtained in 366 fluoroscopic images across 6 cadaveric specimens. In a leave-one-subject-out experiment, networks obtained mean dice coefficients for left and right hemipelves, left and right femurs of 0.86, 0.87, 0.90, and 0.84. The mean 2D landmark error was 5.0 mm. The pelvis was registered within 1 degree for 86% of the images when using the proposed intraoperative approach with an average runtime of 7 seconds. In comparison, an intensity-only approach without manual initialization, registered the pelvis to 1 degree in 18% of images. We have created the first accurately annotated, non-synthetic, dataset of hip fluoroscopy. By using these annotations as training data for neural networks, state of the art performance in fluoroscopic segmentation and landmark localization was achieved. Integrating these annotations allows for a robust, fully automatic, and efficient intraoperative registration during fluoroscopic navigation of the hip.

Integrity signals

Registry

Crossref · 10.1007/s11548-020-02162-7

Cites

40 of 40 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

7 of 12 cited sources checked · 2 supported · 1 flagged · accuracy not yet measured

  • Premisecitation

    Pelvis poses estimated from Bier's annotations produced reprojection errors of 14-34 mm for other landmarks not learned by the network.

    “Pelvis poses were estimated using these annotations, yielding reprojection errors of 14-34 mm for other landmarks not learned by the network.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [19] Bier B, Goldmann F, Zaech JN, Fotouhi J, Hegeman R, Grupp R, Armand M, Osgood G, Navab N, Maier A, Unberath M (2019) Learning to detect anatomical landmarks of the pelvis in X-rays from arbitrary views. Int J Comput Assist Radiol Surg 14(9):1463–1473 source

      The Materials and methods section (‘2D/3D registration’; page unavailable) says detected pelvis landmarks were matched to CT labels to estimate a projection matrix, supporting the pose-estimation part. The Clinical data results (page unavailable) do not report 14–34 mm reprojection errors for landmarks the network did not learn; their stated 2–5 pixel error concerns a different, metal-bead-based reference projection. (LLM, conf=0.87)

      Experimental — accuracy not yet measured
  • Premisecitation

    Applications of CNNs to fluoroscopy have been mostly limited to recognizing surgical instruments and tools.

    “However, since existing large-scale hip datasets have focused on 3D image modalities and pre and postoperative radiography, rather than intraoperative fluoroscopy [12], applications of CNNs to fluoroscopy have been mostly limited to recognizing surgical instruments and tools [13-16].”
    Quote in the paperSubstantial claim2 of 4 sources checked
    • cites [13] Miao S, Wang ZJ, Liao R (2016) A CNN regression approach for real-time 2D/3D registration. IEEE Trans Med Imag 35(5):1352–1363 source
    • cites [14] Ambrosini P, Ruijters D, Niessen WJ, Moelker A, van Walsum T (2017) Fully automatic and real-time catheter segmentation in X-ray fluoroscopy. In: Proceedings of medical image computing and computer-assisted intervention. Springer, pp 577–585 source
    • cites [15] Breininger K, Albarqouni S, Kurzendorfer T, Pfister M, Kowarschik M, Maier A (2018) Intraoperative stent segmentation in X-ray fluoroscopy for endovascular aortic repair. Int J Comput Assist Radiol Surg 13(8):1221–1231 source
    • cites [16] Gao C, Unberath M, Taylor R, Armand M (2019) Localizing dexterous surgical tools in X-ray for image-based navigation. arXiv preprint arXiv:1901.06672 source

      The Methods and Conclusions describe a ConvNet that localizes and segments a surgical tool in X-ray images, supporting this as an example of tool-focused work. The paper does not establish that CNN applications to fluoroscopy have been *mostly limited* to instruments and tools. (LLM, conf=0.82)

      Experimental — accuracy not yet measured
  • Findingown data

    The average landmark 2D localization error was 5.0 ± 5.2 mm in the detector plane.

    “The average landmark 2D localization error was 5.0 ± 5.2 mm in the detector plane.”
    Quote in the paperSubstantial claim
  • Findingown data

    Mean segmentation and landmark detection time per image was 24.0 ± 0.4 milliseconds on a NVIDIA Tesla P100 GPU.

