Catheter Ablation for Ventricular Tachycardia
Open accessAbstract
S ustained ventricular tachycardia (VT) is an important cause of morbidity and sudden death in patients with heart disease. 1 Implantable cardioverter-defibrillators (ICDs) terminate VT episodes, reducing the risk of sudden death.Recurrent VT develops in 40% to 60% of patients who receive an ICD after an episode of spontaneous sustained VT.A first episode of VT occurs in Ϸ20% of patients within 3 to 5 years after ICD implantation for primary prevention of sudden death in high-risk groups.2-4 ICD shocks reduce quality of life and are associated with an increased risk of death.[2][3][4] Antiarrhythmic drug therapy with amiodarone or sotalol reduces VT episodes but with disappointing incidence of side effects and efficacy.2 Catheter ablation is useful for reducing VT episodes and can be life-saving when VT is incessant.1,5,6 Idiopathic VTs occur in patients without structural heart disease and rarely cause sudden death.Electrophysiological study with catheter ablation is often warranted to confirm the diagnosis, to provide further evidence for the absence of ventricular scar or other disease, and often to cure the arrhythmia.Ablation is also an option for symptomatic nonsustained VT and frequent ventricular ectopy in these patients.1 The appearance of the VT on ECG often suggests its likely cause and associated heart disease (Figure 1).Monomorphic VT has the same QRS complex from beat to beat, indicating repetitive ventricular activation from a structural substrate or focus that can be targeted for ablation.Most are due to reentry through regions of ventricular scar.7 Polymorphic VTs have a changing ventricular activation sequence that can be due to functional reentry without a structural target for ablation; myocardial ischemia, acquired long-QT syndrome, and a number of genetic syndromes are causes.Rarely, ablation of the initiating focus is warranted to control frequent episodes.8 Technical Considerations and ComplicationsThe ablation procedure proceeds in steps.After vascular access is obtained, the arrhythmia is induced to confirm the diagnosis and to establish whether abolishing inducible VT is a potential end point for the procedure.Mapping is performed to locate the source, followed by ablation.Testing is then performed to assess the effect of ablation. MappingWhen VT is hemodynamically stable, mapping can be performed during VT.Many patients with heart disease have VTs that are "unmappable" or "unstable" because they require immediate termination because of hemodynamic intolerance, they cannot be reliably induced, or they frequently change from one VT to another.Substrate mapping or multielectrode mapping approaches have been developed to guide ablation of these VTs.Mapping systems that recreate the geometry of the ventricles from point-by-point sampling while providing continuous display of catheter position are in common use.9,10 Electrophysiological data such as the activation time and electrogram amplitude are color-coded for display (Figure 2).Cardiac motion introduces some error.Papillary muscles and other fine anatomic details are not visible.Registration of preacquired computed tomographic or magnetic resonance images on mapping systems shows promise for improving anatomic definition.9,11 Balloon or basket electrode arrays sample from multiple sites simultaneously, potentially allowing activation to be defined from a single beat or brief run of VT.A "noncontact" mapping system mathematically reconstructs potentials from adjacent sites.12,13 Go to Images in Cardiovascular Medicine for related image, "Percutaneous Pericardial Access for Mapping and Ablation of Epicardial Ventricular Tachycardias" by Sosa et al (Circulation.2007;115:e542-e544).
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Zipes, D. P., Camm, A. J., Borggrefe, M., Buxton, A. E., Chaitman, B., Fromer, M., Gregoratos, G., Klein, G., Myerburg, R. J., Quinones, M. A., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.07.010 · Cited by 1,973
Marchlinski, F. E., Callans, D. J., Gottlieb, C. D., & Zado, E. (2000) · Circulation · doi:10.1161/01.cir.101.11.1288 · Cited by 1,220
Prophylactic Catheter Ablation for the Prevention of Defibrillator Therapy
Reddy, V. Y., Reynolds, M. R., Neuzil, P., Richardson, A. W., Taborsky, M., Jongnarangsin, K., Kralovec, S., Sediva, L., Ruskin, J. N., & Josephson, M. E. (2007) · New England Journal of Medicine · doi:10.1056/nejmoa065457 · Cited by 959
Stevenson, W. G., Khan, H., Sager, P., Saxon, L. A., Middlekauff, H. R., Natterson, P. D., & Wiener, I. (1993) · Circulation · doi:10.1161/01.cir.88.4.1647 · Cited by 909
Self-citationSlow conduction in the infarcted human heart. 'Zigzag' course of activation.
de Bakker, J. M., van Capelle, F. J., Janse, M. J., Tasseron, S., Vermeulen, J. T., de Jonge, N., & Lahpor, J. R. (1993) · Circulation · doi:10.1161/01.cir.88.3.915 · Cited by 842
Stevenson, W. G., Wilber, D. J., Natale, A., Jackman, W. M., Marchlinski, F. E., Talbert, T., Gonzalez, M. D., Worley, S. J., Daoud, E. G., Hwang, C., et al. (2008) · Circulation · doi:10.1161/circulationaha.108.788604 · Cited by 732
Self-citationMapping and Ablation of Idiopathic Ventricular Fibrillation
Haïssaguerre, M., Shoda, M., Jaïs, P., Nogami, A., Shah, D. C., Kautzner, J., Arentz, T., Kalushe, D., Lamaison, D., Griffith, M., et al. (2002) · Circulation · doi:10.1161/01.cir.0000027564.55739.b1 · Cited by 686
Schron, E. B., Exner, D. V., Yao, Q., Jenkins, L. S., Steinberg, J. S., Cook, J. R., Kutalek, S. P., Friedman, P. L., Bubien, R. S., Page, R. L., et al. (2002) · Circulation · doi:10.1161/hc0502.103330 · Cited by 624
Moss, A. J., Greenberg, H., Case, R. B., Zareba, W., Hall, W. J., Brown, M. W., Daubert, J. P., McNitt, S., Andrews, M. L., & Elkin, A. D. (2004) · Circulation · doi:10.1161/01.cir.0000150390.04704.b7 · Cited by 597
Cited by
Flexible and Stretchable Electronics for Biointegrated Devices
Kim, D.-H., Ghaffari, R., Lu, N., & Rogers, J. A. (2012) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-071811-150018 · Cited by 749
BUNCH, T. J., MAHAPATRA, S., MURDOCK, D., MOLDEN, J., WEISS, J. P., MAY, H. T., BAIR, T. L., MADER, K. M., CRANDALL, B. G., DAY, J. D., et al. (2011) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2011.03208.x · Cited by 101
Metadata
- Authors
- Stevenson, W. G., & Soejima, K.
- Venue
- Circulation
- Published
- 2007-05-29
- Volume
- 115
- Issue
- 21
- Pages
- 2750-2760
- ISSN
- 0009-7322 (ISSN-L)
- 1524-4539
- OpenAlex
- W2142799840
- Cited by
- 299
- References
- 93
- FWCI
- 12.53
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Catheter Ablation for Ventricular Tachycardia. https://app.cochat.ai/papers/10.1161/circulationaha.106.655720
- The paper
- Stevenson, W. G., & Soejima, K. (2007). Catheter Ablation for Ventricular Tachycardia. Circulation. https://doi.org/10.1161/circulationaha.106.655720
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.