Criteria for the Selection of Materials for Implanted Electrodes
Abstract
There are four criteria that must be considered when choosing material for an implanted electrode: (1) tissue response, (2) allergic response, (3) electrode-tissue impedance, and (4) radiographic visibility. This paper discusses these four criteria and identifies the materials that are the best candidates for such electrodes. For electrodes that make ohmic contact with tissues: gold, platinum, platinum-iridium, tungsten, and tantalum are good candidates. The preferred insulating materials are polyimide and glass. The characteristics of stimulator output circuits and the importance of the bidirectional waveform in relation to electrode decomposition are discussed. The paper concludes with an analysis, the design criteria, and the special properties and materials for capacitive recording and stimulating electrodes.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Good candidates for ohmic electrode-tissue contacts are gold, platinum (Pt), platinum-iridium, tungsten, and tantalum”
From A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems (2005) · cites this paper as Geddes and Roeder, 2003
In the citing paper
“The electrical interface must be stable as a foreign body response over the contacts within excitable tissue will increase the separation between the electrode and the target tissue, leading to a raised stimulation threshold that is proportional to the scar tissue thickness and a drop in recording sensitivities (Geddes and Roeder 2003).”
From Peripheral nerve bionic interface: a review of electrodes (2019) · cites this paper as Geddes and Roeder 2003
In the citing paper
“Encapsulation materials which are conventionally electrical insulators need to be both biocompatible and resistant to bio-dissolution (Geddes and Roeder 2003; Joung 2013).”
From Peripheral nerve bionic interface: a review of electrodes (2019) · cites this paper as Geddes and Roeder 2003
In the citing paper
“Cardiac pacemakers' threshold current is established during surgery, then doubled to account for the expected bioresponse to the implant (Geddes and Roeder 2003).”
From Peripheral nerve bionic interface: a review of electrodes (2019) · cites this paper as Geddes and Roeder 2003
In the citing paper
“The longevity of the implant can be further improved through the operation of the electrodes, stimulating without a DC bias will reduce electro-decomposition of the electrical contacts, whilst bi-phasic pulses with a zero-net-charge have a similar response (Geddes and Roeder 2003).”
From Peripheral nerve bionic interface: a review of electrodes (2019) · cites this paper as Geddes and Roeder 2003
Citations
Cites
Campbell, P. K., Jones, K. E., Huber, R. J., Horch, K. W., & Normann, R. A. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83588 · Cited by 881
Tungsten Microelectrode for Recording from Single Units
Hubel, D. H. (1957) · Science · doi:10.1126/science.125.3247.549 · Cited by 864
Principles of Applied Biomedical Instrumentation.
(1975) · Annals of Internal Medicine · doi:10.7326/0003-4819-83-6-924_2 · Cited by 561
An Integrated-Circuit Approach to Extracellular Microelectrodes
Wise, K. D., Angell, J. B., & Starr, A. (1970) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1970.4502738 · Cited by 544
DETERMINATION OF BIOLOGICAL IMPEDANCES
Schwan, H. P. (1963) · Elsevier eBooks · doi:10.1016/b978-1-4831-6743-5.50013-7 · Cited by 439
Brummer, S. B., & Turner, M. J. (1977) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1977.326218 · Cited by 360
A glass/silicon composite intracortical electrode array