    “The mean time for segmentation and landmark detection per image was 24.0 ± 0.4 milliseconds using a NVIDIA Tesla P100 (PCIe) GPU.”
    Quote in the paperSubstantial claim
  • Findingown data

    Mean 2D landmark localization error reached 5.0 mm.

    “The mean 2D landmark localization error was 5.0 mm.”
    Quote in the paperSubstantial claim
  • May combine two claims. 20
  • Classified as probably not a substantial claim. 4
  • May combine two claims.

    Interpretationown data

    Unlike [16,19,20], the authors train networks on smaller datasets of actual fluoroscopy and achieve state-of-the-art results on clinically relevant data.

    “Contrary to [16,19,20], we train our networks using smaller datasets of actual fluoroscopy and achieve state-of-the-art results on clinically relevant data.”
    Quote in the paperSubstantial claim3 of 3 sources checked
    • cites [16] Gao C, Unberath M, Taylor R, Armand M (2019) Localizing dexterous surgical tools in X-ray for image-based navigation. arXiv preprint arXiv:1901.06672 source
    • cites [19] Bier B, Goldmann F, Zaech JN, Fotouhi J, Hegeman R, Grupp R, Armand M, Osgood G, Navab N, Maier A, Unberath M (2019) Learning to detect anatomical landmarks of the pelvis in X-rays from arbitrary views. Int J Comput Assist Radiol Surg 14(9):1463–1473 source

      Bier et al. (2019) say they trained on synthetic X-rays and then applied the model to real X-rays, contrary to the claim that they trained on actual fluoroscopy. The quoted sentence appears instead in “Automatic Annotation of Hip Anatomy in Fluoroscopy for Robust and Efficient 2D/3D Registration” (unscoped score 0.52). That paper explicitly contrasts its approach with Bier’s, which it cites as [19]; it does not attribute its own training method to Bier. (LLM, conf=0.58)

      Checked by a reasoning model — may be wrong.

      Experimental — accuracy not yet measured
    • cites [20] Esteban J, Grimm M, Unberath M, Zahnd G, Navab N (2019) Towards fully automatic X-ray to CT registration. In: Proceedings of medical image computing and computer-assisted intervention. Springer, pp 631–639 source

      Two searches of the cited paper returned no snippets, so its full text could not be checked. An unscoped search found the sentence nearly verbatim in “Automatic Annotation of Hip Anatomy in Fluoroscopy for Robust and Efficient 2D/3D Registration” (score 0.52), a distinct but closely related paper. That is a strong reason to question the attribution, but the match falls below the rubric’s 0.70 threshold for misattribution, and the cited paper’s abstract was unavailable for comparison. (LLM, conf=0.62)

      Experimental — accuracy not yet measured
  • May combine two claims.

    Premisecitation

    Bier trained CNNs to annotate pelvic anatomical landmarks using simulated fluoroscopy; on five sequences of actual fluoroscopy from two cadaveric specimens, mean annotation errors of 12-24 mm in the detector plane were reported.

    “Using a large collection of simulated fluoroscopy, Bier trained CNNs to annotate anatomical landmarks of the pelvis [19]. When evaluated on five sequences of actual fluoroscopy across two cadaveric specimens, mean annotation errors of 12-24 mm in the detector plane were reported.”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites [19] Bier B, Goldmann F, Zaech JN, Fotouhi J, Hegeman R, Grupp R, Armand M, Osgood G, Navab N, Maier A, Unberath M (2019) Learning to detect anatomical landmarks of the pelvis in X-rays from arbitrary views. Int J Comput Assist Radiol Surg 14(9):1463–1473 source

      Split into 2 atoms — 1 confirmed, 1 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.

      Experimental — accuracy not yet measured
  • May combine two claims.

    Findingown data

    Method 3's performance was robust when only 2, 3, and 4 landmarks were detected, reporting success in 2, 15, and 30 cases, respectively.

