Radiation Dose and Image Quality in Pediatric CT: Effect of Technical Factors and Phantom Size and Shape

Siegel, Marilyn J., Schmidt, Bernhard, Bradley, David, Suess, Christoph, Hildebolt, Charles·2004·Radiology

Abstract

PURPOSE: To evaluate effects of varying tube current and voltage on radiation dose, image noise, and image contrast with different phantom sizes and shapes. MATERIALS AND METHODS: Four round lucite phantoms with 8-32-cm diameters were scanned with multi-detector row computed tomography (CT) and 80-120 kVp. Radiation dose was based on CT dose index, image noise, and iodine contrast and measured with constant and variable tube currents that were age appropriate for each tube voltage. Radiation dose and image noise and contrast were compared in round and oval 24-cm phantoms. For various combinations of technical factors and phantom sizes and shapes, percentage differences were calculated for radiation dose and image noise and contrast. Associations between tube voltage and radiation dose, image noise, and image contrast in round and oval phantoms were determined by fitting second-degree polynomials to data. Differences in radiation dose and image noise and contrast, which were attributable to differences in tube voltage, were tested with paired t tests. RESULTS: With 165-mAs tube current, radiation doses with 140- and 80-kVp tube voltages were 103% ([41.9 mGy - 20.6 mGy]/20.6 mGy) and 58% ([10.2 mGy - 4.2 mGy]/10.1 mGy) higher in the 8-cm phantom than in the 32-cm phantom. When tube current was adapted for phantom size, radiation dose at 80 kVp in the 8-cm phantom was reduced by 82% ([10.1 mGy - 1.8 mGy]/10.1 mGy). In the 8-cm phantom, tube voltage was decreased from 120 to 80 kVp and tube current remained at 165 mAs, resulting in a 68% noise increase ([3.1 HU - 1.8 HU]/1.8 HU). With variable tube current, 80-kVp tube voltage in the 8-cm phantom led to a 138% noise increase ([7.3 HU - 3.1 HU]/3.1 HU). With reduced tube voltage, image contrast increased. In the 8-cm phantom, with a constant 165-mAs tube current and a decrease in tube voltage from 120 to 80 kVp, there was a 35% ([333 HU - 217 HU]/333 HU) increase in contrast. No difference was noted in radiation dose or noise between round and oval phantoms (P = .604 and P = .06, respectively), but a small statistically significant difference (1%) in contrast attenuation was demonstrated (P = .025). CONCLUSION: Reduced tube voltage for pediatric contrast material-enhanced CT reduces radiation dose and maintains image contrast. Image noise increases, but the effect is minimal in smaller phantoms. An additional reduction in tube current further reduces radiation dose.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1148/radiol.2332032107

Cites

27 of 27 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 27 references

Cited by

See all 294

Metadata

Authors
Siegel, M. J., Schmidt, B., Bradley, D., Suess, C., & Hildebolt, C.
Venue
Radiology
Published
2004-11
Volume
233
Issue
2
Pages
515-522
ISSN
  • 0033-8419 (ISSN-L)
  • 1527-1315
OpenAlex
W2147913236
Cited by
294
References
27
FWCI
8.02
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Radiation Dose and Image Quality in Pediatric CT: Effect of Technical Factors and Phantom Size and Shape. https://app.cochat.ai/papers/10.1148/radiol.2332032107
The paper
Siegel, M. J., Schmidt, B., Bradley, D., Suess, C., & Hildebolt, C. (2004). Radiation Dose and Image Quality in Pediatric CT: Effect of Technical Factors and Phantom Size and Shape. Radiology. https://doi.org/10.1148/radiol.2332032107

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.