The effects of skin pigmentation on the accuracy of pulse oximetry in measuring oxygen saturation: a systematic review and meta-analysis
Open accessAbstract
Background: Pulse oximetry was widely used in hospitals and at home to monitor blood oxygen during the COVID-19 pandemic. There have been concerns regarding potential bias in pulse oximetry measurements for people with dark skin. We aimed to assess the effects of skin pigmentation on the accuracy of oxygen saturation measurement by pulse oximetry (SpO 2 ) compared with the gold standard SaO 2 measured by CO-oximetry. Methods: We searched Ovid MEDLINE, Ovid Embase, and EBSCO CINAHL Plus (up to December 2021), as well as ClinicalTrials.gov and World Health Organization International Clinical Trials Registry Platform (up to August 2021). We identified studies comparing SpO 2 values in any population, in any care setting, using any type of pulse oximeter, with SaO 2 by standard CO-oximetry; and measuring the impact of skin pigmentation or ethnicity on pulse oximetry accuracy. We performed meta-analyses for mean bias (the primary outcome in this review) and its standard deviations (SDs) across studies included for each subgroup of level of skin pigmentation and ethnicity. We calculated accuracy root-mean-square ( A rms ) and 95% limits of agreement based on pooled mean bias and pooled SDs for each subgroup. Results: We included 32 studies (6505 participants); 27/32 (84.38%) in hospitals and none in people’s homes. Findings of 14/32 studies (43.75%) were judged, via QUADAS-2, at high overall risk of bias. Fifteen studies measured skin pigmentation and 22 referred only to ethnicity. Compared with standard SaO 2 measurement, pulse oximetry probably overestimates oxygen saturation in people with dark skin (pooled mean bias 1.11%; 95% confidence interval 0.29% to 1.93%) and people described as Black/African American (pooled mean bias 1.52%; 0.95% to 2.09%) (moderate- and low-certainty evidence). These results suggest that, for people with dark skin, pulse oximetry may overestimate blood oxygen saturation by around 1% on average compared with SaO 2 . The bias of pulse oximetry measurements for people with other levels of skin pigmentation, or those from the White/Caucasian group is more uncertain. The data do not suggest overestimation in people from other ethnic groups such as those described as Asian, Hispanic, or mixed ethnicity (pooled mean bias 0.31%, 0.09% to 0.54%), but this evidence is low certainty. Whilst the extent of mean bias is small or negligible for all the subgroups of population evaluated, the associated imprecision is unacceptably large (with the pooled SDs > 1%). Nevertheless, when the extents of measurement bias and precision are considered jointly in A rms , pulse oximetry measurements for all the subgroups appear acceptably accurate (with A rms < 4%). Conclusions: Low-certainty evidence suggests that pulse oximetry may overestimate oxygen saturation in people with dark skin and people whose ethnicity is reported as Black/African American, compared with SaO 2 , although the overestimation may be quite small in hospital settings. The clinical importance of any overestimation will depend on the particular clinical circumstance. Pulse oximetry measurements appear accurate but imprecise for all levels of skin pigmentation. The evidence relates to clinician-measured oximetry in health care environments and may not be reflected in home pulse oximetry where other factors may also influence accuracy. Author summary: Why was this study done?: Pulse oximetry was widely used in hospital and at home to measure blood oxygen levels during the COVID-19 pandemic. There was uncertainty as to whether skin pigmentation affects the accuracy of pulse oximetry measurements. What did the researchers do and find?: We assessed, via systematic review, the effects of skin pigmentation on the accuracy of pulse oximetry measurement (SpO 2 ) compared with SaO 2 measured by standard CO-oximetry. In people with dark skin, oxygen saturation measured in hospital by pulse oximetry may be overestimated by an average of 1% compared with gold standard SaO 2 , however, the evidence is of low certainty. The accuracy of pulse oximetry measured compared with standard SaO 2 is quite uncertain for people with light or medium levels of skin pigmentation and for people from ethnic groups other than those described in papers as Black or African American. Pulse oximetry measurements appear to have acceptable overall accuracy (with A rms < 4%) for all subgroups of population evaluated whilst the variation of oximetry readings appear unacceptably wide (with the pooled SDs > 1%). What do these findings mean?: Hospital measured pulse oximetry may overestimate oxygen saturation in people with dark skin compared with SaO 2 by approximately 1%. The implications of this finding in different clinical scenarios will vary but could be clinically important. Impacts are likely to be at the thresholds of being diagnosed as having hypoxaemia where even a small SpO 2 overestimation could lead to clinically important hypoxaemia remaining undetected and untreated. How these findings extrapolate to community and home care settings is unclear.
Verified 2026-10-09 · 1 reference not checked yet
AI snapshot
Written from the paper’s full text.
TL;DR
This systematic review and meta-analysis found that pulse oximetry probably overestimates oxygen saturation in people with high skin pigmentation and in people described as Black/African American compared with CO-oximetry SaO2. Although mean bias was small and accuracy root-mean-square remained below recommended thresholds, measurement precision was poor across subgroups.
Key findings
- The review included 32 studies with 6505 participants; 15 studies measured skin pigmentation and 22 referred to ethnicity.
- Pooled mean bias was 1.11% (95% CI 0.29 to 1.93%) for high skin pigmentation (moderate-certainty evidence) and 1.52% (95% CI 0.95 to 2.09%) for people described as Black/African American (low-certainty evidence), indicating overestimation by pulse oximetry.
- For medium pigmentation, low pigmentation, Asian/Hispanic/mixed ethnicity, and White/Caucasian groups, evidence was uncertain or did not suggest clinically important systematic overestimation.
- Pooled SDs were generally >1%, indicating unacceptably large imprecision, but calculated Arms values were <4% for all subgroups, which the authors describe as acceptably accurate when bias and precision are considered jointly.
Methods
- Systematic review and meta-analysis of methods-comparison studies comparing SpO2 from commercially available pulse oximeters with SaO2 measured by standard CO-oximetry; searches covered Ovid MEDLINE, Ovid Embase, EBSCO CINAHL, ClinicalTrials.gov, and WHO ICTRP up to December 2021.
- Data were pooled separately for skin pigmentation level and ethnicity using a correlated hierarchical effects model with robust variance estimation; mean bias was the primary outcome, with pooled SDs used to calculate Arms and 95% limits of agreement.
- Risk of bias was assessed with an adapted QUADAS-2, and certainty of evidence for mean bias was assessed using GRADE.
Limitations
- The authors state that some included studies reported only significance tests or diagnostic accuracy proportions rather than SpO2–SaO2 data by subgroup, limiting meta-analysis; they also note that race and ethnicity were combined for simplicity, that Fitzpatrick-type skin pigmentation scales are blunt, that device, age, condition, and pigmentation-measurement differences were not analysed, that only English-language peer-reviewed publications were searched, and that no risk-of-bias or GRADE approach was specific to this topic.
- The paper states that most included studies were hospital-based and had limited information on whether devices were appropriate for home self-monitoring, so findings may not reflect home pulse oximetry by patients or carers.
- Evident from the text: included studies spanned 1985–2021 and some may have used discontinued devices, and meta-analyses showed substantial between-study heterogeneity (e.g., overall I² values often >90% for several pooled estimates).
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 38 cited sources checked · 0 supported · 0 flagged
- Findingown data
For all levels of skin pigmentation, the Arms values are around 2% or lower.
“For all the levels of skin pigmentation, the Arms values are around 2% or lower (95% CI non-estimable), and the pooled SD values are around 1.50% on average (Table 2).”
Quote in the paperNot a substantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
- Findingown data
The extent of mean bias is small or negligible for all subgroups evaluated.
“Whilst the extent of mean bias is small or negligible for all subgroups evaluated, the associated imprecision is unacceptably large (pooled SDs > 1%).”
Quote in the paperSubstantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
- Findingown data
For people with low levels of skin pigmentation, the pooled mean bias is -0.35% (95% CI -1.36 to 0.67).
“The evidence for people with low levels of skin pigmentation does not suggest clinically important systematic bias (pooled mean bias -0.35, 95% CI − 1.36 to 0.67), but the finding is of low certainty.”
Quote in the paperSubstantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
- Findingown data
For all levels of skin pigmentation, the pooled SD values are around 1.50% on average.
“For all the levels of skin pigmentation, the Arms values are around 2% or lower (95% CI non-estimable), and the pooled SD values are around 1.50% on average (Table 2).”
Quote in the paperNot a substantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
- Findingown data
Pulse oximetry probably overestimates oxygen saturation in people described as Black/African American compared with standard SaO2, with a pooled mean bias of 1.52% (95% CI 0.95 to 2.09%).
“and people described as Black/African American (1.52%; 0.95 to 2.09%) (moderate- and low-certainty evidence).”
- Also stated in: § Conclusions · hedged
“Pulse oximetry may overestimate oxygen saturation in people with high levels of skin pigmentation and people whose ethnicity is reported as Black/African American, compared with SaO2.”
Quote in the paperSubstantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
- Also stated in: § Conclusions · hedged
- Findingown data
For all the subgroups of populations evaluated, the degree of mean bias is small or negligible over the ranges of SaO2 reported (median minimum value of 76% and maximum value of 100%).
“For all the subgroups of populations evaluated, whilst the degree of mean bias is small or negligible over the ranges of SaO2 reported (median minimum value of 76% and maximum value of 100%), pulse oximetry readings appear unacceptably imprecise (pooled SDs > the recommended criterion of 1%) [10, 55].”
Quote in the paperSubstantial claim - Findingown data
Oxygen saturation measured for people described in study reports as Black or African American may be overestimated using hospital pulse oximetry compared with standard SaO2 readings, with a mean bias of 1.52% (95% CI 0.95 to 2.09%), based on low-certainty evidence.
“Oxygen saturation measured for people described in study reports as Black or African American may be overestimated using hospital pulse oximetry compared with standard SaO2 readings: mean bias 1.52% (95% CI 0.95 to 2.09%), low-certainty evidence.”
Quote in the paperSubstantial claim - Findingown data
Overall, most devices overestimated oxygen saturation in people described as Black or African American.
“Additional file 19: Figure S8 presents evidence for each type of pulse oximeter evaluated: overall, most devices overestimated oxygen saturation in people described as Black or African American.”
Quote in the paperSubstantial claim - Findingcitation
When measurement bias and precision are considered jointly in Arms, pulse oximetry measurements for all the subgroups appear acceptably accurate, with Arms below the internationally recommended threshold of 4% or the more conservative US FDA threshold of 3%.
“Nevertheless, when the extents of measurement bias and precision are considered jointly in Arms, pulse oximetry measurements for all the subgroups appear acceptably accurate (with Arms < the internationally recommended threshold of 4% [10, 55], or even the more conservative threshold of 3% in the US FDA guidance) [56].”
Quote in the paperSubstantial claim0 of 3 sources checked - Findingown data
Eight of the 15 studies reporting by skin pigmentation (1297 participants) had available data and were included in meta-analyses.
“Eight of these studies (1297 participants) had available data and were included in the meta-analyses: [22, 24, 25, 27, 29, 30, 41, 42, 53]”
Quote in the paperSubstantial claim - Findingown data
For the ethnicity subgroups, Arms values are around 2% or lower (95% CI non-estimable).
“The Arms values are around 2% or lower (95% CI non-estimable) for all these subgroups, and the pooled SD values are around 1.50% on average (Table 2).”
