Towards 2D/3D Registration of the Preoperative MRI to Intraoperative Fluoroscopic Images for Visualisation of Bone Defects
Open accessAbstract
Magnetic Resonance Imaging (MRI) is a medical imaging modality that allows for the evaluation of soft-tissue diseases and the assessment of bone quality. Preoperative MRI volumes are used by surgeons to identify defected bones, perform the segmentation of lesions, and generate surgical plans before the surgery. Nevertheless, conventional intraoperative imaging modalities such as fluoroscopy are less sensitive in detecting potential lesions. In this work, we propose a 2D/3D registration pipeline that aims to register preoperative MRI with intraoperative 2D fluoroscopic images. To showcase the feasibility of our approach, we use the core decompression procedure as a surgical example to perform 2D/3D femur registration. The proposed registration pipeline is evaluated using digitally reconstructed radiographs (DRRs) to simulate the intraoperative fluoroscopic images. The resulting transformation from the registration is later used to create overlays of preoperative MRI annotations and planning data to provide intraoperative visual guidance to surgeons. Our results suggest that the proposed registration pipeline is capable of achieving reasonable transformation between MRI and digitally reconstructed fluoroscopic images for intraoperative visualization applications.
AI snapshot
Written from the paper’s full text.
TL;DR
The paper proposes a 2D/3D registration pipeline that synthesizes CT from preoperative MRI with a gradient-consistency CycleGAN and registers it to simulated intraoperative fluoroscopic images for overlaying bone-defect annotations. In a core-decompression feasibility study using digitally reconstructed radiographs, the pipeline achieved mean femur registration error of 12.95 ± 4.90 mm and rotation error of 4.51 ± 2.67 degrees.
Key findings
- Across 60 simulated DRR projections, the mean femur registration error was 12.95 ± 4.90 mm, with a mean rotation error of 4.51 ± 2.67 degrees.
- The in-plane translation error, which the authors identify as critical for overlay accuracy, was 4.55 ± 1.92 mm; the total translation error was dominated by out-of-plane error.
- The CycleGAN-based MR-to-CT synthesis produced synthetic CT with SSIM 0.3556 ± 0.1238 and MSE 0.0981 ± 0.0216 compared with ground-truth CT.
- The resulting transformation allowed preoperative MRI annotations, including AVN lesion segmentation and planned drilling paths, to be overlaid on DRRs; the authors state this in-plane accuracy may provide useful visual guidance for a 6–10 mm drill, though it does not meet the 2-mm clinical requirement cited for core decompression.
Methods
- The pipeline includes CycleGAN-based MR-to-synthetic-CT generation with gradient consistency loss, automatic bone segmentation with occasional manual touch-ups, ICP initialization using 200 points around the femoral head, and intensity-based 2D/3D registration using Grad-NCC similarity and CMA-ES optimization.
- Data included three preoperative MRI volumes and three paired CT volumes from patients, plus 12 coronal torso CT volumes from the New Mexico Decedent Image Database for CycleGAN training; 60 DRRs were simulated using C-arm geometry approximating a Siemens CIOS Fusion C-Arm.
- Registration accuracy was evaluated against known DRR camera poses, with translation error separated into in-plane and out-of-plane components; synthetic CT quality was assessed with SSIM and MSE.
Limitations
- The paper states that translation error, especially out-of-plane error, is significant for robot-guided surgery and exceeds the 2-mm clinical requirement for core decompression; it also notes that synthetic CT segmentation can be inaccurate because of limited CycleGAN training data.
- The pipeline is semi-automatic, requiring manual clinician annotation of landmarks and occasional manual segmentation touch-ups, and the evaluation uses DRRs rather than real intraoperative fluoroscopy.
- Evident from the text: the clinical MRI sample is very small (three patients), no real intraoperative C-arm data are tested, and the reported SSIM of 0.3556 indicates limited synthetic CT image similarity.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
11 of 29 cited sources checked · 4 supported · 0 flagged · accuracy not yet measured
- Findingown data
The geometry of the synthetic CT femur projections mostly aligns with the DRRs except for the greater trochanter of the femur.
“An example is shown in Figure 8, where the geometry of the sCT femur projections mostly aligns with the DRRs except for the greater trochanter of the femur.”
Quote in the paperSubstantial claim - Findingown data
The 2D annotations of the preoperatively planned drilling trajectory and AVN lesion annotations are not visually affected by the out-of-plane translation errors.
“An example of the overlay is shown in Figure 9, where the 2D annotations of the preoperatively planned drilling trajectory and AVN lesions annotations are not visually affected by the out-of-plane translation errors.”
Quote in the paperSubstantial claim - Findingown data
The in-plane translation error has a mean of 4.55 ± 1.92 mm.
“The accuracy of the preoperative annotations on the intraoperative 2D radiographs is the same as the in-plane translation error acquired in Table 1, which has a mean of 4.55 ± 1.92 mm translation error and the same rotation registration error.”
Quote in the paperNot a substantial claim - Interpretationown data
The 4.55 ± 1.92 mm in-plane translation error and the 4.51 ± 2.67 degree rotation error acquired in this study ensure that the projection of the synthetic CT is positionally close to the target on the 2D imaging plane.
“Although the translation registration error in out-of-plane direction needs to be improved for clinical applications, the 4.55 ± 1.92 mm in-plane translation error and the 4.51 ± 2.67 degree rotation error acquired in this study ensure that the projection of the sCT is positionally close to the target on the 2D imaging plane.”
Quote in the paperSubstantial claim - Interpretationreasoning
Because the synthetic CT is generated from the CycleGAN-based network with limited training size, the border of the proximal femur in the synthetic CT may be affected by soft tissues in the MRI and become less accurate.
“Since the sCT is generated from the CycleGAN-based network with limited training size, the border of the proximal femur in the sCT may be affected by the soft tissues in the MRI and become less accurate.”
Quote in the paperSubstantial claim
- Interpretationown data
The gradient consistency integrated CycleGAN is a feasible method to perform style transfer from MRI to CT even with the limited training size of three MRI volumes.
