Bone Mechanical Properties in Healthy and Diseased States
Open accessThe DOI matches its Crossref record; 173 of the 174 references it cites were found in a registry or an index; 47 of the 225 sources its claims cite were checked for support.
Abstract
Review of bone mechanics across length scales, including cortical bone, trabecular bone, and whole bones, covering elasticity, yield, fracture, fatigue, and damage.
Overview
CoChat Summary
TL;DR
This review synthesizes current knowledge of bone mechanical behavior from the nanoscale to whole-bone scale in healthy, aging, and diseased states. It emphasizes porosity/bone quantity and bone-quality parameters such as cross-linking and protein composition as determinants of cortical, trabecular, and whole-bone stiffness, strength, toughness, fatigue, and damage.
Key findings
- Cortical bone has 5–15% porosity and trabecular bone 40–95%; cortical bone is anisotropic, strongest in longitudinal compression and weakest in transverse tension, with mode I fracture toughness \(K_{IC}\) of 2–6 MPa√m, lower for longitudinal than transverse fracture and lower at high strain rates.
- With aging, cortical tensile/compressive strength declines ~2% per decade from the third decade, tensile ultimate strain declines ~10% per decade from 5% at age 20–30 to <1% above age 80, and fracture toughness declines ~4% per decade; trabecular modulus and strength decline ~10% per decade, and proximal femoral strength declines ~10–12% per decade (~50% from early to late adulthood).
- Human trabecular bone apparent modulus ranges from 10 to 3,000 MPa and strength from 0.1 to 30 MPa; 70–90% of the variance in modulus and strength is explained by volume fraction or apparent density, with power-law exponents of ~2–3 for modulus and ~2 for compressive strength.
- Bone quality matters: nonenzymatic glycation cross-links are associated with increased brittleness, diabetes is associated with lower bone toughness and possibly heightened AGE accumulation, and long-term antiresorptive therapy is associated with decreased toughness.
Methods
- Narrative review of bone mechanics across length scales from hundreds of nanometers to tens of centimeters, covering cortical bone, trabecular bone, whole bones, and material behaviors including elasticity, yield, fracture, fatigue, and damage.
- No primary data collection or systematic search/meta-analysis is described; the review synthesizes in vitro, in vivo, imaging, analytical, and finite element modeling literature.
- It discusses bone quantity (density, porosity) and bone quality (cross-linking, protein composition) and identifies future research avenues.
Limitations
- Stated in the paper: no consensus exists on the most accurate multiaxial yield criterion for trabecular bone, largely because none has been validated with an independent set of specimens; few data are available on whether aging or disease affect trabecular tissue differently than cortical tissue; the mechanistic link between microdamage and increased fracture risk has not been established; and consensus has yet to emerge on Raman spectroscopy readouts as indicators of bone fragility.
- Evident from the text: as a narrative review, it does not report a systematic search strategy, inclusion criteria, or quantitative synthesis, and its conclusions draw on heterogeneous ex vivo, animal, and modeling studies.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
43 Valid · 4 Not valid · 117 Not checked · 68 Own results and interpretations
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“In vitro mechanical tests have found that microdamage is associated with a decrease in modulus (9, 10)”
§ Bone Mechanical Properties in Healthy and Diseased States > 2. MECHANICAL PROPERTIES OF BONE TISSUE > 2.1. Cortical Bone
What the sources say
- Not checked — no full text of the sourcecites [9] Burr DB Turner CH Naick P Forwood MR Ambrosius W 1998 Does microdamage accumulation affect the mechanical properties of bone? J Biomech 31 337 45 9672087 source
- Supportedcites [10] Nicolella DP Ni Q Chan KS 2011 Non-destructive characterization of microdamage in cortical bone using low field pulsed NMR J Mech Behav Biomed Mater 4 383 91 21316626 PMC3073569 source
The Introduction states that microdamage has been directly associated with modulus degradation. The Results compare measured and predicted elastic-modulus degradation in cyclically damaged cortical bone, and the Conclusions say the relationship agrees with experimental observations. Page numbers are not supplied in our copy. (LLM, conf=0.94)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Every source the paper cites for this claim was checked and supports it.
In the paper
“Preliminary values of fracture toughness at the microscale appear to be similar to those measured at the macroscale (41).”
§ Bone Mechanical Properties in Healthy and Diseased States > 2. MECHANICAL PROPERTIES OF BONE TISSUE > 2.1. Cortical Bone
What the sources say
- Supportedcites [41] Kataruka A Mendu K Okeoghene O Puthuvelil J Akono AT 2017 Microscopic assessment of bone toughness using scratch tests Bone Rep 6 17 25 28377977 PMC5365275 source
The Results & discussion section (“Fracture characteristics”) reports microscale scratch-test toughness values of 3.15 ± 0.56 and 3.49 ± 0.80 MPa√m. It explicitly says these fall within the 2.2–4.6 MPa√m range reported for macroscopic fracture tests. (LLM, conf=0.94)
hedged- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“and that this difference may be related to heightened accumulation of AGEs in diabetic bone (75).”
§ Bone Mechanical Properties in Healthy and Diseased States > 2. MECHANICAL PROPERTIES OF BONE TISSUE > 2.1. Cortical Bone
What the sources say
- Supportedcites [75] Rubin MR Paschalis EP Poundarik A Sroga GE McMahon DJ 2016 Advanced glycation endproducts and bone material properties in type 1 diabetic mice PLOS ONE 11 e0154700 27140650 PMC4854398 source
The Abstract reports increased skeletal AGEs and lower bone toughness in a type 1 diabetic mouse model, with higher AGEs correlating with lower macroscopic toughness. The Discussion likewise says AGE accumulation co-exists with bone deterioration, supporting the claim’s tentative wording. (LLM, conf=0.91)
hedgedstates a possible cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“and there is concern that long-term use of antiresorptive therapies can result in decreased toughness (48, 72, 73).”
