A review on wearable photoplethysmography sensors and their potential future applications in health care
Open accessAbstract
Photoplethysmography (PPG) is an uncomplicated and inexpensive optical measurement method that is often used for heart rate monitoring purposes. PPG is a non-invasive technology that uses a light source and a photodetector at the surface of skin to measure the volumetric variations of blood circulation. Recently, there has been much interest from numerous researchers around the globe to extract further valuable information from the PPG signal in addition to heart rate estimation and pulse oxymetry readings. PPG signal's second derivative wave contains important health-related information. Thus, analysis of this waveform can help researchers and clinicians to evaluate various cardiovascular-related diseases such as atherosclerosis and arterial stiffness. Moreover, investigating the second derivative wave of PPG signal can also assist in early detection and diagnosis of various cardiovascular illnesses that may possibly appear later in life. For early recognition and analysis of such illnesses, continuous and real-time monitoring is an important approach that has been enabled by the latest technological advances in sensor technology and wireless communications. The aim of this article is to briefly consider some of the current developments and challenges of wearable PPG-based monitoring technologies and then to discuss some of the potential applications of this technology in clinical settings.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Many sensors lack algorithms to effectively separate weak physiological signals from background noise in darker skin tones, further lowering accuracy.
“Many sensors lack algorithms to effectively separate weak physiological signals from background noise in darker skin tones, further lowering accuracy [20].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [20]
The Introduction (page unavailable) says environmental noise can affect PPG signal acquisition and heart-rate estimation accuracy. The Discussion (page unavailable) describes signal-processing approaches to reducing motion artifacts. Neither passage supports the claim that many sensors lack effective algorithms for darker skin tones or that this specifically lowers accuracy. (LLM, conf=0.83)
Experimental — accuracy not yet measuredPPG detects heart rate by measuring light absorption and reflection occurring within the skin's microvascular bed during blood pulses.
“Photoplethysmography (PPG) detects heart rate by measuring light absorption and reflection occurring within the skin's microvascular bed during blood pulses according to [20].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [20]
Weaker signals and increased scattering from darker skin degrade measurement accuracy.
“This results in weaker signals and increased scattering, which degrade measurement accuracy [19,20].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [20]
The Introduction’s “Factors affecting PPG sensor recordings” section says that anatomy and tissue differences can change how light propagates and alter the signal. But our copy does not link darker skin to weaker signals, increased scattering, or degraded measurement accuracy. Searches for skin pigmentation, melanin, scattering, and measurement accuracy did not find that link. (LLM, conf=0.76)
Experimental — accuracy not yet measuredSkin pigmentation affects light absorption and reflection such that darker skin tones reduce sensor signal detection capabilities and increase misread errors.
“Skin pigmentation affects light absorption and reflection rates such that darker skin tones reduce sensor signal detection capabilities while increasing misread errors.”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [20]
The Introduction’s “PPG sensors” section describes how sensors detect reflected light, and “Factors affecting PPG sensor recordings” discusses other sources of signal variation. Neither section addresses skin pigmentation, darker skin tones, or increased misread errors; searches for those terms found no supporting passage in our copy. No page numbers were available. (LLM, conf=0.72)
Experimental — accuracy not yet measuredSocioeconomic disparities in access to smartwatches may interact with skin-tone-related bias, raising additional equity concerns.
“However, regional variations may exist, and socioeconomic disparities in access to these devices may further interact with skin-tone-related bias, raising additional equity concerns [17-23].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [20]
The Introduction discusses consumer interest in smartwatches, and the 'Factors affecting PPG sensor recordings' section discusses anatomical and movement-related effects; neither addresses socioeconomic access or skin-tone-related bias. Searches for disparities in smartwatch access and for skin pigmentation or skin-tone bias found no passage linking those concerns. Page numbers were not available in our copy. (LLM, conf=0.75)
Experimental — accuracy not yet measured
Differences in the interaction of light with human tissues at differing wavelengths is exploited to provide information including oxygen saturation, blood pressure, and other cardiovascular-related diseases.
