Noise correlation in CBCT projection data and its application for noise reduction in low‐dose CBCT
Abstract
Purpose: To study the noise correlation properties of cone‐beam CT (CBCT) projection data and to incorporate the noise correlation information to a statistics‐based projection restoration algorithm for noise reduction in low‐dose CBCT. Methods: In this study, the authors systematically investigated the noise correlation properties among detector bins of CBCT projection data by analyzing repeated projection measurements. The measurements were performed on a TrueBeam onboard CBCT imaging system with a 4030CB flat panel detector. An anthropomorphic male pelvis phantom was used to acquire 500 repeated projection data at six different dose levels from 0.1 to 1.6 mAs per projection at three fixed angles. To minimize the influence of the lag effect, lag correction was performed on the consecutively acquired projection data. The noise correlation coefficient between detector bin pairs was calculated from the corrected projection data. The noise correlation among CBCT projection data was then incorporated into the covariance matrix of the penalized weighted least‐squares (PWLS) criterion for noise reduction of low‐dose CBCT. Results: The analyses of the repeated measurements show that noise correlation coefficients are nonzero between the nearest neighboring bins of CBCT projection data. The average noise correlation coefficients for the first‐ and second‐order neighbors are 0.20 and 0.06, respectively. The noise correlation coefficients are independent of the dose level. Reconstruction of the pelvis phantom shows that the PWLS criterion with consideration of noise correlation (PWLS‐Cor) results in a lower noise level as compared to the PWLS criterion without considering the noise correlation (PWLS‐Dia) at the matched resolution. At the 2.0 mm resolution level in the axial‐plane noise resolution tradeoff analysis, the noise level of the PWLS‐Cor reconstruction is 6.3% lower than that of the PWLS‐Dia reconstruction. Conclusions: Noise is correlated among nearest neighboring detector bins of CBCT projection data. An accurate noise model of CBCT projection data can improve the performance of the statistics‐based projection restoration algorithm for low‐dose CBCT.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Electronic noise is signal independent and can be modeled as additive Gaussian noise with correlation along rows due to sequential readout [15].”
From DeepDRR – A Catalyst for Machine Learning in Fluoroscopy-Guided Procedures (2018) · cites this paper as [15]
In the citing paper
“Noise in X-ray images is a composite of uncorrelated quantum noise due to photon statistics that becomes correlated due to pixel crosstalk, and correlated readout noise [15].”
From DeepDRR – A Catalyst for Machine Learning in Fluoroscopy-Guided Procedures (2018) · cites this paper as [15]
Citations
Cites
Computed Tomography — An Increasing Source of Radiation Exposure
Brenner, D. J., & Hall, E. J. (2007) · New England Journal of Medicine · doi:10.1056/nejmra072149 · Cited by 8,748
Feldkamp, L. A., Davis, L. C., & Kress, J. W. (1984) · Journal of the Optical Society of America A · doi:10.1364/josaa.1.000612 · Cited by 6,319
LSQR: An Algorithm for Sparse Linear Equations and Sparse Least Squares
Paige, C. C., & Saunders, M. A. (1982) · ACM Transactions on Mathematical Software · doi:10.1145/355984.355989 · Cited by 4,387
Generalized Cross-Validation as a Method for Choosing a Good Ridge Parameter
Golub, G. H., Heath, M., & Wahba, G. (1979) · Technometrics · doi:10.1080/00401706.1979.10489751 · Cited by 3,934
The Use of the L-Curve in the Regularization of Discrete Ill-Posed Problems
Hansen, P. C., & O’Leary, D. P. (1993) · SIAM Journal on Scientific Computing · doi:10.1137/0914086 · Cited by 3,137
Iterative Methods for Total Variation Denoising