“In Figure 6, the gradient consistency integrated CycleGAN has shown to be a feasible method to perform style transfer from the MRI modality to the CT modality even with the limited training size, as we currently only have access to a total of three MRI volumes with small image spacing between the coronal planes.”
Quote in the paperSubstantial claim - Premisecitation
There are theoretical accuracy limitations in a shape-matching process during registration because the estimated femur may be positioned closer or further to the camera along the z-axis of the femoral ACS frame.
“There are theoretical accuracy limitations in a shape-matching process during registration as the estimated femur may be positioned closer or further to the camera along the z-axis of the femoral ACS frame (Fregly et al. 2005).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Fregly et al. 2005 source
- Premisecitation
Direct automatic registration between MRI volumes and 2D radiography such as X-rays and C-arm fluoroscopy is challenging because differences in imaging physics and contrast mechanisms lead to varying tissue contrasts.
“Nevertheless, direct automatic registration between MRI volumes and 2D radiography such as X-rays and C-arm fluoroscopy has been a challenging task, as the differences in their imaging physics and contrast mechanisms lead to varying tissue contrasts (van der Bom et al. 2011).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites van der Bom et al. 2011 source
- Premisecitation
MRI is the most sensitive and specific imaging technology for diagnosis of early-stage and precollapse-stage AVN because it allows surgeons to visualize the size and position of the lesion.
“MRI is the most sensitive and specific imaging technology for the diagnosis of the early-stage and precollapse-stage AVN as it allows the surgeons to visualize the size and the position of the lesion (Saini and Saifuddin 2004).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Saini and Saifuddin 2004 source
- Interpretationreasoning
The misalignment of the greater trochanter may have a negative impact during the optimization step in registration.
“This misalignment may have a negative impact during the optimization step in registration.”
Quote in the paperSubstantial claim - Interpretationown data
The accuracy of preoperative annotations on intraoperative 2D radiographs is the same as the in-plane translation error.
“The accuracy of the preoperative annotations on the intraoperative 2D radiographs is the same as the in-plane translation error acquired in Table 1, which has a mean of 4.55 ± 1.92 mm translation error and the same rotation registration error.”
Quote in the paperNot a substantial claim - Interpretationnone given
The registration pipeline solves the transformation between preoperative MRI and intraoperative radiographs.
“Our registration pipeline solves the transformation between preoperative MRI and intraoperative radiographs.”
Quote in the paperSubstantial claim - Premisecitation
Conventional intraoperative imaging modalities such as fluoroscopy are less sensitive in detecting potential lesions, soft-tissue diseases and bone lesions.
“Nevertheless, conventional intraoperative imaging modalities such as fluoroscopy are less sensitive in detecting potential lesions.”
- Also stated in: § Introduction · plain
“Nevertheless, traditional intraoperative imaging modalities such as fluoroscopy are less sensitive in detecting soft tissues diseases and bone lesions, which makes surgeries that rely heavily on information from preoperative MRIs challenging (Pierce et al. 2015a).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Pierce et al. 2015a
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Also stated in: § Introduction · plain
- Interpretationnone given
The work demonstrates the feasibility of cross-modality registration that has not been investigated in depth in previous studies.
“In this work, we have demonstrated the feasibility of cross-modality registration that has not been investigated in depth in previous studies.”
Quote in the paperNot a substantial claim- Rests on an assumption: No research exists on MRI to 2D radiography registration as a tool for intraoperative visualization of bone defects.
- Rests on an assumption: Generation of DRRs creates a realistic simulation of intraoperative fluoroscopy.
- Rests on an assumption: The performance of the pipeline on DRRs will translate to real intraoperative fluoroscopic images.
- Interpretationown data
The proposed registration pipeline achieves a reasonable transformation between MRI and digitally reconstructed fluoroscopic images.
“The experiment results show that the proposed registration pipeline achieves reasonable transformation between MRI and digitally reconstructed fluoroscopic images.”
Quote in the paperSubstantial claim- Rests on an assumption: Generation of DRRs creates a realistic simulation of intraoperative fluoroscopy.
- Rests on an assumption: The performance of the pipeline on DRRs will translate to real intraoperative fluoroscopic images.
- Interpretationcitation
Applying the Iterative Closest Point algorithm for pose initialization reduces the probability of the 2D/3D registration step trapping in local optima.
“To reduce the probability of the 2D/3D registration step trapping in local optima, we apply Iterative Closest Point (ICP) algorithm (Rusinkiewicz and Levoy 2001) for better pose initialization.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Rusinkiewicz and Levoy 2001
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
Another source of the translation error may come from the synthetic CT segmentation.
“Another source of the translation error may come from the sCT segmentation.”
Quote in the paperNot a substantial claim - Interpretationreasoning
The proposed pipeline could be used to visualize preoperative annotations of segmented bone defects and medical instrument paths on intraoperative fluoroscopic images.
“Although this work uses the core decompression procedure as a surgical example, the proposed pipeline could be used to visualize preoperative annotations of segmented bone defects and medical instrument paths on intraoperative fluoroscopic images.”
Quote in the paperSubstantial claim- Rests on an assumption: Generation of DRRs creates a realistic simulation of intraoperative fluoroscopy.
- Rests on an assumption: The performance of the pipeline on DRRs will translate to real intraoperative fluoroscopic images.
- Interpretationcitation
For a common core-decompression surgery where a 6-10 mm cylindrical drill is used, the overlay accuracy will provide good intraoperative visual guidance for surgeons.
“For a common core-decompression surgery where a 6–10 mm cylindrical drill is used, we believe the overlay accuracy will provide good intraoperative visual guidance for the surgeons (Pierce et al. 2015b), and the accuracy may further improve once the multi-step and multi-view registration is integrated.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Pierce et al. 2015b
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Rests on an assumption: Generation of DRRs creates a realistic simulation of intraoperative fluoroscopy.
- Rests on an assumption: The performance of the pipeline on DRRs will translate to real intraoperative fluoroscopic images.
