THz-TDS Time-Trace Analysis for the Extraction of Material and Metamaterial Parameters
Open accessAbstract
We report on a method to fit time-trace data from a terahertz time-domain-spectroscopy system enabling the extraction of physical parameters from a material or metamaterial. To accomplish this, we developed a Python-based open-source software called Fit@TDS that functions on a personal computer. This software includes commonly used methods where the refractive index is extracted from frequency-domain data. This method has limitations when the signal is too noisy or when an absorption peak saturates the spectrum. Thus, the software also includes a new method where the refractive indices are directly fitted from the time trace. The idea is to model a material or a metamaterial through parametric physical models (Drude-Lorentz model and time-domain coupled mode theory) and implement this in the propagation model to simulate the time trace. Then, an optimization algorithm is used to retrieve the parameters of the model corresponding to the studied material/metamaterial. In this paper, we explain the method and test it on fictitious samples to probe its feasibility and reliability. Finally, we used Fit@TDS on real samples of high-resistivity silicon, lactose, and gold metasurface on quartz to show the capacity of the method.
Overview
CoChat Summary
TL;DR
The paper introduces Fit@TDS, an open-source Python tool that fits THz-TDS time traces directly to parametric material and metamaterial models to extract physical parameters. It validates the method on simulated and real samples, including silicon, lactose, and a gold split-ring metasurface on quartz.
Key findings
- Direct time-domain fitting avoids frequency-domain phase-unwrapping problems and does not require precise sample thickness as an input; for a high-resistivity silicon wafer, the time-domain fit returned a thickness of 4.99908 mm versus a comparator measurement of 5016 ± 4 µm, and refractive-index precision better than 10⁻³.
- In simulated one-oscillator Drude-Lorentz tests, the fit converged to the added noise level; relative parameter error fell below 1% for a noise floor of −40 dB, corresponding to single-shot noise in the authors’ setup (~50 ms acquisition, 20 Hz video-frame rate).
- A six-oscillator fictitious sample was fitted recursively: L2 residual error decreased from 11% with two oscillators to 0.028% with six oscillators, then 0.0269% and 0.0268% after refinement.
- For a lactose monohydrate pellet, the method retrieved four oscillators at 0.53, 1.17, 1.37, and 1.81 THz, plus a fifth oscillator at 5.0764 THz attributed to scattering losses; for a gold SRR metasurface on quartz, the fit gave Q_TOT = 6.11, with external losses ~75% and absorption losses ~25%.
Methods
- Fit@TDS performs a direct time-domain fit by minimizing the L2 norm between the modeled sample pulse and the measured sample pulse; the objective can be computed in the frequency domain via Parseval’s theorem. It uses the PyOpt library with the augmented Lagrangian particle swarm optimizer (ALPSO).
- Models include a multiple Drude-Lorentz permittivity model for bulk solids and a time-domain coupled-mode theory (TDCMT) model for metasurfaces, both embedded in a propagation model with Fresnel coefficients and Fabry-Pérot terms.
- Validation used simulated fictitious samples generated from a measured reference pulse from a Menlo Systems TeraSmart THz-TDS spectrometer (1000 accumulations) with added Gaussian white noise, then real samples: high-resistivity float-zone silicon, lactose monohydrate powder pellet, and a gold split-ring metasurface on quartz.
Limitations
- The paper states that the model assumes a single layer with two interfaces; the metasurface model is suited for infinitely thin metasurfaces, and thick dielectric metasurfaces would require modification. It also assumes the metal does not influence transmission or reflection outside resonance, requiring a very low metal filling factor.
- The authors state that ALPSO is not optimized for computation time, that residual error reaches the THz-TDS setup noise limitation, and that further improvement would require more sensitive THz-TDS systems. They also note that measured sample thickness may not correspond to the average thickness.
- Evident from the text: complex samples can involve more than thirty parameters, requiring recursive fitting to keep the parameter space manageable; resolution is limited by the temporal window and discrete Fourier transform frequency resolution; and high-frequency lactose losses are only tentatively attributed to scattering.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures














Verification
Claims the paper makes
1 Valid · 5 Not valid · 12 Not checked · 100 Own results and interpretations
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“The peaks at 0.53 THz [47, 48], 1.17 THz [49, 50] and 1.37 THz [51, 52] are the characteristic absorption peaks of α-lactose monohydrate [10, 49, 53].”
