Pulse oximetry and postoperative hypothermia: An evaluation of the Nellcor N‐100 in a cardiac surgical intensive care unit
Open accessAbstract
The accuracy of the Nellcor N-100 pulse oximeter was evaluated in hypothermic patients (core temperature less than or equal to 35.0 degrees C) after cardiac surgery. The pulse oximeter overestimated oxygen saturation in comparison to values obtained by direct in vitro oximetry with a mean bias of 0.6% saturation. The 95% predictability limits for individual measurements were +/- 3.9% oxygen saturation. No demonstrable loss of accuracy was caused by skin pigmentation, the concurrent administration of low doses of dopamine or vasodilators, or the intra-operative intravascular administration of Patent Blue V dye.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Some older studies did not identify a considerable bias related to skin pigmentation, suggesting that dark skin pigmentation does not affect the quality of the signal.
“Conversely, some older studies did not identify a considerable bias related to skin pigmentation [44,45,46,56] suggesting that dark skin pigmentation does not affect the quality of the signal [47].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [45]
The skin pigmentation effect on pulse oximetry accuracy has long been studied, reporting some contradictory conclusions.
“The skin pigmentation effect on pulse oximetry accuracy has long been studied reporting some contradictory conclusions.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [45]
Other subjective scales were employed in 13 studies (35.14%).
“and other subjective scales were employed in 13 studies (35.14%) [30,42,43,44,45,50,52,56,72,78,80,87,88].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [45]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Beckett, A. H., Rowland, M., & TURNER, P. (1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,917 · see citing works
Yelderman, M., & New, W. (1983) · Anesthesiology · doi:10.1097/00000542-198310000-00015 · Cited by 488 · see citing works
Fanconi, S., Doherty, P., EDMONDS, J. F., Barker, G. A., & BOHN, D. J. (1985) · The Journal of Pediatrics · doi:10.1016/s0022-3476(85)80507-2 · Cited by 158 · see citing works
Noninvasive Detection of Profound Arterial Desaturations Using a Pulse Oximetry Device
Mihm, F. G., & Halperin, B. D. (1985) · Anesthesiology · doi:10.1097/00000542-198501000-00020 · Cited by 142 · see citing works
The current status of pulse oximetry
Taylor, M. B., & WHITWAM, J. G. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12922.x · Cited by 117 · see citing works
ACCURACY OF A PULSE OXIMETER IN THE CRITICALLY ILL ADULT
Tremper, K. K., Hufstedler, S. M., Barker, S. J., Adams, A. L., Wong, D. H. W., Zaccari, J., Benik, K., & Lemons, V. (1985) · Anesthesiology · doi:10.1097/00000542-198509001-00175 · Cited by 37 · see citing works
The Nellcor N‐101 pulse oximeter
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32 · see citing works
Hanning, C. D. (1985) · British Journal of Anaesthesia · doi:10.1093/bja/57.4.359 · Cited by 24 · see citing works
Normothermia: its place in operative and post-operative care.
VALE, R. J. (1973) · PubMed · Cited by 17 · see citing works
Saloojee, Y., Cole, P. V., & Adam, L. (1981) · Journal of Medical Engineering & Technology · doi:10.3109/03091908109009365 · Cited by 5 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- HANNIX CD. Monitoring oxygcnation during anaesthesia. Britis/i Journul of .4nur.sfhr.riu 1985; 57: 359
- MIHM FG. HALPERIN BD. Noninvasive dctcction o f profound a]-terial desaturations using a pulsc oxinicti-y device. .Anc..sfhc..vi- ologj' 1985; 62: 85
- FANCONI S, DOHERTY P. EDMONDS JF. BARKER G A , BOHY 13652044, 1988, 5, Downloaded from https://associationofanaesthetists-publications.onlinelibrary.wiley.com/doi/10.1111/j.1365-2044.1988.tb09025.x by Mount Vernon Nazarene University, Wiley Online Library on [23/06/2026]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License · DOI
Cited by
Sorted by citations
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N. (1998) · Heart & Lung · doi:10.1016/s0147-9563(98)90086-3 · Cited by 243 · see citing works
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
- Authors
- GABRIELCZYK, M. R., & BUIST, R. J.
- Affiliations
- University College Hospital at Westmoreland Street (GB)
- Venue
- Anaesthesia
- Publisher
- Wiley
- Published
- 1988-05
- Volume
- 43
- Issue
- 5
- Pages
- 402-404
- ISSN
- 0003-2409 (ISSN-L)
- 1365-2044
- OpenAlex
- W1876894315
- Cited by
- 59
- References
- 11
- Citations by year
- 2026: 1 · 2024: 2 · 2023: 1 · 2022: 23
- FWCI
- 4.82
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Anaesthesia (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Thermal Regulation in Medicine
- Keywords
- pulse oximetry, postoperative hypothermia, oxygen saturation measurement, cardiac surgery, measurement accuracy, skin pigmentation, dopamine, vasodilators, Patent Blue V, core body temperature
- MeSH terms
- Aged, Critical Care, Evaluation Studies as Topic, Cardiac Surgical Procedures, Humans, Hypothermia, Middle Aged, Oximetry, Oxygen, Postoperative Complications
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review