Evaluation of an automated algorithm for interpretation of lupus anticoagulant testing
Abstract
INTRODUCTION: Lupus anticoagulant (LAC) testing is a multistep procedure including screening, mixing, and confirmation tests. STA Coag Expert is a software module for STA R Max and STA Compact Max analyzers which includes an on-demand LAC algorithm, based on ISTH guidelines, for automatic interpretation, calculation, and launch of assays in LAC interpretation ("Stago coag algorithm"). MATERIALS AND METHODS: One hundred ninety four patient samples were analyzed in parallel and interpreted manually and automatically by LAC algorithms. LAC algorithms use identical flowcharts and cutoff values as in daily practice. Differently, it only uses index of circulating anticoagulant (ICA), whereas in routine also normalized ratios were assessed for interpretation of mixing tests. Interpretation of dRVVT and aPTT pathways and final conclusions were compared between both approaches. RESULTS: Compared to routine interpretation, LAC algorithm showed a sensitivity of 94% and a specificity of 100% for LAC detection, when discrepancies due to measured clotting times between both analyzers were excluded. Three false negatives were due to different interpretation of dRVVT mixing test. Discrepancies in interpretation of the aPTT mixing test (n = 11) did not result in discrepant final LAC result, all having negative confirmation tests. No false positives were observed. With LAC algorithm, hands-on time reduced from 200 to 80 minutes. CONCLUSION: The LAC algorithm of the STA Coag Expert shows good comparability to the manual interpretation of LAC and may be used to assist laboratories in automatic launching of additional tests and in interpretation of LAC according to ISTH guidelines. This way the STA Coag Expert LAC algorithm may improve interlaboratory and STA comparability of LAC results.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Final conclusions about LA positivity differ depending on how the mixing test is interpreted.
“Final conclusions on LA positivity differ regarding the way the mixing test is interpreted.64”
Cites
MIYAKIS, S., LOCKSHIN, M. D., ATSUMI, T., BRANCH, D. W., BREY, R. L., CERVERA, R., DERKSEN, R. H. W. M., DE GROOT, P. G., KOIKE, T., MERONI, P. L., et al. (2006) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2006.01753.x · Cited by 7,203 · see citing works
Update of the guidelines for lupus anticoagulant detection
PENGO, V., TRIPODI, A., REBER, G., RAND, J. H., ORTEL, T. L., GALLI, M., & DE GROOT, P. G. (2009) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2009.03555.x · Cited by 1,197 · see citing works
Guidelines on the investigation and management of antiphospholipid syndrome
Keeling, D., Mackie, I., Moore, G. W., Greer, I. A., Greaves, M., & British Committee for Standards in Haematology (2012) · British Journal of Haematology · doi:10.1111/j.1365-2141.2012.09037.x · Cited by 740 · see citing works
Antiphospholipid antibody testing and standardization
Devreese, K. M. J. (2014) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.12234 · Cited by 98 · see citing works
Self-citationChallenges in the Diagnosis of the Antiphospholipid Syndrome
Devreese, K. M. J., & Hoylaerts, M. (2010) · Clinical Chemistry · doi:10.1373/clinchem.2009.133678 · Cited by 82 · see citing works
Self-citationStandardization of antiphospholipid antibody assays. Where do we stand?
Devreese, K. (2012) · Lupus · doi:10.1177/0961203312439335 · Cited by 76 · see citing works
Self-citationDetection of lupus anticoagulant in the era of direct oral anticoagulants
Hoxha, A., Banzato, A., Ruffatti, A., & Pengo, V. (2016) · Autoimmunity Reviews · doi:10.1016/j.autrev.2016.12.010 · Cited by 74 · see citing works
Mixing studies in lupus anticoagulant testing are required at least in some type of samples
Devreese, K. M. J., & de Laat, B. (2015) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.13020 · Cited by 63 · see citing works
Self-citationDEVREESE, K. M. J. (2010) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2010.03807.x · Cited by 42 · see citing works
Self-citationMoore, G. W., Culhane, A. P., Daw, C. R., Noronha, C. P., & Kumano, O. (2016) · Thrombosis Research · doi:10.1016/j.thromres.2016.01.019 · Cited by 42 · see citing works
Commonalities and contrasts in recent guidelines for lupus anticoagulant detection
Moore, G. W. (2014) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.12227 · Cited by 41 · see citing works
Current Controversies in Lupus Anticoagulant Detection
Moore, G. (2016) · Antibodies · doi:10.3390/antib5040022 · Cited by 36 · see citing works
Depreter, B., & Devreese, K. M. J. (2016) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-0978 · Cited by 34 · see citing works
Self-citationKumano, O., & Moore, G. W. (2018) · Research and Practice in Thrombosis and Haemostasis · doi:10.1002/rth2.12069 · Cited by 26 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Devreese, K. M. J., de Groot, P. G., de Laat, B., Erkan, D., Favaloro, E. J., Mackie, I., Martinuzzo, M., Ortel, T. L., Pengo, V., Rand, J. H., et al. (2020) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.15047 · Cited by 439 · see citing works
Self-citationLupus Anticoagulant Testing in Challenging Situations: Navigating Complexities
Devreese, K. M. J. (2026) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.14490 · Cited by 2 · see citing works
Self-citation
- Authors
- Florin, L., Desloovere, M., & Devreese, K. M. J.
- Affiliations
- Ghent University Hospital (BE)
- Corresponding author
- Devreese, Katrien M. J.
- Venue
- International Journal of Laboratory Hematology
- Publisher
- Wiley
- Published
- 2019-03-12
- Volume
- 41
- Issue
- 3
- Pages
- 412-417
- ISSN
- 1751-5521 (ISSN-L)
- 1751-553X
- OpenAlex
- W2921042625
- Cited by
- 12
- References
- 14
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 2 · 2022: 2 · 2020: 4
- FWCI
- 1.14
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- International Journal of Laboratory Hematology (publishedVersion)
- PubMed (publishedVersion)
- Ghent University Academic Bibliography (Ghent University) (submittedVersion)
- Topics
- Health Sciences › Medicine › Rheumatology › Systemic Lupus Erythematosus Research
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Hematology › Blood Coagulation and Thrombosis Mechanisms
- Keywords
- lupus anticoagulant, dilute Russell's viper venom time, activated partial thromboplastin time, anticoagulant screening, laboratory automation, assay comparability
- MeSH terms
- Algorithms, Blood Coagulation Tests, Humans, Immunoassay, Partial Thromboplastin Time, Software, Reproducibility of Results, Lupus Coagulation Inhibitor, Antiphospholipid Syndrome, Automation, Laboratory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review