“Differences in the interaction of light with human tissues at differing wavelengths is exploited to provide information including oxygen saturation, blood pressure, and more recently other cardiovascular-related diseases.131”
From Effect of skin color on optical properties and the implications for medical optical technologies: a review (2024) · cites this paper as [131]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
The Internet of Things for Health Care: A Comprehensive Survey
Islam, S. M. R., Kwak, D., Kabir, M. H., Hossain, M., & Kwak, K.-S. (2015) · IEEE Access · doi:10.1109/access.2015.2437951 · Cited by 3,199 · see citing works
A Survey on Wearable Sensor-Based Systems for Health Monitoring and Prognosis
Pantelopoulos, A., & Bourbakis, N. G. (2010) · IEEE Transactions on Systems Man and Cybernetics Part C (Applications and Reviews) · doi:10.1109/tsmcc.2009.2032660 · Cited by 2,296 · see citing works
Systolic Time Intervals in Heart Failure in Man
WEISSLER, A. M., HARRIS, W. S., & SCHOENFELD, C. D. (1968) · Circulation · doi:10.1161/01.cir.37.2.149 · Cited by 1,217 · see citing works
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206 · see citing works
Smart wearable systems: Current status and future challenges
Chan, M., Estève, D., Fourniols, J.-Y., Escriba, C., & Campo, E. (2012) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2012.09.003 · Cited by 953 · see citing works
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · see citing works
Zhang, Z., Pi, Z., & Liu, B. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2359372 · Cited by 759 · see citing works
Determination of age-related increases in large artery stiffness by digital pulse contour analysis
Millasseau, S., Kelly, R. P., Ritter, J., & Chowienczyk, P. (2002) · Clinical Science · doi:10.1042/cs1030371 · Cited by 624 · see citing works
Enhancing the quality of life through wearable technology
Park, S., & Jayaraman, S. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213625 · Cited by 615 · see citing works
Takazawa, K., Tanaka, N., Fujita, M., Matsuoka, O., Saiki, T., Aikawa, M., Tamura, S., & Ibukiyama, C. (1998) · Hypertension · doi:10.1161/01.hyp.32.2.365 · Cited by 589 · see citing works
Introduction to Biomedical Engineering
Goyal, A. S. (2005) · Elsevier eBooks · doi:10.1016/b978-0-12-238662-6.x5000-9 · Cited by 492 · see citing works
Multiparameter Respiratory Rate Estimation From the Photoplethysmogram
Karlen, W., Raman, S., Ansermino, J. M., & Dumont, G. A. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2246160 · Cited by 486 · see citing works
Mobile monitoring with wearable photoplethysmographic biosensors
Asada, H. H., Shaltis, P., Reisner, A., Rhee, S., & Hutchinson, R. C. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213624 · Cited by 466 · see citing works
A wireless wearable ECG sensor for long-term applications
Nemati, E., Deen, M., & Mondal, T. (2012) · IEEE Communications Magazine · doi:10.1109/mcom.2012.6122530 · Cited by 418 · see citing works
Motion Artifact Reduction in Photoplethysmography Using Independent Component Analysis
Kim, B., & Yoo, S. K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.869784 · Cited by 400 · see citing works
Zhang, Z. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2406332 · Cited by 388 · see citing works
A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter
Ram, M. R., Madhav, K. V., Krishna, E. H., Komalla, N. R., & Reddy, K. A. (2011) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2011.2175832 · Cited by 345 · see citing works
Lü, G., Yang, F., Taylor, J. A., & Stein, J. F. (2009) · Journal of Medical Engineering & Technology · doi:10.3109/03091900903150998 · Cited by 324 · see citing works
Stauss, H. M. (2003) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00452.2003 · Cited by 318 · see citing works
Accurate electrocardiographic assessment of the QT interval: Teach the tangent
Postema, P. G., de Jong, J. S. S. G., van der Bilt, I. A. C., & Wilde, A. A. M. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.03.037 · Cited by 317 · see citing works
A Survey on Wireless Body Area Networks for eHealthcare Systems in Residential Environments
Ghamari, M., Janko, B., Sherratt, R. S., Harwin, W., Piechocki, R. J., & Soltanpur, C. (2016) · Sensors · doi:10.3390/s16060831 · Cited by 300 · see citing works
Self-citation27th Annual International Conference of the IEEE Engineering in Medicine and Biology Society
(2005) · IEEE Transactions on NanoBioscience · doi:10.1109/tnb.2005.845677 · Cited by 288 · see citing works
Lu, S. F., Zhao, H., Ju, K., Shin, K., Lee, M., Shelley, K. H., & Chon, K. H. (2007) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-007-9103-y · Cited by 284 · see citing works
Bortolotto, L. A., Blacher, J., Kondo, T., Takazawa, K., & Safar, M. E. (2000) · American Journal of Hypertension · doi:10.1016/s0895-7061(99)00192-2 · Cited by 272 · see citing works