- May combine two claims. 17
- Classified as probably not a substantial claim. 13
May combine two claims.
PremisecitationIntraoperative imaging techniques such as C-arm fluoroscopy or X-ray are important for localizing the greater trochanter, tracking and navigating the drilling and injection device to the femoral head, and positioning the trephine and scraping spoon while removing dead bone.
“Intraoperative imaging techniques such as C-arm fluoroscopy or X-ray have become important for the localization of the greater trochanter, tracking and navigation of the drilling and injection device to the femoral head, and positioning of the trephine and scraping spoon while removing the dead bones (Nolte et al. 2000; Geisert et al. 2017; Tian et al. 2021).”
Quote in the paperSubstantial claim3 of 3 sources checked- cites Nolte et al. 2000 source
The Introduction and Surgical procedure sections describe C-arm fluoroscopy for navigating instruments during spinal pedicle-screw placement. They do not address the greater trochanter, femoral head, drilling or injection devices, or instruments for removing dead bone. Page numbers were not provided for these passages. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites Geisert et al. 2017 source
The Core Decompression section (page not provided) describes C-arm fluoroscopy guiding a pin toward the femoral-head lesion and live fluoroscopy guiding reaming. The Approach to Hip section describes an incision centered over the greater trochanter, but not imaging to localize it. The Osseoscopy section describes camera-guided curette debridement, not fluoroscopic positioning of a trephine or scraping spoon; the paper does not describe an injection device. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites Tian et al. 2021 source
The Surgical Procedure section says C-arm fluoroscopy located the greater trochanter and confirmed the drilling direction, while intraoperative X-rays helped plan the robot-assisted path. It also reports fluoroscopic checks during bone grafting, but does not say imaging positioned the trephine or scraping spoon while dead bone was removed. (LLM, conf=0.86)
Experimental — accuracy not yet measured
May combine two claims.
PremisecitationThe translation component of registration error is significant for purposes such as robot-guided surgery, while a 2-mm error is considered to match the clinical requirement of core decompression.
“In this experiment, the translation component of the registration error is a significant error for purposes such as robot-guided surgery, while a 2-mm error is considered to match the clinical requirement of core decompression (Wang et al. 2017).”
Quote in the paperSubstantial claim1 of 1 sources checked- cites Wang et al. 2017 source
The Discussion (page not provided) says a 1.92-mm error between the needle tip and target was considered to meet the clinical requirement for core decompression, supporting the approximate 2-mm claim. It does not assess the translation component of registration error or its significance in robot-guided surgery; the measured error was drilling accuracy using a navigation template. (LLM, conf=0.88)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationRegistration between breast MRI and X-ray mammography is feasible through a volume-preserving affine transformation model or gradient-based registration methods, reaching accuracy on the order of 10 mm.
“The registration between breast MRI and X-ray mammography has shown to be feasible through a properly defined volume-preserving affine transformation model (Mertzanidou et al. 2012) or through gradient-based registration methods (García et al. 2019). Both algorithms are able to reach registration accuracy in the order of 10mm.”
Quote in the paperNot a substantial claim2 of 2 sources checkedMay combine two claims.
PremisecitationCore decompression requires acquisition of multiple intraoperative radiographs, which increases the risk of surgical complications and increases radiation exposure for both the patient and the surgical team.
“Core decompression procedure requires the acquisition of multiple intraoperative radiographs, which increases the risk of surgical complications and increases the radiation exposure for both the patient and the surgical team (Beckmann et al. 2006; Wang et al. 2017).”
Quote in the paperSubstantial claim1 of 2 sources checked- cites Beckmann et al. 2006 source
- cites Wang et al. 2017 source
The Introduction says conventional core decompression is guided by C-arm X-ray, and the Discussion links multiple drillings to greater radiation exposure for patients and the surgical team and to surgical complications. But the paper does not say every core decompression procedure requires multiple intraoperative radiographs; it describes a template-guided procedure performed without fluoroscopy or C-arm X-ray. (LLM, conf=0.84)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationDue to the lack of distinct features of the proximal femur in DRRs, Gao et al. proposed a two-step registration where the pelvis is first registered for better femoral pose initialization because the pelvis has a rich collection of distinct features leading to accurate registration.
“Due to the lack of distinct features of the proximal femur in the DRRs, Gao et al. have proposed a two-step registration where the pelvis is first registered for a better femoral pose initialization since the pelvis has a rich collection of distinct features that leads to accurate registration accuracy (Gao et al. 2020).”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Gao et al. 2020 source
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationCore decompression has become a preferred procedure to treat early-stage AVN due to its simplicity, low complication rate, and promising efficacy.
“Core decompression has become a preferred procedure to treat early-stage AVN due to its simplicity, low complication rate, and promising efficacy (Lieberman 2004; Veillette et al. 2006; Talmaç 2018; Nori 2015).”
Quote in the paperNot a substantial claim1 of 3 sources checkedClassified as probably not a substantial claim. May combine two claims.
PremisecitationMRI is a medical imaging modality that allows for the evaluation of soft-tissue diseases and the assessment of bone quality.
“Magnetic Resonance Imaging (MRI) is a medical imaging modality that allows for the evaluation of soft-tissue diseases and the assessment of bone quality.”
- Also stated in: § Introduction · plain
“This property has made MRI a good candidate for the evaluation of soft-tissue diseases, the assessment of bone quality, and the detection of bone tumors and tumor-like lesions (Chung et al. 2012; Chang et al. 2017; Nascimento et al. 2014).”
Quote in the paperSubstantial claim2 of 3 sources checked- cites Chung et al. 2012 source
- cites Chang et al. 2017 source
The paper’s Abstract (page not provided) describes MRI as a proposed method for assessing bone quality in vivo. The passages address bone structure and physiology, but do not substantiate the separate claim that MRI evaluates soft-tissue diseases (searched: soft-tissue disease and musculoskeletal disease imaging). (LLM, conf=0.85)
Experimental — accuracy not yet measured - cites Nascimento et al. 2014 source
In “Applying MRI in tumour detection and diagnosis” (page unavailable), the paper says MRI can demonstrate soft-tissue pathology associated with osteochondromas. The Introduction and tumour-detection section also describe MRI’s ability to characterise bone and bone-marrow lesions, but they do not establish that it assesses bone quality. (LLM, conf=0.85)
Experimental — accuracy not yet measured
- Also stated in: § Introduction · plain
May combine two claims.
