Policy Statement of the World Hypertension League on Noninvasive Blood Pressure Measurement Devices and Blood Pressure Measurement in the Clinical or Community Setting
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
Increased blood pressure (BP) is the leading risk factor for death and disability globally,1 with more than 40% of the adult population older than 25 years having hypertension.2 Although much of hypertension is preventable, especially by reducing the amount of salt added to foods, hypertension treatment can also prevent the adverse consequences of stroke, heart attack, and heart and kidney failure.2 Unfortunately, about half of patients with hypertension remain undiagnosed.3 Hence, the World Hypertension League has made the increase in regular BP assessments and encouragement of widespread BP screening programs linked to diagnosis and clinical management of hypertension to be among the highest of priorities. Whether in low-, middle-, or high-resource settings, recommendations for BP assessment are consistent and include a standardized approach to pre-measurement preparation, patient positioning, appropriate cuff selection and placement, measurement technique, and use of accurate BP-measuring devices.4-9 For BP-measuring devices, there is typically a choice between manual devices using the auscultatory technique and either semi-automated (manual inflation) or fully automated (automated inflation) devices using oscillometry.4-8, 10 For decades, the foundation of BP measurement in the community has been manual measurement using the auscultatory technique.4, 5 However, in spite of numerous recommendations on how to perform manual BP measurement accurately, even with training, it is difficult to ensure quality measurements. Moreover, the recommendations are rarely followed outside of high-quality research trials. In fact, a large body of evidence indicates that BP assessments are frequently inaccurate, with a high potential to misdiagnose a large segment of the population.4, 11-19 In particular, manual measurement of BP requires good hearing and extensive training in interpreting Korotkoff sounds.4, 20, 21 Even with training, the use of standardized manual techniques declines rapidly without regular retraining and accuracy testing. The use of automated devices requires much less training and automates many features of the measurement. During the past decade, there have been many advances in technology that have reduced the costs of semi-automated and fully automated devices, improved their accuracy, and allowed them to be used in settings where there is limited access to replacement batteries and limited electricity for recharging (eg, solar panels and use of a cell phone charger for recharging).4, 9, 22-25 Nevertheless, challenges remain. There is a need to select a proper cuff size, and several other technical aspects still rely on the observer's training and skill. In many people, automated devices produce readings that differ from readings with manual techniques and, in these settings, it is unclear which method is a better reflection of arterial BP.26-29 Further, regardless of method used, arrhythmias cause BP to be highly variable and multiple readings are required in these patients to increase accuracy. The use of ambulatory BP measurement during routine daily activity and home/self-measurement of BP have documented benefits in assessing patients with increased BP, and their use has been recommended in many different national and international hypertension guidelines.5, 5, 7, 30-32 The use of these methods, however, requires resources and expertise that are not broadly available throughout much of the globe and are not the focus of this policy statement. In addition, devices that measure BP without an observer present in a clinic environment (automated office BP [AOBP]) also have advantages, including a reduced white-coat effect.33-36 However, these devices are more expensive and not globally accessible at this time, and their role in hypertension management is outside the focus of this policy. The World Health Organization (WHO) has recommended the use of a validated semi-automated oscillometric device in the STEPS survey (http://www.who.int/chp/steps/manual/en/), and a WHO expert meeting also recommended these types of devices for clinics in low-resource settings "given serious inherent inaccuracy of the auscultatory technique."2, 4, 15, 25 Automated devices allow training to be shorter and to focus on fewer essential aspects of BP measurement. Critical, however, is the selection of devices that have been independently assessed and have passed international validation testing and the selection of a proper cuff size (refer to http://www.dableducational.org/ for an up-to-date listing of devices that have passed standards). The World Hypertension League has developed this policy statement to stimulate health and scientific organizations to adopt the WHO recommendation broadly, not only in low- but also in middle- and high-resource settings, to work towards the use of these devices for training healthcare professionals and the public, to advocate that all clinical settings have and use these devices for rou
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationMeasurement of Blood Pressure in the Office: Recognizing the Problem and Proposing the Solution
Myers, M. G., Godwin, M., Dawes, M., Kiss, A., Tobe, S. W., & Kaczorowski, J. (2010) · Hypertension · doi:10.1161/hypertensionaha.109.141879 · Cited by 260
Self-citationStergiou, G. S., Lourida, P., Tzamouranis, D., & Baibas, N. M. (2009) · Journal of Human Hypertension · doi:10.1038/jhh.2009.20 · Cited by 74
Self-citationRecommendations for blood pressure measuring devices for office/clinic use in low resource settings
Parati, G., Mendis, S., Abegunde, D., Asmar, R., Mieke, S., Murray, A., Shengelia, B., Steenvoorden, G., Van Montfrans, G., & O??Brien, E. (2005) · Blood Pressure Monitoring · doi:10.1097/00126097-200502000-00002 · Cited by 60
Self-citation
Cited by
Blood Pressure Assessment in Adults in Clinical Practice and Clinic-Based Research
Muntner, P., Einhorn, P. T., Cushman, W. C., Whelton, P. K., Bello, N. A., Drawz, P. E., Green, B. B., Jones, D. W., Juraschek, S. P., Margolis, K. L., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.10.069 · Cited by 199
“The technician or health care provider remains the most important component of accurate and reliable BP measurement, and standardization of training, retraining, and certification at regular intervals are recommended for everyone who measures BP (6,117).”
BackgroundMethodInfluentialSelf-citationPadwal, R., Campbell, N. R. C., Schutte, A. E., Olsen, M. H., Delles, C., Etyang, A., Cruickshank, J. K., Stergiou, G., Rakotz, M. K., Wozniak, G., et al. (2019) · Journal of Hypertension · doi:10.1097/hjh.0000000000002112 · Cited by 108
Self-citationSão Paulo call to action for the prevention and control of high blood pressure: 2020
Campbell, N. R. C., Schutte, A. E., Varghese, C. V., Ordunez, P., Zhang, X., Khan, T., Sharman, J. E., Whelton, P. K., Parati, G., Weber, M. A., et al. (2019) · The Journal of Clinical Hypertension · doi:10.1111/jch.13741 · Cited by 72
Self-citationCampbell, N. R., Khalsa, T., Executive:, W. H. L., Lackland, D. T., Niebylski, M. L., Nilsson, P. M., Redburn, K. A., Orias, M., Zhang, X., International Society of Hypertension Executive:, et al. (2016) · The Journal of Clinical Hypertension · doi:10.1111/jch.12840 · Cited by 40
Self-citationBoonyasai, R. T., Carson, K. A., Marsteller, J. A., Dietz, K. B., Noronha, G. J., Hsu, Y., Flynn, S. J., Charleston, J. M., Prokopowicz, G. P., Miller, E. R., et al. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13166 · Cited by 34
Metadata
- Authors
- Campbell, N. R. C., Berbari, A. E., Cloutier, L., Gelfer, M., Kenerson, J. G., Khalsa, T. K., Lackland, D. T., Lemogoum, D., Mangat, B. K., Mohan, S., Myers, M. G., Niebylski, M. L., O'Brien, E., Stergiou, G. S., VeIga, E. V., & Zhang, X.
- Venue
- The Journal of Clinical Hypertension
- Publisher
- Wiley
- Published
- 2014-05
- Volume
- 16
- Issue
- 5
- Pages
- 320-322
- ISSN
- 1524-6175 (ISSN-L)
- 1751-7176
- OpenAlex
- W1989834723
- Cited by
- 67
- References
- 6
- FWCI
- 5.90
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.