An ultra-wearable, wireless, low power ECG monitoring system
Park, C., Chou, P. H., Bai, Y., Matthews, R., & Hibbs, A. H. (2006) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2006.4600353 · Cited by 265 · see citing works
Nitzan, M., Babchenko, A., Khanokh, B., & Landau, D. (1998) · Physiological Measurement · doi:10.1088/0967-3334/19/1/008 · Cited by 217 · see citing works
Introduction to Biomedical Engineering
Meldrum, S. J. (2000) · Physiological Measurement · doi:10.1088/0967-3334/21/2/701 · Cited by 216 · see citing works
Estimation of Respiratory Rate From Photoplethysmogram Data Using Time–Frequency Spectral Estimation
Chon, K. H., Dash, S., & Ju, K. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2019766 · Cited by 213 · see citing works
A brief review: history to understand fundamentals of electrocardiography
AlGhatrif, M., & Lindsay, J. M. (2012) · Journal of Community Hospital Internal Medicine Perspectives · doi:10.3402/jchimp.v2i1.14383 · Cited by 211 · see citing works
Motion-tolerant magnetic earring sensor and wireless earpiece for wearable photoplethysmography
Poh, M., Swenson, N. C., & Picard, R. W. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2010.2042607 · Cited by 189 · see citing works
Bolaños, M. J., Nazeran, H., & Haltiwanger, E. P. (2006) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2006.260607 · Cited by 176 · see citing works
Wearable sensors : fundamentals, implementation and applications
Sazonov, E. E., & Neuman, M. R. (2014) · doi:10.1016/c2018-0-01225-6 · Cited by 169 · see citing works
A novel feature ranking algorithm for biometric recognition with PPG signals
Kavsaoğlu, A. R., Polat, K., & Bozkurt, M. R. (2014) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2014.03.005 · Cited by 160 · see citing works
Lee, B., Han, J., Baek, H. J., Shin, J. H., Park, K. S., & Yi, W. J. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/12/003 · Cited by 155 · see citing works
Wei, J. (2014) · IEEE Consumer Electronics Magazine · doi:10.1109/mce.2014.2317895 · Cited by 154 · see citing works
Rompelman, O., Coenen, A. J. R. M., & Kitney, R. I. (1977) · Medical & Biological Engineering & Computing · doi:10.1007/bf02441043 · Cited by 153 · see citing works
Thomas, S. S., Nathan, V., Zong, C., Soundarapandian, K., Shi, X., & Jafari, R. (2015) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2015.2458779 · Cited by 140 · see citing works
The relationship between the photoplethysmographic waveform and systemic vascular resistance
Alian, A. A., Haddadin, A. S., Tantawy, H., Badr, T., Stout, R. G., Silverman, D. G., & Shelley, K. H. (2007) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-007-9097-5 · Cited by 136 · see citing works
Estimating heart rate using wrist-type Photoplethysmography and acceleration sensor while running
Fukushima, H., Kawanaka, H., Bhuiyan, M. S., & OGURI, K. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346570 · Cited by 129 · see citing works
The analysis of PPG morphology: Investigating the effects of aging on arterial compliance
Yousef, Q., Reaz, M. B. I., & Ali, M. A. M. (2012) · Measurement Science Review · doi:10.2478/v10048-012-0036-3 · Cited by 126 · see citing works
Hashimoto, J., Chonan, K., Aoki, Y., Nishimura, T., Ohkubo, T., Hozawa, A., Suzuki, M., Matsubara, M., Michimata, M., Araki, T., et al. (2002) · Journal of Hypertension · doi:10.1097/00004872-200212000-00021 · Cited by 122 · see citing works
Seitsonen, E. R. J., Korhonen, I. K. J., Van Gils, M. J., Huiku, M., Lötjönen, J. M. P., Korttila, K. T., & Yli‐Hankala, A. M. (2005) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2005.00654.x · Cited by 120 · see citing works
Imanaga, I., Hara, H., KOYANAGI, S., & TANAKA, K. (1998) · Japanese Heart Journal · doi:10.1536/ihj.39.775 · Cited by 118 · see citing works
Noninvasive cardiac output estimation using a novel photoplethysmogram index
Wang, L., Pickwell‐MacPherson, E., Liang, Y. P., & Zhang, Y. (2009) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2009.5333091 · Cited by 117 · see citing works
Data Fusion for Improved Respiration Rate Estimation
Nemati, S., Malhotra, A., & Clifford, G. D. (2010) · EURASIP Journal on Advances in Signal Processing · doi:10.1155/2010/926305 · Cited by 115 · see citing works
Lee, C. M., & Zhang, Y. (2004) · doi:10.1109/apbme.2003.1302650 · Cited by 114 · see citing works
Kavsaoğlu, A. R., Polat, K., & Hariharan, M. (2015) · Applied Soft Computing · doi:10.1016/j.asoc.2015.04.008 · Cited by 114 · see citing works
Active noise cancellation using MEMS accelerometers for motion-tolerant wearable bio-sensors
Asada, H. H., Jiang, H.-H., & Gibbs, P. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2004.1403631 · Cited by 113 · see citing works
Zhang, Y., Liu, B., & Zhang, Z. (2015) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.05.006 · Cited by 109 · see citing works
Accuracy of four automatic QT measurement techniques in cardiac patients and healthy subjects.
