The DOI matches its Crossref record; 26 of the 26 references it cites were found in a registry or an index.
Massive training artificial neural network (MTANN) for reduction of false positives in computerized detection of lung nodules in low‐dose computed tomography
Abstract
In this study, we investigated a pattern‐recognition technique based on an artificial neural network (ANN), which is called a massive training artificial neural network (MTANN), for reduction of false positives in computerized detection of lung nodules in low‐dose computed tomography (CT) images. The MTANN consists of a modified multilayer ANN, which is capable of operating on image data directly. The MTANN is trained by use of a large number of subregions extracted from input images together with the teacher images containing the distribution for the “likelihood of being a nodule.” The output image is obtained by scanning an input image with the MTANN. The distinction between a nodule and a non‐nodule is made by use of a score which is defined from the output image of the trained MTANN. In order to eliminate various types of non‐nodules, we extended the capability of a single MTANN, and developed a multiple MTANN (Multi‐MTANN). The Multi‐MTANN consists of plural MTANNs that are arranged in parallel. Each MTANN is trained by using the same nodules, but with a different type of non‐nodule. Each MTANN acts as an expert for a specific type of non‐nodule, e.g., five different MTANNs were trained to distinguish nodules from various‐sized vessels; four other MTANNs were applied to eliminate some other opacities. The outputs of the MTANNs were combined by using the logical AND operation such that each of the trained MTANNs eliminated none of the nodules, but removed the specific type of non‐nodule with which the MTANN was trained, and thus removed various types of non‐nodules. The Multi‐MTANN consisting of nine MTANNs was trained with 10 typical nodules and 10 non‐nodules representing each of nine different non‐nodule types (90 training non‐nodules overall) in a training set. The trained Multi‐MTANN was applied to the reduction of false positives reported by our current computerized scheme for lung nodule detection based on a database of 63 low‐dose CT scans (1765 sections), which contained 71 confirmed nodules including 66 biopsy‐confirmed primary cancers, from a lung cancer screening program. The Multi‐MTANN was applied to 58 true positives (nodules from 54 patients) and 1726 false positives (non‐nodules) reported by our current scheme in a validation test; these were different from the training set. The results indicated that 83% (1424/1726) of non‐nodules were removed with a reduction of one true positive (nodule), i.e., a classification sensitivity of 98.3% (57 of 58 nodules). By using the Multi‐MTANN, the false‐positive rate of our current scheme was improved from 0.98 to 0.18 false positives per section (from 27.4 to 4.8 per patient) at an overall sensitivity of 80.3% (57/71).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Learning representations by back-propagating errors
Rumelhart, D. E., Hinton, G. E., & Williams, R. J. (1986) · Nature · doi:10.1038/323533a0 · Cited by 31,892
Hanley, J. A., & McNeil, B. J. (1983) · Radiology · doi:10.1148/radiology.148.3.6878708 · Cited by 7,016
On the approximate realization of continuous mappings by neural networks
Funahashi, K.-I. (1989) · Neural Networks · doi:10.1016/0893-6080(89)90003-8 · Cited by 4,206
Learning internal representations by error propagation
RUMELHART, D. E., HINTON, G. E., & WILLIAMS, R. J. (1988) · Readings in Cognitive Science · doi:10.1016/b978-1-4832-1446-7.50035-2 · Cited by 3,002
Jemal, A., Thomas, A., Murray, T., & Thun, M. (2002) · CA: A Cancer Journal for Clinicians · doi:10.3322/canjclin.52.1.23 · Cited by 2,917
Universal approximation bounds for superpositions of a sigmoidal function
Barron, A. R. (1993) · IEEE Transactions on Information Theory · doi:10.1109/18.256500 · Cited by 2,804
Early Lung Cancer Action Project: overall design and findings from baseline screening
Henschke, C. I., McCauley, D. I., Yankelevitz, D. F., Naidich, D. P., McGuinness, G., Miettinen, O. S., Libby, D. M., Pasmantier, M. W., Koizumi, J., Altorki, N. K., et al. (1999) · The Lancet · doi:10.1016/s0140-6736(99)06093-6 · Cited by 2,508
ROC Methodology in Radiologic Imaging
METZ, C. E. (1986) · Investigative Radiology · doi:10.1097/00004424-198609000-00009 · Cited by 1,642
Mass screening for lung cancer with mobile spiral computed tomography scanner
Sone, S., Takashima, S., Li, F., Yang, Z., Honda, T., Maruyama, Y., Hasegawa, M., Yamanda, T., Kubo, K., Hanamura, K., et al. (1998) · The Lancet · doi:10.1016/s0140-6736(97)08229-9 · Cited by 931
Self-citationPeripheral lung cancer: screening and detection with low-dose spiral CT versus radiography.