Quote in the paperNot a substantial claim - Findingown data
Fifteen of the 32 studies (1800 participants) reported by level of skin pigmentation.
“Fifteen of the 32 studies (1800 participants) reported by level of skin pigmentation [22, 24, 25, 27–34, 41–44, 53].”
Quote in the paperSubstantial claim - Findingcitation
At least two studies used diagnostic accuracy design and only presented proportions of participants with specific ranges of SpO2 in relation to specific SaO2 values, rather than SpO2 and SaO2 data.
“At least two studies used diagnostic accuracy design and only presented proportions of participants with specific ranges of SpO2 in relation to specific SaO2 values, again rather than SpO2 and SaO2 data [7, 57].”
Quote in the paperNot a substantial claim0 of 2 sources checked - Findingown data
Across devices evaluated and types of participants, included studies were largely consistent in suggesting oxygen saturation overestimation of using pulse oximetry.
“However, we found, across devices evaluated and types of participants, included studies were largely consistent in suggesting oxygen saturation overestimation of using pulse oximetry.”
Quote in the paperSubstantial claim - Findingown data
Some included studies compared SpO2-SaO2 bias data between different subgroups of skin pigmentation or ethnicity and presented only tests of significance results, rather than SpO2 and SaO2 data per se at each subgroup level.
“Firstly, some included studies compared SpO2-SaO2 bias data between different subgroups of skin pigmentation or ethnicity and presented only tests of significance results, rather than SpO2 and SaO2 data per se at each subgroup level.”
Quote in the paperNot a substantial claim - Findingown data
These suggested values are even larger than the average values of our estimates (1% (± 1.5%)) in people with darker skin.
“These suggested values are even larger than the average values of our estimates (1% (± 1.5%)) in people with darker skin.”
Quote in the paperSubstantial claim - Findingnone given
The overestimation of oxygen saturation for darker skin appears consistent in general across most devices evaluated.
“Nevertheless, the overestimation of oxygen saturation for darker skin appears consistent in general across most devices evaluated.”
Quote in the paperSubstantial claim - Findingreasoning
Meta-analyses in this review demonstrated between-studies heterogeneity.
“Because of these, meta-analyses in this review demonstrated between-studies heterogeneity (Table 2).”
Quote in the paperNot a substantial claim - Study design and conductnone given
The evidence for people with medium skin pigmentation is uncertain (very low certainty evidence).
“The evidence for people with medium skin pigmentation is uncertain (very low certainty evidence).”
Quote in the paperNot a substantial claim - Interpretationreasoning
For people with any level of skin pigmentation, about 68% of their pulse oximetry readings would be within ±2% of the CO-oximetry readings.
“This means that, for people with any level of skin pigmentation, about 68% of their pulse oximetry readings would be within ± 2% of the CO-oximetry readings, with one SD indicating a variation around the mean bias of minus 1.50 to plus 1.50%.”
Quote in the paperSubstantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
- Interpretationreasoning
When the extent of measurement bias and precision is considered jointly, pulse oximetry measurements for all the subgroups appear acceptably accurate, with Arms less than 4%.
“When the extent of measurement bias and precision is considered jointly, pulse oximetry measurements for all the subgroups appear acceptably accurate (with Arms < 4%).”
Quote in the paperSubstantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
- Interpretationreasoning
For people with any level of skin pigmentation, one SD indicates a variation around the mean bias of minus 1.50 to plus 1.50%.
“This means that, for people with any level of skin pigmentation, about 68% of their pulse oximetry readings would be within ± 2% of the CO-oximetry readings, with one SD indicating a variation around the mean bias of minus 1.50 to plus 1.50%.”
Quote in the paperSubstantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
- Study design and conductown data
The finding for people with low levels of skin pigmentation is of low certainty.
“The evidence for people with low levels of skin pigmentation does not suggest clinically important systematic bias (pooled mean bias -0.35, 95% CI − 1.36 to 0.67), but the finding is of low certainty.”
Quote in the paperNot a substantial claim - Premisecitation
Two recent diagnostic studies provide evidence on clinical implications resulting from the bias in pulse oximetry for blood oxygen saturation levels.
“Two recent diagnostic studies provide evidence on clinical implications resulting from the bias in pulse oximetry for blood oxygen saturation levels [7, 57].”
Quote in the paperNot a substantial claim0 of 2 sources checked - Study design and conductown data
In 20 studies, there was no blinding information for pulse oximetry SpO2 measurements.
“there was no blinding information for either pulse oximetry SpO2 measurements (20 studies)”
Quote in the paperNot a substantial claim - Interpretationown data
For people with high levels of skin pigmentation and people described in studies as Black or African American, oxygen saturation may be overestimated by pulse oximetry in hospital compared with gold standard SaO2.
“This review suggests that for people with high levels of skin pigmentation and people described in studies as Black or African American, oxygen saturation may be overestimated by pulse oximetry in hospital compared with gold standard SaO2.”
Quote in the paperSubstantial claim - Interpretationreasoning
Overestimation could lead to clinically important hypoxaemia remaining undetected and untreated.
“Overestimation could lead to clinically important hypoxaemia remaining undetected and untreated.”
Quote in the paperSubstantial claim - Study design and conductown data
In 19 studies, the patient selection domain had specific sub-populations inappropriately excluded or selection criteria unclear or not stated.
“for the patient selection domain, where specific sub-populations were inappropriately excluded from a study, or where selection criteria were unclear or not stated (19 studies)”
Quote in the paperNot a substantial claim - Interpretationcitation
Even small amounts of mean bias, when at the margins of diagnostic thresholds, could have an impact on diagnostic accuracy.
“This may suggest that even small amounts of mean bias, when at the margins of diagnostic thresholds, could have an impact on diagnostic accuracy.”
Quote in the paperSubstantial claim - Premisecitation
The specified values of mean bias (SD for precision) (2% (± 1%)) in current standards are consistent with the outdated 1995 Jensen review results.
“The current standards did not point out evidence sources used to underpin such requirements, but the specified values of mean bias (SD for precision) (2% (± 1%)) are consistent with the outdated 1995 Jensen review results [8].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [8] source
Not checked in the free run.
- Interpretationown data
The evidence for the measurement bias identified for other levels of skin pigmentation or ethnicities is more uncertain.
“The evidence for the measurement bias identified for other levels of skin pigmentation or ethnicities is more uncertain.”
Quote in the paperSubstantial claim - Premisecitation
In these studies, people described as Black had a higher risk of occult hypoxemia that was not detected by pulse oximetry compared with those described as White.
“In these studies, people described as Black had a higher risk of ‘occult hypoxemia that was not detected by pulse oximetry’ compared with those described as White [7].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [7] source
Not checked in the free run.
- Interpretationown data
Pulse oximetry may overestimate blood oxygen saturation levels for people with dark skin in hospital settings compared with gold standard SaO2 measures.
“Pulse oximetry may overestimate blood oxygen saturation levels for people with dark skin in hospital settings compared with gold standard SaO2 measures.”
Quote in the paperSubstantial claim
- May combine two claims. 43
- Classified as probably not a substantial claim. 45
May combine two claims.
Findingown dataPulse oximetry probably overestimates oxygen saturation in people with high levels of skin pigmentation compared with standard SaO2, with a pooled mean bias of 1.11% (95% CI 0.29 to 1.93%).
“Compared with standard SaO2 measurement, pulse oximetry probably overestimates oxygen saturation in people with the high level of skin pigmentation (pooled mean bias 1.11%; 95% confidence interval 0.29 to 1.93%)”
- Also stated in: § Pulse oximetry accuracy by levels of skin pigmentation · hedged
“Hospital-based pulse oximetry probably overestimates oxygen saturation for people with high levels of skin pigmentation compared with standard SaO2 (8 studies, 24 comparisons, 3270 SpO2-SaO2 pairs from 221 participants): pooled mean bias 1.11% (95% CI 0.29 to 1.93%), moderate-certainty evidence.”
Quote in the paperSubstantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
- Also stated in: § Pulse oximetry accuracy by levels of skin pigmentation · hedged
May combine two claims.
Findingown dataThe associated imprecision is unacceptably large, with pooled SDs greater than 1%.
“Whilst the extent of mean bias is small or negligible for all subgroups evaluated, the associated imprecision is unacceptably large (pooled SDs > 1%).”
Quote in the paperSubstantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
May combine two claims.
Findingown dataThe bias of pulse oximetry measurements for people with other levels of skin pigmentation or those from other ethnic groups is either more uncertain or suggests no overestimation.
“The bias of pulse oximetry measurements for people with other levels of skin pigmentation or those from other ethnic groups is either more uncertain or suggests no overestimation.”
Quote in the paperSubstantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe pooled SD values are around 1.50% on average for all these subgroups.
“The Arms values are around 2% or lower (95% CI non-estimable) for all these subgroups, and the pooled SD values are around 1.50% on average (Table 2).”
Quote in the paperNot a substantial claimMay combine two claims.
FindingcitationPulse oximetry readings appear unacceptably imprecise for all subgroups, with pooled SDs greater than the recommended criterion of 1%.
“For all the subgroups of populations evaluated, whilst the degree of mean bias is small or negligible over the ranges of SaO2 reported (median minimum value of 76% and maximum value of 100%), pulse oximetry readings appear unacceptably imprecise (pooled SDs > the recommended criterion of 1%) [10, 55].”
Quote in the paperSubstantial claim0 of 2 sources checkedMay combine two claims.
Findingown dataThe review included 32 studies with 6505 participants.
“We included 32 studies (6505 participants): 15 measured skin pigmentation and 22 referred to ethnicity.”
- Also stated in: § Study selection and characteristics · plain
“The 32 studies (6505 participants) reported SpO2-SaO2 comparison evaluations of 54 different pulse oximeters (26 manufacturers) cf. standard SaO2 (Additional file 7: Table S3).”
Quote in the paperSubstantial claim- Also stated in: § Study selection and characteristics · plain
May combine two claims.
Findingown dataSensitivity analyses showed that Arms and SD values were generally consistent, but there was increased uncertainty for mean bias findings.
“We tested the sensitivity of the findings: Arms and SD values were generally consistent but there was increased uncertainty for mean bias findings.”
Quote in the paperNot a substantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataOf the 16 studies reporting SaO2 ranges, four had ranges in line with the recommended range of 70 to 100%, eight had narrower ranges, and four had wider ranges.
“Of the 16 studies, four had SaO2 ranges that were in line with the recommended range of 70 to 100%; eight had narrower ranges such as 80 or 90 to 100%; and four had wider ranges such as 50 or 60 to 100%.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataOf the 32 studies, 16 (50%) reported the ranges of SaO2 over which the accuracy of pulse oximeters was evaluated; the minimum values had a median of 76% and the maximum values had a median of 100%.
“Of the 32 studies, 16 (50%) reported the ranges of SaO2 over which the accuracy of pulse oximeters was evaluated: the minimum values of these ranges had a median of 76% whilst the maximum values had a median of 100%.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingnone givenThe evidence is uncertain for groups described in papers as White/Caucasian, meaning further research is likely to alter findings, with very low certainty evidence.
“The evidence is uncertain for groups described in papers as White/Caucasian, meaning further research is likely to alter findings (very low certainty evidence).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataOf the 32 included studies, 15 measured skin pigmentation and 22 referred to ethnicity.
