MR Imaging and Cardiac Pacemakers: In Vitro Evaluation and in Vivo Studies in 51 Patients at 0.5 T
The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.
Abstract
PURPOSE: To evaluate the safety and feasibility of magnetic resonance (MR) imaging at 0.5 T in patients with implanted cardiac pacemakers. MATERIALS AND METHODS: Twenty-one models of pacemakers and 44 pacemaker electrodes were exposed to in vitro MR imaging with continuous registration of pacemaker output and temperature at the lead tip. Prior to MR imaging examination, pacemakers were programmed to an asynchronous mode (A00, V00, or D00). Pacemakers were examined before and after MR imaging. Forty-four patients with implanted pacemakers underwent 51 MR imaging examinations under cardiologic surveillance, continuous electrocardiography, pulse oximetry, and capnographic monitoring. RESULTS: MR imaging was safely performed in all patients. None of the pacemakers displayed a pacing dysfunction at MR imaging. No changes occurred in the programmed parameters in any device tested in vivo or in vitro. Maximum increases in the temperature at the lead tips were 8.90 degrees C at a specific absorption rate (SAR) of 0.6 W/kg and 23.50 degrees C under a worst-case radio-frequency (RF) heating condition with an SAR of 1.3 W/kg. CONCLUSION: MR imaging at 0.5 T can be safely performed in patients with implanted pacemakers in carefully selected clinical circumstances when appropriate strategies (programming to an asynchronous mode, adequate monitoring techniques, limited RF exposure) are used.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Gradient coil design: A review of methods
Turner, R. (1993) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(93)90209-v · Cited by 418
Recording of EEG during fMRI experiments: Patient safety
Lemieux, L., Allen, P. J., Franconi, F., Symms, M. R., & Fish, D. K. (1997) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910380614 · Cited by 311
Safety considerations in MR imaging.
Kanal, E., Shellock, F. G., & Talagala, L. (1990) · Radiology · doi:10.1148/radiology.176.3.2202008 · Cited by 302
Effects of magnetic resonance imaging on cardiac pacemakers and electrodes
Achenbach, S. S., Moshage, W. E. L., Diem, B., Bieberlea, T., Schibgilla, V., & Bachmann, K. F. (1997) · American Heart Journal · doi:10.1016/s0002-8703(97)70083-8 · Cited by 254
Radio frequency electromagnetic exposure: Tutorial review on experimental dosimetry
Chou, C., Bassen, H., Osepchuk, J. M., Balzano, Q., Petersen, R., Meltz, M. L., Cleveland, R. F., Lin, J. C.-J., & Heynick, L. N. (1996) · Bioelectromagnetics · doi:10.1002/(sici)1521-186x(1996)17:3<195::aid-bem5>3.0.co;2-z · Cited by 206
Temperature monitoring during radiofrequency catheter ablation of accessory pathways.
Langberg, J. J., Calkins, H., el-Atassi, R., Borganelli, M., Leon, A., Kalbfleisch, S. J., & Morady, F. (1992) · Circulation · doi:10.1161/01.cir.86.5.1469 · Cited by 179
Estimating radiofrequency power deposition in body NMR imaging
Bottomley, P. A., Redington, R. W., Edelstein, W. A., & Schenck, J. F. (1985) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910020404 · Cited by 171
Power deposition in whole-body NMR imaging
Bottomley, P. A., & Edelstein, W. A. (1981) · Medical Physics · doi:10.1118/1.595000 · Cited by 147
Safe Performance of Magnetic Resonance Imaging on Five Patients with Permanent Cardiac Pacemakers
Gimbel, J., JOHNSON, D., LEVINE, P. A., & Wilkoff, B. L. (1996) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1996.tb03387.x · Cited by 145
Effect of magnetic resonance imaging on DDD pacemakers
Erlebacher, J. A., Cahill, P. T., Pannizzo, F., & Knowles, R. (1986) · The American Journal of Cardiology · doi:10.1016/0002-9149(86)90768-x · Cited by 138
Cited by
WRITING COMMITTEE MEMBERS, Hundley, W. G., Bluemke, D. A., Finn, J. P., Flamm, S. D., Fogel, M. A., Friedrich, M. G., Ho, V. B., Jerosch-Herold, M., Kramer, C. M., et al. (2010) · Circulation · doi:10.1161/cir.0b013e3181d44a8f · Cited by 821
ACCF/ACR/AHA/NASCI/SCMR 2010 Expert Consensus Document on Cardiovascular Magnetic Resonance
Hundley, W. G., Bluemke, D. A., Finn, J. P., Flamm, S. D., Fogel, M. A., Friedrich, M. G., Ho, V. B., Jerosch-Herold, M., Kramer, C. M., Manning, W. J., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.11.011 · Cited by 593
MR Procedures: Biologic Effects, Safety, and Patient Care
Shellock, F. G., & Crues, J. V. (2004) · Radiology · doi:10.1148/radiol.2323030830 · Cited by 563
Levine, G. N., Gomes, A. S., Arai, A. E., Bluemke, D. A., Flamm, S. D., Kanal, E., Manning, W. J., Martin, E. T., Smith, J. M., Wilke, N., et al. (2007) · Circulation · doi:10.1161/circulationaha.107.187256 · Cited by 489
Indik, J. H., Gimbel, J. R., Abe, H., Alkmim-Teixeira, R., Birgersdotter-Green, U., Clarke, G. D., Dickfeld, T.-M. L., Froelich, J. W., Grant, J., Hayes, D. L., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.04.025 · Cited by 438
Metadata
- Authors
- Sommer, T., Vahlhaus, C., Lauck, G., Smekal, A. V., Reinke, M., Hofer, U., Block, W., Träber, F., Schneider, C., Gieseke, J., Jung, W., & Schild, H.
- Venue
- Radiology
- Published
- 2000-06
- Volume
- 215
- Issue
- 3
- Pages
- 869-879
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W2013634519
- Cited by
- 244
- References
- 10
- FWCI
- 4.46
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.