Findingown dataThe final SSIM result has a mean of 0.3556 and a standard deviation of 0.1238, while the MSE result has a mean of 0.0981 and a standard deviation of 0.0216.
“The final SSIM result has a mean of 0.3556 and a standard deviation of 0.1238; while the MSE result has a mean of 0.0981 and a standard deviation of 0.0216.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe generated synthetic CT volumes preserve compact bone structures from the input MRI volumes and map them to high CT intensity, while visible structures in MRI such as muscles, urinary bladder, and sciatic nerves are mapped to low intensities in the synthetic CT.
“The generated sCT volumes preserve compact bone structures from the input MRI volumes and map them to a high CT intensity; while visible structures in MRI such as muscles, urinary bladder, and sciatic nerves are mapped to low intensities in the sCT.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingnone givenThe left or right femur does not demonstrate a significant impact on the registration result.
“The left or right femur does not demonstrate a significant impact on the registration result.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataThe in-plane translation error in the first subject is slightly smaller than in the other two subjects, while the rotation error remains similar among all three subjects.
“As shown in Figure 10, the in-plane translation error in the first subject is slightly smaller than the other two subjects, while the rotation error remains similar among all three subjects.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe mean femur registration error is 12.95 ± 4.90 mm, and the rotation error is 4.51 ± 2.67 degrees.
“The mean femur registration error is 12.95 ± 4.90 mm, and the rotation error is 4.51 ± 2.67 degrees.”
- Also stated in: § Discussion · plain
“The registration result of the femur bone between the sCT and the DRR has a mean translation error of 12.95 ± 4.90 mm with a mean rotation error of 4.51 ± 2.67 degrees.”
Quote in the paperSubstantial claim- Also stated in: § Discussion · plain
Classified as probably not a substantial claim.
Interpretationnone givenThe proposed 2D/3D registration pipeline registers preoperative MRI and intraoperative fluoroscopic images without requiring preoperative CT scans from the patient.
“In this paper, we propose a novel 2D/3D registration pipeline that aims to register preoperative MRI and intraoperative fluoroscopic images without the requirement of preoperative CT scans from the patient.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningBecause slight movement along the z-axis maintains the projected shape of the femur in synthetic CT, it has a smaller impact on similarity scores during registration, leading to low rotation registration errors but less ideal translation registration results.
“Since slight movement along this axis maintains the projected shape of the femur in sCT, it has a smaller impact on the calculation of similarity scores during registration, which leads to low rotation registration errors but less ideal translation registration results.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable). May combine two claims.
InterpretationreasoningWith improved intraoperative visualization of bone defects, the proposed registration pipeline could improve the performance of surgical procedures and potentially reduce the number of fluoroscopic images required during surgery.
“With the improved intraoperative visualization of bone defects, the proposed registration pipeline could improve the performance of surgical procedures and potentially reduce the number of fluoroscopic images required during surgery.”
Quote in the paperNot a substantial claim- Rests on an assumption: Generation of DRRs creates a realistic simulation of intraoperative fluoroscopy.
- Rests on an assumption: The performance of the pipeline on DRRs will translate to real intraoperative fluoroscopic images.
May combine two claims.
InterpretationreasoningIn this study, the translation error mostly originates from the out-of-plane translation error and the less accurate synthetic CT model used for registration.
“In this study, the translation error mostly originates from the out-of-plane translation error and the less accurate sCT model used for registration.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
PremisecitationDuring core decompression surgery, the precision of drilling is essential to accurately decompress targeted lesions while avoiding femoral cartilage damage.
“During the surgery, the precision of the drilling is essential to accurately decompress the targeted lesions while avoiding any femoral cartilage damage (Beckmann et al. 2006).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Beckmann et al. 2006 source
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationAdditional loss functions such as gradient consistency loss and structure-consistency loss have shown promising results for avoiding geometric distortions in MR-to-CT synthesis.
“To avoid geometric distortions, additional loss functions such as gradient consistency loss (Hiasa et al. 2018) and structure-consistency loss (Yang et al. 2020) have shown promising results.”
Quote in the paperNot a substantial claim0 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationCore decompression creates a drilling hole from the distal end of the trochanter to stimulate blood supply to the necrotic area, diminish femoral head pressure, and delay progression of osteonecrosis.
“The core decompression method is a surgical procedure that creates a drilling hole made from the distal end of the trochanter to stimulate the blood supply to the necrotic area, diminish the femoral head pressure, and delay the progression of osteonecrosis (Lieberman 2004; Yu et al. 2018).”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Lieberman 2004 source
- cites Yu et al. 2018
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Interpretationnone givenWith a trained network model, the generated synthetic CT will detect and preserve bone structures with high image intensities and reduce the intensity impact of soft tissues from the input MRI.
“With a trained network model, the generated sCT will detect and preserve the bone structures with high image intensities and reduce the intensity impact of the soft tissues from the input MRI.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Interpretationnone givenAlthough the experiment is restricted to core decompression procedures, no assumptions were made that would preclude application of the proposed registration pipeline to other surgical procedures.
“Although the experiment carried out in this paper is restricted to core decompression procedures, no assumptions were made that would preclude the application of the proposed registration pipeline to other surgical procedures.”
Quote in the paperNot a substantial claim- Rests on an assumption: The core decompression procedure is a suitable surgical example for evaluating the proposed pipeline.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationPerforming MR-to-CT synthesis through the naive CycleGAN often results in inconsistent structures between input and synthetic images.
“However, performing MR-to-CT synthesis through the naive CycleGAN often results in inconsistent structures between input and synthetic images.”