§ Bone Mechanical Properties in Healthy and Diseased States > 2. MECHANICAL PROPERTIES OF BONE TISSUE > 2.1. Cortical Bone
What the sources say
- Supportedcites [48] Lloyd AA Gludovatz B Riedel C Luengo EA Saiyed R 2017 Atypical fracture with long-term bisphosphonate therapy is associated with altered cortical composition and reduced fracture resistance PNAS 114 8722 27 28760963 PMC5565436 source
The Conclusion states that long-term antiresorptive treatment can affect bone toughness through reductions in intrinsic and extrinsic toughening mechanisms. The Loss of Intrinsic Toughening section reports reduced fracture toughness in cortical bone from bisphosphonate-treated patients, supporting the citing sentence’s hedged concern. (LLM, conf=0.92)
- Supportedcites [72] Allen MR Burr DB 2011 Bisphosphonate effects on bone turnover, microdamage, and mechanical properties: what we think we know and what we know that we don’t know Bone 49 56 65 20955825 source
The mechanical-properties discussion (page not provided) says bisphosphonate treatment consistently reduces bone toughness and that toughness declines with longer treatment. The paper discusses this as a concern, while noting that whether the change occurs in human bone remains to be determined. (LLM, conf=0.90)
- Not checked — no full text of the sourcecites [73] Chapurlat RD Delmas PD 2009 Bone microdamage: a clinical perspective Osteoporos Int 20 1299 308 19291343 source
plainstates a possible cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“The mechanical effects of osteoporosis therapies, however, differ among the types of therapy (70, 71)”
§ Bone Mechanical Properties in Healthy and Diseased States > 2. MECHANICAL PROPERTIES OF BONE TISSUE > 2.1. Cortical Bone
What the sources say
- Not checked — no full text of the sourcecites [70] Campbell GM Tiwari S Picke AK Hofbauer C Rauner M 2016 Effects of insulin therapy on porosity, non-enzymatic glycation and mechanical competence in the bone of rats with type 2 diabetes mellitus Bone 91 186 93 27497735 source
- Supportedcites [71] Brock GR Chen JT Ingraffea AR MacLeay J Pluhar GE 2015 The effect of osteoporosis treatments on fatigue properties of cortical bone tissue Bone Rep 2 8 13 25642445 PMC4306187 source
The Discussion and conclusions state that differences in cortical bone fatigue life depended on treatment class and drug: alendronate reduced fatigue life, PTH increased it, and raloxifene and zoledronate did not change it. This supports the cited claim that mechanical effects differ among osteoporosis therapies. (LLM, conf=0.93)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
165 more claims · 23 with their cited sources checked
Create a free account to see all 232 claims and to verify the 14 references still open.
- Probably not a substantial claim. 25
- It may state two separate claims as one. 41
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Bone Mechanical Properties in Healthy and Diseased States > 2. MECHANICAL PROPERTIES OF BONE TISSUE > 2.1. Cortical Bone
“When loaded in tension along the longitudinal direction, cortical bone exhibits a bilinear stress–strain response in which a distinct yield point separates a linearly elastic region and a region of linear hardening that ends abruptly at a fracture strain of less than 3% (Figure 1a).”
§ Bone Mechanical Properties in Healthy and Diseased States > 2. MECHANICAL PROPERTIES OF BONE TISSUE > 2.1. Cortical Bone
“Tensile ultimate strain decreases by approximately 10% per decade, from a high of 5% strain at age 20–30 years to a low of less than 1% strain above age 80 years.”
Citations
Cites
The compressive behavior of bone as a two-phase porous structure
Carter, D., & Hayes, W. (1977) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-197759070-00021 · Cited by 2,001
The elastic and ultimate properties of compact bone tissue
Reilly, D. T., & Burstein, A. H. (1975) · Journal of Biomechanics · doi:10.1016/0021-9290(75)90075-5 · Cited by 1,533
Hildebrand, T., Laib, A., Müller, R., Dequeker, J., & Rüegsegger, P. (1999) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.1999.14.7.1167 · Cited by 1,303 · cited 1×
“Measures of trabecular architecture include trabecular thickness, trabecular spacing, trabecular number, connectivity density, structure model index (SMI; a measure of how rod-like versus plate-like the architecture is), and the degree of anisotropy (83).”
Young's modulus of trabecular and cortical bone material: Ultrasonic and microtensile measurements
Rho, J. Y., Ashman, R. B., & Turner, C. H. (1993) · Journal of Biomechanics · doi:10.1016/0021-9290(93)90042-d · Cited by 1,283 · cited 1×
“Tensile tests (136), ultrasonic measurements (137), nanoindentation (138), and scanning acoustic microscopy (139) techniques indicate that the elastic modulus of trabecular tissue is around 10–20 GPa.”
Trabecular bone modulus–density relationships depend on anatomic site
Morgan, E. F., Bayraktar, H. H., & Keaveny, T. M. (2003) · Journal of Biomechanics · doi:10.1016/s0021-9290(03)00071-x · Cited by 1,209 · cited 3×
“Numerous studies have indicated that the power-law exponent for human trabecular bone is between two and three for modulus (91) and two for compressive strength (Figure 4) (82).”
