Extraction of Heart Rate Variability from Smartphone Photoplethysmograms
Open accessThe DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.
Abstract
Heart rate variability (HRV) is a useful clinical tool for autonomic function assessment and cardiovascular diseases diagnosis. It is traditionally calculated from a dedicated medical electrocardiograph (ECG). In this paper, we demonstrate that HRV can also be extracted from photoplethysmograms (PPG) obtained by the camera of a smartphone. Sixteen HRV parameters, including time-domain, frequency-domain, and nonlinear parameters, were calculated from PPG captured by a smartphone for 30 healthy subjects and were compared with those derived from ECG. The statistical results showed that 14 parameters (AVNN, SDNN, CV, RMSSD, SDSD, TP, VLF, LF, HF, LF/HF, nLF, nHF, SD1, and SD2) from PPG were highly correlated (r>0.7,P<0.001) with those from ECG, and 7 parameters (AVNN, TP, VLF, LF, HF, nLF, and nHF) from PPG were in good agreement with those from ECG within the acceptable limits. In addition, five different algorithms to detect the characteristic points of PPG wave were also investigated: peak point (PP), valley point (VP), maximum first derivative (M1D), maximum second derivative (M2D), and tangent intersection (TI). The results showed that M2D and TI algorithms had the best performance. These results suggest that the smartphone might be used for HRV measurement.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Heart Rate Variability: Standards of Measurement, Physiological Interpretation, and Clinical Use
Electrophysiology, T. F. O. T. E. S. O. C. T. N. A. (1996) · Circulation · doi:10.1161/01.cir.93.5.1043 · Cited by 17,159
Heart rate variability: Standards of measurement, physiological interpretation, and clinical use
Malik, M., Bigger, J. T., Camm, A. J., Kleiger, R. E., Malliani, A., Moss, A. J., & Schwartz, P. J. (1996) · European Heart Journal · doi:10.1093/oxfordjournals.eurheartj.a014868 · Cited by 15,289
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855 · cited 2×
“The ECG signals were first passed through a finite impulse response (FIR) low-pass filter with cutoff frequency of 11Hz and then a FIR high-pass filter with cutoff frequency of 5Hz to reduce most of the noise and interference [26].”
MethodAssessment of autonomic function in humans by heart rate spectral analysis
Pomeranz, B., Macaulay, R. J., Caudill, M. A., Kutz, I., Adam, D., Gordon, D., Kilborn, K. M., Barger, A. C., Shannon, D. C., Cohen, R. J., et al. (1985) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1985.248.1.h151 · Cited by 2,876 · cited 2×
“It reflects the tenseness and the balance of the sympathetic and the vagus nerve activities and their effects on cardiovascular motion [1].”
BackgroundMethodIstepanian, R. S. H., Jovanov, E., & Zhang, Y. T. (2004) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2004.840019 · Cited by 766 · cited 1×
“Mobile phones have already shown promising applications in healthcare service [15].”
BackgroundSelf-citationHow accurate is pulse rate variability as an estimate of heart rate variability?
Schäfer, A., & Vagedes, J. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.03.119 · Cited by 661 · cited 2×
“Agreements were ranked as good (BAR < 10%), moderate (10% ≤ BAR < 20%), or insufficient (BAR ≥ 20%) [28].”
BackgroundGil, E., Orini, M., Bailón, R., Vergara, J. M., Mainardi, L., & Laguna, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/9/015 · Cited by 461
Physiological parameter monitoring from optical recordings with a mobile phone
Scully, C. G., Lee, J., Meyer, J., Gorbach, A. M., Granquist-Fraser, D., Mendelson, Y., & Chon, K. H. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2163157 · Cited by 452 · cited 1×
“PPG signals can thus be obtained by calculating the intensity changes of the frames, and then several physiological parameters such asHR [17–21], respiratory rate [19], pulse volume [21], and oxygen saturation [19] can be extracted from the PPG signals.”
BackgroundThe effect of local mild cold exposure on pulse transit time
Zhang, X.-Y., & Zhang, Y.-T. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/7/008 · Cited by 40
Cited by
Reminder: RMSSD and SD1 are identical heart rate variability metrics
Ciccone, A. B., Siedlik, J. A., Wecht, J. M., Deckert, J. A., Nguyen, N. D., & Weir, J. P. (2017) · Muscle & Nerve · doi:10.1002/mus.25573 · Cited by 173
“Indeed, the inferential statistics show identical P-values,(19,20,26) effect sizes,(27,28) and Pearson r correlation coefficients.(17,19,29) Anecdotally, we have been queried at a conference by multiple attendees as to why we did not report both RMSSD and SD1 in our analysis.”
BackgroundMejía-Mejía, E., Budidha, K., Abay, T. Y., May, J. M., & Kyriacou, P. A. (2020) · Frontiers in Physiology · doi:10.3389/fphys.2020.00779 · Cited by 171
An Ultra Low Power Personalizable Wrist Worn ECG Monitor Integrated With IoT Infrastructure
Beach, C., Krachunov, S., Pope, J., Fafoutis, X., Piechocki, R. J., Craddock, I., & Casson, A. J. (2018) · IEEE Access · doi:10.1109/access.2018.2864675 · Cited by 91
Analysis of a Pulse Rate Variability Measurement Using a Smartphone Camera
Bánhalmi, A., Borbás, J., Fidrich, M., Bilicki, V., Gingl, Z., & Rudas, L. (2018) · Journal of Healthcare Engineering · doi:10.1155/2018/4038034 · Cited by 69
“Another goal of this study was to compare our PRV measurements with those HRV parameters computed from the gold standard ECG signals and also to investigate our comparison results among other experiment results like those of [12, 13, 21].”
BackgroundMethodResultInfluentialSurgeons' and surgical trainees' acute stress in real operations or simulation: A systematic review
Georgiou, K., Larentzakis, A., & Papavassiliou, A. G. (2017) · The Surgeon · doi:10.1016/j.surge.2017.06.003 · Cited by 69
Metadata
- Authors
- Peng, R.-C., Zhou, X.-L., Lin, W.-H., & Zhang, Y.-T.
- Venue
- Computational and Mathematical Methods in Medicine
- Publisher
- Wiley
- Published
- 2015
- Volume
- 2015
- Pages
- 1-11
- ISSN
- 1748-670X (ISSN-L)
- 1748-6718
- OpenAlex
- W1968864692
- Cited by
- 149
- References
- 10
- FWCI
- 6.61
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Extraction of Heart Rate Variability from Smartphone Photoplethysmograms. https://app.cochat.ai/papers/10.1155/2015/516826
- The paper
- Peng, R.-C., Zhou, X.-L., Lin, W.-H., & Zhang, Y.-T. (2015). Extraction of Heart Rate Variability from Smartphone Photoplethysmograms. Computational and Mathematical Methods in Medicine. https://doi.org/10.1155/2015/516826
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.