Quote in the paperNot a substantial claim0 of 2 sources checkedMay combine two claims.
PremisecitationAvascular necrosis of the femoral head is visible in preoperative MRI but hardly detectable intraoperatively; radiographs have low sensitivity for early-stage AVN, while MRI is sensitive to necrotic tissues but hard to acquire intraoperatively.
“A clear example of a lesion that is visible in preoperative MRI but hardly detectable intraoperatively is avascular necrosis (AVN).”
- Also stated in: § Introduction · boosted
“However, studies have shown that radiographs exhibit low sensitivity for the early stages of AVN; while MRI scans are particularly sensitive to the detection of necrotic tissues, MRI scans are hard to acquire intraoperatively (Malizos et al. 2007).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Malizos et al. 2007 source
- Also stated in: § Introduction · boosted
The paper relies on this without support.
No research exists on MRI to 2D radiography registration as a tool for intraoperative visualization of bone defects.
“To our knowledge, there is no research on MRI to 2D radiography registration as a tool for intraoperative visualization of bone defects.”
The paper relies on this without support.
Generation of DRRs creates a realistic simulation of intraoperative fluoroscopy.
“The generation of DRR creates a realistic simulation of the intraoperative fluoroscopy, and it also enables us to perform a more comprehensive 2D/3D registration analysis with the increased number of projections available in various projecting directions.”
The paper relies on this without support.
The core decompression procedure is a suitable surgical example for evaluating the proposed pipeline.
“In this paper, we use the core decompression procedure as a surgical example for the proposed pipeline.”
The paper relies on this without support.
The performance of the pipeline on DRRs will translate to real intraoperative fluoroscopic images.
“The generated DRRs (Im) are used to simulate intraoperative radiographs and are used in the 2D/3D registration step.”
The paper relies on this without support.
Manual annotation of three anatomical landmarks reliably defines the femoral anatomical coordinate system.
“To construct the femoral ACS, three anatomical landmarks are manually annotated in the CT volumes prior to the experiments: the center of the femoral head (FH), the tip of the supero-posterior facet of the greater trochanter (GTR), and the tip of the lesser trochanter (LTR) (Domb and Carreira 2013).”
The paper relies on this without support.
The 12 coronal torso CT volumes from the New Mexico Decedent Image Database are representative for training the MR-to-CT synthesis network.
“a total of 12 coronal torso CT volumes from the New Mexico Decedent Image Database (NMDID) are used as our CT dataset.”
Cites
Unpaired image-to-image translation using cycle-consistent adversarial networks
Zhu, J.-Y., Park, T., Isola, P., & Efros, A. A. (2017) · 2017 IEEE International Conference on Computer Vision (ICCV) · doi:10.1109/iccv.2017.244 · Cited by 23,602 · see citing works
Efficient variants of the ICP algorithm
Rusinkiewicz, S., & Levoy, M. (2002) · Proceedings Third International Conference on 3-D Digital Imaging and Modeling · doi:10.1109/im.2001.924423 · Cited by 3,945 · cited 2× · see citing works
“To reduce the probability of the 2D/3D registration step trapping in local optima, we apply Iterative Closest Point (ICP) algorithm (Rusinkiewicz and Levoy 2001) for better pose initialization.”
MethodHansen, N., Müller, S. D., & Koumoutsakos, P. (2003) · Evolutionary Computation · doi:10.1162/106365603321828970 · Cited by 2,601 · see citing works
The Insight ToolKit image registration framework
Avants, B. B., Tustison, N. J., Stauffer, M., Song, G., Wu, B., & Gee, J. C. (2014) · Frontiers in Neuroinformatics · doi:10.3389/fninf.2014.00044 · Cited by 658 · cited 1× · see citing works
“The ground truth CT is acquired by performing registration through advanced normalisation Tools (ANTs) (Avants et al. 2014).”
MethodMedical Image Synthesis with Deep Convolutional Adversarial Networks
Nie, D., Trullo, R., Lian, J., Wang, L., Petitjean, C., Ruan, S., Wang, Q., & Shen, D. G. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2814538 · Cited by 626 · cited 2× · see citing works
“As MR-to-CT synthesis could be considered a style-transfer problem, algorithms based on Generative Adversarial Net (GAN) and its variants have been utilized to perform cross-modality image synthesis in medical images (Nie et al. 2018; Brou Boni et al. 2020).”
MethodOsteonecrosis of the femoral head: Etiology, imaging and treatment
Malizos, K. N., Karantanas, A. H., Varitimidis, S. E., Dailiana, Z. H., Bargiotas, K., & Maris, T. (2007) · European Journal of Radiology · doi:10.1016/j.ejrad.2007.03.019 · Cited by 316 · cited 1× · see citing works
“However, studies have shown that radiographs exhibit low sensitivity for the early stages of AVN; while MRI scans are particularly sensitive to the detection of necrotic tissues, MRI scans are hard to acquire intraoperatively (Malizos et al. 2007).”
BackgroundLeynes, A. P., Yang, J., Wiesinger, F., Kaushik, S. S., Shanbhag, D., Seo, Y., Hope, T. A., & Larson, P. E. Z. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.117.198051 · Cited by 250 · cited 1× · see citing works
“…the development of learning-based algorithms allows the estimation of non-linear mappings between MRI and CT. Self-optimising algorithms such as random forest (Andreasen et al. 2016) and convolutional neural networks were used to generate synthetic CT of the pelvis from MRI (Leynes et al. 2017).”
MethodHiasa, Y., Otake, Y., Takao, M., Matsuoka, T., Takashima, K., Carass, A., Prince, J. L., Sugano, N., & Sato, Y. (2018) · Lecture Notes in Computer Science · doi:10.1007/978-3-030-00536-8_4 · Cited by 221 · cited 4× · see citing works
“To avoid geometric distortions, additional loss functions such as gradient consistency loss (Hiasa et al. 2018) and structureconsistency loss (Yang et al. 2020) have shown promising results.”
