Criteria for the Selection of Materials for Implanted Electrodes
Abstract
There are four criteria that must be considered when choosing material for an implanted electrode: (1) tissue response, (2) allergic response, (3) electrode-tissue impedance, and (4) radiographic visibility. This paper discusses these four criteria and identifies the materials that are the best candidates for such electrodes. For electrodes that make ohmic contact with tissues: gold, platinum, platinum-iridium, tungsten, and tantalum are good candidates. The preferred insulating materials are polyimide and glass. The characteristics of stimulator output circuits and the importance of the bidirectional waveform in relation to electrode decomposition are discussed. The paper concludes with an analysis, the design criteria, and the special properties and materials for capacitive recording and stimulating electrodes.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Good candidates for ohmic electrode-tissue contacts are gold, platinum (Pt), platinum-iridium, tungsten, and tantalum”
From A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems (2005) · cites this paper as Geddes and Roeder, 2003
In the citing paper
“The electrical interface must be stable as a foreign body response over the contacts within excitable tissue will increase the separation between the electrode and the target tissue, leading to a raised stimulation threshold that is proportional to the scar tissue thickness and a drop in recording sensitivities (Geddes and Roeder 2003).”
From Peripheral nerve bionic interface: a review of electrodes (2019) · cites this paper as Geddes and Roeder 2003
In the citing paper
“Encapsulation materials which are conventionally electrical insulators need to be both biocompatible and resistant to bio-dissolution (Geddes and Roeder 2003; Joung 2013).”
From Peripheral nerve bionic interface: a review of electrodes (2019) · cites this paper as Geddes and Roeder 2003
In the citing paper
“Cardiac pacemakers' threshold current is established during surgery, then doubled to account for the expected bioresponse to the implant (Geddes and Roeder 2003).”
From Peripheral nerve bionic interface: a review of electrodes (2019) · cites this paper as Geddes and Roeder 2003
In the citing paper
“The longevity of the implant can be further improved through the operation of the electrodes, stimulating without a DC bias will reduce electro-decomposition of the electrical contacts, whilst bi-phasic pulses with a zero-net-charge have a similar response (Geddes and Roeder 2003).”
From Peripheral nerve bionic interface: a review of electrodes (2019) · cites this paper as Geddes and Roeder 2003
Citations
Cites
Campbell, P. K., Jones, K. E., Huber, R. J., Horch, K. W., & Normann, R. A. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83588 · Cited by 881
Tungsten Microelectrode for Recording from Single Units
Hubel, D. H. (1957) · Science · doi:10.1126/science.125.3247.549 · Cited by 864
Principles of Applied Biomedical Instrumentation.
(1975) · Annals of Internal Medicine · doi:10.7326/0003-4819-83-6-924_2 · Cited by 561
An Integrated-Circuit Approach to Extracellular Microelectrodes
Wise, K. D., Angell, J. B., & Starr, A. (1970) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1970.4502738 · Cited by 544
DETERMINATION OF BIOLOGICAL IMPEDANCES
Schwan, H. P. (1963) · Elsevier eBooks · doi:10.1016/b978-1-4831-6743-5.50013-7 · Cited by 439
Brummer, S. B., & Turner, M. J. (1977) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1977.326218 · Cited by 360
A glass/silicon composite intracortical electrode array
Jones, K. E., Campbell, P. K., & Normann, R. A. (1992) · Annals of Biomedical Engineering · doi:10.1007/bf02368134 · Cited by 354
Brief, Noninjurious Electric Waveform for Stimulation of the Brain
Lilly, J. C., Hughes, J. R., Alvord, E. C., & Galkin, T. W. (1955) · Science · doi:10.1126/science.121.3144.468 · Cited by 287
Electrodes and the measurement of bioelectric events
Geddes, L. A. (1972) · Medical Entomology and Zoology · Cited by 276
Self-citationHistorical evolution of circuit models for the electrode-electrolyte interface
Geddes, L. A. (1997) · Annals of Biomedical Engineering · doi:10.1007/bf02738534 · Cited by 222
Self-citationBiocompatibility of silicon‐based electrode arrays implanted in feline cortical tissue
Schmidt, S., Horch, K., & Normann, R. A. (1993) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820271106 · Cited by 182
Histopathological evaluation of materials implanted in the cerebral cortex
Stensaas, S. S., & Stensaas, L. J. (1978) · Acta Neuropathologica · doi:10.1007/bf00689766 · Cited by 177
Capacitive Electrocardiographic and Bioelectric Electrodes
López, A., & Richardson, P. C. (1969) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1969.4502613 · Cited by 143
Brain tissue reaction to some chronically implanted metals
Dymond, A. M., Le, K., Jurist, J. M., & Crandall, P. H. (1970) · Journal of neurosurgery · doi:10.3171/jns.1970.33.5.0574 · Cited by 140
The impedance of stainless-steel electrodes
Geddes, L. A., Costa, C., & Wise, G. L. (1971) · Medical & Biological Engineering & Computing · doi:10.1007/bf02474708 · Cited by 112
Self-citationThe polarization impedance of common electrode metals operated at low current density
Ragheb, T., & Geddes, L. A. (1991) · Annals of Biomedical Engineering · doi:10.1007/bf02368466 · Cited by 87
Self-citationHistopathological Changes Produced by Implanted Electrodes in Cat Brains
Collias, J. C., & Manuelidis, E. E. (1957) · Journal of neurosurgery · doi:10.3171/jns.1957.14.3.0302 · Cited by 68
An Evaluation of the Resistance to Electrolysis of Metals for Use in Biostimulation Microprobes
White, R. L., & Gross, T. J. (1974) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1974.324339 · Cited by 57
Phagocytic and metabolic reactions to chronically implanted metal brain electrodes.