McLaughlin, N. B., Campbell, R. W., & Murray, A. (1996) · Heart · doi:10.1136/hrt.76.5.422 · Cited by 109 · see citing works
Allen, J., Overbeck, K., Nath, A. F., Murray, A., & Stansby, G. P. (2008) · Journal of Vascular Surgery · doi:10.1016/j.jvs.2007.11.057 · Cited by 105 · see citing works
Chan, K. W., & Zhang, Y. (2003) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2002.1037314 · Cited by 104 · see citing works
Otsuka, T., Kawada, T., Katsumata, M., & Ibuki, C. (2006) · Circulation Journal · doi:10.1253/circj.70.304 · Cited by 103 · see citing works
Spīgulis, J., Gailite, L., Lihachev, A., & Erts, R. (2007) · Applied Optics · doi:10.1364/ao.46.001754 · Cited by 98 · see citing works
Allen, J., & Murray, A. (2000) · Physiological Measurement · doi:10.1088/0967-3334/21/3/303 · Cited by 93 · see citing works
Shelley, K. H., Tamai, D., Jablonka, D., Gesquiere, M., Stout, R. G., & Silverman, D. G. (2005) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000145063.01043.4b · Cited by 91 · see citing works
An Algorithm for Real-Time Pulse Waveform Segmentation and Artifact Detection in Photoplethysmograms
Fischer, C., Dömer, B., Wibmer, T., & Penzel, T. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2518202 · Cited by 91 · see citing works
Nilsson, L. M., Johansson, A., & Kálmán, S. (2005) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2005.00721.x · Cited by 78 · see citing works
Human photoplethysmogram: new insight into chaotic characteristics
Sviridova, N., & Sakai, K. (2015) · Chaos Solitons & Fractals · doi:10.1016/j.chaos.2015.05.005 · Cited by 75 · see citing works
Mendelson, Y., & Pujary, C. J. (2004) · doi:10.1109/iembs.2003.1280775 · Cited by 74 · see citing works
Detection of c, d, and e waves in the acceleration photoplethysmogram
Elgendi, M. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2014.08.001 · Cited by 73 · see citing works
Estimation of respiration rate from ECG, BP and PPG signals using empirical mode decomposition
Madhav, K. V., Ram, M. R., Krishna, E. H., Komalla, N. R., & Reddy, K. A. (2011) · IEEE International Instrumentation and Measurement Technology Conference · doi:10.1109/imtc.2011.5944249 · Cited by 70 · see citing works
Adaptive Noise Cancellation Using Accelerometers for the PPG Signal from Forehead
Kim, S. H., Ryoo, D. W., & Bae, C. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4352852 · Cited by 70 · see citing works
Second Derivative of Photoplethysmography for Estimating Vascular Aging
Baek, H. J., Kim, J. S., Kim, Y. S., Lee, H. B., & Park, K. S. (2007) · IEEE International Conference on Information Technology and Applications in Biomedicine · doi:10.1109/itab.2007.4407346 · Cited by 69 · see citing works
Budidha, K., & Kyriacou, P. A. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/2/111 · Cited by 67 · see citing works
Continuous Blood Pressure Monitoring using ECG and Finger Photoplethysmogram
Chua, C. P., & Heneghan, C. (2006) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2006.259612 · Cited by 65 · see citing works
Sleep apnea: a review of diagnostic sensors, algorithms, and therapies