BackgroundMethodResultInfluentialUnsupervised MR-to-CT Synthesis Using Structure-Constrained CycleGAN
Yang, H., Sun, J., Carass, A., Zhao, C., Lee, J., Prince, J. L., & Xu, Z. (2020) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2020.3015379 · Cited by 204 · cited 2× · see citing works
“2018) and structure-consistency loss (Yang et al. 2020) have shown promising results.”
BackgroundResultA new approach to computer-aided spine surgery: fluoroscopy-based surgical navigation
Nolte, L. P., Slomczykowski, M. A., Berlemann, U., Strauss, M. J., Hofstetter, R., Schlenzka, D., Laine, T., & Lund, T. (2000) · European Spine Journal · doi:10.1007/pl00010026 · Cited by 159 · cited 2× · see citing works
“…X-ray have become important for the localisation of the greater trochanter, tracking and navigation of the drilling and injection device to the femoral head, and positioning of the trephine and scraping spoon while removing the dead bones (Nolte et al. 2000; Geisert et al. 2017; Tian et al. 2021).”
BackgroundFregly, B. J., Rahman, H. A., & Banks, S. A. (2005) · Journal of Biomechanical Engineering · doi:10.1115/1.1933949 · Cited by 155 · cited 1× · see citing works
“There are theoretical accuracy limitations in a shape-matching process during registration as the estimated femur may be positioned closer or further to the camera along the z-axis of the femoral ACS frame (Fregly et al. 2005).”
BackgroundMRI assessment of bone structure and microarchitecture
Chang, G., Boone, S., Martel, D., Rajapakse, C. S., Hallyburton, R. S., Valko, M., Honig, S., & Regatte, R. R. (2017) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25647 · Cited by 151 · cited 2× · see citing works
“This property has made MRI a good candidate for the evaluation of soft-tissue diseases, the assessment of bone quality, and the detection of bone tumours and tumour-like lesions (Chung et al. 2012; Nascimento et al. 2014; Chang et al. 2017).”
BackgroundVEILLETTE, C. J. H., MEHDIAN, H., SCHEMITSCH, E. H., & MCKEE, M. D. (2006) · The Journal of Bone and Joint Surgery-American Volume · doi:10.2106/00004623-200611001-00008 · Cited by 140 · see citing works
A current review of core decompression in the treatment of osteonecrosis of the femoral head
Pierce, T. P., Jauregui, J. J., Elmallah, R. K., et al. (2015) · Current Reviews in Musculoskeletal Medicine · doi:10.1007/s12178-015-9280-0 · Cited by 140 · cited 3× · see citing works
“(Pierce et al. 2015b), and the accuracy may further improve once the multi-step and multi-view registration is integrated.”
BackgroundSaini, A., & Saifuddin, A. (2004) · Clinical Radiology · doi:10.1016/j.crad.2004.04.014 · Cited by 130 · cited 2× · see citing works
“MRI is the most sensitive and specific imaging technology for the diagnosis of the early-stage and precollapse-stage AVN as it allows the surgeons to visualize the size and the position of the lesion (Saini and Saifuddin 2004).”
BackgroundChung, W. J., Chung, H. W., Shin, M. J., Lee, S. H., Lee, M. H., Lee, J. S., Kim, M.-J., & Lee, W. K. (2012) · The British Journal of Radiology · doi:10.1259/bjr/27487871 · Cited by 114 · cited 1× · see citing works
“This property has made MRI a good candidate for the evaluation of soft-tissue diseases, the assessment of bone quality, and the detection of bone tumours and tumour-like lesions (Chung et al. 2012; Nascimento et al. 2014; Chang et al. 2017).”
BackgroundCore Decompression for Osteonecrosis of the Hip
Lieberman, J. R. (2004) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-200401000-00006 · Cited by 111 · cited 2× · see citing works
“The core decompression method is a surgical procedure that creates a drilling hole made from the distal end of the trochanter to stimulate the blood supply to the necrotic area, diminish the femoral head pressure, and delay the progression of osteonecrosis (Lieberman 2004; Yu et al. 2018).”
BackgroundMethodThe role of magnetic resonance imaging in the evaluation of bone tumours and tumour-like lesions
Nascimento, D., Suchard, G., Hatem, M., & de Abreu, A. (2014) · Insights into Imaging · doi:10.1007/s13244-014-0339-z · Cited by 106 · cited 2× · see citing works
“This property has made MRI a good candidate for the evaluation of soft-tissue diseases, the assessment of bone quality, and the detection of bone tumours and tumour-like lesions (Chung et al. 2012; Nascimento et al. 2014; Chang et al. 2017).”
BackgroundFully automatic and fast segmentation of the femur bone from 3D-CT images with no shape prior
Krčah, M., Székely, G., & Blanc, R. (2011) · IEEE International Symposium on Biomedical Imaging: From Nano to Macro (ISBI) · doi:10.1109/isbi.2011.5872823 · Cited by 98 · cited 1× · see citing works
“To perform the ICP algorithm, both sCT and rCT volumes are first segmented through the automatic bone segmentation method introduced by Krcah et al. (Krčah et al. 2011).”
MethodImaging evaluation of patients with osteonecrosis of the femoral head
Pierce, T. P., Jauregui, J. J., Cherian, J. J., et al. (2015) · Current Reviews in Musculoskeletal Medicine · doi:10.1007/s12178-015-9279-6 · Cited by 77 · cited 3× · see citing works
“lesions, which makes surgeries that rely heavily on information from preoperative MRIs challenging (Pierce et al. 2015a).”
BackgroundBoni, K. B., Klein, J., Vanquin, L., Wagner, A., Lacornerie, T., Pasquier, D., & Reynaert, N. (2020) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab7633 · Cited by 75 · cited 1× · see citing works
“As MR-to-CT synthesis could be considered a style-transfer problem, algorithms based on Generative Adversarial Net (GAN) and its variants have been utilised to perform crossmodality image synthesis in medical images (Nie et al. 2018; Brou Boni et al. 2020).”