Babb, T. L., & Kupfer, W. (1984) · Experimental Neurology · doi:10.1016/0014-4886(84)90179-1 · Cited by 51
An approach to corrosion control during electrical stimulation
McHardy, J., Geller, D., & Brummer, S. B. (1977) · Annals of Biomedical Engineering · doi:10.1007/bf02364014 · Cited by 46
Histopathological studies of tissue reactions to various metals implanted in cat brains.
Robinson, F. R., & Johnson, M. T. (1961) · PubMed · Cited by 41
Bickford, R. G., FISCHER, G., & Sayre, G. P. (1957) · PubMed · Cited by 41
Schaidach, M. (1971) · PubMed · Cited by 38
Geddes, L. A., & Roeder, R. (2001) · Annals of Biomedical Engineering · doi:10.1114/1.1349699 · Cited by 37
Self-citationRadovsky, A. S., & Vleet, J. F. V. (1989) · American Heart Journal · doi:10.1016/0002-8703(89)90408-0 · Cited by 37
Some new electrode techniques for long-term physiologic monitoring.
Pc, R., Coombs, F. K., & Adams, R. M. (1968) · PubMed · Cited by 31
Titannitrid-Herzschrittmacher-Elektroden - Pacemaker Electrodes Made of Titanium Nitride
Schaldach, M., Hubmann, M., Hardt, R., & Weikl, A. (1989) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmte.1989.34.7-8.185 · Cited by 20
Aronson, S., & Geddes, L. A. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325650 · Cited by 16
Self-citationTHE PAUCITY OF DEMONSTRABLE BRAIN DAMAGE FOLLOWING INTRACEREBRAL ELECTROGRAPHY: REPORT OF CASE
Dodge, H. W., Petersen, M. C., Sem-Jacobsen, C. W., Sayre, G. P., & Bickford, R. G. (1955) · Mayo Clinic Proceedings · doi:10.1016/s0025-6196(26)01696-4 · Cited by 14
Stimulation with capacitor electrodes
Geddes, L. A., Hinds, M., & Foster, K. S. (1987) · Medical & Biological Engineering & Computing · doi:10.1007/bf02447440 · Cited by 2
Self-citation
Cited by
Sorted by citations
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845
Next-generation probes, particles, and proteins for neural interfacing
Rivnay, J., Wang, H., Fenno, L., Deisseroth, K., & Malliaras, G. G. (2017) · Science Advances · doi:10.1126/sciadv.1601649 · Cited by 514
Boehler, C., Carli, S., Fadiga, L., Stieglitz, T., & Asplund, M. (2020) · Nature Protocols · doi:10.1038/s41596-020-0389-2 · Cited by 324
Prasad, A., Xue, Q.-S., Sankar, V., Nishida, T., Shaw, G., Streit, W. J., & Sanchez, J. C. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/5/056015 · Cited by 298
Organic electrode coatings for next-generation neural interfaces
Robles, U. A. A., Woolley, A. J., Poole-Warren, L. A., Lovell, N. H., & Green, R. A. (2014) · Frontiers in Neuroengineering · doi:10.3389/fneng.2014.00015 · Cited by 266
Metadata
- Authors
- Geddes, L. A., & Roeder, R.
- Affiliations
- Purdue University West Lafayette (US)
- Venue
- Annals of Biomedical Engineering
- Published
- 2003-07
- Volume
- 31
- Issue
- 7
- Pages
- 879-890
- ISSN
- 0090-6964 (ISSN-L)
- OpenAlex
- W1978696561
- Cited by
- 240
- References
- 31
- Citations by year
- 2026: 5 · 2025: 5 · 2024: 12 · 2023: 8 · 2022: 13 · 2021: 18 · 2020: 19 · 2019: 21 · 2018: 20 · 2017: 15 · 2016: 19 · 2015: 11 · 2014: 18 · 2013: 7 · 2012: 8
- FWCI
- 0.92
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Annals of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Medicine › Surgery › Orthopaedic implants and arthroplasty
- Keywords
- implanted electrodes, electrode materials, tissue reaction, allergic reactions, gold, platinum, tungsten, tantalum, polyimide, glass insulators
- MeSH terms
- Biocompatible Materials, Decision Support Techniques, Electrodes, Implanted, Electrophysiology, Equipment Design, Hypersensitivity, Materials Testing, Microelectrodes, Technology Assessment, Biomedical, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review