Shokoueinejad, M., Fernandez, C., Carroll, E., Wang, F., Levin, J., Rusk, S., Glattard, N., Mulchrone, A., Zhang, X., Xie, A., et al. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa6ec6 · Cited by 64 · see citing works
Zahedi, E., Chellappan, K., Ali, M. A. M., & Singh, H. (2007) · Cardiovascular Engineering An International Journal · doi:10.1007/s10558-007-9037-5 · Cited by 60 · see citing works
New aging index using signal features of both photoplethysmograms and acceleration plethysmograms
Ahn, J. M. (2017) · Healthcare Informatics Research · doi:10.4258/hir.2017.23.1.53 · Cited by 58 · see citing works
Effects of filtering on multi-site photoplethysmography pulse waveform characteristics
Allen, J., & Murray, A. (2005) · Computers in cardiology · doi:10.1109/cic.2004.1442980 · Cited by 56 · see citing works
Montebugnoli, L., Servidio, D., MIATON, R. A., et al. (2004) · The Journal of the American Dental Association · doi:10.14219/jada.archive.2004.0125 · Cited by 53 · see citing works
Chen, W., Lei, S., Guo, L., Yuquan, C., & Pan, M. (2011) · International Congress on Image and Signal Processing (CISP) · doi:10.1109/cisp.2011.6100581 · Cited by 53 · see citing works
Tautan, A.-M., Young, A., Wentink, E., & Wieringa, F. (2015) · 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7319795 · Cited by 47 · see citing works
Monitoring heart and respiratory rates at radial artery based on PPG
Wang, C., Li, Z., & Wei, X. (2013) · Optik · doi:10.1016/j.ijleo.2012.11.044 · Cited by 44 · see citing works
Development of a wristwatch-type PPG array sensor module
Lee, Y., Shin, H., Jo, J. H., & Lee, Y.-K. (2011) · IEEE International Conference on Consumer Electronics - Berlin (ICCE-Berlin) · doi:10.1109/icce-berlin.2011.6031811 · Cited by 41 · see citing works
Reducing motion artifacts in photoplethysmograms by using relative sensor motion: Phantom study
Wijshoff, R. W. C. G. R., Mischi, M., Veen, J., & Van Der Lee, A. M. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.11.117007 · Cited by 40 · see citing works
Ghamari, M., Arora, H., Sherratt, R. S., et al. (2015) · doi:10.1109/i4ct.2015.7219525 · Cited by 38 · see citing works
Self-citationLai, P., & Kim, I.-S. (2015) · Healthcare Technology Letters · doi:10.1049/htl.2014.0097 · Cited by 37 · see citing works
Ushiroyama, T., Kajimoto, Y., Sakuma, K., & Ueki, M. (2005) · Institutional Repositories DataBase (IRDB) · doi:10.57371/00000405 · Cited by 37 · see citing works
Toe Pulse Wave Analysis in Obliterating Atherosclerosis
Oliva, I. O., & Roztočil, K. (1983) · Angiology · doi:10.1177/000331978303400906 · Cited by 35 · see citing works
A New Probe for Ankle Systolic Pressure Measurement Using Photoplethysmography (PPG)
J�nsson, B., Laurent, C. M., Skau, T., & Lindberg, L. (2005) · Annals of Biomedical Engineering · doi:10.1007/s10439-005-8982-7 · Cited by 31 · see citing works
An overview of non-contact photoplethysmography
Demirezen, H., & Erdem, Ç. E. (2017) · Signal Processing and Communications Applications Conference (SIU) · doi:10.1109/siu.2017.7960176 · Cited by 31 · see citing works
Arterial Stiffness Alterations during Hemodialysis: The Role of Dialysate Calcium
Kyriazis, J., Katsipi, I., Stylianou, K., Jenakis, N., Karida, A., & Daphnis, E. K. (2007) · Nephron Clinical Practice · doi:10.1159/000101482 · Cited by 30 · see citing works
