Comparison of Heart-Rate-Variability Recording With Smartphone Photoplethysmography, Polar H7 Chest Strap, and Electrocardiography
Abstract
Purpose: To establish the validity of smartphone photoplethysmography (PPG) and heart-rate sensor in the measurement of heart-rate variability (HRV). Methods: 29 healthy subjects were measured at rest during 5 min of guided breathing and normal breathing using smartphone PPG, a heart-rate chest strap, and electrocardiography (ECG). The root mean sum of the squared differences between R–R intervals (rMSSD) was determined from each device. Results: Compared to ECG, the technical error of estimate (TEE) was acceptable for all conditions (average TEE CV% [90% CI] = 6.35 [5.13; 8.5]). When assessed as a standardized difference, all differences were deemed “trivial” (average standard difference [90% CI] = 0.10 [0.08; 0.13]). Both PPG- and heart-rate-sensor-derived measures had almost perfect correlations with ECG (R = 1.00 [0.99; 1.00]). Conclusion: Both PPG and heart-rate sensors provide an acceptable agreement for the measurement of rMSSD when compared with ECG. Smartphone PPG technology may be a preferred method of HRV data collection for athletes due to its practicality and ease of use in the field.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Heart rate variability: Standards of measurement, physiological interpretation, and clinical use
Malik, M., Bigger, J. T., Camm, A. J., Kleiger, R. E., Malliani, A., Moss, A. J., & Schwartz, P. J. (1996) · European Heart Journal · doi:10.1093/oxfordjournals.eurheartj.a014868 · Cited by 15,289
Monitoring training status with HR measures: do all roads lead to Rome?
Buchheit, M. (2014) · Frontiers in Physiology · doi:10.3389/fphys.2014.00073 · Cited by 879
How accurate is pulse rate variability as an estimate of heart rate variability?
Schäfer, A., & Vagedes, J. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.03.119 · Cited by 661 · cited 1×
“When we contrast other studies that have compared HRV values measured through PPG against ECG it is currently difficult, as differences between experimental settings and/or methods of analysis are apparent.(19) Furthermore, many of these studies have been carried out using a variety of “clip-on” devices (e.”
ResultPlews, D. J., Laursen, P. B., Stanley, J., Kilding, A. E., & Buchheit, M. (2013) · Sports Medicine · doi:10.1007/s40279-013-0071-8 · Cited by 617
Self-citationSmart wearable body sensors for patient self-assessment and monitoring
Appelboom, G., Camacho, E., Abraham, M. E., Bruce, S. S., Dumont, E. L., Zacharia, B. E., D’Amico, R., Slomian, J., Reginster, J. Y., Bruyère, O., et al. (2014) · Archives of Public Health · doi:10.1186/2049-3258-72-28 · Cited by 448
Plews, D. J., Laursen, P. B., Kilding, A. E., & Buchheit, M. (2012) · European Journal of Applied Physiology · doi:10.1007/s00421-012-2354-4 · Cited by 269
Self-citationComparison of methods for removal of ectopy in measurement of heart rate variability
Lippman, N., Stein, K. M., & Lerman, B. B. (1994) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1994.267.1.h411 · Cited by 220
Monitoring Training With Heart-Rate Variability: How Much Compliance Is Needed for Valid Assessment?
Plews, D. J., Laursen, P. B., Meur, Y. L., Hausswirth, C., Kilding, A. E., & Buchheit, M. (2014) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.2013-0455 · Cited by 156
Self-citationEvaluating Training Adaptation With Heart-Rate Measures: A Methodological Comparison
Plews, D. J., Laursen, P. B., Kilding, A. E., & Buchheit, M. (2013) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.8.6.688 · Cited by 151
Self-citationA Real Time Analysis of PPG Signal for Measurement of SpO2 and Pulse Rate
Bagha, S., & Shaw, L. (2011) · International Journal of Computer Applications · Cited by 143
Cited by
RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise
Gilgen-Ammann, R., Schweizer, T., & Wyss, T. (2019) · European Journal of Applied Physiology · doi:10.1007/s00421-019-04142-5 · Cited by 438
Miller, D. J., Sargent, C., & Roach, G. D. (2022) · Sensors · doi:10.3390/s22166317 · Cited by 328
“The passive collection of these measures is less intrusive compared to laboratory ‐ based assessments and is therefore of significant interest to ath ‐ letes and support staff [12].”
Can Wearable Devices Accurately Measure Heart Rate Variability? A Systematic Review
Georgiou, K., Larentzakis, A. V., Khamis, N. N., Alsuhaibani, G. I., Alaska, Y. A., & Giallafos, E. J. (2018) · Folia Medica · doi:10.2478/folmed-2018-0012 · Cited by 244
Blum, J., Rockstroh, C., & Göritz, A. S. (2019) · Frontiers in Psychology · doi:10.3389/fpsyg.2019.02172 · Cited by 183
Li, K., Cardoso, C., Moctezuma-Ramirez, A., Elgalad, A., & Perin, E. (2023) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph20247146 · Cited by 160
Metadata
- Authors
- Plews, D. J., Scott, B., Altini, M., Wood, M., Kilding, A. E., & Laursen, P. B.
- Venue
- International Journal of Sports Physiology and Performance
- Publisher
- Human Kinetics
- Published
- 2017-11-01
- Volume
- 12
- Issue
- 10
- Pages
- 1324-1328
- ISSN
- 1555-0265 (ISSN-L)
- 1555-0273
- OpenAlex
- W2606585248
- Cited by
- 312
- References
- 15
- FWCI
- 10.94
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Comparison of Heart-Rate-Variability Recording With Smartphone Photoplethysmography, Polar H7 Chest Strap, and Electrocardiography. https://app.cochat.ai/papers/10.1123/ijspp.2016-0668
- The paper
- Plews, D. J., Scott, B., Altini, M., Wood, M., Kilding, A. E., & Laursen, P. B. (2017). Comparison of Heart-Rate-Variability Recording With Smartphone Photoplethysmography, Polar H7 Chest Strap, and Electrocardiography. International Journal of Sports Physiology and Performance. https://doi.org/10.1123/ijspp.2016-0668
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.