“We included 32 studies (6505 participants): 15 measured skin pigmentation and 22 referred to ethnicity.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe evidence for people described in studies as Asian, Hispanic or of mixed ethnicity does not indicate a clinically important systematic bias, with a mean bias of 0.31% (95% CI 0.09 to 0.54%), but it is of low certainty.
“The evidence for people described in studies as Asian, Hispanic or of mixed ethnicity does not indicate a clinically important systematic bias (mean bias 0.31%, 0.09 to 0.54%), but it is of low certainty.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe 32 studies reported SpO2-SaO2 comparison evaluations of 54 different pulse oximeters from 26 manufacturers.
“The 32 studies (6505 participants) reported SpO2-SaO2 comparison evaluations of 54 different pulse oximeters (26 manufacturers) cf. standard SaO2 (Additional file 7: Table S3).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe remaining seven studies (503 participants) were excluded from meta-analysis due to lack of mean bias data by levels of skin pigmentation.
“The remaining seven studies (503 participants) were excluded from meta-analysis due to lack of mean bias data by levels of skin pigmentation (Additional file 9: Table S5).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe review identified 33 publications of 32 studies published between 1985 and 2021 as eligible for inclusion.
“Of these records, we identified 33 publications of 32 studies—published between 1985 and 2021—as eligible for inclusion (Fig. 1) [21–53].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [21–53]
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataBefore this review, a scoping exercise using a simple search of Ovid Medline with pulse oximetry terms identified one systematic review in this area published by Jensen and colleagues in 1995.
“Before this review, our scoping exercise using a simple search of Ovid Medline with ‘pulse oximetry’ terms identified one systematic review in this area published by Jensen and colleagues in 1995 [8].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [8] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingnone givenA recent rapid review by the NHS Race and Health Observatory had an unclear methodology.
“We also identified a recent rapid review by the NHS Race and Health Observatory that had an unclear methodology [6].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [6] NHS Race and Health Observatory. Pulse oximetry and racial bias: Recommendations for national healthcare, regulatory and research bodies. March 2021. https://www.nhsrho.org/wp-content/uploads/2021/03/Pulse-oximetry-racial-bias-report.pdf
Not checked in the free run.
May combine two claims.
FindingcitationThe calculated Arms values are generally around 2% or less over the ranges of SaO2 reported, far below the Arms threshold of 4% required by current international and UK standards.
“It is interesting to note that, despite clinically important mean bias and unacceptably large imprecision identified, the calculated Arms values are generally around 2% or less over the ranges of SaO2 reported, that is, the Arms values are far below the Arms threshold of 4% required by the current international and UK standards [10, 55].”
Quote in the paperSubstantial claim0 of 2 sources checkedMay combine two claims.
FindingreasoningWhilst the extent of measurement bias and overall accuracy meet current international thresholds, the variation of pulse oximetry measurements appears unacceptably wide.
“Whilst the extent of measurement bias and overall accuracy meet current international thresholds, the variation of pulse oximetry measurements appears unacceptably wide.”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenMost included studies in this review were based in hospital settings and had limited information whether the pulse oximeters evaluated were appropriate for home self-monitoring.
“However, most included studies in this review were based in hospital settings and had limited information whether the pulse oximeters evaluated were appropriate for home self-monitoring.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
FindingreasoningThe included studies were published between 1985 and 2021 and some of the older studies may have used discontinued devices.
“The included studies were published between 1985 and 2021 and some of the older studies may have used discontinued devices.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone givenThe extent of overestimation may be small in hospital settings but unknown in community settings.
“The extent of overestimation may be small in hospital settings but unknown in community settings.”
Quote in the paperNot a substantial claim- Rests on an assumption: SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataThe review was registered with the Open Science Framework.
“The review was registered with the Open Science Framework (https://osf.io/gm7ty).”
- Also stated in: § Search strategy and selection criteria · plain
“The methods used were described in the registered protocol (https://osf.io/gm7ty).”
Quote in the paperNot a substantial claim- Also stated in: § Search strategy and selection criteria · plain
May combine two claims.
Study design and conductown dataUsing QUADAS-2, 14/32 studies (43.75%) were at unclear risk of bias for all four domains or high risk of bias for at least one domain, and the remaining 18 (56.25%) were at low risk of bias for at least one of the four domains.
“Using QUADAS-2, we considered 14/32 studies (43.75%) to be at unclear risk of bias for all four domains or high risk of bias for at least one domain, and the remaining 18 (56.25%) to be at low risk of bias for at least one of the four domains (Fig. 2).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationThe review is reported in accordance with the PRISMA statement.
“We report this review in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement [9].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [9] Higgins JPT , Thomas J , Chandler J , Cumpston M , Li T , Page MJ , Welch VA (editors). Cochrane Handbook for Systematic Reviews of Interventions version 6.2 (updated February 2021). Cochrane, 2021. Available from http://www.training.cochrane.org/handbook.
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Interpretationnone givenThese results are for clinician-measured oximetry in controlled clinical environments and do not necessarily reflect the measurement bias of home pulse oximetry by patients or carers.
“These results are for clinician-measured oximetry in controlled clinical environments and do not necessarily reflect the measurement bias of home pulse oximetry by patients or carers.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenPre-planned sensitivity analyses were undertaken excluding studies with similar skin pigmentation or same ethnicity, studies with no data available for meta-analysis without transformation, and studies at high overall risk of bias.
“We undertook pre-planned sensitivity analyses through (1) excluding studies where all participants had similar skin pigmentation or the same ethnicity, (2) excluding studies with no data available for meta-analysis without transformation, and (3) excluding studies at high overall risk of bias.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningPulse oximetry for people with other levels of skin pigmentation is less likely to be overestimated, but the evidence is uncertain.
“Pulse oximetry for people with other levels of skin pigmentation is less likely to be overestimated but the evidence is uncertain.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenSeparate analysis of studies reporting level of skin pigmentation and ethnicity was pre-specified.
“We pre-specified separate analysis of studies reporting level of skin pigmentation and ethnicity.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationOne reviewer assessed the certainty of evidence on mean bias using the GRADE approach, checked by another reviewer.
“One reviewer (CS) assessed the certainty of evidence on mean bias using the GRADE approach developed for the test accuracy topic, checked by another reviewer (GN) [18, 19].”
Quote in the paperNot a substantial claim0 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataTwo reviewers independently assessed titles and abstracts and full texts, with disagreements resolved through discussion or involving a third reviewer.
“Two reviewers (CS and MG, or JH, OH) independently assessed titles and abstracts of the search results for relevance and the full texts of all potentially eligible studies for inclusion, with disagreements resolved through discussion or involving a third reviewer (GN) where necessary.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataIn 2 studies, the flow and timing domain had time intervals between SpO2 readings and arterial blood sampling for SaO2 measurement that were too long or participants excluded from analysis without rationale.
“for the flow and timing domain, where the time intervals between SpO2 readings and the arterial blood sampling for SaO2 measurement were too long or participants were excluded from the analysis without rationale (2 studies).”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductown dataApplicability concern was judged high for one study, moderate for 13 studies, and low for the remaining 18 studies.
“We judged the applicability concern as high for one study, moderate for 13 studies, and low in terms of all three applicability considerations for the remaining 18 studies.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningEven though the internationally recommended thresholds were met in terms of measurement bias, the relatively small amount of mean bias identified could impact on clinical decision-making at threshold values for diagnosis of hypoxaemia.
“Even though our estimates suggest that the internationally recommended thresholds were met in terms of measurement bias [10, 55], the relatively small amount of mean bias identified could impact on clinical decision-making at threshold values for diagnosis of hypoxaemia.”
Quote in the paperSubstantial claim0 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationPublication bias was assessed following a qualitative approach.
“We assessed publication bias following a qualitative approach given funnel plots or Egger’s tests were not considered appropriate for this review [20].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [20] 2017, A framework for the meta-analysis of Bland–Altman studies based on a limits of agreement approach, Stat Med source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataOne reviewer extracted data and assessed risk of bias using an adapted QUADAS-2, all checked by another reviewer.
“One reviewer (CS, or OH or JH) independently extracted data from included studies for items in Additional file 3: Box S2 and assessed the risk of bias for the included studies using an adapted QUADAS-2 (Additional file 4: Box S3) [12], all checked by another reviewer (JH, MG, OH, GN).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [12] British Standards Institution (BSI). BS EN ISO 80601-2-61:2019: Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment (ISO 80601-2-61:2017, Corrected version 2018-02). London: British Standards Institution. 2019.
Not checked in the free run.
Classified as probably not a substantial claim.
PremisecitationA recent US study found that people described as Black had nearly three times the frequency of occult hypoxemia not detected by pulse oximetry as those described as White.
“Results showed people described as Black had ‘nearly three times the frequency of occult hypoxemia that was not detected by pulse oximetry’ as those described as White [7].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [7] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataIn 30 studies, there was no blinding information for CO-oximeter SaO2 readings.
“or CO-oximeter SaO2 readings (30 studies)”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationreasoningUnderestimated SpO2 readings could also be harmful, resulting in unnecessary treatment with oxygen (and the risk of hyperoxaemia) and wider impacts such as delayed hospital discharge.
“Underestimated SpO2 readings could also be harmful, resulting in unnecessary treatment with oxygen (and the risk of hyperoxaemia) and wider impacts such as delayed hospital discharge.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationA 1995 systematic review included three studies that explicitly considered the impact of skin pigmentation on pulse oximetry accuracy and suggested that pulse oximeters may overestimate blood oxygen saturation in people with dark skin.
“The only current relevant systematic review, published in 1995, included three studies that explicitly considered the impact of skin pigmentation on pulse oximetry accuracy [8].”
- Also stated in: § Background · hedged
“The review suggested that pulse oximeters may overestimate blood oxygen saturation in people with dark skin [8].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [8] source
Not checked in the free run.
- Also stated in: § Background · hedged
Classified as probably not a substantial claim.
InterpretationreasoningThe low certainty for much of the data presented means that further research could overturn these conclusions.
“The low certainty for much of the data presented means that further research could overturn these conclusions.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductown dataSensitivity analyses showed Arms and SD values were generally consistent, but there was increased uncertainty for mean bias findings.
“We tested the sensitivity of the findings: Arms and SD values were generally consistent but there was increased uncertainty for mean bias findings.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe recent rapid review by the NHS Race and Health Observatory came to similar conclusions but used a non-systematic review process.
“The recent rapid review by the NHS Race and Health Observatory came to similar conclusions but used a non-systematic review process, i.e., no comprehensive search, risk of bias assessment or meta-analysis [6].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [6] NHS Race and Health Observatory. Pulse oximetry and racial bias: Recommendations for national healthcare, regulatory and research bodies. March 2021. https://www.nhsrho.org/wp-content/uploads/2021/03/Pulse-oximetry-racial-bias-report.pdf
Not checked in the free run.
Classified as probably not a substantial claim.
Study design and conductcitationThis approach deals with correlations of multiple effect size estimates within a repeated-measures design study.
“This approach deals with correlations of multiple effect size estimates within a repeated-measures design study [14, 15].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [14] 2004, Skin color is not the same thing as race, Arch Dermatol
Not checked in the free run.
- cites [15] source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable). May combine two claims.
InterpretationreasoningCurrently recommended thresholds may need re-evaluation, and use of the more conservative criterion of 3% applied by the US FDA guidance may have merit.