Self-citationComparison of the elastic and yield properties of human femoral trabecular and cortical bone tissue
Bayraktar, H. H., Morgan, E. F., Niebur, G. L., Morris, G. E., Wong, E. K., & Keaveny, T. M. (2004) · Journal of Biomechanics · doi:10.1016/s0021-9290(03)00257-4 · Cited by 1,142 · cited 1×
“Results from this approach place the elastic modulus of human trabecular tissue at approximately 10% less, tensile yield strain at 15% less, and tensile/compressive tissue strength at 25% less than that of cortical bone (142).”
Self-citationBiomechanics of cellular solids
Gibson, L. J. (2005) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2004.09.027 · Cited by 961 · cited 1×
“Using this approach, investigators attributed the experimentally observed dependence of Young’s modulus and compressive strength on apparent density raised to the second power to bendingdominated linear elastic behavior and failure by buckling (127).”
Riggs, B. L., Melton, L. J., Robb, R. A., Camp, J. J., Atkinson, E. J., Peterson, J. M., Rouleau, P. A., McCollough, C. H., Bouxsein, M. L., & Khosla, S. (2004) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.040916 · Cited by 925 · cited 2×
“For the vertebrae, an increase in the cross-sectional area with aging has been reported for both men and women (154), and one study found that this increase was three times greater in men than in women (158).”
Elastic properties of human cortical and trabecular lamellar bone measured by nanoindentation
Rho, J.-Y., Tsui, T. Y., & Pharr, G. M. (1997) · Biomaterials · doi:10.1016/s0142-9612(97)00073-2 · Cited by 924 · cited 1×
“Elastic moduli measured by nanoindentation are on the order of 23 GPa and 26 GPa for osteonal and interstitial lamellae, respectively, in the longitudinal direction (36, 37), and are approximately 45% lower in the transverse direction (37).”
van Rietbergen, B., Weinans, H., Huiskes, R., & Odgaard, A. (1995) · Journal of Biomechanics · doi:10.1016/0021-9290(95)80008-5 · Cited by 865
Cited by
Items 501–600 of 1,021
Chutani, K., Rai, N., Sardar, A., Yadav, A., Rai, D., Raj, A., Maji, B., Verma, S., Tripathi, A. K., Dhaniya, G., et al. (2025) · Frontiers in Endocrinology · doi:10.3389/fendo.2025.1540237 · Cited by 6
“Such properties and their ef fi cacy in estrogen-de fi cient states indicate that FWA (10% w/w) may be versatile therapeutic agent for generalised clinical use in managing bone metabolic disorders.”
The biomechanics of metaphyseal cone augmentation in revision knee replacement
Hu, J., Gundry, M., Zheng, K., Zhong, J., Hourigan, P. G., Meakin, J. R., Winlove, C. P., Toms, A. D., KNAPP, K. M., & Chen, J. (2022) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2022.105233 · Cited by 5
Nanomaterial-reinforced composites for biomedical implant applications: a mini-review
Kumar, M., Kumar, R., & Kumar, S. (2022) · Bioinspired Biomimetic and Nanobiomaterials · doi:10.1680/jbibn.21.00061 · Cited by 6
Higa, M., Nakayama, H., Kanto, R., Onishi, S., Yoshiya, S., Tachibana, T., & Iseki, T. (2023) · Journal of Orthopaedics · doi:10.1016/j.jor.2023.07.017 · Cited by 6
Solid-state foaming of poly(ether-ether-ketone)/hydroxyapatite composites
Zhu, H., Morquecho, E. V., & Li, W. (2023) · Journal of Cellular Plastics · doi:10.1177/0021955x231215762 · Cited by 5
Aurégan, J.-C., Bosser, C., Bachy, M., Bensidhoum, M., & Hoc, T. (2023) · Bioengineering · doi:10.3390/bioengineering10121338 · Cited by 5
Zhu, Y., Naseri, A. B., Brake, M. R. W., et al. (2024) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2024.1420870 · Cited by 5
“First, because of limitations in computational speed, we chose to model bone as a linear elastic material, even though bone has been shown to behave as a nonlinear viscoelastic material (Morgan et al., 2018).”
BackgroundAssessing osteoporosis and bone mineral density through 18F-NaF uptake at lumbar spine
Huang, J., Li, J., Li, Z., et al. (2024) · Annals of Nuclear Medicine · doi:10.1007/s12149-024-01982-w · Cited by 5
Ceddia, M., Romasco, T., Bortoli, N. D., et al. (2024) · Materials · doi:10.3390/ma17235680 · Cited by 5
Badami, A., Esmaeili, J., & Mirtalaie, H. (2025) · Gels · doi:10.3390/gels11010028 · Cited by 6
“Considering PCL, a concentration lower than 90% seemed to result in lower CMS and BM. Human bone tissues have different mechanical qualities based on things like age, gender, and where the bone is located [31].”
Kazemi, S., Jirofti, N., Arabzadeh, S., Mosaffa, F., Hashemi, M., & Kalalinia, F. (2025) · International Journal of Biological Macromolecules · doi:10.1016/j.ijbiomac.2025.146848 · Cited by 11
Antiepileptic Stiripentol May Influence Bones
Matuszewska, A., Nowak, B., Nikodem, A., Merwid‐Ląd, A., Wiatrak, B., Tomkalski, T., Jędrzejuk, D., Szeląg, E., Sozański, T., Danielewski, M., et al. (2021) · International Journal of Molecular Sciences · doi:10.3390/ijms22137162 · Cited by 7
“Whole-bone mechanical behavior depends on bone mass, size, geometry and the structure of bone [63].”
