Technical validation of real-world monitoring of gait: a multicentric observational study

Open access
Mazzà, Claudia, Alcock, Lisa, Aminian, Kamiar, Becker, Clemens, Bertuletti, Stefano, Bonci, Tecla, Brown, Philip, Brozgol, Marina, Buckley, Ellen, Carsin, Anne-Elie, Caruso, Marco, Caulfield, Brian, Cereatti, Andrea, Chiari, Lorenzo, Chynkiamis, Nikolaos, Ciravegna, Fabio, Del Din, Silvia, Eskofier, Björn, Evers, Jordi, Garcia Aymerich, Judith, Gazit, Eran, Hansen, Clint, Hausdorff, Jeffrey M, Helbostad, Jorunn L, Hiden, Hugo, Hume, Emily, Paraschiv-Ionescu, Anisoara, Ireson, Neil, Keogh, Alison, Kirk, Cameron, Kluge, Felix, Koch, Sarah, Küderle, Arne, Lanfranchi, Vitaveska, Maetzler, Walter, Micó-Amigo, M Encarna, Mueller, Arne, Neatrour, Isabel, Niessen, Martijn, Palmerini, Luca, Pluimgraaff, Lucas, Reggi, Luca, Salis, Francesca, Schwickert, Lars, Scott, Kirsty, Sharrack, Basil, Sillen, Henrik, Singleton, David, Soltani, Abolfazi, Taraldsen, Kristin, Ullrich, Martin, Van Gelder, Linda, Vereijken, Beatrix, Vogiatzis, Ioannis, Warmerdam, Elke, Yarnall, Alison, Rochester, Lynn · 2021 · BMJ Open
Cited by 118

The DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.

Abstract

Introduction: Existing mobility endpoints based on functional performance, physical assessments and patient self-reporting are often affected by lack of sensitivity, limiting their utility in clinical practice. Wearable devices including inertial measurement units (IMUs) can overcome these limitations by quantifying digital mobility outcomes (DMOs) both during supervised structured assessments and in real-world conditions. The validity of IMU-based methods in the real-world, however, is still limited in patient populations. Rigorous validation procedures should cover the device metrological verification, the validation of the algorithms for the DMOs computation specifically for the population of interest and in daily life situations, and the users’ perspective on the device. Methods and analysis: This protocol was designed to establish the technical validity and patient acceptability of the approach used to quantify digital mobility in the real world by Mobilise-D, a consortium funded by the European Union (EU) as part of the Innovative Medicine Initiative, aiming at fostering regulatory approval and clinical adoption of DMOs. After defining the procedures for the metrological verification of an IMU-based device, the experimental procedures for the validation of algorithms used to calculate the DMOs are presented. These include laboratory and real-world assessment in 120 participants from five groups: healthy older adults; chronic obstructive pulmonary disease, Parkinson’s disease, multiple sclerosis, proximal femoral fracture and congestive heart failure. DMOs extracted from the monitoring device will be compared with those from different reference systems, chosen according to the contexts of observation. Questionnaires and interviews will evaluate the users’ perspective on the deployed technology and relevance of the mobility assessment. Ethics and dissemination: The study has been granted ethics approval by the centre’s committees (London—Bloomsbury Research Ethics committee; Helsinki Committee, Tel Aviv Sourasky Medical Centre; Medical Faculties of The University of Tübingen and of the University of Kiel). Data and algorithms will be made publicly available. Trial registration number: ISRCTN (12246987).

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1136/bmjopen-2021-050785

Cites

12 of 12 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 12 references

Cited by

See all 118

Metadata

Authors
Mazzà, C., Alcock, L., Aminian, K., Becker, C., Bertuletti, S., Bonci, T., Brown, P., Brozgol, M., Buckley, E., Carsin, A.-E., Caruso, M., Caulfield, B., Cereatti, A., Chiari, L., Chynkiamis, N., Ciravegna, F., Del Din, S., Eskofier, B., Evers, J., Garcia Aymerich, J., Gazit, E., Hansen, C., Hausdorff, J. M., Helbostad, J. L., Hiden, H., Hume, E., Paraschiv-Ionescu, A., Ireson, N., Keogh, A., Kirk, C., Kluge, F., Koch, S., Küderle, A., Lanfranchi, V., Maetzler, W., Micó-Amigo, M. E., Mueller, A., Neatrour, I., Niessen, M., Palmerini, L., Pluimgraaff, L., Reggi, L., Salis, F., Schwickert, L., Scott, K., Sharrack, B., Sillen, H., Singleton, D., Soltani, A., Taraldsen, K., Ullrich, M., Van Gelder, L., Vereijken, B., Vogiatzis, I., Warmerdam, E., Yarnall, A., & Rochester, L.
Venue
BMJ Open
Publisher
BMJ
Published
2021-12
Volume
11
Issue
12
Pages
e050785
ISSN
2044-6055 (ISSN-L)
OpenAlex
W3216794426
Cited by
118
References
12
FWCI
31.92
Type
journal article
Language
en
Open access
gold

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Technical validation of real-world monitoring of gait: a multicentric observational study. https://app.cochat.ai/papers/10.1136/bmjopen-2021-050785
The paper
Mazzà, C., Alcock, L., Aminian, K., Becker, C., Bertuletti, S., Bonci, T., Brown, P., Brozgol, M., Buckley, E., Carsin, A.-E., Caruso, M., Caulfield, B., Cereatti, A., Chiari, L., Chynkiamis, N., Ciravegna, F., Del Din, S., Eskofier, B., Evers, J., . . . Rochester, L. (2021). Technical validation of real-world monitoring of gait: a multicentric observational study. BMJ Open. https://doi.org/10.1136/bmjopen-2021-050785

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.