“Given these, currently recommended thresholds may need re-evaluation, and use of the more conservative criterion of 3% applied by the US FDA guidance may have merit [56].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [56] 2018, Comparison of nasal and forehead oximetry accuracy and pressure injury in critically ill patients, Heart Lung source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationcitationUsing scales like the Fitzpatrick scale to measure levels of skin pigmentation is a limitation; such scales are criticised as being too blunt, which impacts on the findings of this review.
“We also acknowledge the limitation of using scales like the Fitzpatrick scale to measure levels of skin pigmentation [59]. Such scales are criticised as being too blunt—an issue that impacts on the findings of this review and should be considered in future research.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [59] 1994, Accuracy of pulse oximetry during hypoxemia, South Med J source
Not checked in the free run.
- The authors state this as an assumption: Scales such as the Fitzpatrick scale used in the included studies provide valid measures of levels of skin pigmentation for assessing pulse oximetry bias.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenWe did not consider the differences between specific pulse oximeter devices, the differences between children and adults and their health conditions or the difference between skin pigmentation measurement methods.
“Thirdly, we did not consider the differences between specific pulse oximeter devices, the differences between children and adults and their health conditions or the difference between skin pigmentation measurement methods.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationreasoningThis review only addresses skin pigmentation and ethnicity; therefore, little is known for pulse oximetry undertaken by untrained people at home where other factors such as movement need to be considered.
“This review only addresses skin pigmentation and ethnicity. Therefore, little is known for the case of pulse oximetry undertaken by untrained people at home where other factors such as movement need to be considered.”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductcitationWe developed a correlated hierarchical effects model and used the novel RVE approaches to meta-analyse not only independent data (of 11 studies) but also data from studies (n = 21) with repeated-measures design.
“We developed a correlated hierarchical effects model and used the novel RVE approaches to meta-analyse not only independent data (of 11 studies) but also data from studies (n = 21) with repeated-measures design [15].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [15] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisenone givenNo available approach to risk of bias and GRADE assessment is specific to the topic of this review.
“Finally, no available approach to risk of bias and GRADE assessment is specific to the topic of this review.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationIn the NHS rapid review, a summary of narrative findings suggested the overestimation of blood oxygen saturation levels in people with darker skin.
“In this rapid review, a summary of narrative findings suggested the overestimation of blood oxygen saturation levels in people with darker skin.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [6] NHS Race and Health Observatory. Pulse oximetry and racial bias: Recommendations for national healthcare, regulatory and research bodies. March 2021. https://www.nhsrho.org/wp-content/uploads/2021/03/Pulse-oximetry-racial-bias-report.pdf
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningFindings support calls for better calibrating algorithms used in oximeter device software to inherently address possible measurement bias.
“Findings also support calls for better calibrating algorithms used in oximeter device software to inherently address possible measurement bias.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductreasoningWe were only able to use the relevant approaches developed for the test accuracy topic, and the GRADE approach used was only applicable to assess the certainty of evidence for mean bias, rather than precision, Arms and limits of agreement.
“We were only able to use the relevant approaches developed for the test accuracy topic, and the GRADE approach used was only applicable to assess the certainty of evidence for mean bias, rather than precision, Arms and limits of agreement.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenWe considered the possible publication bias in assessing the certainty of evidence using GRADE approach.
“We considered the possible publication bias in assessing the certainty of evidence using GRADE approach.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductnone givenWe used the gold standard CO-oximetry as the comparator for pulse oximetry, and accuracy outcomes as recommended in the British Standards Institution standards for pulse oximetry.
“We used the gold standard CO-oximetry as the comparator for pulse oximetry, and accuracy outcomes as recommended in the British Standards Institution standards for pulse oximetry.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable), S4 (based).
Interpretationnone givenSuch a small overestimation may be crucial for some patients, particularly at the threshold that informs clinical decision-making.
“Such a small overestimation may be crucial for some patients: particularly at the threshold that informs clinical decision-making.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningOur findings indicate that sub-population specific recommendations would be needed for future updates.
“Our findings indicate that sub-population specific recommendations would be needed for future updates.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Study design and conductnone givenOur search for research is comprehensive and identified more studies.
“Our search for research is comprehensive and identified more studies.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenWe only searched for English language peer-reviewed publications, without considering preprints.
“Fourthly, we only searched for English language peer-reviewed publications, without considering preprints.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe Jensen review included only one study with data on the impact of skin pigmentation, and findings were inconclusive.
“It included only one study with data on the impact of skin pigmentation, and findings were inconclusive.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [8] source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningIf more data were received, then the review results could change.
“If more data were received, then the review results could change.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
PremisecitationThe 2021 WHO living guidance recommends using pulse oximetry monitoring at home as part of care package for symptomatic people in community settings but does not note the potential impact of level of skin pigmentation.
“In terms of COVID-19 management, the 2021 WHO living guidance recommends using pulse oximetry monitoring at home as part of care package for symptomatic people in community settings but does not note the potential impact of level of skin pigmentation [3].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [3] World Health Organization (WHO). Covid-19 clinical management: living guideline (updated 25.1.21). 2021. https://apps.who.int/iris/handle/10665/338882
Not checked in the free run.
May combine two claims.
PremisecitationOf the nine studies identified in the NHS rapid review, seven had appropriate SpO2-SaO2 comparison data but the other two used inappropriate designs for the question being addressed.
“Of the nine studies identified in this rapid review, seven had appropriate SpO2–SaO2 comparison data but the other two used inappropriate designs for the question being addressed.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [6] NHS Race and Health Observatory. Pulse oximetry and racial bias: Recommendations for national healthcare, regulatory and research bodies. March 2021. https://www.nhsrho.org/wp-content/uploads/2021/03/Pulse-oximetry-racial-bias-report.pdf
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based). May combine two claims.
Interpretationnone givenIf we had treated race and ethnicity data separately, the evidence base would change; however, we would not expect the overall conclusion to change.
“If we had treated race and ethnicity data separately, the evidence base would change; however, we would not expect the overall conclusion to change.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationFor simplicity, we chose to use the term of ethnicity throughout this review given race and ethnicity are context/country-specific concepts and there is no globally accepted classification approach to distinguishing them.
“For simplicity, we chose to use the term of ‘ethnicity’ throughout this review given race and ethnicity are context/country-specific concepts and there is no globally accepted classification approach to distinguishing them [58].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [58] source
Not checked in the free run.
- The authors state this as an assumption: Ethnicity categories used in the included studies are valid proxies for levels of skin pigmentation when pooling data on pulse oximetry bias.
May combine two claims.
PremisecitationThere are probably no major differences between summary treatment effects in English-language restricted meta-analyses and other language-inclusive meta-analyses, and the exclusion of non-English language publications from systematic reviews had no impacts on overall findings.
“However, there are probably no major differences between summary treatment effects in English-language restricted meta-analyses and other language-inclusive meta-analyses [60], and the exclusion of non-English language publications from systematic reviews had no impacts on overall findings [61].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [60] 2021, Racial bias in pulse oximetry measurement among patients about to undergo ECMO in 2019-2020, a retrospective cohort study, Chest
Not checked in the free run.
- cites [61] 2021, Racial discrepancy in pulse oximeter accuracy in preterm infants, J Perinatol
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
Study design and conductreasoningFollowing prespecified methods to minimise the risk of bias in the review process, this review has important strengths.
“Following prespecified methods to minimise the risk of bias in the review process, this review has important strengths.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Interpretationnone givenThere may be differences between devices for the use of health professionals in hospitals and those for home self-monitoring.
“Regarding pulse oximeters evaluated, there may be differences between devices for the use of health professionals in hospitals and those for home self-monitoring.”
Quote in the paperNot a substantial claim
The paper relies on this without support.
SaO2 measured by CO-oximetry is the gold standard for blood oxygen saturation.
“The gold standard measure of blood oxygen saturation levels (SaO2) requires a sample of arterial blood and measurement using CO-oximetry.”
The paper relies on this without support.
Pulse oximetry works by beaming light through skin into the blood and inferring an SpO2 reading from the amount of light absorbed.
“Pulse oximetry works by beaming light through skin into the blood and inferring an SpO2 reading from the amount of light absorbed.”
The authors state this as an assumption.
Ethnicity categories used in the included studies are valid proxies for levels of skin pigmentation when pooling data on pulse oximetry bias.
“For simplicity, we chose to use the term of ‘ethnicity’ throughout this review given race and ethnicity are context/country-specific concepts and there is no globally accepted classification approach to distinguishing them [58].”
The authors state this as an assumption.
Scales such as the Fitzpatrick scale used in the included studies provide valid measures of levels of skin pigmentation for assessing pulse oximetry bias.
“We also acknowledge the limitation of using scales like the Fitzpatrick scale to measure levels of skin pigmentation [59]. Such scales are criticised as being too blunt—an issue that impacts on the findings of this review and should be considered in future research.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR or pulse oximeter SpO2.
“There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR [35] or pulse oximeter SpO2 [36-38].”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [36]
Only 1 meta-analysis focused on pulse oximetry for this topic is available, and an additional 7 studies have been published recently, providing an additional 50,980 participants and 182,369 paired observations.
“Only 1 meta-analysis focused on pulse oximetry for this topic is available [36], and an additional 7 studies have been published recently, providing an additional 50,980 participants and 182,369 paired observations.”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [36]
No established methodologies exist for providing pooled 95% LoA and Arms, so the study followed an analytical approach used in prior meta-analyses.
“When it comes to providing pooled 95% LoA and Arms, no established methodologies exist, so we followed an analytical approach used in prior meta-analyses within the discipline [36].”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [36]
A recent comprehensive review comparing SpO2 measurements with arterial blood gas analysis SaO2 in different patients and conditions reported accurate but imprecise measurements across all levels of skin pigmentation.
“A recent comprehensive review of different studies comparing SpO2 measurements with arterial blood gas analysis SaO2 in different patients and conditions has reported accurate but imprecise measurements across all levels of skin pigmentation [61].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [61]
Several health organizations and healthcare professionals have raised growing concerns about the clinical accuracy of pulse oximetry in dark-skinned patients, particularly in detecting occult hypoxemia.
“Due to this increasing evidence, several health organizations and healthcare professionals have raised growing concerns about the clinical accuracy of pulse oximetry in dark-skinned patients [20,57,58,59,60,61], particularly in detecting occult hypoxemia [54,55].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [61]
Low-certainty evidence about overestimations in dark-skinned patients was reported.
“Although not all the reported studies assess specifically the influence of skin pigmentation, low-certainty evidence about overestimations in dark-skinned patients was reported.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [61]
Ancestry descriptors were used in 11 studies (29.73%).
“On the other hand, ancestry descriptors were used in 11 studies (29.73%) [17,21,23,24,41,51,53,61,75,82,93]”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [61]
Objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) and 4/37 studies (10.81%), respectively.
“We find that objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) [61,69,70,71,76,77] and 4/37 studies (10.81%) [47,48,74,89], respectively.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [61]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Jensen review publication year differs from its reference
MinorThe text dates the Jensen review to 1995, whereas reference 8 gives its publication year as 1998.
“one systematic review in this area published by Jensen and colleagues in 1995 [8].”
“8. Jensen LA Onyskiw JE Prasad NG Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults Heart Lung 1998 27 6 387 408”
Preprint cited among included studies despite stated exclusion
MinorThe paper says it did not consider preprints, but its list of included ethnicity studies cites reference 26, which the references explicitly label a preprint.