BackgroundKusins, J. R., Knowles, N. K., Targosinski, J., Columbus, M. P., Athwal, G. S., & Ferreira, L. M. (2021) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2021.104922 · Cited by 5
Laurent, C. P., Marano, A., Baldit, A., et al. (2022) · Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · doi:10.1177/09544119221109019 · Cited by 5
Rathor, S., Uddanwadiker, R., Saryam, N., & Apte, A. (2022) · Indian Journal of Orthopaedics · doi:10.1007/s43465-022-00795-1 · Cited by 5
Emerging hybrid particle-reinforced high-density polyethylene nanocomposite for bone replacement
Bello, S. A., Egbanubi, O. E., & Alabi, A. G. F. (2023) · Polymer Bulletin · doi:10.1007/s00289-023-04791-9 · Cited by 9
“Moreover, reports show that tensile yield stress of the cortical femoral bone is 71.56 ± 10.19 Nmm −2 , while its ultimate tensile strength varies from 92.95 ± 10.07 to 135 Nmm −2 [40–42], yield and ultimate tensile strains are 0.67 ± 0.04% and 1.9 ± 0.6% [40, 41], respectively.”
InfluentialJemaa, H., Eisenburger, M., & Greuling, A. (2023) · Computer Methods in Biomechanics & Biomedical Engineering · doi:10.1080/10255842.2023.2257345 · Cited by 5
Bazaluk, O., Chuzhak, A., Sulyma, V., et al. (2022) · Applied Sciences · doi:10.3390/app12104903 · Cited by 5
Santichatngam, P., Chongmuenwai, A., & Rooppakhun, S. (2024) · Scientific Reports · doi:10.1038/s41598-024-68662-7 · Cited by 4
“Generally, the push-off loading generates the highest implant stress with 115% of body weight applied 16,17 .”
BackgroundPrajapati, A. K., Varma, P. R. H., Kumar, G. S., et al. (2024) · Asian Spine Journal · doi:10.31616/asj.2024.0220 · Cited by 4
“Bone quality can be evaluated by measuring the vertebral body’s bone density [15,16] and the anatomic characteristics of the pedicles [17].”
BackgroundMethodInfluentialAuxetic Meta-Biomaterials: Computer Simulation and Experimental Results
Farber, E., Orlov, A. V., & Popovich, A. (2025) · Metals · doi:10.3390/met15030241 · Cited by 4
Dong, J., Li, R., Chen, Y., Zhu, G., & Liang, X. (2025) · Frontiers in Cell and Developmental Biology · doi:10.3389/fcell.2025.1607337 · Cited by 7
“3 The position-specific function of the Piezo family in bone Bone tissues are highly mechanosensitive and undergo adaptive remodeling in response to mechanical stimuli such as exercise and gravity (Morgan et al., 2018).”
Gabrieli, R., Schiavi, A., Pennecchi, F. R., Alidoost, D., Schwentenwein, M., D’Andrea, L., Vena, P., Verné, E., & Baino, F. (2025) · Journal of the American Ceramic Society · doi:10.1111/jace.70104 · Cited by 4
“53 The pore diameter of each TPMS type reveals that the pore sizes are in the range of few hundred micrometres.”
Structure–property–biocompatibility interplay in heat-treated WE43 magnesium alloy
Raj, V., Sabarinath, S., Vignesh, R. V., et al. (2025) · Materials Chemistry and Physics · doi:10.1016/j.matchemphys.2025.131623 · Cited by 4
Hybrid diacrylate resin-gelatin methacryloyl composite with bone-to-brain stiffness range
Zadeh, M. N., Patel, K. D., Gosden, D., Smith, J. A., Gates, P. J., Qi, Q., Scarpa, F. L., Conn, A. T., Perriman, A. W., & Rossiter, J. M. (2025) · Communications Materials · doi:10.1038/s43246-025-00931-y · Cited by 7
“On the stiff end of the composite, which is pure resin, its average elastic modulus of 1.4 GPa covers that of the trabecular bone and is close to the range of cortical bone ( ≈ 17.9 GPa) 52,53 .”