MethodEndoscopic Repair of Full‐Thickness Gluteus Medius Tears
Domb, B. G., & Carreira, D. S. (2013) · Arthroscopy Techniques · doi:10.1016/j.eats.2012.11.005 · Cited by 65 · cited 1× · see citing works
“To construct the femoral ACS, three anatomical landmarks are manually annotated in the CT volumes prior to the experiments: the centre of the femoral head (FH), the tip of the superoposterior facet of the greater trochanter (GTR), and the tip of the lesser trochanter (LTR) (Domb and Carreira 2013).”
MethodFu, J. Y., Singhrao, K., Cao, M., et al. (2020) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/ab6e1f · Cited by 54 · see citing works
Yu, X., Zhang, D., Chen, X., et al. (2017) · Journal of Orthopaedic Science · doi:10.1016/j.jos.2017.12.004 · Cited by 52 · see citing works
Kalantar, R., Messiou, C., Winfield, J. M., et al. (2021) · Frontiers in Oncology · doi:10.3389/fonc.2021.665807 · Cited by 48 · cited 1× · see citing works
“CycleGAN-based frameworks have been used in MR-to-CT synthesis for different anatomical structures (Fu et al. 2020; Kalantar et al. 2021).”
MethodPose Estimation of Periacetabular Osteotomy Fragments With Intraoperative X-Ray Navigation
Grupp, R. B., Hegeman, R. A., Murphy, R. J., Alexander, C. P., Otake, Y., McArthur, B. A., Armand, M., & Taylor, R. H. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2019.2915165 · Cited by 38 · cited 3× · see citing works
“Registration pipelines between CT and 2D radiographs have been proposed for surgical treatments around the hip region such as periacetabular osteotomy (Grupp et al. 2020) and femoroplasty (Gao et al. 2020).”
BackgroundMethodSelf-citationFiducial-Free 2D/3D Registration for Robot-Assisted Femoroplasty
Gao, C., Farvardin, A., Grupp, R. B., Bakhtiarinejad, M., Ma, L., Thies, M., Unberath, M., Taylor, R. H., & Armand, M. (2020) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2020.3012460 · Cited by 36 · cited 3× · see citing works
“Registration pipelines between CT and 2D radiographs have been proposed for surgical treatments around the hip region such as periacetabular osteotomy (Grupp et al. 2020) and femoroplasty (Gao et al. 2020).”
BackgroundSelf-citationAndreasen, D., Edmund, J., Zografos, V., et al. (2016) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2216924 · Cited by 33 · cited 2× · see citing works
“…the development of learning-based algorithms allows the estimation of non-linear mappings between MRI and CT. Self-optimising algorithms such as random forest (Andreasen et al. 2016) and convolutional neural networks were used to generate synthetic CT of the pelvis from MRI (Leynes et al. 2017).”
MethodPrecision of computer-assisted core decompression drilling of the femoral head
Beckmann, J., Goetz, J., Baethis, H., Kalteis, T., Grifka, J., & Perlick, L. (2006) · Archives of Orthopaedic and Trauma Surgery · doi:10.1007/s00402-006-0155-0 · Cited by 27 · cited 2× · see citing works
“Core decompression procedure requires the acquisition of multiple intraoperative radiographs, which increases the risk of surgical complications and increases the radiation exposure for both the patient and the surgical team (Beckmann et al. 2006; Wang et al. 2017).”
BackgroundBreast MRI and X-ray mammography registration using gradient values
García, E., Diez, Y., Diaz, O., Lladó, X., Gubern-Mérida, A., Martí, R., Martí, J., & Oliver, A. (2019) · Medical Image Analysis · doi:10.1016/j.media.2019.02.013 · Cited by 26 · cited 1× · see citing works
“2012) or through gradient-based registration methods (Garćıa et al. 2019).”
MethodMRI to X-ray mammography registration using a volume-preserving affine transformation
Mertzanidou, T., Hipwell, J., Cardoso, M. J., Zhang, X., Tanner, C., Ourselin, S., Bick, U., Huisman, H., Karssemeijer, N., & Hawkes, D. (2012) · Medical Image Analysis · doi:10.1016/j.media.2012.03.001 · Cited by 24 · cited 2× · see citing works
“The registration between breast MRI and X-ray mammography has shown to be feasible through a properly defined volume-preserving affine transformation model (Mertzanidou et al. 2012) or through gradient-based registration methods (Garćıa et al.”
MethodGrupp, R. B., Armand, M., & Taylor, R. H. (2018) · Lecture Notes in Computer Science · doi:10.1007/978-3-030-01201-4_17 · Cited by 23 · cited 2× · see citing works
“The normalized gradient cross-correlation (Grad-NCC) method is used to calculate the similarity scores between the image patches (Grupp et al. 2018).”
MethodSelf-citationRegistration of 2D x-ray images to 3D MRI by generating pseudo-CT data
van der Bom, M. J., Pluim, J. P. W., Gounis, M. J., van de Kraats, E. B., Sprinkhuizen, S. M., Timmer, J., Homan, R., & Bartels, L. W. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/4/010 · Cited by 21 · see citing works
Assessment of spline‐based 2D–3D registration for image‐guided spine surgery
Zhang, X., Zheng, G., Langlotz, F., et al. (2006) · Minimally Invasive Therapy & Allied Technologies · doi:10.1080/13645700600769474 · Cited by 20 · cited 1× · see citing works
“Performing a 2D/3D registration between preoperative volumes and intraoperative projection images has made significant improvements to orthopaedic surgeries (Zhang et al. 2006).”
BackgroundRocchi, M., Piccolo, N. D., Mazzotta, A., et al. (2020) · Hip International · doi:10.1177/1120700020964996 · Cited by 20 · cited 1× · see citing works
“Surgical treatments of AVN during the early stages are crucial as they result in better treatment outcomes (Rocchi et al. 2020).”
