Laboratory-Based Non-Clinical Comparison of Occlusion Rates Using Three Rapid-Acting Insulin Analogs in Continuous Subcutaneous Insulin Infusion Catheters Using Low Flow Rates
Abstract
Background: Rapid-acting analog insulin is used increasingly for continuous subcutaneous insulin infusion therapy (CSII). As the choice of insulin may be a determinant of catheter occlusion, we compared rates of early and late occlusion of a standard CSII catheter with three insulin analogs in a laboratory-based setting. Methods: Twenty-four pumps were used for the study. Each insulin analog (glulisine, lispro, and aspart) was assigned to eight pumps in a randomized order for each of nine runs of 5-day duration. Pumps were primed to receive a basal dose of 0.1 IU/h with a bolus dose of 2 IU given three times each day. Pumps were placed in an incubator to maintain temperature in the range of 32 to 36 °C. Results: Over the entire study period, there were 48 occlusions. Early occlusions (within 72 hours) occurred during five of the nine runs with no evidence of any difference between insulins ( p = .27); there were no occlusions before 48 hours. Over the whole of the 5-day infusion period, the probabilities of overall occlusion for each insulin were 40.9% [28 to 55%, 95% confidence interval (CI)] for glulisine, 9.2% (4 to 19.5%, 95% CI) for aspart, and 15.7% (8.1 to 28.1%, 95% CI) for lispro. All occlusions, except for three, occurred during a bolus infusion. Conclusions: During CSII under laboratory conditions, early catheter occlusions (within 72 hours) are rare and independent of the choice of insulin analog. For patients using insulin pump therapy, the importance of catheter change within 72 hours should be emphasized irrespective of the insulin used. Beyond 72 hours, the risk of occlusion differs between insulins, being more common with glulisine.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Standard insulin infusion sets should be changed every 2 to 3 days to prevent or reduce adverse events.
“Standard IISs should be changed every 2 to 3 days,1–3 regardless of rapid-acting insulin used,4 to prevent or reduce adverse events (e.g., site failure resulting in hyperglycemia, skin inflammation, infection or pain at the IIS site, adhesive patch failure, etc.).”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [4]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Bode, B., Weinstein, R., Bell, D., McGill, J., Nadeau, D., Raskin, P., Davidson, J., Henry, R., Huang, W.-C., & Reinhardt, R. R. (2002) · Diabetes Care · doi:10.2337/diacare.25.3.439 · Cited by 148 · see citing works
Flagged first, then by citations
Research works printed but not linked: 8
- [1]1. - Plank J, Siebenhofer A, Berghold A, Jeitler K, Horvath K, Mrak P, Pieber TR. Systematic review and meta-analysis of short-acting insulin analogues in patients with diabetes mellitus. Arch Intern Med. 2005;165(12):1337-44.
- [3]3. - Guilhem I, Leguerrier AM, Lecordier F, Poirier JY, Maugendre D. Technical risks with subcutaneous insulin infusion. Diabetes Metab. 2006;32(3):279-84.
- [4]4. - Weissberg-Benchell J, Antisdel-Lomaglio J, Seshadri R. Insulin pump therapy: a meta-analysis. Diabetes Care. 2003;26(4):1079-87.
- [6]6. - Wolpert HA, Faradji RN, Bonner-Weir S, Lipes MA. Metabolic decompensation in pump users due to lispro insulin precipitation. BMJ. 2002;324(7348):1253.
- [7]7. - Hoogma RP, Schumicki D. Safety of insulin glulisine when given by continuous subcutaneous infusion using an external pump in patients with type 1 diabetes. Horm Metab Res. 2006;38(6):429-33.
- [8]8. - DeFelippis MR, Bell MA, Heyob JA, Storms SM. In vitro stability of insulin lispro in continuous subcutaneous insulin infusion. Diabetes Technol Ther. 2006;8(3):358-68.
- [9]9. - Senstius J, Harboe E, Westermann H. In vitro stability of insulin aspart in simulated continuous subcutaneous insulin infusion using a MiniMed 508 insulin pump. Diabetes Technol Ther. 2007;9(1):75-9.
- [10]10. - Brange J, Andersen L, Laursen ED, Meyn G, Rasmussen E. Toward understanding insulin fibrillation. J Pharma Sci. 1997;86(5):517-25. 11.
Cited by
Sorted by citations
Insulin analogues in type 1 diabetes mellitus: Getting better all the time
Mathieu, C., Gillard, P., & Benhalima, K. (2017) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2017.39 · Cited by 248 · see citing works
Insulin Infusion Set: The Achilles Heel of Continuous Subcutaneous Insulin Infusion
Heinemann, L., & Krinelke, L. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600429 · Cited by 148 · see citing works
Nonmetabolic Complications of Continuous Subcutaneous Insulin Infusion: A Patient Survey
Pickup, J. C., Yemane, N., Brackenridge, A., & Pender, S. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0192 · Cited by 123 · see citing works
Schmid, V., Hohberg, C., Borchert, M., Forst, T., & Pfützner, A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400429 · Cited by 97 · see citing works
van Bon, A. C., Bode, B. W., Sert-Langeron, C., DeVries, J. H., & Charpentier, G. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0224 · Cited by 96 · see citing works
Pfützner, A., Sachsenheimer, D., Grenningloh, M., Heschel, M., Walther-Johannesen, L., Gharabli, R., & Klonoff, D. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815604438 · Cited by 53 · see citing works
Brazg, R., Garg, S. K., Bhargava, A., Thrasher, J. R., Latif, K., Bode, B. W., Bailey, T. S., Horowitz, B. S., Cavale, A., Kudva, Y. C., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0540 · Cited by 28 · see citing works
- Authors
- Kerr, D., Morton, J., Whately-Smith, C., Everett, J., & Begley, J. P.
- Affiliations
- Bournemouth Diabetes and Endocrine Centre, Royal Bournemouth Hospital NHS Foundation Trust, Dorset, United Kingdom;
- Department of Clinical Biochemistry, Royal Bournemouth Hospital NHS Foundation Trust, Dorset, United Kingdom;
- Whately-Smith Ltd., Hertfordshire, United Kingdom;
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2008-05
- Volume
- 2
- Issue
- 3
- Pages
- 450-455
- ISSN
- 1932-2968
- Cited by
- 55
- References
- 1
- Language
- en
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review