Sex‐ and site‐dependent variations in the thickness and mechanical properties of human skin in vivo
Abstract
Synopsis A new method for the in vivo characterization of the physical properties of skin is presented. This comprises an ultrasound device to measure the vertical displacement of the surface of the skin, as well as its thickness and that of the hypodermis under suction. In combination with this, a mathematical model is used to calculate the following skin parameters: Young’s modulus, the initial stress and an index of non‐elasticity. These parameters were evaluated from the volar forearm and the forehead of 30 male and 30 females, of similar ages (28 ± 6‐years‐old). The sensitivity of the testing procedure, allowing the characterization of the mechanical parameters of the skin, easily differentiated these two sites, and in some cases, differences between women and men were demonstrated. The main results showed for both sexes that the thickness (P = 0.0001), Young’s modulus (P = 0.0001), and the index of non‐elasticity (P = 0.0001) were greater for the forehead than for the ventral forearm, but that the initial stress was lower (P = 0.0001). The results show that the skin is thicker, stiffer and less tense and elastic on the forehead than on the ventral forearm, suggesting structural differences between these two sites (collagen fibre network, elastic fibres, epidermis, stratum corneum, microvascularization, actinic damage, presence of sebaceous glands, etc.). It is hoped that this device will be useful for the evaluation of certain skin disorders (scleroderma, Ehlers–Danlos syndrome, cutis laxa, oedema, etc.) and their therapy, as well as being a useful tool in skin ageing and cosmetic product assessment. Résumé Nous présentons un nouveau dispositif qui permet in vivo de caractériser les propriétés mécaniques de la peau. Ce système est composé d’une part, d’un appareillage ultrasonore qui permet de mesurer le déplacement vertical de la surface cutanée, et les épaisseurs de l’épiderme‐derme et de l’hypoderme sollicités par une dépression, et d’autre part, d’un modèle biomécanique qui associe la peau à une membrane élastique tendue. Ce dispositif permet de déterminer l’épaisseur cutanée, le module d’Young, la contrainte initiale et un indice de non élasticité de la peau. Ces paramètres ont étéétudiés sur la face antérieure de l’avant‐bras, et sur le front de deux groupes de 30 sujets des deux sexes et de même classe d’âge (28 ± 6 ans). La grande sensibilité de la méthode a permis de différencier nettement ces deux sites et dans certains cas de montrer des différences entre les femmes et les hommes. Les résultats les plus significatifs montrent pour les femmes comme pour les hommes, que l’épaisseur cutanée (P = 0.0001), le module d’Young (P = 0.0001), et l’indice de nonélasticité (P = 0.0001) sont plus importants sur le front que sur la face antérieure de l’avant‐bras, alors que la contrainte initiale est plus faible (P = 0.0001). Ces résultats montrent que le tissu cutané sur le front est plus épais, plus raide, moins élastique et moins tendu que sur l’avant‐bras face antérieure, suggérant des differences de structures (réseaux de fibres de collagène, fibres élastiques, épiderme, stratum corneum, micro‐vascularisation…) entre ces deux sites. On espère maintenant, avec un tel dispositif étudier certaines pathologies cutanées (maladies scléreuses, syndrome d’Ehler–Danlos, cutis laxas, maladies oedemateuses, etc.), leur thérapie, le vieillissement cutané, et l’efficacité des produits cosmétiques.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“the modulii of the dermis and subcutaneous tissue range from 56–260 kPa and 0.12–23 kPa, respectively,165–168”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [167]
Citations
Cites
Biomechanics. Mechanical Properties of Living Tissues
Fung, Y. C., & Skalak, R. (1982) · Journal of Applied Mechanics · doi:10.1115/1.3162171 · Cited by 6,040
Biomechanics: Mechanical Properties of Living Tissues, 2nd ed.
Fung, Y. C., & Cowin, S. C. (1994) · Journal of Applied Mechanics · doi:10.1115/1.2901550 · Cited by 745
Mechanical properties and Young's modulus of human skin in vivo
Agache, P. G., Monneur, C., Lévêque, J., & de Rigal, J. (1980) · Archives of Dermatological Research · doi:10.1007/bf00406415 · Cited by 742
Age-Related Mechanical Properties of Human Skin: An In Vivo Study.
