Size‐dependent ferrohydrodynamic relaxometry of magnetic particle imaging tracers in different environments
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Purpose: Magnetic particle imaging (MPI) is a recently developed imaging technique that seeks to provide ultrahigh resolution and tracer sensitivity with positive contrast directly originated from superparamagnetic iron oxide nanoparticles (NPs). MPI signals can be generated from a combination of Néel relaxation, Brownian rotational diffusion, and hysteretic reversal mechanisms of NPs in response to applied magnetic fields. When specific targeting of organs, such as carcinoma and endothelial cardiovascular cells, is needed, different behavior may be expected in immobilized NPs, due to complete or partial elimination of the Brownian motion. Here, the authors present an experimental investigation of the MPI spatial resolution and signal intensities as a function of a wide range of median core sizes of NPs under four representative conditions, including after immobilization in a tissue equivalent medium. Methods: Monodisperse hydrophobic NPs with median core diameters (d 0) ranging from 7 to 22 nm were synthesized in organic media and subsequently dispersed in aqueous solution after a facile surface modification. Morphology, median size, size distribution, and magnetic properties of the NPs were investigated. Hydrophobic and hydrophilic NPs with various core sizes were immobilized in trioctyl phosphine oxide and agarose gel, respectively. Their size‐dependent performance as MPI tracers for system matrix and x‐space image reconstruction was evaluated using magnetic particle spectrometry (MPS) and compared with the free rotating counterparts. Results: Immobilized NPs with core diameters smaller than ∼20 nm have similar spatial resolution, but lower signal intensities when compared with their free rotating counterparts. Compared to their performance in solution, spatial resolution was improved, but signal intensity was lower, when larger NPs with core size of 22 nm were immobilized in agarose. Same trends were observed in signal intensities, when considering either system matrix or x‐space approaches. The harmonic and dm/dH signal intensities changed linearly and the spatial resolution did not change with decreasing NP concentration up to 15 μ g/ml. Conclusions: The results show that the MPI signal is very sensitive to both NP size and environment. The authors’ calculations show that Brownian rotational diffusion is slower than the field switching cycle and, therefore, it has minimal influence on MPS signals. dm/dH analyses show that Néel relaxation is the dominant mechanism determining MPI response in smaller NPs (d 0 < ∼20 nm). Larger NPs show hysteretic reversal when the applied field amplitude is large enough to overcome the coercivity. Linear variation of the MPS signal intensity with iron concentration but with uniform spatial resolution enables quantitative imaging for a range of applications, from high‐concentration bolus chase imaging to low‐concentration molecular imaging (while the authors’ instrument is noise‐limited to ∼millimolar iron concentrations, nanomolar sensitivity is expected for MPI, theoretically). These results pave the way for future application of the authors’ synthesized tracers for immobilized or in vivo targeted MPI of tissues.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Thermal Fluctuations of a Single-Domain Particle
Brown, W. F. (1963) · Physical Review · doi:10.1103/physrev.130.1677 · Cited by 3,374
Three-dimensional real-time in vivo magnetic particle imaging
Weizenecker, J., Gleich, B., Rahmer, J., Dahnke, H., & Borgert, J. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/5/l01 · Cited by 651
Monodisperse magnetite nanoparticle tracers for in vivo magnetic particle imaging
Khandhar, A. P., Ferguson, R. M., Arami, H., & Krishnan, K. M. (2013) · Biomaterials · doi:10.1016/j.biomaterials.2013.01.087 · Cited by 152
Self-citationPoint spread function analysis of magnetic particles
Schmale, I., Rahmer, J., Gleich, B., Borgert, J., & Weizenecker, J. (2012) · Springer proceedings in physics · doi:10.1007/978-3-642-24133-8_46 · Cited by 17
Cited by
The Effect of Nanoparticle Size on In Vivo Pharmacokinetics and Cellular Interaction
Hoshyar, N., Gray, S., Han, H., & Bao, G. (2016) · Nanomedicine · doi:10.2217/nnm.16.5 · Cited by 1,990
In vivo delivery, pharmacokinetics, biodistribution and toxicity of iron oxide nanoparticles
Arami, H., Khandhar, A., Liggitt, D., & Krishnan, K. M. (2015) · Chemical Society Reviews · doi:10.1039/c5cs00541h · Cited by 830
Self-citationMagnetic nanoparticles in nanomedicine: a review of recent advances
Wu, K., Su, D., Liu, J., Saha, R., & Wang, J.-P. (2019) · Nanotechnology · doi:10.1088/1361-6528/ab4241 · Cited by 509
Advances in the Application of Magnetic Nanoparticles for Sensing
Gloag, L., Mehdipour, M., Chen, D., Tilley, R. D., & Gooding, J. J. (2019) · Advanced Materials · doi:10.1002/adma.201904385 · Cited by 364
Synthesis of phase-pure and monodisperse iron oxide nanoparticles by thermal decomposition
Hufschmid, R., Arami, H., Ferguson, R. M., Gonzales, M., Teeman, E., Brush, L. N., Browning, N. D., & Krishnan, K. M. (2015) · Nanoscale · doi:10.1039/c5nr01651g · Cited by 341
“Particle size, phase, morphology, and interactions with immediate environment determine SPION magnetic properties and must be tailored for the specific application.(21,22) For example, 16 nm SPIONs have been shown to be optimal for MFH (with a 373 kHz, 14 kA/m field), with broader distributions decreasing heating effi…”
Self-citation
Metadata
- Authors
- Arami, H., Ferguson, R. M., Khandhar, A. P., & Krishnan, K. M.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2013-06-19
- Volume
- 40
- Issue
- 7
- Pages
- 071904-071904
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2035026160
- Cited by
- 90
- References
- 4
- FWCI
- 4.61
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.