    “Method 3’s performance was robust when only 2, 3, and 4 landmarks were detected; reporting success in 2, 15, and 30 cases, respectively.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Depth estimation inaccuracies accounted for nearly all pelvis translation error, with mean errors about the X, Y, and Z axes of 0.1 ± 0.1 mm, 0.1 ± 0.1 mm, and 1.4 ± 2.0 mm, respectively, for method 3.

    “Depth estimation inaccuracies accounted for nearly all of the pelvis translation error, with mean errors about the X, Y, and Z axes of 0.1 ± 0.1 mm, 0.1 ± 0.1 mm and 1.4 ± 2.0 mm, respectively for method 3.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    Although the naive registration succeeded in only 66 cases, its mean femur rotation errors were about 1° smaller than those of methods 2 and 3.

    “Although the naive registration only succeeded in 66 cases, the mean femur rotation errors were about 1° smaller than those of methods 2 and 3.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    The proposed intraoperative approach registered the pelvis within 1° in 86% of images, with an average runtime of 7 seconds.

    “The pelvis was registered within 1° for 86% of the images when using the proposed intraoperative approach with an average runtime of 7 seconds.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Across all test images, mean dice coefficients were 0.86 ± 0.20, 0.87 ± 0.18, 0.90 ± 0.24, 0.84 ± 0.31, 0.74 ± 0.19, and 0.63 ± 0.13 for the left hemipelvis, right hemipelvis, left femur, right femur, vertebra, and upper sacrum, respectively.

    “Across all test images, mean dice coefficients of 0.86 ± 0.20, 0.87 ± 0.18, 0.90 ± 0.24, 0.84 ± 0.31, 0.74 ± 0.19, and 0.63 ± 0.13 were obtained for the left hemipelvis, right hemipelvis, left femur, right femur, vertebra, and upper sacrum, respectively.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Methods 2 and 3 were also successful in 62 and 64 of method 1's successes, each with a mean femur rotation error of 0.8° ± 0.5°.

    “However, methods 2 and 3 were also successful in 62 and 64 of method l’s successes, each with a mean femur rotation error of 0.8° ± 0.5°.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    With a 0.005 threshold, a significant difference was found between pelvis rotation errors of methods 2 and 3 (p < 0.001), while no significant differences were found for pelvis translation errors (p = 0.045), femur rotation errors (p = 0.089), or femur translation errors (p = 0.268).

    “Using a 0.005 threshold, a significant difference was found between the pelvis rotation errors (p < 0.001), while no significant differences were found between pelvis translation errors (p = 0.045), femur rotation errors (p = 0.089), and femur translation errors (p = 0.268).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    For method 2, correlation coefficients between dice scores and pelvis rotation and translation errors were −0.32 and −0.29, respectively; for method 3, they were −0.31 and −0.33.

    “For method 2, the correlation coefficients for the pelvis rotation and translation errors were −0.32 and −0.29, respectively, and −0.31 and −0.33, respectively for method 3.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    When testing on specimen 6, average landmark localization error was 10.0 mm, false negative rate was 30%, and successful pelvis registration rate was 29% for both methods 2 and 3.

    “When testing on spec. 6, average landmark localization error was 10.0 mm, with a false negative rate of 30%, and successful pelvis registration rate of 29% for both methods 2 and 3.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    CNNs have excelled at landmark detection and semantic segmentation when sufficiently large annotated training datasets are available.

    “Convolutional neural networks (CNNs) have excelled at detecting landmarks and performing semantic segmentation when sufficiently large annotated datasets are available for training [10,11].”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites [10] Newell A, Yang K, Deng J (2016) Stacked hourglass networks for human pose estimation. In: Proceedings of European conference on computer vision. Springer, pp 483–499 source
    • cites [11] Shelhamer E, Long J, Darrell T (2017) Fully convolutional networks for semantic segmentation. IEEE Trans Pattern Anal Mach Intell 4:640–651 source
  • May combine two claims.

    Interpretationnone given

    The paper's online intraoperative registration strategy, leveraging image intensities and CNN features, is fully automatic, requires no user initialization, and completes in seconds.