Wearable Photoplethysmographic Sensor based on Different LED Light Intensities
Lee, H., Ko, H., Jeong, C., & Lee, J. (2016) · IEEE Sensors Journal · doi:10.1109/jsen.2016.2633575 · Cited by 30 · see citing works
A computer based photoplethysmographic vascular analyzer through derivatives
Gonzalez, R., Manzo-Martínez, A., Delgado, J., et al. (2008) · Computers in cardiology · doi:10.1109/cic.2008.4749006 · Cited by 27 · see citing works
Ghamari, M., Soltanpur, C., Cabrera, S. D., Romero, R., Martínek, R., & Nazeran, H. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591842 · Cited by 27 · see citing works
Self-citationValidation of a wrist monitor for accurate estimation of RR intervals during sleep
Renevey, P., Solà, J., Theurillat, P., et al. (2013) · doi:10.1109/embc.2013.6610793 · Cited by 26 · see citing works
Heart Rate Variability and the Acceleration Plethysmogram Signals Measured at Rest
Elgendi, M., Jonkman, M., & Boer, F. D. (2011) · Communications in computer and information science · doi:10.1007/978-3-642-18472-7_21 · Cited by 25 · see citing works
Rozi, R. M., Usman, S., Ali, M. A. M., & Reaz, M. B. I. (2012) · IEEE-EMBS Conference on Biomedical Engineering and Sciences · doi:10.1109/iecbes.2012.6498064 · Cited by 25 · see citing works
Lu, G.-H. N., Yang, F., Taylor, J. A., et al. (2009) · Journal of Medical Engineering & Technology · doi:10.1080/03091900903150998 · Cited by 24 · see citing works
HEART RATE VARIABILITY MEASUREMENT USING THE SECOND DERIVATIVE PHOTOPLETHYSMOGRAM
Elgendi, M., Jonkman, M., & Boer, F. D. (2010) · International Conference on Bio-inspired Systems and Signal Processing · doi:10.5220/0002743300820087 · Cited by 22 · see citing works
Yao, J., & Warren, S. (2005) · doi:10.1109/iembs.2004.1403630 · Cited by 18 · see citing works
Applying the APG to measure Heart Rate Variability
Elgendi, M., Jonkman, M., & Boer, F. D. (2010) · doi:10.1109/iccae.2010.5452042 · Cited by 16 · see citing works
Bolaños, M. E., Nazeran, H., & Haltiwanger, E. P. (2006) · Conference proceedings · doi:10.1109/iembs.2006.4398399 · Cited by 15 · see citing works
Measurement of a-a Intervals at Rest in the Second Derivative Plethysmogram
Elgendi, M., Jonkman, M., & Boer, F. D. (2009) · Cited by 13 · see citing works
A Forecast of the Adoption of Wearable Technology
Page, T. (2018) · Wearable Technologies · doi:10.4018/978-1-5225-5484-4.ch063 · Cited by 6 · see citing works
Continuous Blood Pressure Monitoring using ECG and Finger Photoplethysmogram
Chua, C. P., & Heneghan, C. (2006) · Conference proceedings · doi:10.1109/iembs.2006.4398605 · Cited by 3 · see citing works
Detection of c, d, and e Waves in the Acceleration PhotoplethysmogramPreprint
Elgendi, M. (2013) · viXra · Cited by 1 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- [1]Rawassizadeh R, Price BA, Petre M. Wearables: has the age of smartwatches finally arrived? Commun ACM. 2015; 58(1): 45– 47.
- [45]Lemay M, Bertschi M, Sola J, Chapter 2.3. Application of Optical Heart Rate Monitoring. 2014.
- [81]Hyun JB, Jung SK, Yun SK, Second derivative of Photoplethysmography for estimating vascular aging. Proc IEEE/EMBS Reg 8 Int Conf Inf Technol Appl Biomed. 2008: 70– 72.