“we only searched for English language peer-reviewed publications, without considering preprints.”
“26. [preprint] Brooks JC, Raman S, Gibbons K, et al. Transcutaneous oxygen saturation accuracy in critically ill children. Research Square 2020.”
Cites
R: A Language and Environment for Statistical Computing
R Core Team (2000) · doi:10.32614/r.manuals · Cited by 355,706 · see citing works
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Conducting Meta-Analyses in <i>R</i> with the <b>metafor</b> Package
Viechtbauer, W. (2010) · Journal of Statistical Software · doi:10.18637/jss.v036.i03 · Cited by 20,464 · see citing works
QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies
Whiting, P. F., Rutjes, A. W. S., Westwood, M. E., Mallett, S., Deeks, J. J., Reitsma, J. B., Leeflang, M. M. G., Sterne, J. A. C., Bossuyt, P. M. M., & the QUADAS-2 Group* (2011) · Annals of Internal Medicine · doi:10.7326/0003-4819-155-8-201110180-00009 · Cited by 14,199 · see citing works
International Organization for Standardization
(2011) · Springer eBooks · doi:10.1007/springerreference_16847 · Cited by 12,502 · see citing works
Page, M. J., Moher, D., Bossuyt, P. M., Boutron, I., Hoffmann, T. C., Mulrow, C. D., Shamseer, L., Tetzlaff, J. M., Akl, E. A., Brennan, S. E., et al. (2021) · BMJ · doi:10.1136/bmj.n160 · Cited by 11,883 · see citing works
Wan, X., Wang, W., Liu, J., & Tong, T. (2014) · BMC Medical Research Methodology · doi:10.1186/1471-2288-14-135 · Cited by 11,735 · cited 1× · see citing works
“from reported 95% limits of agreement to standard deviation (SD) [13].”
BackgroundSynthesis without meta-analysis (SWiM) in systematic reviews: reporting guideline
Campbell, M., McKenzie, J. E., Sowden, A., Katikireddi, S. V., Brennan, S. E., Ellis, S., Hartmann-Boyce, J., Ryan, R., Shepperd, S., Thomas, J., et al. (2020) · BMJ · doi:10.1136/bmj.l6890 · Cited by 4,628 · see citing works
The validity and practicality of sun-reactive skin types I through VI
Fitzpatrick, T. B. (1988) · Archives of Dermatology · doi:10.1001/archderm.1988.01670060015008 · Cited by 4,339 · see citing works
Robust variance estimation in meta‐regression with dependent effect size estimates
Hedges, L. V., Tipton, E., & Johnson, M. C. (2010) · Research Synthesis Methods · doi:10.1002/jrsm.5 · Cited by 2,086 · see citing works
THE ROYAL AUSTRALIAN COLLEGE OF GENERAL PRACTITIONERS
(1974) · The Medical Journal of Australia · doi:10.5694/j.1326-5377.1974.tb93279.x · Cited by 1,335 · see citing works
Morrison, A., Polisena, J., Husereau, D., Moulton, K., Clark, M., Fiander, M., Mierzwinski-Urban, M., Clifford, T., Hutton, B., & Rabb, D. (2012) · International Journal of Technology Assessment in Health Care · doi:10.1017/s0266462312000086 · Cited by 1,257 · see citing works
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114 · see citing works
Meta-analysis with robust variance estimation: Expanding the range of working models
Pustejovsky, J. E., & Tipton, E. (2022) · Prevention Science · doi:10.1007/s11121-021-01246-3 · Cited by 737 · cited 2× · see citing works
“mean bias and SD) of a repeated-measures design study in meta-analysis even when the dependence structure is unknown.[19-21] We used Tau2 (95% CI), I2 and the Q statistic and the related Chi2 test to fully assess heterogeneity in meta-analysis.”
MethodQUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies
Whiting, P. F., Rutjes, A. W., Westwood, M. E., Mallett, S., Deeks, J. J., Reitsma, J. B., Leeflang, M., Sterne, J. A. C., & Bossuyt, P. M. (2011) · Data Archiving and Networked Services (DANS) · doi:10.1059/0003-4819-155-8-201110180-00009 · Cited by 601 · see citing works
Onukwugha, E., Yong, C., Hussain, A., Seal, B., & Mullins, C. D. (2014) · BMC Medical Research Methodology · doi:10.1186/1471-2288-14-1 · Cited by 589 · see citing works
Deploying an interactive machine learning system in an evidence-based practice center: abstrackr
Wallace, B. C., Small, K., Brodley, C. E., Lau, J., & Trikalinos, T. A. (2012) · Proceedings of the 2nd ACM SIGHIT International Health Informatics Symposium · doi:10.1145/2110363.2110464 · Cited by 532 · see citing works
Nussbaumer-Streit, B., Klerings, I., Dobrescu, A. I., Persad, E., Stevens, A., Garritty, C., Kamel, C., Affengruber, L., King, V. J., & Gartlehner, G. (2020) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2019.10.011 · Cited by 440 · see citing works
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427 · see citing works
Schünemann, H. J., Mustafa, R. A., Brozek, J., Steingart, K. R., Leeflang, M., Murad, M. H., Bossuyt, P. M., Glasziou, P., Jaeschke, R., Lange, S., et al. (2020) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2019.12.021 · Cited by 416 · see citing works
Schünemann, H. J., Mustafa, R. A., Brożek, J., Steingart, K. R., Leeflang, M., Murad, M. H., Bossuyt, P. M., Glasziou, P., Jaeschke, R., Lange, S., et al. (2020) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2019.12.020 · Cited by 416 · see citing works
Feiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400 · see citing works
The Reporting of Race and Ethnicity in Medical and Science Journals
Flanagin, A., Frey, T., Christiansen, S. L., & Bauchner, H. (2021) · JAMA · doi:10.1001/jama.2021.2104 · Cited by 304 · see citing works
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · see citing works
Wong, A.-K. I., Charpignon, M., Kim, H., Josef, C., de Hond, A. A. H., Fojas, J. J., Tabaie, A., Liu, X., Mireles-Cabodevila, E., Carvalho, L., et al. (2021) · JAMA Network Open · doi:10.1001/jamanetworkopen.2021.31674 · Cited by 268 · see citing works
Luks, A. M., & Swenson, E. R. (2020) · Annals of the American Thoracic Society · doi:10.1513/annalsats.202005-418fr · Cited by 263 · see citing works
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N. (1998) · Heart & Lung · doi:10.1016/s0147-9563(98)90086-3 · Cited by 243 · see citing works
History of blood gas analysis. VII. Pulse oximetry
Severinghaus, J. W., & Honda, Y. (1987) · The Journal of Clinical Monitoring · doi:10.1007/bf00858362 · Cited by 180 · see citing works
Schünemann, H. J., Mustafa, R. A., Brożek, J., Steingart, K. R., Leeflang, M., Murad, M. H., Bossuyt, P. M., Glasziou, P., Jaeschke, R., Lange, S., et al. (2020) · LSTM Online Archive (Liverpool School of Tropical Medicine) · doi:10.7892/boris.140684 · Cited by 156 · see citing works
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145 · see citing works
Accuracy of Pulse Oximetry in Children
Ross, P. A., Newth, C. J. L., & Khemani, R. G. (2014) · Pediatrics · doi:10.1542/peds.2013-1760 · Cited by 129 · see citing works
Utility of pulse oximetry in the detection of arterial hypoxemia in liver transplant candidates
Abrams, G. A., Sanders, M. K., & Fallon, M. B. (2002) · Liver Transplantation · doi:10.1053/jlts.2002.32252 · Cited by 127 · see citing works
Racial discrepancy in pulse oximeter accuracy in preterm infants
Vesoulis, Z., Tims, A., Lodhi, H., Lalos, N., & Whitehead, H. (2022) · Journal of Perinatology · doi:10.1038/s41372-021-01230-3 · Cited by 120 · see citing works
Valbuena, V. S. M., Barbaro, R. P., Claar, D. D., Valley, T. S., Dickson, R. P., Gay, S. E., Sjoding, M. W., & Iwashyna, T. J. (2021) · CHEST Journal · doi:10.1016/j.chest.2021.09.025 · Cited by 114 · see citing works
Reliability of noninvasive oximetry in black subjects during exercise and hypoxia
Zeballos, R. J., & Weisman, I. M. (1991) · American Review of Respiratory Disease · doi:10.1164/ajrccm/144.6.1240 · Cited by 108 · see citing works
(2020) · PubMed · Cited by 108 · see citing works
Accuracy of two ear oximeters at rest and during exercise in pulmonary patients.
Ries, A. L., Farrow, J. T., & Clausen, J. L. (1985) · PubMed · doi:10.1164/arrd.1985.132.3.685 · Cited by 106 · see citing works
Sotgiu, G., Barassi, A., Miozzo, M. R., Saderi, L., Piana, A. F., Orfeo, N. V., Colosio, C., Felisati, G., Davì, M., Gerli, A. G., et al. (2020) · BMC Pulmonary Medicine · doi:10.1186/s12890-020-01237-0 · Cited by 105 · see citing works
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102 · see citing works
Effect of nail polish on oxygen saturation determined by pulse oximetry in critically ill patients
Hinkelbein, J., Genzwuerker, H. V., Sogl, R., & Fiedler, F. (2006) · Resuscitation · doi:10.1016/j.resuscitation.2006.06.024 · Cited by 100 · see citing works
The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia
Foglia, E. E., Whyte, R. K., Chaudhary, A., Mott, A., Chen, J., Propert, K. J., & Schmidt, B. (2017) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2016.11.043 · Cited by 96 · see citing works
Ebmeier, S. J., Barker, M., Bacon, M., Beasley, R. C., Bellomo, R., Chong, C. K., Eastwood, G. M., Gilchrist, J., Kagaya, H., Pilcher, J., et al. (2018) · Anaesthesia and Intensive Care · doi:10.1177/0310057x1804600307 · Cited by 92 · see citing works
Ear oximetry during combined hypoxia and exercise
Smyth, R. J., D'Urzo, A. D., Slutsky, A. S., Galko, B. M., & Rebuck, A. S. (1986) · Journal of Applied Physiology · doi:10.1152/jappl.1986.60.2.716 · Cited by 80 · cited 4× · see citing works
“75%) to be at unclear risk of bias for all four domains or high risk of bias for at least one domain,[34,38,44,47,48,53,54,56-61,63] and the remaining 18 (56.”