InfluentialGuo, X., Su, Y., Wu, Z., Wu, Z., Liu, K., Yue, J., Qu, Y., Wang, K., He, J., Ma, R., et al. (2025) · Chemical Engineering Journal · doi:10.1016/j.cej.2025.170004 · Cited by 5
Schöttner, J., Zeller‐Plumhoff, B., Hagenmuller, P., Weißgraeber, P., Rosendahl, P. L., Löwe, H., Schweizer, J., & van Herwijnen, A. (2025) · Acta Materialia · doi:10.1016/j.actamat.2025.121657 · Cited by 14
Maughan, J., Kaur, H., Prendeville, L., Carey, T., O’Connor, C., Synnatschke, K., Chávez, J. C. P., Woods, I., O’Brien, F. J., & Coleman, J. N. (2025) · Small Science · doi:10.1002/smsc.202500409 · Cited by 3
Wu, Z., Hou, Q., Liu, Y., Chen, T., Yang, K., Li, L., & Wang, L. Q. (2025) · Advanced Science · doi:10.1002/advs.202517885 · Cited by 6
Suitability of Mg-Nd and Mg-Zn Alloys to Obtain Biodegradable Structures for Bone Defects
Mănescu, V., Antoniac, A., Moraru, M.-C., Antoniac, I., Cotruț, C. M., Grădinaru, S., Dreancă, A., Sevastre, B., Pop, R., Tăbăran, F., et al. (2025) · Journal of Functional Biomaterials · doi:10.3390/jfb16110423 · Cited by 4
Kim, M., Nawaz, Q., Hartmann, B., Braun, P. H. D. R., Eulenkamp, C., Torgersen, J., Clausen‐Schaumann, H., Boccaccini, A. R., Holzapfel, B. M., & Mayer‐Wagner, S. (2026) · Chemical Engineering Journal · doi:10.1016/j.cej.2026.174600 · Cited by 5
Mechanical and biological properties of dual-topology fusion orthotropic bone scaffold
Tian, S., Zhang, Y., Gong, H., Zhang, X., Gao, J., & Zhou, L. (2026) · Biomaterials Advances · doi:10.1016/j.bioadv.2026.214825 · Cited by 6
Tantalum granules with hierarchical pore structure for bone regeneration
Pan, P., Hao, L., Tang, J., et al. (2024) · Journal of Materials Research and Technology · doi:10.1016/j.jmrt.2024.09.044 · Cited by 4
Integrated analysis of clinical indicators and mechanical properties in cancellous bone
He, J., Pan, Z., Zhou, G., Yu, J., & Li, D. (2024) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2024.104245 · Cited by 4
Evolution of Metals and Alloys in Orthopedics with Their Relevance in Osteoporosis
Mohapatra, N. C., Rana, R., & Sahoo, S. K. (2025) · Indian Journal of Orthopaedics · doi:10.1007/s43465-025-01344-2 · Cited by 4
Rocha, M. S., Marques, C. F., Pina, S., Oliveira, J. M., Reis, R. L., & Silva, T. H. (2025) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.4c01926 · Cited by 6
3D printed metal structures with inherent micro pores to enhance biodegradability and absorbability
Xie, D., Zhou, Q., He, Y., Lei, W., Liu, A., Sohan, H. M. S., OuYang, S., Zheng, Y., Jin, L., & Wen, P. (2025) · Additive manufacturing · doi:10.1016/j.addma.2025.104837 · Cited by 5
Xu, X., Golz, B. T., Flannery, B. T., Gallant, M. A., Bullock, W. A., Bellido, T., Nauman, E. A., Voytik‐Harbin, S. L., Little, D., & Main, R. P. (2025) · Biomaterials · doi:10.1016/j.biomaterials.2025.123451 · Cited by 6
A critical review on bone plates: state of the art and future directions
Sardana, G., Ghosh, S. B., & Bandyopadhyay‐Ghosh, S. (2025) · Journal of Biomaterials Science Polymer Edition · doi:10.1080/09205063.2025.2517717 · Cited by 4
Islam, M., Ordoño, J., Tang, W., Monclús, M. A., Echeverry‐Rendón, M., & Wang, D. (2025) · Small Structures · doi:10.1002/sstr.202500290 · Cited by 5
Jayash, S. N., Duff, T., Tanveer, Q., Promruk, W., & Farquharson, C. (2025) · Bone Reports · doi:10.1016/j.bonr.2025.101863 · Cited by 3
“In contrast, bone is the major reservoir for inorganic phosphate (P i ) where approximately 85 % of Pi is stored as calcium phosphate crystals (Morgan et al., 2018).”
Christensen, A., Rickman, J., & Berryman, H. (2021) · Forensic Anthropology · doi:10.5744/fa.2020.0041 · Cited by 4
Boffito, M., Servello, L., Arango‐Ospina, M., Miglietta, S., Tortorici, M., Sartori, S., Ciardelli, G., & Boccaccini, A. R. (2021) · Polymers · doi:10.3390/polym14010151 · Cited by 4
Smart Biomechanical Adaptation Revealed by the Structure of Ostrich Limb Bones
Conti, S., Sala, G., & Mateus, O. (2023) · Biomimetics · doi:10.3390/biomimetics8010098 · Cited by 8
“Bone, as a material, can be differentiated into the cortical bone (also known as compact bone), having a porosity of 5–15%, and the spongeous bone (also known as trabecular bone or cancellous bone), having a porosity of 40–95% [19].”
BackgroundFinite element analysis of proximal femur in sideways fall under quasi-static loading
Subramaniam, D., Mat, F., Majid, M. S. A., & Basaruddin, K. S. (2023) · Journal of Mechanical Science and Technology · doi:10.1007/s12206-023-2210-y · Cited by 4
“3318 ties is a key stage in a simulation in which bone is modelled as an inhomogeneous and isotropic material [24, 26, 32].”
BackgroundWu, H. C., Guo, Y., & Guo, W. (2023) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2023.1243936 · Cited by 5
“The most common drawbacks are the heavy weight due to high density, and the signi fi cant difference in elastic modulus between the metal implants (ranging from 100 to 200 GPa) and human bones (3 – 20 GPa).”
BackgroundInfluentialSatish, P., Hadagalli, K., Praveen, L. L., et al. (2023) · Inorganics · doi:10.3390/inorganics11110427 · Cited by 4
Alruqi, A. B., & Ongwen, N. O. (2024) · Inorganics · doi:10.3390/inorganics12020043 · Cited by 4
Zhou, Y., Helgason, B., Ferguson, S. J., et al. (2024) · Biomedical Engineering Advances · doi:10.1016/j.bea.2024.100115 · Cited by 4
Cioangher, M., Amarande, L., Stan, G. E., Nedelcu, L., Pasuk, I., Leonat, L., Popa, A. C., Miclea, L. C., Savopol, T., Moisescu, M. G., et al. (2024) · Ceramics International · doi:10.1016/j.ceramint.2024.05.268 · Cited by 4
Degradation of the mechanical properties of cortical bone due to long duration storage
Daras, N., Nurick, G. N., & Cloete, T. J. (2024) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2024.106632 · Cited by 4
Liu, X., Pei, J., Zhao, D., et al. (2024) · Biomedical Materials · doi:10.1088/1748-605x/ad5e2a · Cited by 4
“The compressive strength of S4 reached about 87.2 MPa (figure 5(A)), close to the bottom limit of human cortical bone strength (90–209 MPa) [28].”