Vogel, C. R., & Oman, M. E. (1996) · SIAM Journal on Scientific Computing · doi:10.1137/0917016 · Cited by 1,076
A three‐dimensional statistical approach to improved image quality for multislice helical CT
Thibault, J., Sauer, K. D., Bouman, C. A., & Hsieh, J. (2007) · Medical Physics · doi:10.1118/1.2789499 · Cited by 937
Penalized weighted least-squares image reconstruction for positron emission tomography
Fessler, J. A. (1994) · IEEE Transactions on Medical Imaging · doi:10.1109/42.293921 · Cited by 611
Penalized weighted least-squares image reconstruction for dual energy X-ray transmission tomography
Sukovic, P., & Clinthorne, N. H. (2000) · IEEE Transactions on Medical Imaging · doi:10.1109/42.896783 · Cited by 552
Wang, J., Li, T., Lu, H., & Liang, Z. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.882141 · Cited by 533
Self-citation
Research works printed but not linked: 1
- Hsieh J., 2003, Computer Tomography, Principles, Design, Artifacts, and Recent Advances
Cited by
DeepDRR – A Catalyst for Machine Learning in Fluoroscopy-Guided Procedures
Unberath, M., Zaech, J.-N., Lee, S. C., Bier, B., Fotouhi, J., Armand, M., & Navab, N. (2018) · Lecture Notes in Computer Science · doi:10.1007/978-3-030-00937-3_12 · Cited by 145
“Since real flat panel detectors suffer from pixel crosstalk, we correlate the quantum noise of neighboring pixels by convolving the noise signal with a blurring kernel [15].”
BackgroundMethodInfluentialIterative CT Reconstruction Using Shearlet-Based Regularization
Vandeghinste, B., Goossens, B., Holen, R. V., Vanhove, C., Pižurica, A., Vandenberghe, S., & Staelens, S. G. (2013) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2013.2275994 · Cited by 67
Enabling machine learning in X-ray-based procedures via realistic simulation of image formation
Unberath, M., Zaech, J.-N., Gao, C., Bier, B., Goldmann, F., Lee, S. C., Fotouhi, J., Taylor, R., Armand, M., & Navab, N. (2019) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-019-02011-2 · Cited by 63
Tilley, S., Siewerdsen, J. H., & Stayman, J. W. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/1/296 · Cited by 60
Lanis, A., & del Canto, O. A. (2015) · The International Journal of Prosthodontics · doi:10.11607/ijp.4148 · Cited by 50
Metadata
- Authors
- Zhang, H., Ouyang, L., Ma, J., Huang, J., Chen, W., & Wang, J.
- Affiliations
- Southern Medical University (CN)
- The University of Texas Southwestern Medical Center (US)
- Corresponding author
- Ma, Jianhua
- Wang, Jing
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2014-02-19
- Volume
- 41
- Issue
- 3
- Pages
- 031906-031906
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2014598003
- Cited by
- 47
- References
- 32
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 4 · 2023: 3 · 2022: 2 · 2021: 1 · 2020: 4 · 2019: 6 · 2018: 8 · 2017: 3 · 2016: 4 · 2015: 5 · 2014: 2
- FWCI
- 3.65
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- onlinelibrary.wiley.com/termsAndConditions#vor (vor)
- doi.wiley.com/10.1002/tdm_license_1.1 (tdm, 559 d)
- Locations
- Medical Physics (publishedVersion)
- PubMed (publishedVersion)
- Funders
- NNSFC (81371544, 81000613, 81101046)
- National Natural Science Foundation of China (81371544, 81000613, 81101046)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Image and Signal Denoising Methods
- Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
- Keywords
- cone-beam computed tomography, noise correlation, low-dose CT, noise reduction, penalized weighted least squares, covariance matrix, noise models, pelvis phantom
- MeSH terms
- Algorithms, Anthropometry, Humans, Image Processing, Computer-Assisted, Male, Radiation Dosage, Radiographic Image Enhancement, Radiographic Image Interpretation, Computer-Assisted, Reproducibility of Results, Models, Statistical, Poisson Distribution, Least-Squares Analysis, Phantoms, Imaging, Cone-Beam Computed Tomography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.