Jones, K. E., Campbell, P. K., & Normann, R. A. (1992) · Annals of Biomedical Engineering · doi:10.1007/bf02368134 · Cited by 354
Brief, Noninjurious Electric Waveform for Stimulation of the Brain
Lilly, J. C., Hughes, J. R., Alvord, E. C., & Galkin, T. W. (1955) · Science · doi:10.1126/science.121.3144.468 · Cited by 287
Electrodes and the measurement of bioelectric events
Geddes, L. A. (1972) · Medical Entomology and Zoology · Cited by 276
Self-citationHistorical evolution of circuit models for the electrode-electrolyte interface
Geddes, L. A. (1997) · Annals of Biomedical Engineering · doi:10.1007/bf02738534 · Cited by 222
Self-citation
Cited by
Sorted by citations
Items 201–240 of 240
Guimarães, J. S., Lemos, N. A. M., Freire, M. A. M., Pereira, A., & Ribeiro, S. T. G. (2022) · Experimental Brain Research · doi:10.1007/s00221-022-06434-3 · Cited by 2
Cortical surface recurrent brain-computer interfaces
Zanos, S. P. (2013) · ResearchWorks at the University of Washington (University of Washington) · Cited by 1
DESIGN, FABRICATION, AND CHARACTERIZATION OF MICROELECTRODES FOR BRAIN-MACHINE INTERFACES By
Patrick, E. E., & Patrick, C. E. (2010) · Cited by 1
Characterisation of wearable and implantable physiological measurement devices
Riistama, J. M. (2010) · Tampere University Institutional Repository (Tampere University) · Cited by 1
L methodology for high precision fabrication
Dario, P., Menciassi, A., Pensabene, V., & Raffa, V. (2006) · IEEE/RAS-EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob.2006.1639131
Gut, N. K. (2014) · St Andrews Research Repository (St Andrews Research Repository)
Rastogi, A. (2009) · Texas ScholarWorks (Texas Digital Library)
Surface modification of recording electrodes
Pereira, I. M., Carvalho, S. M., Oréfice, R. L., Fontes, M. B. A., Arantes, L. M., Figueiró, N. C. G., Leite, M. F., & Neves, H. P. (2013) · Polímeros · doi:10.4322/polimeros.2014.009
Chen, Y., Chen, B., Tsai, D., Huang, C., Yu, J., Shih, W., & Lin, C. (2014) · Micro & Nano Letters · doi:10.1049/mnl.2014.0447
The Enhancement of a Chronically Implanted Microwire Electrode Performance
Shi, L., Chen, J., & Wang, S. W. (2014) · Advanced materials research · doi:10.4028/www.scientific.net/amr.960-961.689
Mapping of Eloquent Cortex in Focal Epilepsy with Intracranial Electrodes
Kanner, A. M., & Morcillo, M. C. A. (2015) · Springer eBooks · doi:10.1007/978-3-319-23383-3_7
Method of Fabrication for Nerve Cuff Electrodes for Use in Animal Models
Sanner, B. M. (2015) · SMARTech Repository (Georgia Institute of Technology)
Fabrication and testing of a multi-electrode spiral nerve cuff
Pečlin, P., & Rozman, J. (2017) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-981-10-4166-2_17
Study of kernel machines towards Brain-computer Interfaces
Tian, X. (2012) · HAL
Surface characterization of platinum stimulating electrodes using an electrochemical scanning method
Mehle, A., Rozman, J., Šala, M., Ribarič, S., & Pečlin, P. (2017) · Materiali in tehnologije · doi:10.17222/mit.2017.044
Arhatari, B. D., & Abbey, B. (2018) · Nanophotonics Australasia · doi:10.1117/12.2282937
An insect-computer hybrid walking robot
Cao, F. (2018) · doi:10.32657/10356/74213
Direct Probing of Neural Stem Cell Interactions on Hydrogel Electrodes using Atomic Force Microscopy
Puckert, C. (2018) · Research Online (University of Wollongong)
A system for the multi-channel transcutaneous stimulation of the auricular nerve
Rozman, J., Bartolić, A., Maiga, A. I., & Ribarič, S. (2019) · Materiali in tehnologije · doi:10.17222/mit.2018.233
Shoffstall, A. J., & Capadona, J. R. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816137-1.00073-8