BackgroundMethodInfluentialReliability of Pulse Oximetry During Exercise in Pulmonary Patients
Escourrou, P., Delaperche, M. F., & Visseaux, A. (1990) · CHEST Journal · doi:10.1378/chest.97.3.635 · Cited by 74 · see citing works
Artificial acrylic finger nails may alter pulse oximetry measurement
Hinkelbein, J., Koehler, H., Genzwuerker, H. V., & Fiedler, F. (2007) · Resuscitation · doi:10.1016/j.resuscitation.2006.11.018 · Cited by 69 · see citing works
Swigris, J. J., Zhou, X., Wamboldt, F. S., du Bois, R. M., Keith, R., Fischer, A., Cosgrove, G. P., Frankel, S. K., Curran‐Everett, D., & Brown, K. K. (2009) · Thorax · doi:10.1136/thx.2008.111393 · Cited by 65 · see citing works
Alexandre, F., Héraud, N., Tremey, E., et al. (2020) · BMC Pulmonary Medicine · doi:10.1186/s12890-019-1042-0 · Cited by 61 · see citing works
GABRIELCZYK, M. R., & BUIST, R. J. (1988) · Anaesthesia · doi:10.1111/j.1365-2044.1988.tb09025.x · Cited by 59 · see citing works
Accuracy and reliability of pulse oximetry at different arterial carbon dioxide pressure levels
Muñoz, X., Torres, F., Sampol, G., Ríos, J., Martí, S., & Escrich, E. (2008) · European Respiratory Journal · doi:10.1183/09031936.00126507 · Cited by 57 · see citing works
Misconceptions in Reporting Oxygen Saturation
Toffaletti, J. G., & Zijlstra, W. G. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000278741.29274.e1 · Cited by 57 · see citing works
Accuracy of Pulse Oximeters Intended for Hypoxemic Pediatric Patients
Harris, B. U., Char, D. S., Feinstein, J. A., Verma, A., Shiboski, S. C., & Ramamoorthy, C. (2016) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0000000000000660 · Cited by 57 · see citing works
A framework for the meta‐analysis of Bland–Altman studies based on a limits of agreement approach
Tipton, E., & Shuster, J. J. (2017) · Statistics in Medicine · doi:10.1002/sim.7352 · Cited by 57 · see citing works
Inaccuracy of pulse oximetry in patients with severe tricuspid regurgitation
STEWART, K. G., & Rowbottom, S. J. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09720.x · Cited by 55 · see citing works
McGovern, J., Sasse, S. A., Stansbury, D. W., Causing, L. A., & Light, R. W. (1996) · CHEST Journal · doi:10.1378/chest.109.5.1151 · Cited by 55 · see citing works
Harskamp, R. E., Bekker, L., Himmelreich, J. C. L., Clercq, L. D., Karregat, E. P. M., Sleeswijk, M. E., & Lucassen, W. A. M. (2021) · BMJ Open Respiratory Research · doi:10.1136/bmjresp-2021-000939 · Cited by 55 · see citing works
Accuracy of pulse oximetry in pigmented patients.
Bothma, P. A., Joynt, G. M., Lipman, J., Hon, H., Mathala, B., Scribante, J., & Kromberg, J. G. R. (1996) · PubMed · Cited by 50 · see citing works
Pilcher, J., Ploen, L., McKinstry, S., Bardsley, G., Chien, J., Howard, L., Lee, S., Beckert, L., Swanney, M. P., Weatherall, M., et al. (2020) · BMC Pulmonary Medicine · doi:10.1186/s12890-019-1007-3 · Cited by 50 · see citing works
Accuracy of Pulse Oximetry During Hypoxemia
Thrush, D., & HODGES, M. R. (1994) · Southern Medical Journal · doi:10.1097/00007611-199404000-00019 · Cited by 49 · see citing works
Wiles, M. D., El‐Nayal, A., Elton, G., Malaj, M., Winterbottom, J., Gillies, C., Moppett, I. K., & Bauchmuller, K. (2022) · Anaesthesia · doi:10.1111/anae.15581 · Cited by 45 · see citing works
Accuracy of a portable pulse oximeter in monitoring hypoxemic infants with cyanotic heart disease
Harris, B. U., Stewart, S. E., Verma, A., Hoen, H. M., Stein, M. L., Wright, G. E., & Ramamoorthy, C. (2019) · Cardiology in the Young · doi:10.1017/s1047951119001355 · Cited by 37 · see citing works
Smith, R., & Hofmeyr, R. (2019) · South African Medical Journal · doi:10.7196/samj.2019.v109i3.13633 · Cited by 34 · see citing works
Lee, K. H., Hui, K. P., Tan, W. C., & Lim, T. K. (1993) · PubMed · Cited by 30 · see citing works
Comparison of nasal and forehead oximetry accuracy and pressure injury in critically ill patients
Schallom, M., Prentice, D., Sona, C., Arroyo, C., & Mazuski, J. (2018) · Heart & Lung · doi:10.1016/j.hrtlng.2017.12.002 · Cited by 30 · see citing works
Skin Color Is Not the Same Thing as Race
Silver, S. E. (2004) · Archives of Dermatology · doi:10.1001/archderm.140.3.361-a · Cited by 27 · see citing works
A Cost-Utility Analysis of Remote Pulse-Oximetry Monitoring of Patients With COVID-19
Padula, W. V., Miano, M., Kelley, M. A., Crawford, S. A., Choy, B. H., Hughes, R. M., Grosso, R., & Pronovost, P. J. (2021) · Value in Health · doi:10.1016/j.jval.2021.09.008 · Cited by 27 · see citing works
Wiles, M. D., El‐Nayal, A., Elton, G., Malaj, M., Winterbottom, J., Gillies, C., Moppett, I. K., & Bauchmüller, K. (2022) · Anaesthesia · doi:10.1111/anae.15656 · Cited by 26 · see citing works
Avant, M. G., Lowe, N., & Torres, A. (1997) · Respiratory Care · doi:10.1177/002013249704200705 · Cited by 3 · see citing works
Transcutaneous Oxygen Saturation Accuracy in Critically Ill ChildrenPreprint
Brooks, J. C., Raman, S., GIBBONS, K. S., Williams, T., Dunster, K. R., Pegg, D., Harkins, M., Alphonso, N., Pham, T. M. T., & Schibler, A. (2020) · doi:10.21203/rs.2.21938/v1 · Cited by 2 · see citing works
Flagged first, then by citations
Research works printed but not linked: 2
- 2016, An investigation into the bias of pulse oximeters, Respirology, 21, 6
- 2021, Accuracy of pulse oximetry (SpO2) with different oximeters. Oxygap study, Intensive Care Medicine Experimental, 9, 51
Cited by
Sorted by citations
Berg, K. M., Bray, J., Ng, K.-C., Liley, H. G., Greif, R., Carlson, J. N., Morley, P. T., Drennan, I. R., Smyth, M. A., Scholefield, B. R., et al. (2023) · Circulation · doi:10.1161/cir.0000000000001179 · Cited by 177 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
A review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119 · see citing works
Berg, K. M., Bray, J., Ng, K.-C., Liley, H. G., Greif, R., Carlson, J. N., Morley, P. T., Drennan, I. R., Smyth, M. A., Scholefield, B. R., et al. (2023) · Resuscitation · doi:10.1016/j.resuscitation.2023.109992 · Cited by 117 · see citing works
Consumer Wearable Health and Fitness Technology in Cardiovascular Medicine
Petek, B. J., Al‐Alusi, M. A., Moulson, N., Grant, A. J., Besson, C., Guseh, J. S., Wasfy, M. M., Grémeaux, V., Churchill, T. W., & Baggish, A. L. (2023) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2023.04.054 · Cited by 106 · see citing works
Dcosta, J. V., Ochoa, D., & Sanaur, S. (2023) · Advanced Science · doi:10.1002/advs.202302752 · Cited by 86 · see citing works
Graham, H., King, C., Rahman, A. E., Kitutu, F. E., Greenslade, L., Aqeel, M., Baker, T., Brito, L. F. D. M., Campbell, H. J., L, K. C., et al. (2025) · The Lancet Global Health · doi:10.1016/s2214-109x(24)00496-0 · Cited by 88 · see citing works
Martín, D., Johns, C., Sorrell, L., Healy, E., Phull, M., Olusanya, O., Peters, M. J., & Fabes, J. (2024) · British Journal of Anaesthesia · doi:10.1016/j.bja.2024.01.023 · Cited by 80 · see citing works
Validating cuffless continuous blood pressure monitoring devices
Hu, J., Martin, G., Iyengar, S., Kovell, L. C., Plante, T. B., van Helmond, N., Dart, R. A., Brady, T. M., Turkson‐Ocran, R., & Juraschek, S. P. (2023) · Cardiovascular Digital Health Journal · doi:10.1016/j.cvdhj.2023.01.001 · Cited by 74 · see citing works
Cao, D. J., Arens, A. M., Chow, S. L., Easter, S. R., Hoffman, R. S., Lagina, A. T., Lavonas, E. J., Patil, K. D., Sutherland, L. D., Tijssen, J. A., et al. (2025) · Circulation · doi:10.1161/cir.0000000000001380 · Cited by 64 · see citing works
2024 American Heart Association and American Red Cross Guidelines for First Aid
Brumberg, E. K. H., Douma, M. J., Alibertis, K., Charlton, N. P., Goldman, M. P., Harper-Kirksey, K., Hawkins, S. C., Hoover, A. V., Kule, A., Leichtle, S. W., et al. (2024) · Circulation · doi:10.1161/cir.0000000000001281 · Cited by 72 · see citing works
Leeb, G., Auchus, I., Law, T. J., Bickler, P. E., Feiner, J., Hashi, S., Monk, E. P., Igaga, E. N., Bernstein, M., Chou, Y., et al. (2024) · EBioMedicine · doi:10.1016/j.ebiom.2024.105051 · Cited by 43 · see citing works
Dezfulian, C., McCallin, T., Bierens, J., Dunne, C. L., Idris, A. H., Kiragu, A. W., Mahgoub, M., Shenoi, R., Szpilman, D., Terry, M., et al. (2024) · Circulation · doi:10.1161/cir.0000000000001274 · Cited by 39 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
“To reduce heterogeneity, this meta-analysis reclassified race, ethnicity, and skin tone into a universal schema for skin pigmentation based on the system used by Shi et al [36].”