Cited by
Sorted by citations
Advances in healthcare wearable devices
Iqbal, S. M. A., Mahgoub, I., Du, E., Leavitt, M. A., & Asghar, W. (2021) · npj Flexible Electronics · doi:10.1038/s41528-021-00107-x · Cited by 714 · see citing works
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · see citing works
“Another suspected measurement error in wrist-worn devices is motion artifact, which is typically caused by displacement of the PPG sensor over the skin, changes in skin deformation, blood flow dynamics, and ambient temperature.(33,34) Motion artifacts may manifest as missing or false beats which result in incorrect HR…”
BackgroundFlexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics
Heng, W., Solomon, S., & Gao, W. (2022) · Advanced Materials · doi:10.1002/adma.202107902 · Cited by 609 · see citing works
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Machine Learning for Healthcare Wearable Devices: The Big Picture
Sabry, F., Eltaras, T., Labda, W., Alzoubi, K., & Malluhi, Q. (2022) · Journal of Healthcare Engineering · doi:10.1155/2022/4653923 · Cited by 321 · see citing works
Photoplethysmography based atrial fibrillation detection: a review
Pereira, T., Tran, N., Gadhoumi, K., Pelter, M. M., Do, D. H., Lee, R. J., Colorado, R., Meisel, K., & Hu, X. (2020) · npj Digital Medicine · doi:10.1038/s41746-019-0207-9 · Cited by 300 · see citing works
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250 · see citing works
Li, K., Cardoso, C., Moctezuma-Ramirez, A., Elgalad, A., & Perin, E. (2023) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph20247146 · Cited by 160 · see citing works
A Review of Wearable Multi-Wavelength Photoplethysmography
Ray, D., Collins, T., Woolley, S., & Ponnapalli, P. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3121476 · Cited by 155 · see citing works
Limiting racial disparities and bias for wearable devices in health science research
Colvonen, P. J., DeYoung, P. N., Bosompra, N.-O. A., & Owens, R. L. (2020) · Sleep · doi:10.1093/sleep/zsaa159 · Cited by 137 · see citing works
PPG2ABP: Translating photoplethysmogram (PPG) signals to arterial blood pressure (ABP) waveforms
Ibtehaz, N., Mahmud, S., Chowdhury, M. E. H., Khandakar, A., Salman Khan, M., Ayari, M. A., Tahir, A. M., & Rahman, M. S. (2022) · Bioengineering · doi:10.3390/bioengineering9110692 · Cited by 133 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
Wearables in cardiology: Here to stay
Dagher, L., Shi, H., Zhao, Y., & Marrouche, N. F. (2020) · Heart Rhythm · doi:10.1016/j.hrthm.2020.02.023 · Cited by 107 · see citing works
Heart rate tracking in photoplethysmography signals affected by motion artifacts: a review
Ismail, S., Akram, U., & Siddiqi, I. (2021) · EURASIP Journal on Advances in Signal Processing · doi:10.1186/s13634-020-00714-2 · Cited by 101 · see citing works
Taylor, K. I., Staunton, H., Lipsmeier, F., Nobbs, D., & Lindemann, M. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0305-8 · Cited by 85 · see citing works
The Current State of Optical Sensors in Medical Wearables
Vavrinsky, E., Esfahani, N. E., Hausner, M., Kuzma, A., Rezo, V., Donoval, M., & Kosnacova, H. (2022) · Biosensors · doi:10.3390/bios12040217 · Cited by 103 · see citing works
Home monitoring in interstitial lung diseases
Wijsenbeek, M. S., Moor, C. C., Johannson, K. A., Jackson, P. D., Khor, Y. H., Kondoh, Y., Rajan, S. K., Tabaj, G. C., Varela, B. E., van der Wal, P., et al. (2023) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(22)00228-4 · Cited by 98 · see citing works
Signal quality and patient experience with wearable devices for epilepsy management
Nasseri, M., Nurse, E., Glasstetter, M., Böttcher, S., Gregg, N. M., Laks Nandakumar, A., Joseph, B., Pal Attia, T., Viana, P. F., Bruno, E., et al. (2020) · Epilepsia · doi:10.1111/epi.16527 · Cited by 90 · see citing works
Papini, G. B., Fonseca, P., van Gilst, M. M., Bergmans, J. W. M., Vullings, R., & Overeem, S. (2020) · Scientific Reports · doi:10.1038/s41598-020-69935-7 · Cited by 90 · see citing works