Probing nanoscale properties of organic conducting polymer interfaces using atomic force microscopy
Gelmi, A. A. (2012) · Research Online (University of Wollongong)
Design of Low Cost Bio-impedance Measuring Instrument
Birok, R., & Kapoor, R. (2022) · International Journal of Advanced Computer Science and Applications · doi:10.14569/ijacsa.2022.0130186
McLean, L., & Reeves, J. E. (2022) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2022.102661
López-Castro, A., Ángeles-Valdéz, D., Rojas‐Piloni, G., & Garza‐Villarreal, E. A. (2022) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2022.05.16.492128
Bradley, K. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-817000-7.00009-0
Lansdown, A. B. G. (2023) · Wiley eBooks · doi:10.1002/0471125474.tox116
Tashiro, H., Herr, H., Popović, M. B., Iramina, K., Terasawa, Y., & Ohta, J. (2024) · Elsevier eBooks · doi:10.1016/b978-0-443-13862-1.00017-5
Biomaterials for Bionic Implants
Shokuhfar, T. (2025) · Wiley eBooks · doi:10.1002/9781119851004.ch8
Effect of heat treatment on microstructure and mechanical properties of the Pt–10Ir alloy wires
Sha, F., Jiang, L., Fu, Q., Luo, D., Wu, Z., Peng, B., & Yang, J. (2025) · Journal of Materials Research and Technology · doi:10.1016/j.jmrt.2025.10.020
Gholston, A. K., Sanchez, R. A., Bigger, T. B., Averbeck, S. R., Rieth, L. W., Ludwig, K. A., & Trevathan, J. K. (2025) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2025.12.08.691908
Мамлеев, А. Р., Suchkov, D., Malyshev, E. I., Vorobyev, A., Sitdikova, V., Силаева, В. М., Логашкин, А. Е., Kireev, A., Sorokina, M. A., Mitin, D. M., et al. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.03.05.709791
Nanostructuration d'électrodes en interface avec le vivant
Bibari, O. (2012) · theses.fr (ABES) · doi:10.70675/8daf86fbz85fcz4007zb642zf9eb0cb769ed
Study of the role of the lateral habenula in fear conditioning and stress response in rat
Durieux, L. (2021) · theses.fr (ABES) · doi:10.70675/6aef2b06ze4d6z47dezbb31z9f954c98768d
Poulkouras, R. (2022) · theses.fr (ABES) · doi:10.70675/18e3c078z78a1z4e5czbd7dz301fcc213f06
Rozman, J., Bizjak, J., Smrkolj, T., Ribarič, S., & Hawlina, S. (2026) · Research Square · doi:10.21203/rs.3.rs-9440557/v1
Gonzalez, J. G., Spurling, D., O’Connor, C., Caffrey, E., McGuire, T. K., Maughan, J., Carey, T., Munuera, J., Oikonomou, I. M., Carolan, J., et al. (2026) · Small · doi:10.1002/smll.75583
Fahmy, M. A., & Subhi, F. A. (2026) · F1000Research · doi:10.12688/f1000research.183372.1
Metadata
- Authors
- Geddes, L. A., & Roeder, R.
- Affiliations
- Purdue University West Lafayette (US)
- Venue
- Annals of Biomedical Engineering
- Published
- 2003-07
- Volume
- 31
- Issue
- 7
- Pages
- 879-890
- ISSN
- 0090-6964 (ISSN-L)
- OpenAlex
- W1978696561
- Cited by
- 240
- References
- 31
- Citations by year
- 2026: 5 · 2025: 5 · 2024: 12 · 2023: 8 · 2022: 13 · 2021: 18 · 2020: 19 · 2019: 21 · 2018: 20 · 2017: 15 · 2016: 19 · 2015: 11 · 2014: 18 · 2013: 7 · 2012: 8
- FWCI
- 0.92
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Annals of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Medicine › Surgery › Orthopaedic implants and arthroplasty
- Keywords
- implanted electrodes, electrode materials, tissue reaction, allergic reactions, gold, platinum, tungsten, tantalum, polyimide, glass insulators
- MeSH terms
- Biocompatible Materials, Decision Support Techniques, Electrodes, Implanted, Electrophysiology, Equipment Design, Hypersensitivity, Materials Testing, Microelectrodes, Technology Assessment, Biomedical, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review