    “An online, intraoperative, registration strategy, leveraging image intensities and CNN-features, that is fully automatic, requires no initialization from a user, and completes in an order of seconds.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationown data

    The naive intraoperative registration performed poorly, succeeding in only 18% of trials, while methods leveraging CNN annotations succeeded more than four times as often.

    “The naive intraoperative registration performed poorly, succeeding in only 18% of trials, while the methods leveraging CNN annotations succeeded over 4 times as often.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: Human-assisted annotations of the intraoperative image are typically required to efficiently solve complex registration problems during navigation.
    • Rests on an assumption: Manual initialization interferes with the surgical workflow and diminishes advantages gained from navigation.
  • May combine two claims.

    Interpretationreasoning

    Method 3 was robust to poor initializations, which most likely led to the larger number of successful pelvis registrations compared with method 2.

    “Method 3 was robust to poor initializations, which most likely resulted in the larger number of successful pelvis registrations compared to those resulting from method 2.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: Large errors in pose estimates may occur when a registration is not initialized sufficiently close to the actual pose of an object.
  • Classified as probably not a substantial claim. May combine two claims.

    Study design and conductown data

    Test data sets consisted of images from a single specimen, and no images from that specimen were present in the training and validation data.

    “Test data sets are comprised of images collected from a single specimen, and no images derived from this specimen are present in the training and validation data.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationown data

    Despite a fairly large false-negative detection rate of 17%, an average of 7 landmarks per image were detected, allowing methods 2 and 3 to perform well.

    “Despite the fairly large false-negative detection rate of 17%, an average of 7 landmarks per image were detected, allowing methods 2 and 3 to perform well.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductnone given

    Manual inspection was performed to prune images corresponding to failed registrations from the training dataset.

    “Manual inspection is performed so that images corresponding to failed registrations are pruned from the dataset.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Study design and conductnone given

    The 2D labels and landmarks for each projection were manually inspected and verified.

    “The 2D labels and landmarks for each projection are manually inspected and verified.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationreasoning

    The pelvis's fluoroscopic appearance is more sensitive to rotational changes than the femur's, consistent with the smaller rotation errors observed for pelvis registrations.

    “The pelvis’ fluoroscopic appearance is thus more sensitive to rotational changes than that of the femur, which is consistent with the smaller rotation errors observed with registrations of the pelvis and shown in Table 2.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: The pelvis is a larger, more complex shape than the femur, occupying larger regions of the fluoroscopic views.
    • Rests on an assumption: The pelvis's fluoroscopic appearance is more sensitive to rotational changes than that of the femur.
  • May combine two claims.

    Interpretationreasoning

    The influence of pelvis translation errors on these estimates was minimal because nearly all pelvis translation error was in the depth direction, which has minor impact on single-view appearance.

    “The influence of pelvis translation errors on these estimates was minimal, since nearly all of the pelvis translation error was found in the depth direction, which has a minor impact on single-view appearance.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationreasoning

    This prior knowledge is incorporated into the registration strategies and causes mostly small femoral translations to be reported, resulting in the small translation errors for femur registrations.

    “This prior knowledge is incorporated into the registration strategies and causes mostly small femoral translations to be reported, which results in the small translation errors for femur registrations listed in Table 2.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: For intact hip anatomy, connective tissues joining each femoral head to the acetabular regions of the pelvis cause each femur and the pelvis to mostly translate together.
    • Rests on an assumption: As a result, the relative pose of a femur with respect to the pelvis contains very little translation.
  • May combine two claims.

    Interpretationreasoning

    This robustness removes the need for manual initialization and allows navigation to be more naturally integrated into existing surgical workflows.

    “This is a clinically relevant result, as this robustness removes the need for manual initialization and allows navigation to be more naturally integrated into existing surgical workflows.”
    Quote in the paperSubstantial claim
23 of 23
  • The paper relies on this without support.

    Human-assisted annotations of the intraoperative image are typically required to efficiently solve complex registration problems during navigation.

    “In order to efficiently solve the complex registration problems presented during navigation, human-assisted annotations of the intraoperative image are typically required.”
  • The paper relies on this without support.

    Manual initialization interferes with the surgical workflow and diminishes advantages gained from navigation.