BackgroundPatient-specific core decompression surgery for early-stage ischemic necrosis of the femoral head
Wang, W., Hu, W., Yang, P., Dang, X. Q., Li, X. H., & Wang, K. Z. (2017) · PLOS ONE · doi:10.1371/journal.pone.0175366 · Cited by 14 · cited 3× · see citing works
“DRRs from a single patient, used as a replacement for 2D fluoroscopy. guided surgery, while a 2-mm error is considered to match the clinical requirement of core decompression (Wang et al. 2017).”
BackgroundMethodMRI Evaluation of Post Core Decompression Changes in Avascular Necrosis of HiP
Nori, M. (2015) · JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH · doi:10.7860/jcdr/2015/13995.6967 · Cited by 9 · cited 2× · see citing works
“Core decompression has become a preferred procedure to treat early-stage AVN due to its simplicity, low complication rate, and promising efficacy (Lieberman 2004; Veillette et al. 2006; Nori 2015; Talma¸c 2018).”
BackgroundThe Results Of Core Decompression Treatment In Femoral Head Avascular Necrosis
Talmaç, M. A. (2018) · SiSli Etfal Hastanesi Tip Bulteni / The Medical Bulletin of Sisli Hospital · doi:10.14744/semb.2018.47135 · Cited by 9 · see citing works
Geisert, W. C., Perdue, A. M., & Ozer, K. (2017) · Arthroscopy Techniques · doi:10.1016/j.eats.2017.01.004 · Cited by 5 · cited 2× · see citing works
“…X-ray have become important for the localisation of the greater trochanter, tracking and navigation of the drilling and injection device to the femoral head, and positioning of the trephine and scraping spoon while removing the dead bones (Nolte et al. 2000; Geisert et al. 2017; Tian et al. 2021).”
BackgroundTian, R., Luo, S., Wang, C., Wang, K., & Yang, P. (2021) · doi:10.21203/rs.3.rs-1053942/v1 · Cited by 1 · cited 2× · see citing works
“…X-ray have become important for the localisation of the greater trochanter, tracking and navigation of the drilling and injection device to the femoral head, and positioning of the trephine and scraping spoon while removing the dead bones (Nolte et al. 2000; Geisert et al. 2017; Tian et al. 2021).”
Background
Flagged first, then by citations
3 of the 43 references couldn’t be matched to a known paper.
Research works printed but not linked: 1
- Goodfellow I, 2014, Advances in Neural Information Processing Systems, 27, 2672
Cited by
Sorted by citations
Burton, W. S., Myers, C. A., Stefanovic, M., Shelburne, K. B., & Rullkoetter, P. J. (2024) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2024.112066 · Cited by 9 · see citing works
Huang, Y., Zhang, X., Hu, Y., Johnston, A., Jones, C. K., Zbijewski, W. B., Siewerdsen, J. H., Helm, P. A., Witham, T. F., & Uneri, A. (2024) · Computerized Medical Imaging and Graphics · doi:10.1016/j.compmedimag.2024.102365 · Cited by 8 · see citing works
Nagarathna, P., Jeelani, A., Fiza, S., et al. (2024) · Automatika · doi:10.1080/00051144.2024.2353543 · Cited by 5 · see citing works
Burton, W. S., Myers, C. A., Stefanovic, M., Shelburne, K. B., & Rullkoetter, P. J. (2024) · Annals of Biomedical Engineering · doi:10.1007/s10439-024-03473-5 · Cited by 3 · see citing works
Chiang, L.-H., Chen, Y.-C., Huang, G., et al. (2023) · Quantitative Imaging in Medicine and Surgery · doi:10.21037/qims-23-490 · Cited by 1 · see citing works
“Multimodality image fusion guidance could be the next new alternative to conventional angiogram road-mapping in endovascular procedures, which can effectively facilitate the vascular embolization and reduce radiation dose (13,14).”
BackgroundResultReliable 2D–3D Registration in Dynamic Stereo-Radiography With Energy Barrier Constraints
Burton, W. S., Myers, C. A., Clary, C. W., & Rullkoetter, P. J. (2024) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2024.3522200
“Previous investigations performed template matching [22], [23], [24], pose regression [25], or numerical optimization [26] to acquire preliminary pose estimates in radiographs.”
Ku, P., Liu, M., Grupp, R. B., et al. (2025) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-025-03426-w
“Recently, the development of learning-based generation models has been employed to transfer image intensities from MRI to CT, where encoder–decoder networks and generative adversarial networks (GANs) are often used [13, 14].”
Self-citationSu, L., Guan, D., Xie, P., et al. (2026) · Current Problems in Surgery · doi:10.1016/j.cpsurg.2026.102062
- Authors
- Ku, P.-C., Martin-Gomez, A., Gao, C., Grupp, R., Mears, S. C., & Armand, M.
- Affiliations
- Johns Hopkins University (US)
- Biomechanical- and Image-Guided Surgical Systems (BIGSS), Laboratory for Computational Sensing and Robotics, Johns Hopkins University, Baltimore, MD, USA
- University of Arkansas for Medical Sciences (US)
- Corresponding author
- Ku, Ping-Cheng
- Venue
- Computer Methods in Biomechanics and Biomedical Engineering: Imaging & Visualization
- Publisher
- Informa UK Limited
- Published
- 2022-12-15
- Volume
- 11
- Issue
- 4
- Pages
- 1096-1105
- ISSN
- 2168-1163 (ISSN-L)
- 2168-1171
- PMC
- PMC10406464
- OpenAlex
- W4311755298
- Semantic Scholar
- bb129dc415fbdde9f0413e393182d97deb026ac5
- Cited by
- 8
- References
- 43
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 5 · 2023: 1
- FWCI
- 0.57
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Computer Methods in Biomechanics and Biomedical Engineering Imaging & Visualization (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (submittedVersion)
- Funders
- National Institutes of Health
- Topics
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Medical Image Segmentation Techniques
- Physical Sciences › Engineering › Biomedical Engineering › Anatomy and Medical Technology
- Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
- Keywords
- 2D-3D registration, preoperative MRI, intraoperative fluoroscopy, bone defects, core decompression, femur, digitally reconstructed radiographs, intraoperative visualization, surgical planning, bone quality assessment
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-14.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-14.