Kaneko, M., Eguchi, K., Ohmatsu, H., Kakinuma, R., Naruke, T., Suemasu, K., & Moriyama, N. (1996) · Radiology · doi:10.1148/radiology.201.3.8939234 · Cited by 831
Metz, C. E., Herman, B. A., & Shen, J.-H. (1998) · Statistics in Medicine · doi:10.1002/(sici)1097-0258(19980515)17:9<1033::aid-sim784>3.0.co;2-z · Cited by 793
Sone, S., Li, F., Yang, Z.-G., Honda, T., Maruyama, Y., Takashima, S., Hasegawa, M., Kawakami, S., Kubo, K., Haniuda, M., et al. (2001) · British Journal of Cancer · doi:10.1054/bjoc.2000.1531 · Cited by 495
Self-citationComputerized Detection of Pulmonary Nodules on CT Scans
Armato, S. G., Giger, M. L., Moran, C. J., Blackburn, J. T., Doi, K., & MacMahon, H. (1999) · Radiographics · doi:10.1148/radiographics.19.5.g99se181303 · Cited by 380
Self-citationFree-response methodology: alternate analysis and a new observer-performance experiment.
Chakraborty, D. P., & Winter, L. H. (1990) · Radiology · doi:10.1148/radiology.174.3.2305073 · Cited by 363
Armato, S. G., Li, F., Giger, M. L., MacMahon, H., Sone, S., & Doi, K. (2002) · Radiology · doi:10.1148/radiol.2253011376 · Cited by 280
Self-citationAutomated detection of lung nodules in CT scans: Preliminary results
Armato, S. G., Giger, M. L., & MacMahon, H. (2001) · Medical Physics · doi:10.1118/1.1387272 · Cited by 250
Self-citationLi, F., Sone, S., Abe, H., MacMahon, H., Armato, S. G., & Doi, K. (2002) · Radiology · doi:10.1148/radiol.2253011375 · Cited by 201
Self-citationComputerized Detection of Pulmonary Nodules in Computed Tomography Images
GIGER, M. L., BAE, K. T., & MacMAHON, H. (1994) · Investigative Radiology · doi:10.1097/00004424-199404000-00013 · Cited by 200
Non-small-cell lung cancer: results of the New York screening program.
Heelan, R. T., Flehinger, B. J., Melamed, M. R., Zaman, M., Perchick, W. B., Caravelli, J. F., & Martini, N. (1984) · Radiology · doi:10.1148/radiology.151.2.6324279 · Cited by 152
A new class of nonlinear filters-neural filters
Yin, L., Astola, J. T., & Neuvo, Y. A. (1993) · IEEE Transactions on Signal Processing · doi:10.1109/78.205724 · Cited by 88
A Simple Neural Network Pruning Algorithm with Application to Filter Synthesis
Suzuki, K., Horiba, I., & Sugie, N. (2001) · Neural Processing Letters · doi:10.1023/a:1009639214138 · Cited by 84
Self-citationEfficient approximation of neural filters for removing quantum noise from images
Suzuki, K., Horiba, I., & Sugie, N. (2002) · IEEE Transactions on Signal Processing · doi:10.1109/tsp.2002.1011218 · Cited by 80
Self-citationSuzuki, K., Horiba, I., Sugie, N., & Nanki, M. (2002) · Tokyo Tech Research Repository (Tokyo Institute of Technology) · Cited by 48
Self-citationEdge detection from noisy images using a neural edge detector
Suzuki, K., Horiba, I., & Sugie, N. (2002) · Machine learning for signal processing ... · doi:10.1109/nnsp.2000.890125 · Cited by 12
Self-citationTraining under achievement quotient criterion
Suzuki, K., Horiba, I., & Sugie, N. (2002) · Machine learning for signal processing ... · doi:10.1109/nnsp.2000.890132 · Cited by 9
Self-citationNeural Edge Detector - A Good Mimic of Conventional One Yet Robuster against Noise
Suzuki, K., Horiba, I., & Sugie, N. (2001) · Lecture notes in computer science · doi:10.1007/3-540-45723-2_36 · Cited by 7
Self-citation
Cited by
Overview of deep learning in medical imaging
Suzuki, K. (2017) · Radiological Physics and Technology · doi:10.1007/s12194-017-0406-5 · Cited by 1,087
Self-citationDeep learning in medical imaging and radiation therapy
Sahiner, B., Pezeshk, A., Hadjiiski, L. M., Wang, X., Drukker, K., Cha, K. H., Summers, R. M., & Giger, M. L. (2019) · Medical Physics · doi:10.1002/mp.13264 · Cited by 740
Computer analysis of computed tomography scans of the lung: a survey
Sluimer, I., Schilham, A., Prokop, M., & van Ginneken, B. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2005.862753 · Cited by 556
Current status and future potential of computer-aided diagnosis in medical imaging
Doi, K. (2005) · The British Journal of Radiology · doi:10.1259/bjr/82933343 · Cited by 444
Self-citationGiger, M. L., Chan, H., & Boone, J. (2008) · Medical Physics · doi:10.1118/1.3013555 · Cited by 311
Metadata
- Authors
- Suzuki, K., Armato, S. G., Li, F., Sone, S., & Doi, K.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2003-06-20
- Volume
- 30
- Issue
- 7
- Pages
- 1602-1617
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2023522838
- Cited by
- 262
- References
- 26
- FWCI
- 10.25
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.