InfluentialIbrahim, H. K., Abolarin, M. S., Abdulrahman, A. S., et al. (2024) · International Journal on Interactive Design and Manufacturing (IJIDeM) · doi:10.1007/s12008-024-01968-y · Cited by 4
“Human cortical bones have a compressive yield strength that ranges from 100 to 130 MPa, according to Morgan et al. [22].”
BackgroundYau, H., Chen, Q.-F., Huang, H., & Vu‐Dinh, H. (2024) · Materials & Design · doi:10.1016/j.matdes.2024.113277 · Cited by 5
Kalsi, S., Singh, J., Sharma, N. K., Kumar, R., Khatibi, A., Kedia, A., Mann, V. S., Bhowmik, A., & Santhosh, A. J. (2024) · Results in Engineering · doi:10.1016/j.rineng.2024.102979 · Cited by 5
Li, Y., Ma, J., Wang, J., Kong, Y., Wang, F., Zhang, P., & Fan, Y. (2024) · Biomedical Materials · doi:10.1088/1748-605x/ada241 · Cited by 6
“The porosity of cortical bone in the human body ranges from 5% to 15%, while that of cancellous bone ranges from 40% to 95% [48].”
Mukhtar, S., Kamran, M., Tayyeb, A., Hussain, F., Ishtiaq, M., Riaz, F., & Asghar, W. (2025) · Journal of Biological Physics · doi:10.1007/s10867-025-09672-y · Cited by 4
“Alloy A has depicted the average Young’s modulus of 34.60 GPa, which exceeds that of bone indicating significant stiffness [55].”
InfluentialMenshikh, K. A., Gobbo, V. A., Nascimben, M., Hannula, M., Cochis, A., Serra, T., Massera, J., Pandit, A. S., & Rimondini, L. (2025) · Materials Today Bio · doi:10.1016/j.mtbio.2025.101766 · Cited by 8
Radwan-Pragłowska, J., Śmietana, J., Janus, Ł., Sierakowska-Byczek, A., Łysiak, K., & Kuźmiak, K. (2025) · Processes · doi:10.3390/pr13072144 · Cited by 5
Femur morphology in healthy infants and young children
Bertocci, G. E., Brown, N. P., Thompson, A. K., et al. (2021) · Clinical Anatomy · doi:10.1002/ca.23825 · Cited by 4
“Currently, radiologists typically distinguish between healthy and unhealthy bone using subjective assessment of bone morphology and quality from radiographs (Mulugeta et al., 2011).”
BackgroundPatil, K. A., & Prashant, K. V. M. (2021) · IEEE Bombay Section Signature Conference (IBSSC) · doi:10.1109/ibssc53889.2021.9673249 · Cited by 4
Assessing the Elasticity of Child Cortical Bone
Baron, C., Follet, H., Pithioux, M., Payan, C., & Lasaygues, P. (2021) · Advances in experimental medicine and biology · doi:10.1007/978-3-030-91979-5_14 · Cited by 4
Mphaphuli, M. T., Sithole, M. N., Kumar, P., Kondiah, P. P. D., Mabrouk, M., & Choonara, Y. E. (2023) · Journal of Drug Delivery Science and Technology · doi:10.1016/j.jddst.2023.105039 · Cited by 4
Weerasooriya, T., & Alexander, S. (2021) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2021.104599 · Cited by 7
Cedeño-Viveros, L. D., Olivas-Alanis, L. H., Lopez-Botello, O. E., Rodriguez, C. A., Vázquez, E., & García‐López, E. (2022) · Engineering Science and Technology an International Journal · doi:10.1016/j.jestch.2022.101096 · Cited by 6
Studies on the Stress-Strain Relationship Bovine Cortical Bone Based on Ramberg–Osgood Equation
Sharma, N. K., Sarker, M. D., Naghieh, S., & Chen, D. (2019) · Journal of Biomechanical Engineering · doi:10.1115/1.4042901 · Cited by 4
The use of contravariant tensor invariants to model damage in anisotropic soft tissues
Basak, A., Rajagopal, A., & Basappa, U. (2021) · Mechanics of Advanced Materials and Structures · doi:10.1080/15376494.2021.1963019 · Cited by 4
“For instance it has been seen that the strength of bones vary largely in healthy and diseased or aged state [7].”
BackgroundPradhan, V., Ramachandra, R., Stammen, J. A., et al. (2024) · SAE International Journal of Transportation Safety · doi:10.4271/09-12-03-0014 · Cited by 3
Hodaei, M., Mandelis, A., & Cheung, A. M. (2025) · The Journal of the Acoustical Society of America · doi:10.1121/10.0036142 · Cited by 4
3D printing of an anatomically shaped bone model inspired by vascularized tubular bone structure
Ghobadi, E., Yahay, Z., Nouri, N., Karamali, F., & Masaeli, E. (2025) · Biomaterials Advances · doi:10.1016/j.bioadv.2025.214348 · Cited by 8
Zojaji, M., Yang, B., & Ploeg, H. (2025) · Journal of Biomechanical Engineering · doi:10.1115/1.4069110 · Cited by 3
Advances in bone organoids research and future perspectives
Liu, J., Li, J. Y., Chen, Y., Xu, F., Liu, W., Deng, 志. Z. 邓., Zhong, H., & Yang, H. (2025) · Biofabrication · doi:10.1088/1758-5090/adffb9 · Cited by 6
“Trabecular bone, by contrast, has a porosity of 40%–95%, which enhances shock absorption and stress distribution [21–23].”