InfluentialNinety years of pulse oximetry: history, current status, and outlook
Quaresima, V., Ferrari, M., & Scholkmann, F. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33307 · Cited by 28 · see citing works
Khor, Y. H., Poberezhets, V., Buhr, R. G., Chalmers, J. D., Choi, H., Fan, V. S., George, M., Holland, A. E., Pinnock, H., Ryerson, C. J., et al. (2024) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.202410-2080st · Cited by 28 · see citing works
The critical role of technologies in neonatal care
Taha, S., Simpson, R., & Sharkey, D. (2023) · Early Human Development · doi:10.1016/j.earlhumdev.2023.105898 · Cited by 25 · see citing works
European Resuscitation Council Guidelines 2025 First Aid
Djärv, T., Rogers, J., Semeraro, F. F., Brädde, L., Cassan, P., Cimpoesu, D. C., van Goor, S., Klaassen, B., Laermans, J., Meyrán, D., et al. (2025) · Resuscitation · doi:10.1016/j.resuscitation.2025.110752 · Cited by 28 · see citing works
Parr, N. J., Beech, E. H., Young, S., & Valley, T. S. (2024) · Journal of General Internal Medicine · doi:10.1007/s11606-024-08852-1 · Cited by 21 · see citing works
The Case for Globalizing Fairness: A Mixed Methods Study on Colonialism, AI, and Health in Africa
Asiedu, M., Dieng, A., Haykel, I., Rostamzadeh, N., Pfohl, S., Nagpal, C., Nagawa, M., Oppong, A., Koyejo, S., & Heller, K. (2024) · ACM Conference on Equity and Access in Algorithms, Mechanisms, and Optimization (EAAMO) · doi:10.1145/3689904.3694708 · Cited by 26 · see citing works
Utility of Skin Tone on Pulse Oximetry in Critically Ill Patients: A Prospective Cohort Study
Hao, S., Dempsey, K., de Matos, J. C. R. G., Cox, C. E., Rotemberg, V. M., Gichoya, J. W., Kibbe, W. A., Hong, C., & Wong, A.-K. I. (2024) · Critical Care Explorations · doi:10.1097/cce.0000000000001133 · Cited by 18 · see citing works
Consumer Wearable Sleep Trackers
Chiang, A. A.-P., & Khosla, S. (2023) · Sleep Medicine Clinics · doi:10.1016/j.jsmc.2023.05.005 · Cited by 19 · see citing works
Gueye‐Ndiaye, S., & Redline, S. S. (2024) · Annual Review of Medicine · doi:10.1146/annurev-med-070323-103130 · Cited by 21 · see citing works
Herbst, A., Goel, S., Beane, A., Brotherton, B. J., Dula, D., Ely, E. W., Gordon, S. B., Haniffa, R., Hedt‐Gauthier, B., Limbani, F., et al. (2023) · Frontiers in Medicine · doi:10.3389/fmed.2023.1148334 · Cited by 17 · see citing works
Determining factors of pulse oximetry accuracy: a literature review
Valladares, D. L., Mateu, L. A. B., Cortés-Carmona, N., Fuentes-Lizana, G., Cabanas, A. M., Latorre-Progulakis, K., Fuentes‐Guajardo, M., & Martín-Escudero, P. (2024) · Revista Clínica Española (English Edition) · doi:10.1016/j.rceng.2024.04.005 · Cited by 18 · see citing works
Pulse oximetry in pediatric care: Balancing advantages and limitations
Al‐Beltagi, M., Saeed, N. K., Bediwy, A. S., & Elbeltagi, R. M. (2024) · World Journal of Clinical Pediatrics · doi:10.5409/wjcp.v13.i3.96950 · Cited by 19 · see citing works
Identifying and reducing barriers to having a racially and ethnically diverse research population
Harris, M. T., Statler, M. C., Montague, E. N. H., Oduor, K. F., & Stonewall, J. H. (2023) · International Journal of Industrial Ergonomics · doi:10.1016/j.ergon.2023.103527 · Cited by 13 · see citing works
Garnet, B., Lankenau, R. D., Jean, E., & Campos, M. (2023) · Annals of the American Thoracic Society · doi:10.1513/annalsats.202209-837oc · Cited by 15 · see citing works
Pulse Oximeter Bias and Inequities in Retrospective Studies––Now What?
Moore, K., Gudelunas, K., Lipnick, M. S., Bickler, P. E., & Hendrickson, C. M. (2022) · Respiratory Care · doi:10.4187/respcare.10654 · Cited by 14 · see citing works
Haxha, S., Nwibor, C., Ali, M., Sakel, M., Saunders, K., Dyo, V., & Nabakooza, S. (2024) · Sensors · doi:10.3390/s24113301 · Cited by 12 · see citing works
Dual-ratio approach to pulse oximetry and the effect of skin tone
Blaney, G. P., Frias, J., Tavakoli, F., Sassaroli, A., & Fantini, S. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33311 · Cited by 12 · see citing works
Wearable Prophylaxis Tool for AI-Driven Identification of Early Warning Patterns of Pressure Ulcers
Grünerbel, L., Heinrich, F., Böhlhoff-Martin, J., Röper, L., Machens, H., Gruenerbel, A., Schillinger, M., Kist, A. M., Wenninger, F., Richter, M., et al. (2023) · Bioengineering · doi:10.3390/bioengineering10101125 · Cited by 11 · see citing works
Skin pigmentation bias in regional brain oximetry measurements?
Quaresima, V., Scholkmann, F., & Ferrari, M. (2023) · Critical Care · doi:10.1186/s13054-022-04295-4 · Cited by 10 · see citing works
Blaney, G. P., Sassaroli, A., & Fantini, S. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33313 · Cited by 10 · see citing works
Al-Halawani, R., Qassem, M., & Kyriacou, P. A. (2025) · Sensors · doi:10.3390/s25020372 · Cited by 11 · see citing works
Elia, A., & Lemaître, F. (2025) · European Journal of Applied Physiology · doi:10.1007/s00421-025-05752-y · Cited by 12 · see citing works
Assessment of skin pigmentation-related bias in pulse oximetry readings among adults
Khanna, A. K., Beard, J., Lamminmäki, S., Närväinen, J., Antaki, N., & Yapici, H. O. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-023-01095-1 · Cited by 12 · see citing works
Factores determinantes de la precisión de la oximetría de pulso: revisión bibliográfica
Valladares, D. L., Mateu, L. A. B., Cortés-Carmona, N., Fuentes-Lizana, G., Cabanas, A. M., Latorre-Progulakis, K., Fuentes‐Guajardo, M., & Martín-Escudero, P. (2024) · Revista Clínica Española · doi:10.1016/j.rce.2024.03.001 · Cited by 11 · see citing works
Jones, G. A. L., Wiegand, M., Ray, S., Gould, D. W., Agbeko, R., Giallongo, E., Charles, W. N., Orzol, M., O’Neill, L., Lampro, L., et al. (2024) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0000000000003583 · Cited by 12 · see citing works
A Hyperspectral Space of Skin Tones for Inverse Rendering of Biophysical Skin Properties
Aliaga, C., Xia, M., Xie, H., Jarabo, A., Braun, G. J., & Hery, C. (2023) · Computer Graphics Forum · doi:10.1111/cgf.14887 · Cited by 9 · see citing works
Oxygen Extraction Ratios to Guide Red Blood Cell Transfusion
Hess, A. S. (2024) · Transfusion Medicine Reviews · doi:10.1016/j.tmrv.2024.150834 · Cited by 9 · see citing works
Lynn, H., & Wilson, M. A. (2023) · Pulmonary Therapy · doi:10.1007/s41030-022-00213-9 · Cited by 8 · see citing works
Subhi, R., McLeod, L., Ayede, A. I., Dedeke, I. O. F., Risikat, Q., Alao, R. A., Fasasi, A. B., Bakare, A. A., Adeniyi, O. H., Akinrinoye, O. O., et al. (2025) · The Lancet Global Health · doi:10.1016/s2214-109x(24)00458-3 · Cited by 8 · see citing works
Lo, S. B., Svensson, A. D., Presley, C. J., & Andersen, B. L. (2023) · Palliative & Supportive Care · doi:10.1017/s1478951523000640 · Cited by 8 · see citing works
Discrimination by Design: Is It Time to Recalibrate Interpretation of Pulse Oximetry?*
Barreto, J. A., & Moynihan, K. M. (2023) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0000000000003229 · Cited by 8 · see citing works
Skin Phototype Classification with Machine Learning Based on Broadband Optical Measurements
Yu, X., Ong, K. G., & McGeehan, M. A. (2024) · Sensors · doi:10.3390/s24227397 · Cited by 8 · see citing works
Non-Invasive Hemodynamic Monitoring in Critically Ill Patients: A Guide for Emergency Physicians
Beltrame, M., Bellan, M., Patrucco, F., & Gavelli, F. (2025) · Journal of Clinical Medicine · doi:10.3390/jcm14197002 · Cited by 12 · see citing works
James, A., Petit, M., Biancale, F., Bouglé, A., Degos, V., Monsel, A., Vieillard‐Baron, A., & Constantin, J. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-022-00959-2 · Cited by 7 · see citing works
The digital divide in cardiovascular care: who gets left behind?
Hon, J. J., & Carr‐White, G. (2026) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztag011 · Cited by 9 · see citing works
The benefits and drawbacks of home oxygen therapy for COPD: what’s next?
Khor, Y. H., & Ekström, M. P. (2024) · Expert Review of Respiratory Medicine · doi:10.1080/17476348.2024.2379459 · Cited by 5 · see citing works
Yuri, E., Chung, H.-Y., & Chen, F. (2025) · Frontiers in Physiology · doi:10.3389/fphys.2025.1667238 · Cited by 5 · see citing works
Law, T. J., Leeb, G., van Wijk, X. M. R., Kopotic, R. J., Hendrickson, C. M., Behnke, E., Bickler, P. E., Feiner, J., Zamora, R. V., & Lipnick, M. S. (2025) · Respiratory Care · doi:10.1177/19433654251376285 · Cited by 5 · see citing works
Polarimetric BSSRDF Acquisition of Dynamic Faces
Ha, H., Hwang, I., Monzon, N., Cho, J.-S., Kim, D., Baek, S., Muñoz, A., Gutiérrez, D., & Kim, M. H. (2024) · ACM Transactions on Graphics · doi:10.1145/3687767 · Cited by 5 · see citing works
Health Implications of Colorism: A Narrative Review of the Literature
Patterson, J., & Grob, K. L. (2025) · Journal of Racial and Ethnic Health Disparities · doi:10.1007/s40615-025-02369-x · Cited by 5 · see citing works
Proposal of a diagnostic algorithm based on the use of pulse oximetry in obstructive sleep apnea
Borsini, E. E., & Nigro, C. A. (2022) · Sleep And Breathing · doi:10.1007/s11325-022-02757-1 · Cited by 5 · see citing works
Sun, H., Dang, R., Haack, M., Hauser, K., Scott‐Sutherland, J., Westover, M. B., Parthasarathy, S., Redline, S. S., THOMAS, R. J., & Mullington, J. M. (2025) · SLEEP Advances · doi:10.1093/sleepadvances/zpaf017 · Cited by 5 · see citing works
Open Access Dataset and Common Data Model for Pulse Oximeter Performance Data
Fong, N., Lipnick, M. S., Behnke, E., Chou, Y., Elmankabadi, S., Ortiz, L., Almond, C. S. D., Auchus, I., Burnett, G. W., Bisegerwa, R., et al. (2025) · Scientific Data · doi:10.1038/s41597-025-04870-8 · Cited by 6 · see citing works
Biuzzi, C., Modica, E., Filippis, N. D., Pizzirani, D., Galgani, B., Chiaro, A. D., Marianello, D., Franchi, F., Taccone, F. S., & Scolletta, S. (2025) · Diagnostics · doi:10.3390/diagnostics15151930 · Cited by 4 · see citing works
A Continuous and Noninvasive Method to Estimate Pao 2/Fio 2 Ratio
Beauchamp, F.-O., & Sauthier, M. (2024) · Critical Care Explorations · doi:10.1097/cce.0000000000001174 · Cited by 4 · see citing works
Aldana, M. C. S., Matava, C. T., & Blake, C. M. (2025) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000007401 · Cited by 4 · see citing works
