The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Modeling and incorporation of system response functions in 3-D whole body PET
Abstract
Appropriate application of spatially variant system models can correct for degraded resolution response and mispositioning errors. This paper explores the detector blurring component of the system model for a whole body positron emission tomography (PET) system and extends this factor into a more general system response function to account for other system effects including the influence of Fourier rebinning (FORE). We model the system response function as a three-dimensional (3-D) function that blurs in the radial and axial dimension and is spatially variant in radial location. This function is derived from Monte Carlo simulations and incorporates inter-crystal scatter, crystal penetration, and the blurring due to the FORE algorithm. The improved system model is applied in a modified ordered subsets expectation maximization (OSEM) algorithm to reconstruct images from rebinned, fully 3-D PET data. The proposed method effectively removes the spatial variance in the resolution response, as shown in simulations of point sources. Furthermore, simulation and measured studies show the proposed method improves quantitative accuracy with a reduction in tumor bias compared to conventional OSEM on the order of 10%-30% depending on tumor size and smoothing parameter.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Fully 3-D PET reconstruction with system matrix derived from point source measurements
Panin, V. Y., Kehren, F., Michel, C., & Casey, M. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.876171 · Cited by 536
Cited by
Fundamental limits of spatial resolution in PET
Moses, W. W. (2011) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/j.nima.2010.11.092 · Cited by 565
Physical Performance of the new hybrid PET/CT Discovery-690
Bettinardi, V., Presotto, L., Rapisarda, E., Picchio, M., Gianolli, L., & Gilardi, M. C. (2011) · Medical Physics · doi:10.1118/1.3635220 · Cited by 380
Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PET
Rapisarda, E., Bettinardi, V, Thielemans, K., & Gilardi, M. C. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/14/012 · Cited by 142
A method for model-free partial volume correction in oncological PET
Hofheinz, F., Langner, J., Petr, J., Beuthien-Baumann, B., Oehme, L., Steinbach, J., Kotzerke, J., & van den Hoff, J. (2012) · EJNMMI Research · doi:10.1186/2191-219x-2-16 · Cited by 55
Risk Factors Associated with Canine Heartworm Infection
Selby, L. A., Corwin, R. M., & Hayes, H. M. (1980) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.1980.176.01.33 · Cited by 49
Metadata
- Authors
- Alessio, A. M., Kinahan, P. E., & Lewellen, T. K.
- Venue
- IEEE Transactions on Medical Imaging
- Published
- 2006-07
- Volume
- 25
- Issue
- 7
- Pages
- 828-837
- ISSN
- 0278-0062 (ISSN-L)
- OpenAlex
- W2116181343
- Cited by
- 176
- References
- 1
- FWCI
- 8.82
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Modeling and incorporation of system response functions in 3-D whole body PET. https://app.cochat.ai/papers/10.1109/tmi.2006.873222
- The paper
- Alessio, A. M., Kinahan, P. E., & Lewellen, T. K. (2006). Modeling and incorporation of system response functions in 3-D whole body PET. IEEE Transactions on Medical Imaging. https://doi.org/10.1109/tmi.2006.873222
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.