Neufurth, M., Molter, D., La, X., Wu, C., Ushijima, H., Schröder, H. C., Wang, X., & Müller, W. E. G. (2025) · Biomedical Materials · doi:10.1088/1748-605x/ae084b · Cited by 4
“The elastic modulus of the polyP-κ -Car-Xan- β -TCP material was found to be ∼ 54.4 MPa, which is within the broad range observed for different skeletal regions [102].”
InfluentialKim, M., Cheers, G. M., Hartmann, B., Clausen‐Schaumann, H., Boccaccini, A. R., Holzapfel, B. M., & Mayer‐Wagner, S. (2025) · ACS Applied Bio Materials · doi:10.1021/acsabm.5c01304 · Cited by 5
Lei, P., Wang, S., Chen, K., Wang, L., Shen, Y., Zhong, H., Chen, S., Su, X., Zhao, Y., Wang, H., et al. (2025) · Materials Today Bio · doi:10.1016/j.mtbio.2025.102730 · Cited by 5
“In clinical practice, when fractures occur in weight-bearing bones, metal internal fixation devices remain essential for maintaining structural stability due to the significant load-bearing demands [77 – 79].”
Ganji, F., Mirzagoli, M., & Tayebi, L. (2026) · ADMET & DMPK · doi:10.5599/admet.3037 · Cited by 4
“Excessive loading and unloading may cause permanent deformation or fractures [9].”
Recent advances in 3D printing of metallic component-containing bone repairing materials
Xiong, M., Zeng, L., Ruan, W., Qiu, Y., Ye, X., Zhang, W., Wang, C., & Wang, M. (2026) · Metals Advances · doi:10.1016/j.metadv.2026.02.009 · Cited by 3
Enzyme-induced mineralization of calcium carbonate in 3D printable granular hydrogels
Bono, F., Puiggalí-Jou, A., Lucherini, L., et al. (2026) · Advanced Composites and Hybrid Materials · doi:10.1007/s42114-026-01662-5 · Cited by 3
Porous Zn scaffolds for bone regeneration: A synergy of experiment and artificial intelligence
Zhou, Q., Saberi, A., Shirmohammadi, Z., Campusano, C. F., Chai, Y., Wang, T., Sonsare, P. M., Gheisari, M., & Din, S. (2026) · Journal of Materials Research and Technology · doi:10.1016/j.jmrt.2026.03.078 · Cited by 3
Thinner femoral cortical thickness in patients with destructive rheumatoid arthritis of the knee
Kakutani, R., Kondo, N., Yamako, G., et al. (2023) · Journal of Orthopaedic Surgery and Research · doi:10.1186/s13018-023-04340-0 · Cited by 3
“However, bone strength is defined by BMD and bone quality, which includes structural and material characteristics of the trabecular and cortical bone [7, 8].”
BackgroundBiomechanical and histomorphological analysis of the mandible in rats with chronic kidney disease
Hu, T., Chen, J., Shao, S.-Q., et al. (2023) · Scientific Reports · doi:10.1038/s41598-023-49152-8 · Cited by 3
“The two groups of rats were subcutaneously injected with calcein (10 mg/kg) labeling on Days 13, 12, 3, and 2 before necropsy to allow fluorescent labeling of bone tissue for measurement of dynamic parameters 18–20 .”
MethodTanguay, A. P., Menon, N. G., Boudreau, M. H., Jastrzebski, S. L., Woods, P. S., Doyle, E. A., Edwards, W. B., Jay, G. D., Deymier, A. C., Lorenzo, J., et al. (2024) · Journal of Orthopaedic Research® · doi:10.1002/jor.25772 · Cited by 3
The anti-HIV drug abacavir stimulates β-catenin activity in osteoblast lineage cells
Olali, A. Z., Wallace, J., Gonzalez, H., Carpenter, K. A., Patel, N., Winchester, L. C., Podany, A. T., Venkatesh, I., Narasipura, S. D., Al‐Harthi, L., et al. (2024) · JBMR Plus · doi:10.1093/jbmrpl/ziae037 · Cited by 4
“The tissue density of adult bone is approximately 2.0 g/cm 3 or 2.0 g/mL. (34) Therefore,the tissue concentrations of each ARV tested are converted to g/mL and presented in Supplementary Table 1.”
BackgroundInfluentialZhu, Z., Yang, Y., Li, L., Zhu, S. J., & Zhang, L. (2024) · PLoS ONE · doi:10.1371/journal.pone.0299996 · Cited by 3
Diederichs, E. V., Mondal, D., & Willett, T. L. (2024) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2024.106694 · Cited by 3
Metzner, K. L., Fang, Q., Sanderson, R. W., Yeow, Y. L., Green, C., Abdul-Aziz, F., Hamzah, J., Mowla, A., & Kennedy, B. F. (2024) · APL Bioengineering · doi:10.1063/5.0220309 · Cited by 3
Does the Proximal Humerus Nail with 2 Distal Screws Provide Sufficient Rotational Stability?