What the sources say
- Not checked — no full text of the sourcecites [47] [47] M. Walther, M. R. Freeman, and F. A. Hegmann, 'Metal-wire terahertz time-domain spectroscopy,' Applied Physics Letters , vol. 87, no. 26, p. 261107, 2005. [Online]. Available: https://doi.org/10.1063/1.2158025 - source
- Not checked — no full text of the sourcecites [48] [48] D. Allis, A. Fedor, T. Korter, J. Bjarnason, and E. Brown, 'Assignment of the lowest-lying thz absorption signatures in biotin and lactose monohydrate by solid-state density functional theory,' Chemical Physics Letters , vol. 440, no. 4, pp. 203 - 209, 2007. [Online]. Available: http://www.sciencedirect.com/science/article/pii/S0009261407004794 - source
- Could not verifycites [49] [49] W. Withayachumnankul, B. M. Fischer, and D. Abbott, 'Material thickness optimization for transmission-mode terahertz time-domain 18. spectroscopy,' Opt. Express , vol. 16, no. 10, pp. 7382-7396, May 2008. [Online]. Available: http://www.opticsexpress.org/abstract.cfm?URI=oe-1610-7382 - sourceExperimental — accuracy not yet measured
Our copy does not discuss α-lactose monohydrate or identify its characteristic absorption peaks (searched for the compound and its absorption peaks). Figure 6 in §4.2.2 (p. 7) identifies a resonance at 1.17 THz as water vapour, but that does not establish whether lactose also absorbs at that frequency. (LLM, conf=0.75)
- Not checked — no full text of the sourcecites [50] [50] B. B. Jin, Z. X. Chen, Z. Li, J. L. Ma, R. Fu, C. H. Zhang, J. Chen, and P. H. Wu, 'Mode assignment of terahertz spectrum of alpha-lactose monohydrate,' in 2009 34th International Conference on Infrared, Millimeter, and Terahertz Waves , Sept 2009, pp. 1-2. - source
- Not checked — no full text of the sourcecites [51] [51] J. A. Zeitler, K. Kogermann, J. Rantanen, T. Rades, P. F. Taday, M. Pepper, J. Aaltonen, and C. J. Strachan, 'Drug hydrate systems and dehydration processes studied by terahertz pulsed spectroscopy,' International journal of pharmaceutics , vol. 334, no. 1, pp. 78-84, 2007. - source
- Not checked — no full text of the sourcecites [52] [52] Y. Kawada, T. Yasuda, A. Nakanishi, K. Akiyama, and H. Takahashi, 'Single-shot terahertz spectroscopy using pulse-front tilting of an ultra-short probe pulse,' Opt. Express , vol. 19, no. 12, pp. 11228-11235, Jun 2011. [Online]. Available: http://www.opticsexpress.org/abstract.cfm?URI=oe-1912-11228 - source
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“After step #4, a residual error remains in the region around 1 THz.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Here, one can see that the residual error decreases step by step with additional oscillators.”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Moreover, in FIG. 8 (B) it is clear that the addition of a new oscillator decreases the residual error in the frequency range where the oscillator is added.”
boostedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Secondly, the relative error follows a trend between proportional to the amplitude and proportional to the energy, as indicated by the solid gray lines in FIG. 5.”
hedgedstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
58 more claims
Create a free account to see all 118 claims and to verify the 1 reference still open.
- Probably not a substantial claim. 23
- It may state two separate claims as one. 39
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Other approaches have emerged that improve on the standard procedures [14-16]”
From Maximum-likelihood parameter estimation in terahertz time-domain spectroscopy (2021) · cites this paper as [16]
The cited source says what the claim attributes to it.
In the citing paper
“It provides a more precise measurement of the thickness of the material, a higher precision on the refractive index retrieved, a better interpretation of the results and experiments, and a reliable and consistent retrieval of material parameters using permittivity models.”
From A New Metric for the Comparison of Permittivity Models in Terahertz Time-Domain Spectroscopy (2024) · cites this paper as [29]
The cited source says what the claim attributes to it.
In the citing paper
“It provides a more precise measurement of the thickness of the material, a higher precision on the refractive index retrieved, a better interpretation of the results and experiments, and a reliable and consistent retrieval of material parameters using permittivity models.”
