A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter
Abstract
The performance of pulse oximeters is highly influenced by motion artifacts (MAs) in photoplethysmographic (PPG) signals. In this paper, we propose a simple and efficient approach based on adaptive step-size least mean squares (AS-LMS) adaptive filter for reducing MA in corrupted PPG signals. The presented method is an extension to our prior work on efficient use of adaptive filters for reduction of MA in PPG signals. The novelty of the method lies in the fact that a synthetic noise reference signal for an adaptive filtering process, representing MA noise, is generated internally from the MA-corrupted PPG signal itself instead of using any additional hardware such as accelerometer or source-detector pair for acquiring noise reference signal. Thus, the generated noise reference signal is then applied to the AS-LMS adaptive filter for artifact removal. While experimental results proved the efficacy of the proposed scheme, the merit of the method is clearly demonstrated using convergence and correlation analysis, thus making it best suitable for present-day pulse oximeters utilizing PPG sensor head with a single pair of source and detector, which does not have any extra hardware meant for capturing noise reference signal. In addition to arterial oxygen saturation estimation, the artifact reduction method facilitated the waveform contour analysis on artifact-reduced PPG, and the conventional parameters were evaluated for assessing the arterial stiffness.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1997) · Institute of Physics eBooks · doi:10.1887/0750304677 · Cited by 453
Motion Artifact Reduction in Photoplethysmography Using Independent Component Analysis
Kim, B. S., & Yoo, S. K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.869784 · Cited by 399
(1997) · Series in Medical Physics and Biomedical Engineering · doi:10.1201/9781420050790 · Cited by 351
Masimo Signal Extraction Pulse Oximetry
Goldman, J. M., Petterson, M. T., Kopotic, R. J., & Barker, S. J. (2000) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1011493521730 · Cited by 267
Two-Stage Approach for Detection and Reduction of Motion Artifacts in Photoplethysmographic Data
Krishnan, R., Natarajan, B., & Warren, S. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2039568 · Cited by 263
Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography
Mendelson, Y., & Ochs, B. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.7286 · Cited by 251
Artifact-resistant power-efficient design of finger-ring plethysmographic sensors
Rhee, S., Yang, B.-H., & Asada, H. H. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.930904 · Cited by 235
Signal processing methods for pulse oximetry
Rusch, T. L., Sankar, R., & Scharf, J. E. (1996) · Computers in Biology and Medicine · doi:10.1016/0010-4825(95)00049-6 · Cited by 185
Reddy, K. A., George, B., & Kumar, V. J. (2009) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2008.2009136 · Cited by 165
Self-citationReduction of motion artifact in pulse oximetry by smoothed pseudo Wigner-Ville distribution
Yan, Y. S., Poon, C. C. Y., & Zhang, Y. (2005) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-2-3 · Cited by 159
Biomedical Signal Processing and Signal Modeling
Bruce, E. N. (2000) · SDÜ Bilgi Merkezi · Cited by 140
A new method for pulse oximetry possessing inherent insensitivity to artifact
Hayes, M. J., & Smith, P. R. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.915711 · Cited by 122
Lee, C. M., & Zhang, Y. (2004) · IEEE EMBS Asian-Pacific Conference on Biomedical Engineering · doi:10.1109/apbme.2003.1302650 · Cited by 121
Active noise cancellation using MEMS accelerometers for motion-tolerant wearable bio-sensors
Asada, H. H., Jiang, H.-H., & Gibbs, P. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2004.1403631 · Cited by 114
Chan, K. W., & Zhang, Y. (2003) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2002.1037314 · Cited by 103
Design of filter to reject motion artifact of pulse oximetry
Lee, J., Jung, W.-G., Kang, I., Kim, Y., & Lee, G.-K. (2003) · Computer Standards & Interfaces · doi:10.1016/s0920-5489(03)00077-1 · Cited by 91
A Novel Calibration-Free Method of Measurement of Oxygen Saturation in Arterial Blood
Reddy, K. A., George, B., Mohan, N. M., & Kumar, V. J. (2009) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2009.2012934 · Cited by 84
Self-citationYao, J., & Warren, S. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2005.1617256 · Cited by 83
Motion Artifact Reduction in Photoplethysmographic Signals using Singular Value Decomposition
Reddy, K. A., & Kumar, V. J. (2007) · Conference proceedings - IEEE Instrumentation/Measurement Technology Conference · doi:10.1109/imtc.2007.379467 · Cited by 83
Self-citationArtifact reduction in photoplethysmography
Hayes, M. J., & Smith, P. R. (1998) · Applied Optics · doi:10.1364/ao.37.007437 · Cited by 72
Foo, J. Y. A. (2006) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2006.01.001 · Cited by 71
Gibbs, P., & Asada, H. H. (2005) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2005.1470193 · Cited by 61
Characterization and adaptive filtering of motion artifacts in pulse oximetry using accelerometers
Relente, A. R., & Sison, L. G. (2002) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2002.1106644 · Cited by 60
Noise-resistant pulse oximetry using a synthetic reference signal
Coetzee, F. M., & Elghazzawi, Z. F. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.855928 · Cited by 53
Wu, H., Liu, C.-C., Lin, P.-H., Chung, H.-M., Liu, M.-C., Yip, H., Liu, A., & Sun, C. (2010) · Atherosclerosis · doi:10.1016/j.atherosclerosis.2010.08.075 · Cited by 40
Adaptive reduction of motion artifacts from PPG signals using a synthetic noise reference signal
Ram, M. R., Madhav, K. V., Krishna, E. H., Reddy, K. N., & Reddy, K. A. (2010) · IEEE-EMBS Conference on Biomedical Engineering and Sciences · doi:10.1109/iecbes.2010.5742252 · Cited by 22
Self-citationBarreto, A. B., Vicente, L., & Persad, I. K. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1995.579392 · Cited by 18
Ram, M. R., Madhav, K. V., Krishna, E. H., Reddy, K. N., & Reddy, K. A. (2011) · International Conference on Communications and Signal Processing (ICCSP) · doi:10.1109/iccsp.2011.5739353 · Cited by 13
Self-citation
Cited by
Ghamari, M. (2018) · International Journal of Biosensors & Bioelectronics · doi:10.15406/ijbsbe.2018.04.00125 · Cited by 945
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
“(or even lower) Hz) prior to beat detection and make reference to several PPG beat detection and artifact handling methods [51]–[56], [140], [141].”
BackgroundWearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912
“A simple, efficient approach, based on the AS-LMS adaptive filter, was applied to reduce motion artifacts in corrupted PPG signals [59].”
MethodDeep PPG: Large-Scale Heart Rate Estimation with Convolutional Neural Networks
Reiss, A., Indlekofer, I., Schmidt, P., & Van Laerhoven, K. (2019) · Sensors · doi:10.3390/s19143079 · Cited by 377
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348
Metadata
- Authors
- Ram, M. R., Madhav, K. V., Krishna, E. H., Komalla, N. R., & Reddy, K. A.
- Venue
- IEEE Transactions on Instrumentation and Measurement
- Published
- 2012-05
- Volume
- 61
- Issue
- 5
- Pages
- 1445-1457
- ISSN
- 0018-9456 (ISSN-L)
- 1557-9662
- OpenAlex
- W2103675325
- Cited by
- 345
- References
- 28
- FWCI
- 3.65
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.