Data quality evaluation in wearable monitoring
Böttcher, S., Vieluf, S., Bruno, E., Joseph, B., Epitashvili, N., Biondi, A., Zabler, N., Glasstetter, M., Dümpelmann, M., Van Laerhoven, K., et al. (2022) · Scientific Reports · doi:10.1038/s41598-022-25949-x · Cited by 91 · see citing works
Mobile health applications for the detection of atrial fibrillation: a systematic review
Lopez Perales, C. R., Van Spall, H. G. C., Maeda, S., Jimenez, A., Laţcu, D. G., Milman, A., Kirakoya-Samadoulougou, F., Mamas, M. A., Muser, D., & Casado Arroyo, R. (2021) · EP Europace · doi:10.1093/europace/euaa139 · Cited by 77 · see citing works
Setchfield, K., Gorman, A., Simpson, A. H. R. W., Somekh, M. G., & Wright, A. J. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.1.010901 · Cited by 72 · see citing works
Scardulla, F., D’Acquisto, L., Colombarini, R., Hu, S., Pasta, S., & Bellavia, D. (2020) · Sensors · doi:10.3390/s20185052 · Cited by 72 · see citing works
Response To: Investigating sources of inaccuracy in wearable optical heart rate sensors
Colvonen, P. J. (2021) · npj Digital Medicine · doi:10.1038/s41746-021-00408-5 · Cited by 65 · see citing works
Xu, K., Jiang, X., Lin, S., Dai, C., & Chen, W. (2020) · IEEE Transactions on Industrial Informatics · doi:10.1109/tii.2020.2988097 · Cited by 37 · see citing works
Siddi, S., Bailon, R., Giné-Vázquez, I., Matcham, F., Lamers, F., Kontaxis, S., Laporta, E., Garcia, E., Lombardini, F., Annas, P., et al. (2023) · Psychological Medicine · doi:10.1017/s0033291723001034 · Cited by 31 · see citing works
Duarah, R., Torné‐Morató, H., Zhang, G., Amin, Y., Pabast, M., Sharma, N., He, S., Das, M. R., & Pompa, P. P. (2026) · Advanced Healthcare Materials · doi:10.1002/adhm.202504419 · Cited by 24 · see citing works
Validity of heart rate measurements in wrist-based monitors across skin tones during exercise
Hung, S. H., Serwa, K., Rosenthal, G., & Eng, J. J. (2025) · PLOS ONE · doi:10.1371/journal.pone.0318724 · Cited by 21 · see citing works
When two hearts beat as one: Heart-rate synchrony in social anxiety disorder
Asher, M., Barthel, A. L., Hofmann, S. G., Okon-Singer, H., & Aderka, I. M. (2021) · Behaviour Research and Therapy · doi:10.1016/j.brat.2021.103859 · Cited by 17 · see citing works
El Atrache, R., Tamilia, E., Mohammadpour Touserkani, F., Hammond, S., Papadelis, C., Kapur, K., Jackson, M., Bucciarelli, B., Tsuboyama, M., Sarkis, R. A., et al. (2020) · Epilepsia · doi:10.1111/epi.16621 · Cited by 16 · see citing works
Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring
Asif, S., AlSaafeen, A., Nadar, S., Nambiar, S., Dannawi, J., Korrapati, H., & Wilkhoo, H. S. (2025) · Cureus · doi:10.7759/cureus.94074 · Cited by 9 · see citing works
- Authors
- Ghamari, M.
- Affiliations
- Pennsylvania State University (US)
- Corresponding author
- Ghamari, Mohammad
- ORCID
- Ghamari, Mohammad
- Venue
- International Journal of Biosensors & Bioelectronics
- Publisher
- MedCrave Group Kft.
- Published
- 2018
- Volume
- 4
- Issue
- 4
- Pages
- 195-202
- ISSN
- 2573-2838 (ISSN-L)
- PMC
- PMC6426305
- OpenAlex
- W2894956427
- Cited by
- 945
- References
- 101
- Citations by year
- 2026: 60 · 2025: 141 · 2024: 168 · 2023: 184 · 2022: 152 · 2021: 136 · 2020: 80 · 2019: 22 · 2018: 1
- FWCI
- 23.87
- Type
- journal article
- Language
- en
- Open access
- diamond
- doi.org
- Locations
- International Journal of Biosensors & Bioelectronics (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion, cc-by-nc)
- PubMed Central (acceptedVersion, cc-by-nc)
- Funders
- National Institute of General Medical Sciences (UL1GM118970, TL4GM118971, RL5GM118969)
- National Institute on Minority Health and Health Disparities (RL5MD009592)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- photoplethysmography, wearable sensors, heart rate monitoring, pulse oximetry, atherosclerosis, arterial stiffness, early detection of cardiovascular disease, wireless health monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review