CoChat

Feasibility of Long‐Term Electrocardiographic Monitoring with an Implanted Device for Syncope Diagnosis

MURDOCK, CHALLON J., KLEIN, GEORGE J., YEE, RAYMOND, LEITCH, JAMES W., TEO, WEE SIONG, NORRIS, CARO · 1990 · Pacing and Clinical Electrophysiology
Cited by 28

The DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.

Abstract

Patients with infrequent recurrent syncope undiagnosed after extensive noninvasive and invasive testing pose a diagnostic and therapeutic dilemma. The purpose of this pilot study was to assess the feasibility of using an implanted, long‐term monitor as an aid to diagnosis in these patients. This was done using commercially available pacemakers with monitoring functions. Sixteen patients (eight males and eight femaiesj, aged 59.7 ± 17 years who had unexplained syncope despite a 12‐lead electrocardiogram, repeated Holter monitoring, exercise testing, echocardiography, an eJectro‐physioJogical study, and a tiJf test (n = 6), were entered into the study. Patients had experienced a mean of 3.3 ± 1 episodes of syncope in the 12 months prior to the study. All provided a history suggestive of Stokes‐Adams attacks and were referred for consideration of pacemaker implantation. Two patients had ischemic heart disease and one patient had a long QT interval. Patients had an Intermedics Nova II or Medtronic Quintech DPG pulse generator capable of recording sensed and paced events implanted with a single right ventricular lead. Syncope or presyncope occurred in ten patients (62%) 4.9 ± 4.2 months after pacemaker implantation. Bradycardia was detected in six patients and four patients had no arrhythmia. In addition to bradycardia, one patient also had tachycardia detected. Pacing therapy resulted in symptom relief in all six patients with syncope or presyncope due to bradycardia. Complications of pacemaker implantation (lead insulation failure) occurred in two patients. One of these patients subsequently had an infection of the generator pocket with associated systemic sepsis. Although the devices used in this study have serious limitations, the results suggest that continuous long‐term electrocardiographic monitoring can be useful in patients with recurrent syncope and a negative diagnostic work‐up. A smaller device with self‐contained electrodes designed for this purpose would be required for wider applicability.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1111/j.1540-8159.1990.tb04011.x

Cites

6 of 6 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 6 references

Cited by

See all 28

Metadata

Authors
MURDOCK, C. J., KLEIN, G. J., YEE, R., LEITCH, J. W., TEO, W. S., & NORRIS, C.
Venue
Pacing and Clinical Electrophysiology
Publisher
Wiley
Published
1990-11
Volume
13
Issue
11
Pages
1374-1378
ISSN
  • 0147-8389 (ISSN-L)
  • 1540-8159
OpenAlex
W2061008133
Cited by
28
References
6
FWCI
3.44
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Feasibility of Long‐Term Electrocardiographic Monitoring with an Implanted Device for Syncope Diagnosis. https://app.cochat.ai/papers/10.1111/j.1540-8159.1990.tb04011.x
The paper
MURDOCK, C. J., KLEIN, G. J., YEE, R., LEITCH, J. W., TEO, W. S., & NORRIS, C. (1990). Feasibility of Long‐Term Electrocardiographic Monitoring with an Implanted Device for Syncope Diagnosis. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/j.1540-8159.1990.tb04011.x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.