From A New Metric for the Comparison of Permittivity Models in Terahertz Time-Domain Spectroscopy (2024) · cites this paper as [29]
The cited source says what the claim attributes to it.
In the citing paper
“It provides a more precise measurement of the thickness of the material, a higher precision on the refractive index retrieved, a better interpretation of the results and experiments, and a reliable and consistent retrieval of material parameters using permittivity models.”
From A New Metric for the Comparison of Permittivity Models in Terahertz Time-Domain Spectroscopy (2024) · cites this paper as [29]
The cited source says what the claim attributes to it.
In the citing paper
“It provides a more precise measurement of the thickness of the material, a higher precision on the refractive index retrieved, a better interpretation of the results and experiments, and a reliable and consistent retrieval of material parameters using permittivity models.”
From A New Metric for the Comparison of Permittivity Models in Terahertz Time-Domain Spectroscopy (2024) · cites this paper as [29]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ C. Resolution test · p. 8
“The recursive approach we used for this fictitious sample allowed keeping a reasonable computational load while adding up to seven oscillators.”
§ C. Resolution test · p. 7
“| Number of oscillators | 2 | 3 | 4 | 5 | 6 | 6 | 6 |”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ C. Implementation · p. 5
“Since the problem is to find a global maximum rather than a local one, we implement the optimized particle swarm routine ALPSO amongst the proposed algorithms.”
§ II. OPTIMIZATION PROBLEM · p. 3
“Then, the objective function to minimize is set as the L 2 norm (square root of the sum of the square of the differences) of the difference between the modelled pulse and the measured (sample) one:”
Citations
Cites
Convergence Properties of the Nelder--Mead Simplex Method in Low Dimensions
Lagarias, J. C., Reeds, J. A., Wright, M. H., & Wright, P. E. (1998) · SIAM Journal on Optimization · doi:10.1137/s1052623496303470 · Cited by 7,447 · cited 1×
“For example, one can use the simplex method (gradient free method) [20] or a quasi-Newton algorithm [21], [22].”
MethodEmpirical study of particle swarm optimization
Shi, Y., & Eberhart, R. C. (2003) · Proceedings of the 1999 Congress on Evolutionary Computation-CEC99 (Cat. No. 99TH8406) · doi:10.1109/cec.1999.785511 · Cited by 3,945 · cited 1×
“The particle swarm algorithm is a versatile metaheuristic method that makes no or few assumptions about the problem being optimized, and can search very large spaces of candidate solutions [38], [39].”
MethodFar-infrared time-domain spectroscopy with terahertz beams of dielectrics and semiconductors
Grischkowsky, D., Keiding, S., van Exter, M., & Fattinger, C. (1990) · Journal of the Optical Society of America B · doi:10.1364/josab.7.002006 · Cited by 2,239
Picosecond photoconducting Hertzian dipoles
Auston, D. H., Cheung, K. P., & Smith, P. R. (1984) · Applied Physics Letters · doi:10.1063/1.95174 · Cited by 1,037
A reliable method for extraction of material parameters in terahertz time-domain spectroscopy
Duvillaret, L., Garet, F., & Coutaz, J.-L. (1996) · IEEE Journal of Selected Topics in Quantum Electronics · doi:10.1109/2944.571775 · Cited by 994 · cited 3×
“spectra is called the complex transmission coefficient and can be written as [16], [17]”
BackgroundMethodVariable Metric Method for Minimization
Davidon, W. C. (1991) · SIAM Journal on Optimization · doi:10.1137/0801001 · Cited by 892
Using selection to improve particle swarm optimization
Angeline, P. J. (2002) · 1998 IEEE International Conference on Evolutionary Computation Proceedings. IEEE World Congress on Computational Intelligence (Cat. No.98TH8360) · doi:10.1109/icec.1998.699327 · Cited by 864 · cited 1×
“The particle swarm algorithm is a versatile metaheuristic method that makes no or few assumptions about the problem being optimized, and can search very large spaces of candidate solutions [38], [39].”