An, K.-Y., Park, T. G., & Kim, M. Y. (2024) · Clinics in Orthopedic Surgery · doi:10.4055/cios23150 · Cited by 3
Soft hydrogel-embedded ceramic skeleton mimicking bone structure via sacrificial bond concept
Sukamto, S., Lama, M., Gong, J. P., & Nonoyama, T. (2024) · Soft Matter · doi:10.1039/d4sm01205d · Cited by 4
Influence of Implant Geometry on the Surface Strain Behavior of Peri‐Implant Bone: A 3D Analysis
Löhlein, M., Motel, C., Wichmann, M. G., et al. (2025) · Clinical Implant Dentistry and Related Research · doi:10.1111/cid.70003 · Cited by 3
Bio-inspired generative design of contact interfaces subjected to time-dependent loading conditions
Hernandez–Aristizabal, D., Arroyave-Tobón, S., Márquez-Flórez, K. M., et al. (2025) · Results in Engineering · doi:10.1016/j.rineng.2025.104340 · Cited by 3
Hernández, O. D., Baumer, V., Marrero, G., Karumanchi, S., & Prawel, D. A. (2025) · Journal of Functional Biomaterials · doi:10.3390/jfb16040119 · Cited by 5
“Future work should include in vitro cell studies and in vivo implantation to evaluate biomedical suitability.”
Bone Modelling and Remodelling in Cold Environment
Xue, L., Guan, Q., & Zhang, L. (2025) · Biomolecules · doi:10.3390/biom15040564 · Cited by 3
“Cortical porosity is negatively correlated with Young’s modulus [42,43].”
Nelea, V., Ittah, E., McKee, M. D., & Reznikov, N. (2025) · Acta Biomaterialia · doi:10.1016/j.actbio.2025.05.016 · Cited by 4
Zahedi, M., Taherkhani, A., Baserinia, R., Zahedi, S. A., Ali, H. M. A., & Abdi, M. (2025) · Macromolecular Materials and Engineering · doi:10.1002/mame.202500073 · Cited by 4
Levy, H. A., Magera, C. A., Messer, C. J., Potes, M. D. A., Allen, T., Kumar, J., Hamouda, A. M., Bydon, M., Fogelson, J. L., Elder, B. D., et al. (2025) · Global Spine Journal · doi:10.1177/21925682251356986 · Cited by 5
“33 This transition from predominately trabecular to predominately cortical bone loss is consistent with the increased risk for trabecular followed by cortical fractures in osteoporotic patients.”
Seifi, E., Mohseni, M., Samson, K. D. G., Cavelier, S., Ayyachi, T., & Hutmacher, D. W. (2025) · Journal of the American Ceramic Society · doi:10.1111/jace.70069 · Cited by 4
Putra, N. E., Xu, J., Leeflang, M. A., et al. (2025) · Materials Today Bio · doi:10.1016/j.mtbio.2025.102123 · Cited by 3
“More-over, after 16 weeks in vivo , the elastic modulus and yield strength of the FeMn-20Ak scaffolds were preserved and remained in a similar range to those of the human trabecular bone [43].”
Modelling the flexoelectric effect in human bone—A micromorphic approach
Titlbach, A., Papastavrou, A., McBride, A. T., & Steinmann, P. (2025) · Computer Methods in Applied Mechanics and Engineering · doi:10.1016/j.cma.2025.118234 · Cited by 3
Study on the morphological and metabolic changes of femur in laying hens with hypophosphatemia
Yang, N., Luo, D., Teng, X., Liu, Z., Wang, H., Ge, M., & Zhang, R. (2020) · Research in Veterinary Science · doi:10.1016/j.rvsc.2020.12.004 · Cited by 4
Rickman, J. M., Painter, J. D., & Hazael, R. (2021) · International Journal of Legal Medicine · doi:10.1007/s00414-021-02743-4 · Cited by 3
“Thus, whilst low strain rates allow bone to deform in the plastic part of the stress–strain curve, high strain rates apply energy too quickly for the viscous mechanisms to respond and the bone fractures in brittle fashion [2, 32, 43, 44].”
Background
Metadata
- Authors
- Morgan, E. F., Unnikrisnan, G. U., & Hussein, A. I.
- Affiliations
- Boston University (US)
- Corresponding author
- Morgan, Elise F.
- Venue
- Annual Review of Biomedical Engineering
- Publisher
- Annual Reviews
- Published
- 2018-06-04
- Volume
- 20
- Issue
- 1
- Pages
- 119-143
- ISSN
- 1523-9829 (ISSN-L)
- 1545-4274
- PMC
- PMC6053074
- OpenAlex
- W2807165335
- Semantic Scholar
- a169125cebb03ba7d2bb7a6ffa25b4106fcdfc55
- Cited by
- 1,026
- References
- 174
- Citations by year
- 2026: 163 · 2025: 223 · 2024: 194 · 2023: 150 · 2022: 136 · 2021: 114 · 2020: 37 · 2019: 8 · 2018: 1
- FWCI
- 8.21
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Annual Review of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- OpenBU (Boston University) (submittedVersion, cc-by)
- Funders
- National Science Foundation (1266243)
- National Institutes of Health (AR054620)
- National Institute of Arthritis and Musculoskeletal and Skin Diseases (R56AR054620, R01AR054620)
- Division of Civil, Mechanical and Manufacturing Innovation
- Topics
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
- Physical Sciences › Engineering › Biomedical Engineering › Bone Tissue Engineering Materials
- Physical Sciences › Engineering › Biomedical Engineering › Elasticity and Material Modeling
- Keywords
- bone mechanical properties, bone mechanics, osteoporosis, diabetes mellitus, cortical bone, trabecular bone, elasticity, yield, fracture, fatigue (material), damages, bone mineral density, porosity, crosslinking, aging
- MeSH terms
- Aging, Bone and Bones, Bone Diseases, Diabetes Mellitus, Elasticity, Humans, Models, Theoretical, Osteoporosis, Quality of Life, Stress, Mechanical, Viscosity, Bone Density, Porosity, Anisotropy, Fractures, Bone
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.