    “This manual initialization interferes with the surgical workflow and diminishes any advantages gained from navigation.”
  • The paper relies on this without support.

    Manual initialization of processing may interrupt surgical workflows and negatively affect patient outcomes through increased operating time or blood loss.

    “Although these systems report favorable navigation-related results, the manual initialization of processing may interrupt surgical workflows and negatively affect patient outcomes due to increased operating time or blood loss.”
  • The paper relies on this without support.

    Large errors in pose estimates may occur when a registration is not initialized sufficiently close to the actual pose of an object.

    “Large errors in pose estimates may occur when a registration is not initialized sufficiently close to the actual pose of an object.”
  • The paper relies on this without support.

    Quality initializations are derived from manual human input, often through annotated landmark locations in the fluoroscopic image.

    “Quality initializations are derived from some manual human input, often through annotated landmark locations in the fluoroscopic image.”
  • The paper relies on this without support.

    For intact hip anatomy, connective tissues joining each femoral head to the acetabular regions of the pelvis cause each femur and the pelvis to mostly translate together.

    “For the intact hip anatomy that is considered in this paper, the connective tissues joining each femoral head to the acetabular regions of the pelvis cause each femur and the pelvis to mostly translate together.”
  • The paper relies on this without support.

    As a result, the relative pose of a femur with respect to the pelvis contains very little translation.

    “As a result, the relative pose of a femur with respect to the pelvis contains very little translation.”
  • The paper relies on this without support.

    The pelvis is a larger, more complex shape than the femur, occupying larger regions of the fluoroscopic views.

    “In comparison to the femur, the pelvis is a larger, more complex shape, which occupies larger regions of the fluoroscopic views.”
  • The paper relies on this without support.

    The pelvis's fluoroscopic appearance is more sensitive to rotational changes than that of the femur.

    “The pelvis’ fluoroscopic appearance is thus more sensitive to rotational changes than that of the femur”
9 of 9

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Landmark regularization achieved an SMSR of 1.7%.

    “Landmark regularization [5], an extension of xReg that uses predicted 2D landmarks as an additional loss term during optimization from the fixed initialization, achieved an SMSR of 1.7%.”

    From Rapid patient-specific neural networks for intraoperative X-ray to volume registration (2025) · cites this paper as [5]

    The Methods (Intraoperative Registration; page unavailable) describe using predicted 2D landmarks in a reprojection regularizer for intensity-based registration. Neither that passage nor the Results (Abstract; page unavailable) report an SMSR of 1.7%; the Results instead report registration within 1° for 86% of images, a different measure. (LLM, conf=0.86)

    Experimental — accuracy not yet measured
  • In the Femur and Ljubljana datasets, there are no 2D segmentation masks from which to regress annotated fiducials, so landmark localization is not feasible with the UNet model proposed in [5].

    “For example, in the Femur and Ljubljana datasets, there are no 2D segmentation masks from which to regress annotated fiducials, so landmark localization is not feasible with the UNet model proposed in [5].”

    From Rapid patient-specific neural networks for intraoperative X-ray to volume registration (2025) · cites this paper as [5]

    The Methods sections on Network Architecture and Training and Extracting Landmark Locations describe a U-Net with segmentation and landmark-heatmap paths that uses both outputs to locate landmarks. The paper does not discuss the Femur or Ljubljana datasets, establish whether they lack 2D masks, or conclude that landmark localization on them is infeasible. Page numbers are unavailable in our copy. (LLM, conf=0.86)

    Experimental — accuracy not yet measured
  • Neural networks trained on small datasets for 2D/3D registration fail on new patients or require impractical landmark supervision.

    “neural networks trained on small datasets fail on new patients or require impractical landmark supervision”

    From Intraoperative 2D/3D Image Registration via Differentiable X-Ray Rendering (2024) · cites this paper as [25]

    The Discussion and Conclusion (page unavailable) reports poor registration on one held-out specimen whose C-arm poses differed from those in training, supporting a limited generalization concern—not a general claim that networks trained on small datasets fail on new patients. The Methods, Training Dataset Creation (page unavailable) describes automatically generating landmark labels and manually verifying them; it does not establish that landmark supervision is impractical. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • Landmark-based localization requires expert knowledge of landmarks visible on X-ray and manual annotation of 3D landmarks for every new CT volume.