Prosperi, P., Verratti, V., Bondi, D., & Spacone, A. (2023) · Respiratory Physiology & Neurobiology · doi:10.1016/j.resp.2023.104111 · Cited by 4 · see citing works
Gaps in patient safety: Areas that need our attention
Wu, A. W., Vincent, C. A., Øvretveit, J., Mair, A., Buckle, P., García‐Elorrio, E., Bellandi, T., Letaief, M., Ushiro, S., & Koizumi, S. (2023) · Journal of Patient Safety and Risk Management · doi:10.1177/25160435231218489 · Cited by 4 · see citing works
Preoperative assessment of patients at risk of postoperative respiratory depression
Assadi, A., Chung, F. F., & Yadollahi, A. (2025) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2025.109805 · Cited by 3 · see citing works
Bias-Mitigated AI as a Foundation for Resilient and Effective Health Systems
Júnior, J. B. D. S., Birminghamm, M., Rivière-Cinnamond, A., Boisson, E., Valdez, M. L., García-Saisó, S., Souza, J. P., Martí, M., Richards, L.-M. L.-M., Guzman, J., et al. (2025) · JMIR Public Health and Surveillance · doi:10.2196/88457 · Cited by 3 · see citing works
Accuracy and Bias of Pulse Oximetry in the Intensive Care Unit: A Prospective Observational Study
de Barros, K. B., Pinheiro, J. S., Sampaio, M. S., da Cruz, L. B., Ferreira, F. F., Basile-Filho, A., Guirro, R. R. D. J., & Bachmann, L. (2026) · Nursing in Critical Care · doi:10.1111/nicc.70433 · Cited by 3 · see citing works
Correlation between periodontal phenotype and sinus membrane thickness: A systematic review
Díaz, L., Fan, S., Urrutia, P., Uriarte, X., Fodich, I., Torres, A., & Sáenz‐Ravello, G. (2023) · Clinical Oral Implants Research · doi:10.1111/clr.14121 · Cited by 3 · see citing works
Peck, R. B., Storey, H. L. G., Barney, B., Israeli, S., Halas, O. R., Oroszlan, D., Brodsky, S., Agarwal, N., Murphy, E., Sagalovsky, M., et al. (2023) · PLOS Global Public Health · doi:10.1371/journal.pgph.0002044 · Cited by 3 · see citing works
Assessing Clinical Deterioration in Children With Dark‐Coloured Skin: A Scoping Review
Kelly, C., Leslie, G. D., Laird, P. J., Stokes, S., & Gill, F. J. (2025) · Journal of Clinical Nursing · doi:10.1111/jocn.70060 · Cited by 3 · see citing works
Katz, S. L., Barwell, T., Bijelić, V., Barrowman, N. J., Blinder, H., Dussah, N., Shamsi, R., Leitman, A. R., & Ersu, R. H. (2025) · Pediatric Pulmonology · doi:10.1002/ppul.71228 · Cited by 3 · see citing works
Correctly Establishing and Interpreting Oxygenation Status in Sickle Cell Disease
Lucas, F. M., Connell, N. T., & Tolan, N. V. (2023) · The Journal of Applied Laboratory Medicine · doi:10.1093/jalm/jfac096 · Cited by 3 · see citing works
Ahmed, W., Hardey, M., Winters, B. D., & Sarwal, A. (2024) · Journal of Medical Internet Research · doi:10.2196/56034 · Cited by 4 · see citing works
Marlar, A. I., Knabe, B. K., Taghikhan, Y., Applegate, R. L., & Fleming, N. (2024) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-024-01211-9 · Cited by 3 · see citing works
Grigoli, L., Marocchi, M., Venditto, L., Piazza, M., Tenero, L., PIACENTINI, G. L., Zaffanello, M., & Ferrante, G. (2025) · Expert Review of Respiratory Medicine · doi:10.1080/17476348.2025.2451960 · Cited by 3 · see citing works
Leeb, G., Auchus, I., Law, T. J., Bickler, P. E., Feiner, J., Hashi, S., Monk, E. P., Igaga, E. N., Bernstein, M., Chou, Y., et al. (2023) · SSRN Electronic Journal · doi:10.2139/ssrn.4578562 · Cited by 3 · see citing works
Who “Needs” Long-Term Oxygen? How Little We Really Know
Casaburi, R. H., & Lacasse, Y. (2025) · Annals of the American Thoracic Society · doi:10.1513/annalsats.202501-101ed · Cited by 3 · see citing works
Harrison, J. (2025) · Paramedicine · doi:10.1177/27536386251335652 · Cited by 3 · see citing works
Georgiev, Y. H., Neunhoeffer, F., Hofbeck, M., & Michel, J. (2024) · Pediatric Anesthesia · doi:10.1111/pan.15017 · Cited by 2 · see citing works
Ruppel, H., Huang, L., Petit, C. J., Qureshi, A. M., Bonafide, C. P., Foglia, E. E., Feldman, J., Lane-Fall, M. B., Lynch, J. M., Chien, A. L.-L., et al. (2025) · BMJ Open · doi:10.1136/bmjopen-2025-100391 · Cited by 2 · see citing works
Improving PAP adherence through consumer wearables: a promising step forward
Sharafkhaneh, A., Benjafield, A., & Penzel, T. (2026) · SLEEP · doi:10.1093/sleep/zsag007 · Cited by 2 · see citing works
Castrejón-Peralta, C., Montiel-Pérez, J. Y., Gante-Díaz, S. A., Vázquez, J. A. C., Rubín-Alvarado, A. A., Reyes-Vera, Z., Torres-Delgadillo, J. M., Sossa, H., Villegas, O. O. V., & Sánchez, V. G. C. (2026) · Applied Sciences · doi:10.3390/app16021126 · Cited by 2 · see citing works
Zhao, Q., Jiang, W., Zhang, C., Ouyang, M., Wu, N., & Guo, Y. (2026) · Frontiers in Public Health · doi:10.3389/fpubh.2026.1863050 · Cited by 2 · see citing works
Sarraf, É. (2024) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-024-01131-8 · Cited by 2 · see citing works
Ramprasad, A., Ezekwe, A., Lee, B. R., Balasubramanian, S., & Jones, B. L. (2024) · Clinical and Translational Science · doi:10.1111/cts.13777 · Cited by 2 · see citing works
Real-Time ECG Analysis with the ArdMob-ECG: A Comparative Assessment
Möller, T. J., Wunderwald, M., & Tünte, M. R. (2024) · Springer eBooks · doi:10.1007/978-3-031-51893-5_6 · Cited by 2 · see citing works
Díaz–González, C. D. L. M., Pérez-Bello, C., la Rosa-Hormiga, M. D., González, J. J., & Reyes-Noha, M. D. L. M. (2024) · Healthcare · doi:10.3390/healthcare12090940 · Cited by 2 · see citing works
Vagedes, J., Sobh, M., Islam, M. O. A., & Poets, C. F. (2024) · Nature and Science of Sleep · doi:10.2147/nss.s460231 · Cited by 2 · see citing works
Yang, T., Liu, Y., Cai, F., Li, Y., & Mudabbar, M. S. (2024) · BMC Anesthesiology · doi:10.1186/s12871-024-02823-z · Cited by 2 · see citing works
Nelson, B. W., Saeb, S., Barman, P., Verma, N., Allen, H., de Zambotti, M., Baker, F. C., Arra, N., Sridhar, N., Sullivan, S. S., et al. (2025) · Journal of Sleep Research · doi:10.1111/jsr.70036 · Cited by 2 · see citing works
Spencer, S. A., Malowa, F., McCarty, D., Joekes, E. C., Phulusa, J., Chinoko, B., Kaimba, S., Keyala, L., Mandala, P., Mkandawire, M., et al. (2025) · Thorax · doi:10.1136/thorax-2025-223623 · Cited by 3 · see citing works
Self-citationIoT-based COVID-19 Realtime Monitoring and Alerting System in Brunei
Hassan, N. F. B. A. H., & Kumar, P. R. (2022) · International Conference on Digital Transformation and Intelligence (ICDI) · doi:10.1109/icdi57181.2022.10007157 · Cited by 2 · see citing works
Promoting inclusivity in nursing education
Mizerek, E. (2023) · Teaching and learning in nursing · doi:10.1016/j.teln.2023.08.007 · Cited by 2 · see citing works
Hooli, S., Colbourn, T., Shah, M. I., Murray, K. O., Mandalakas, A. M., & McCollum, E. D. (2024) · American Journal of Tropical Medicine and Hygiene · doi:10.4269/ajtmh.23-0656 · Cited by 3 · see citing works
EquiOx: A Prospective study of pulse oximeter bias and skin pigmentation in critically-ill adultsPreprint
Hendrickson, C. M., Lipnick, M. S., Chen, D. D. Y., Chen, D., Law, T. J., Pirracchio, R., Feiner, J., Behnke, E., Chou, Y. (., Elmankabadi, S., et al. (2025) · medRxiv · doi:10.1101/2025.10.06.25337217 · Cited by 2 · see citing works
Silverston, V. P. P., Ferrari, M., & Quaresima, V. (2023) · InnovAiT Education and inspiration for general practice · doi:10.1177/17557380231174050 · Cited by 1 · see citing works
Chen, Q.-C., Shen, J., Huang, Y.-L., Kong, R., Xie, Y.-M., & Wang, S. (2024) · Heliyon · doi:10.1016/j.heliyon.2024.e28336 · Cited by 1 · see citing works
Rubinstein, F., Williams, R., Dumville, J. C., Kane, B., William, W., Bower, P., & Kontopantelis, E. (2024) · PLoS ONE · doi:10.1371/journal.pone.0310822 · Cited by 1 · see citing works
Self-citationParent’s experiences of the impact of ethnicity and skin pigmentation on perinatal care.
Fair, F. J., Furness, A., Higginbottom, G. A., Oddie, S., & Soltani, H. (2025) · Ethnicity and Health · doi:10.1080/13557858.2025.2488893 · Cited by 1 · see citing works
Choosing wisely: Improving decision-making for deployment of diagnostics in primary care settings
Shridhar, V., Kannan, V., Prasad, R., & Mohan, P. (2025) · Journal of Family Medicine and Primary Care · doi:10.4103/jfmpc.jfmpc_1455_24 · Cited by 1 · see citing works
Skidmore, K. L., Le, H., Segredo, V., Boullion, J. A., Daniel, C. P., Smith, V., Varrassi, G., Shekoohi, S., & Kaye, A. D. (2025) · Cureus · doi:10.7759/cureus.87774 · Cited by 1 · see citing works
Wang, Y., Shu, Z., Zhang, J., Zhang, L., Yang, Q., He, Y., & Sun, G. (2025) · European journal of medical research · doi:10.1186/s40001-025-03223-w · Cited by 1 · see citing works
- Authors
- Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., Lightbody, C. E., Dark, P., & Cullum, N.
- Affiliations
- Manchester Academic Health Science Centre (GB)
- University of Liverpool (GB)
- University of Lancashire (GB)
- University of Manchester (GB)
- Corresponding author
- Shi, Chunhu
- ORCID
- Venue
- BMC Medicine
- Publisher
- openRxiv
- Published
- 2022-02-17
- Volume
- 20
- Issue
- 1
- Pages
- 267-267
- ISSN
- 1741-7015 (ISSN-L)
- PMC
- PMC9377806
- OpenAlex
- W4291754627
- Cited by
- 158
- References
- 68
- Citations by year
- 2026: 32 · 2025: 46 · 2024: 41 · 2023: 33 · 2022: 6
- FWCI
- 2.81
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- BMC Medicine (publishedVersion, cc-by)
- medRxiv (acceptedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Research Explorer (The University of Manchester) (publishedVersion, cc-by)
- CLOK (University of Central Lancashire) (acceptedVersion)
- Research Explorer (The University of Manchester) (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Edge Hill University Research Information Repository (Edge Hill University) (submittedVersion, cc-by)
- Funders
- National Institute for Health and Care Research
- Department of Health and Social Care
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Chronic Obstructive Pulmonary Disease (COPD) Research
- Keywords
- pulse oximetry, oxygen saturation, skin pigmentation, measurement accuracy, co-oximetry, arterial oxygen saturation, SpO2, measurement bias, COVID-19 pandemic
- MeSH terms
- COVID-19, Oxygen Saturation, Humans, Oximetry, Oxygen, Skin Pigmentation, Pandemics
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review