Biomechanics of the Sensor-Tissue Interface—Effects of Motion, Pressure, and Design on Sensor Performance and the Foreign Body Response—Part I: Theoretical Framework
Open accessThe DOI matches its Crossref record; 151 of the 151 references it cites were found in a registry or an index.
Abstract
The importance of biomechanics in glucose sensor function has been largely overlooked. This article is the first part of a two-part review in which we look beyond commonly recognized chemical biocompatibility to explore the biomechanics of the sensor-tissue interface as an important aspect of continuous glucose sensor biocompatibility. Part I provides a theoretical framework to describe how biomechanical factors such as motion and pressure (typically micromotion and micropressure) give rise to interfacial stresses, which affect tissue physiology around a sensor and, in turn, impact sensor performance. Three main contributors to sensor motion and pressure are explored: Applied forces, sensor design, and subject/patient considerations. We describe how acute forces can temporarily impact sensor signal and how chronic forces can alter the foreign body response and Inflammation around an implanted sensor, and thus impact sensor performance. The importance of sensor design (e.g., size, shape, modulus, texture) and specific implant location on the tissue response are also explored. In Part II: Examples and Application (a sister publication), examples from the literature are reviewed, and the application of biomechanical concepts to sensor design are described. We believe that adding biomechanical strategies to the arsenal of material compositions, surface modifications, drug elution, and other chemical strategies will lead to improvements in sensor biocompatibility and performance.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“At the macro level (>100 μm in size), it is generally held that thicker materials produce a proportionally higher magnitude of foreign body responses and fibrosis3, 21.”
From Size- and shape-dependent foreign body immune response to materials implanted in rodents and non-human primates (2015) · cites this paper as [21]
In the citing paper
“however there is still no consensus on an ideal geometry21.”
From Size- and shape-dependent foreign body immune response to materials implanted in rodents and non-human primates (2015) · cites this paper as [21]
In the citing paper
“The subcutaneous nature of the sensors likely minimized any micromotion effects associated with the foreign body response.32”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [32]
Citations
Cites
1 retracted
Jayant, R. D., McShane, M. J., & Srivastava, R. (2010) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2010.10.035 · Cited by 41
Retracted(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Tissue Cells Feel and Respond to the Stiffness of Their Substrate
Discher, D. E., Janmey, P., & Wang, Y.-L. (2005) · Science · doi:10.1126/science.1116995 · Cited by 6,382
Foreign body reaction to biomaterials
Anderson, J. M., Rodriguez, A., & Chang, D. T. (2008) · Seminars in Immunology · doi:10.1016/j.smim.2007.11.004 · Cited by 5,138
On the mechanisms of biocompatibility
Williams, D. F. (2008) · Biomaterials · doi:10.1016/j.biomaterials.2008.04.023 · Cited by 2,772
Bell, E., Ivarsson, B., & Merrill, C. (1979) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.76.3.1274 · Cited by 2,067
TENSEGRITY: THE ARCHITECTURAL BASIS OF CELLULAR MECHANOTRANSDUCTION
Ingber, D. E. (1997) · Annual Review of Physiology · doi:10.1146/annurev.physiol.59.1.575 · Cited by 1,566
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Biomaterials: Where We Have Been and Where We Are Going
Ratner, B. D., & Bryant, S. J. (2004) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.6.040803.140027 · Cited by 1,507