Escoffier, C., de Rigal, J., Rochefort, A., Vasselet, R., Lévêque, J.-L., & Agache, P. G. (1989) · Journal of Investigative Dermatology · doi:10.1016/0022-202x(89)90058-4 · Cited by 492
STRUCTURE AND FUNCTION OF MAMMALIAN TENDON
Elliott, D. H. (1965) · Biological reviews/Biological reviews of the Cambridge Philosophical Society · doi:10.1111/j.1469-185x.1965.tb00808.x · Cited by 392
Elastic properties of human skin: relation to age, sex, and anatomical region
Cua, A. B., Wilhelm, K., & Maibach, H. I. (1990) · Archives of Dermatological Research · doi:10.1007/bf00375720 · Cited by 383
In vivo model of the mechanical properties of the human skin under suction
Diridollou, S., Patat, F., Gens, F., Vaillant, L., Black, D., Lagarde, J., Gall, Y., & Berson, M. (2000) · Skin Research and Technology · doi:10.1034/j.1600-0846.2000.006004214.x · Cited by 350
Self-citationSkin thickness measurement by pulsed ultrasound: its reproducibility, validation and variability
Tan, C. Y., Statham, B. N., Marks, R., & Payne, P. A. (1982) · British Journal of Dermatology · doi:10.1111/j.1365-2133.1982.tb14702.x · Cited by 346
Determining Skin Thickness with Pulsed Ultra Sound
Alexander, H., & Miller, D. L. (1979) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12530104 · Cited by 342
Age-related changes in the elastic properties and thickness of human facial skin
Takema, Y., Yorimoto, Y., Kawai, M., & Imokawa, G. (1994) · British Journal of Dermatology · doi:10.1111/j.1365-2133.1994.tb04975.x · Cited by 340
Research works printed but not linked: 13
- Dupuytren G., 1835, Traite theorique et pratique des blessures par armes de guerre., Maison Paris, 1, 1
- Langer AK., 1861, Zur anatomie und Physiologie der haut., Sber. Akad. Wiss. Wien, 44, 19
- Bramont C., 1988, Mechanical properties of the skin in Marfan’s syndrome and Ehlers–Danlos syndrome., Bioeng. Skin, 4, 217
- Cook T.The mechanical characterization of human skinin vivo. Thesis Stevens Institute of Technology New Jersey (1975).
- Vassellet R.Etudein vitrodes propriétés mécaniques du stratum corneum humain à partir des essais de traction‐relaxation et fluage: identification de son comportement viscoélastique à un modèle rhéo‐logique non‐linéaire avec écrouissage.Thesis University Franche Comté Besançon France (1987).
- Lévêque J.L, 1987, In vivo methods for measuring the viscoelastic properties of the skin., Bioeng. Skin, 3, 375
- Vogel H.G., 1987, Age dependence of mechanical and biochemical properties of human skin. Part II. Hysteresis, relaxation, creep and repeated strain experiments., Bioeng. Skin, 3, 141
- Panisset F.Le stratum corneum: sa place dans la fonsction mécanique de la peau humainein vivo. Thesis Université de Franche‐Comté Faculté de médecine et de Pharmacie Besançon France (1992).
- Diridollou S. Patat F.&Gens Fet al.. In vivomodel of the mechanical properties of the skin under suction.Skin Res. Tech.in press.
- Feuillard G., 1994, High resolution B‐Scan imaging of the skin using a 50 Mhz P(VDF‐TrFE)‐based ultrasonic transducer., Eur. J. Ultrasound, 1, 183
- Escoffier C., 1986, In vitro study of the velocity of ultrasound in the skin., Bioeng. Skin, 2, 87
- Canfield T.R.&Dobrin P.B.Static elastic properties of blood vessels.Handbook of bioengineering(R. Skalak and S. Chien eds) pp.1–16.28.McGraw‐Hill New York (1987).
- Finlay J.B., 1971, The torsional characterictics of human skin in vivo., J. Biomed. Eng, 6, 567
Cited by
Sorted by citations
Lab-on-Skin: A Review of Flexible and Stretchable Electronics for Wearable Health Monitoring
Liu, Y., Pharr, M., & Salvatore, G. A. (2017) · ACS Nano · doi:10.1021/acsnano.7b04898 · Cited by 1,725
Dagdeviren, C., Shi, Y., Joe, P., Ghaffari, R., Balooch, G., Usgaonkar, K., Gur, O., Tran, P. L., Crosby, J. R., Meyer, M., et al. (2015) · Nature Materials · doi:10.1038/nmat4289 · Cited by 513
Biomechanical Properties of In Vivo Human Skin From Dynamic Optical Coherence Elastography
Liang, X., & Boppart, S. A. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2033464 · Cited by 470
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347
Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500317 · Cited by 125
Metadata
- Authors
- Diridollou, S., Black, D., Lagarde, J. M., Gall, Y., Berson, M., Vabre, V., Patat, F., & Vaillant, L.
- Affiliations
- Inserm (FR)
- Centre Hospitalier Universitaire de Tours (FR)
- Corresponding author
- Diridollou, S.
- Venue
- International Journal of Cosmetic Science
- Publisher
- Wiley
- Published
- 2000-12
- Volume
- 22
- Issue
- 6
- Pages
- 421-435
- ISSN
- 0142-5463 (ISSN-L)
- 1468-2494
- OpenAlex
- W2014197576
- Cited by
- 99
- References
- 58
- Citations by year
- 2026: 2 · 2025: 5 · 2023: 9 · 2022: 4 · 2021: 4 · 2020: 3 · 2019: 5 · 2018: 3 · 2017: 12 · 2016: 3 · 2015: 4 · 2014: 4 · 2013: 6 · 2012: 5
- FWCI
- 1.39
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- International Journal of Cosmetic Science (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Elasticity and Material Modeling
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Ultrasound Imaging and Elastography
- Health Sciences › Medicine › Dermatology › Skin Protection and Aging
- Keywords
- skin biomechanics, skin thickness, Young's modulus, skin elasticity, sex differences, frons, volar forearm, elastic membrane model, skin aging, skin diseases, systemic sclerosis, Ehlers-Danlos syndrome, cutis laxa, edema
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review