MethodMaterial parameter estimation with terahertz time-domain spectroscopy
Dorney, T. D., Baraniuk, R. G., & Mittleman, D. M. (2001) · Journal of the Optical Society of America A · doi:10.1364/josaa.18.001562 · Cited by 807
Coupling of modes analysis of resonant channel add-drop filters
Manolatou, C., Khan, M. J., Fan, S., Villeneuve, P. R., Haus, H. A., & Joannopoulos, J. D. (1999) · IEEE Journal of Quantum Electronics · doi:10.1109/3.784592 · Cited by 654
Numerical Optimization, Theoretical and Practical Aspects—
(2006) · IEEE Transactions on Automatic Control · doi:10.1109/tac.2005.864181 · Cited by 565
Research works printed but not linked: 1
- jin, 0, Mode assignment of terahertz spectrum of alpha-lactose monohydrate, 34th Int Conf Infrared Millim and THz Waves, 1
Cited by
Tayvah, U., Spies, J. A., Neu, J., & Schmuttenmaer, C. A. (2021) · Analytical Chemistry · doi:10.1021/acs.analchem.1c02132 · Cited by 41
Zhang, X. Q., Wang, Y., Luo, F., Zhang, X., Chen, X., Ru, Y., Song, B., Cui, Z., & Zhang, K. (2024) · Carbon · doi:10.1016/j.carbon.2024.119273 · Cited by 42
Imaging of THz Photonic Modes by Scattering Scanning Near-Field Optical Microscopy
Thomas, L., Hannotte, T., Santos, C. N., Walter, B., Lavancier, M., Eliet, S., Faucher, M., Lampin, J., & Peretti, R. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c01871 · Cited by 30
Self-citationTrofimov, V. A., & Varentsova, S. A. (2019) · Sensors · doi:10.3390/s19102365 · Cited by 34
The dotTHz Project: A Standard Data Format for Terahertz Time-Domain Data
Lee, J., Leung, C. K., Ma, M., Ward-Berry, J., Santitewagun, S., & Zeitler, J. A. (2023) · Journal of Infrared, Millimeter, and Terahertz Waves · doi:10.1007/s10762-023-00947-w · Cited by 28
Metadata
- Authors
- Peretti, R., Mitryukovskiy, S., Froberger, K., Mebarki, M. A., Eliet, S., Vanwolleghem, M., & Lampin, J.-F.
- Affiliations
- Centre National de la Recherche Scientifique (FR)
- Venue
- IEEE Transactions on Terahertz Science and Technology
- Published
- 2019-03
- Volume
- 9
- Issue
- 2
- Pages
- 136-149
- ISSN
- 2156-342X (ISSN-L)
- 2156-3446
- OpenAlex
- W2906032000
- Cited by
- 78
- References
- 69
- Citations by year
- 2026: 5 · 2025: 9 · 2024: 12 · 2023: 10 · 2022: 14 · 2021: 12 · 2020: 9 · 2019: 7
- FWCI
- 2.68
- Type
- journal article
- Language
- en
- Open access
- green
- arxiv.org
- Licence
- Locations
- IEEE Transactions on Terahertz Science and Technology (publishedVersion)
- arXiv (Cornell University) (submittedVersion)
- LillOA (Université de Lille (University Of Lille)) (submittedVersion, other-oa)
- HAL (submittedVersion, cc-by-nc-sa)
- Figshare (submittedVersion, public-domain)
- arXiv (Cornell University)
- Funders
- International Chair of Excellence
- Welcome Talent Grant NeFiStoV from the European Metropole of Lille
- French Government through the National Research Agency (ANR 11-EQPX-0015)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Terahertz technology and applications
- Physical Sciences › Materials Science › Electronic, Optical and Magnetic Materials › Metamaterials and Metasurfaces Applications
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Microwave and Dielectric Measurement Techniques
- Keywords
- terahertz time-domain spectroscopy, parameter extraction, metamaterial characterization, refractive index retrieval, Drude-Lorentz model, parameter optimization, gold metasurfaces, high-resistivity silicon, lactose
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: THz-TDS Time-Trace Analysis for the Extraction of Material and Metamaterial Parameters. https://app.cochat.ai/papers/10.1109/tthz.2018.2889227
- The paper
- Peretti, R., Mitryukovskiy, S., Froberger, K., Mebarki, M. A., Eliet, S., Vanwolleghem, M., & Lampin, J.-F. (2019). THz-TDS Time-Trace Analysis for the Extraction of Material and Metamaterial Parameters. IEEE Transactions on Terahertz Science and Technology. https://doi.org/10.1109/tthz.2018.2889227
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.