    “Such methods require expert knowledge of landmarks that are visible on X-ray for the specific surgical application and manual annotation of 3D landmarks for every new CT volume [25].”

    From Intraoperative 2D/3D Image Registration via Differentiable X-Ray Rendering (2024) · cites this paper as [25]

    Methods > Data Preprocessing reports that 14 anatomical landmarks were manually annotated in 3D, and Experiments and Results > Data Collection and Training Dataset Creation says landmarks were manually digitized for each CT scan in the study. These passages support the manual-annotation part, but do not establish that landmark-based methods generally require expert knowledge of X-ray-visible landmarks for each surgical application or manual annotation for every new CT volume. (LLM, conf=0.83)

    Experimental — accuracy not yet measured
  • The CMAES benchmark method for pose estimation of bone anatomy and surgical devices has been tested to meet clinical requirements in various orthopedic applications.

    “Due to its well-known robustness to local minima, the CMAES benchmark method for pose estimation of bone anatomy and surgical devices has been tested to meet clinical requirements in various orthopedic applications, including osteotomy [2], [13], femoroplasty [41], [4], core decompression of the hip [42], and transforaminal lumbar epidural injections [43], etc.”

    From A Fully Differentiable Framework for 2D/3D Registration and the Projective Spatial Transformers (2024) · cites this paper as [13]

    The Methods (Training Dataset Creation and Intraoperative Registration; pages unavailable) use CMA-ES for pelvis and femur registration, and the Results evaluate registration on cadaveric hip fluoroscopy. The paper does not test CMA-ES for surgical-device pose estimation or show that it meets clinical requirements across the listed orthopedic applications; the Discussion describes registration of instruments as future research. (LLM, conf=0.88)

    Experimental — accuracy not yet measured

Cites

Flagged first, then by citations

4 of the 44 references couldn’t be matched to a known paper.

Cited by

Sorted by citations

Authors
Grupp, R. B., Unberath, M., Gao, C., Hegeman, R. A., Murphy, R. J., Alexander, C. P., Otake, Y., McArthur, B. A., Armand, M., & Taylor, R. H.
Affiliations
  • Johns Hopkins University (US)
  • Johns Hopkins University Applied Physics Laboratory (US)
  • Auris Health (United States) (US)
  • Nara Institute of Science and Technology (JP)
  • The University of Texas at Austin (US)
Corresponding author
Grupp, Robert B.
Venue
International Journal of Computer Assisted Radiology and Surgery
Published
2020-04-24
Volume
15
Issue
5
Pages
759-769
ISSN
  • 1861-6410 (ISSN-L)
  • 1861-6429
OpenAlex
W3017446891
Cited by
81
References
44
Citations by year
2026: 14 · 2025: 15 · 2024: 16 · 2023: 16 · 2022: 4 · 2021: 9 · 2020: 6
FWCI
6.94
Type
journal article
Language
en
Open access
Locations
  • International Journal of Computer Assisted Radiology and Surgery (publishedVersion)
  • PubMed (publishedVersion)
  • arXiv (Cornell University) (submittedVersion)
  • PubMed Central (submittedVersion)
  • arXiv (Cornell University)
Funders
  • National Institute of Biomedical Imaging and Bioengineering (R01EB006839, R21EB020113)
  • Johns Hopkins University Applied Physics Laboratory (FNACCX24)
Topics
  • Health Sciences › Medicine › Surgery › Orthopaedic implants and arthroplasty
  • Health Sciences › Medicine › Surgery › Pelvic and Acetabular Injuries
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
Keywords
2D-3D registration, intraoperative registration, computer-assisted navigation, hip surgery, neural network segmentation, anatomical landmark detection, femur segmentation, annotated dataset
MeSH terms
Algorithms, Femur, Fluoroscopy, Humans, Pelvis

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-09-29.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-09-29.

Cited sources are checked by gpt-6-sol.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review