Self-citationBiological Responses to Materials
Anderson, J. M. (2001) · Annual Review of Materials Research · doi:10.1146/annurev.matsci.31.1.81 · Cited by 1,465
Research works printed but not linked: 15
- Clark LC, 1988, ASAIO Trans., 34, 259
- 10.1177/193229680800200628 · DOI · Paper page
- 10.2337/diacare.23.2.208 · DOI · Paper page
- Morhenn VB, 2002, Dermatol Surg., 28, 484
- 10.1002/(sici)1097-4636(199603)30:3<305::aid-jbm5>3.0.co;2-u · doi:10.1002/(sici)1097-4636(199603)30:3<305::aid-jbm5>3.0.co;2-u
- Von Recum A, 1990, New aspects of biocompatibility: Motion at the interface
- Lambert CA, 1992, Lab Invest., 66, 444
- Sanders JE, 1995, J Rehabil Res Dev., 32, 214
- Kvist PH, 2006, In vivo., 20, 195
- Von Recum AF, 1981, Crit Rev Bioeng., 5, 37
- Marshall AJ, 2004, Polym Prepr., 45, 100
- Perez R, 2008, Wounds., 20, 3
- Morris GM, 1990, Cell Tissue Kinet., 23, 271
- Tan CY, 1982, Br J Dermatol., 106, 657
- Marshall AJ, 2004, Porous hydrogels with well-defined pore structure for biomaterials applications [doctoral dissertation]
Cited by
Veiseh, O., Doloff, J. C., Ma, M., Vegas, A. J., Tam, H. H., Bader, A. R., Li, J., Langan, E., Wyckoff, J., Loo, W. S., et al. (2015) · Nature Materials · doi:10.1038/nmat4290 · Cited by 897
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347
Zhang, D., Chen, Q., Shi, C., Chen, M., Ma, K., Wan, J., & Liu, R. (2021) · Advanced Functional Materials · doi:10.1002/adfm.202007226 · Cited by 303
State-of-the-art MEMS and microsystem tools for brain research
Seymour, J. P., Wu, F., Wise, K. D., & Yoon, E. (2017) · Microsystems & Nanoengineering · doi:10.1038/micronano.2016.66 · Cited by 216
Infiltration of chitin by protein coacervates defines the squid beak mechanical gradient
Tan, Y., Hoon, S., Guerette, P. A., Wei, W., Ghadban, A., Hao, C., Miserez, A., & Waite, J. H. (2015) · Nature Chemical Biology · doi:10.1038/nchembio.1833 · Cited by 207
Metadata
- Authors
- Helton, K. L., Ratner, B. D., & Wisniewski, N. A.
- Affiliations
- University of Washington (US)
- Profusa (United States) (US)
- Corresponding author
- Wisniewski, Natalie A.
- ORCID
- Ratner, Buddy D.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2011-05
- Volume
- 5
- Issue
- 3
- Pages
- 632-646
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3192629
- OpenAlex
- W2102068807
- Cited by
- 125
- References
- 149
- Citations by year
- 2026: 6 · 2025: 7 · 2024: 2 · 2023: 8 · 2022: 9 · 2021: 8 · 2020: 10 · 2019: 9 · 2018: 13 · 2017: 12 · 2016: 11 · 2015: 14 · 2014: 2 · 2013: 8 · 2012: 3
- Type
- conference abstract
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Keywords
- continuous glucose monitoring, glucose sensor, biomechanics, foreign body reaction, inflammation, micromotion, interfacial stress, sensor design, implantation site, biocompatibility, sensor performance
- MeSH terms
- Animals, Biocompatible Materials, Biomechanical Phenomena, Blood Glucose, Diabetes Mellitus, Disease Models, Animal, Electrochemistry, Equipment Design, Foreign-Body Reaction, Humans, Motion, Pressure, Stress, Mechanical, Tensile Strength, Wound Healing, Blood Glucose Self-Monitoring, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Biomechanics of the Sensor-Tissue Interface—Effects of Motion, Pressure, and Design on Sensor Performance and the Foreign Body Response—Part I: Theoretical Framework. https://app.cochat.ai/papers/10.1177/193229681100500317
- The paper
- Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011). Biomechanics of the Sensor-Tissue Interface—Effects of Motion, Pressure, and Design on Sensor Performance and the Foreign Body Response—Part I: Theoretical Framework. Journal of Diabetes Science and Technology. https://doi.org/10.1177/193229681100500317
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.