Pilot Study for Assessment of Optimal Frequency for Changing Catheters in Insulin Pump Therapy—Trouble Starts on Day 3

Open access
Schmid, Volkmar, Hohberg, Cloth, Borchert, Marcus, Forst, Thomas, Pfützner, Andreas·2010·Journal of Diabetes Science and Technology

Abstract

Background: Continuous subcutaneous insulin infusion (CSII) by means of insulin pump devices is considered to be one of the most optimal therapies to achieve treatment targets in patients with diabetes mellitus. In CSII, the insulin is delivered through Teflon catheters or steel needle infusion sets, which need to be renewed on a regular basis. This pilot study was performed to investigate the optimal change frequency in daily practice and to explore potential problems that may occur when the sets are used for a more prolonged time than the recommended up to 72 hours of usage (Teflon catheters). Method: Twelve patients with type 1 diabetes participated in the trial [age (mean ± STD): 40.3 ± 12.6 years, body mass index: 26.2 ± 3.3 kg/m 2 , hemoglobin A1c: 6.7 ± 0.6%]. They were asked to wear their infusion set (Comfort™ or Silhouette®) for increasing periods of 1, 2, 3, 4, and 5 days. After each use, patients completed standardized questionnaires regarding technical and medical issues associated with infusion set use. A health care professional investigated the infusion sites and infusion sets and completed an “infusion set inspection” questionnaire. Blood glucose was measured and recorded to assess a potential Influence of duration of catheter use on glycemic control. Results: Infusion set and injection site problems (itching, bruising, swelling, and pain) started to occur in measurable amounts on the 3rd day of catheter use, and about 40% of patients reported significant issues when using a catheter for 5 days. In parallel, there was a consistent increase in mean daily blood glucose levels that correlated with the number of days of catheter use (e.g., day 1: 7. 5 ± 3.8 mmol/liter, day 3: 8.4 ± 4.2 mmol/liter, day 5: 9.0 ± 4.0 mmol/liter, day 7: 11.6 ± 2.2 mmol/liter, p < 0.05 vs day 1). Conclusions: Using the catheters for 2 days resulted in a safe and well-tolerated therapy. Clinically relevant adverse events started to occur during the 3rd day and their incidence increased constantly with longer use. This was associated with undesired changes in mean glycemic control. Data support the recommendation by the drug and device manufacturers that insulin pump catheters should only be used for 48–72 hours to avoid adverse events and potential metabolic deterioration.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/193229681000400429

Cites

24 of 24 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • IIS failure is most often attributed to cannula-associated issues (kinking, occlusion, and dislodgement) or insertion site inflammation/infection.

    “IIS failure is most often attributed to cannula-associated issues (kinking, occlusion, and dislodgement) or insertion site inflammation/infection.2,6,7”

    From Frequency and Detection of Insulin Infusion Site Failure in the Type 1 Diabetes Exchange Online Community (2023) · cites this paper as [6]

    The Discussion lists kinking, pulling out, occlusion, and skin irritation among infusion-set problems, and says infection can occur. But it does not establish that the claimed categories are the most frequent causes of IIS failure; it calls infection rare, and the study’s most frequent reason for premature set changes was leakage (Patient Characteristics; page unavailable). (LLM, conf=0.86)

    Experimental — accuracy not yet measured
  • Despite recent improvements in CSII catheter design and insertion method, the CSII catheter remains the weak link of an insulin infusion system for the management of type 1 diabetes.

    “Despite recent improvements in CSII catheter design and method of insertion, the CSII catheter remains the weak link of an insulin infusion system for the management of type 1 diabetes.”

    From Detailed Analysis of Insulin Absorption Variability and the Tissue Response to Continuous Subcutaneous Insulin Infusion Catheter Implantation in Swine (2017) · cites this paper as [6]

    The Abstract (page unavailable) reports that infusion-set and injection-site problems began to occur in measurable amounts on day 3 in patients with type 1 diabetes. The Discussion (page unavailable) also describes catheter-related disruptions to insulin delivery. Our copy does not establish that catheter design or insertion methods had recently improved, or compare the catheter with other components to call it the system’s weak link. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • Patients routinely insert a new CSII catheter every 2-3 days because insulin absorption into the circulation becomes more variable and less reliable over time.

    “Patients routinely insert a new catheter every 2–3 days due to increasing variability of insulin absorption over time.”

    From Detailed Analysis of Insulin Absorption Variability and the Tissue Response to Continuous Subcutaneous Insulin Infusion Catheter Implantation in Swine (2017) · cites this paper as [6]

    The Introduction recommends changing infusion sets every 48–72 hours, and the Discussion says prolonged use appeared to change insulin absorption and reduce glycemic control. But the study measured blood glucose, not variability in insulin absorption, and does not establish that increasing absorption variability is why patients routinely change catheters. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • Patients commonly experience an insulin bolus that fails to produce the expected glucose lowering effect, leading to persistent hyperglycemia.

    “Patients commonly experience an insulin bolus that fails to produce the expected glucose lowering effect, leading to persistent hyperglycemia.6,19”

    From Detailed Analysis of Insulin Absorption Variability and the Tissue Response to Continuous Subcutaneous Insulin Infusion Catheter Implantation in Swine (2017) · cites this paper as [6]

    The Discussion (page unavailable) says disrupted pump delivery or prolonged catheter use can impair insulin absorption, and reports rising mean blood glucose with longer catheter use. It does not establish that a bolus commonly fails to lower glucose as expected or that this leads to persistent hyperglycemia. (LLM, conf=0.82)

    Experimental — accuracy not yet measured
  • Infusion set failure can occur when there is local inflammation at the infusion site, when the catheter is occluded, kinked or has simply dislodged from the subcutaneous space.

    “Infusion set failure can occur when there is local inflammation at the infusion site, when the catheter is occluded, kinked or has simply dislodged from the subcutaneous space.3,4”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 3

  1. Pfützner A, Berger S, Spinas GA. Aktueller stellenwert der kontinuierlichen subkutanen Insulininfusion (CSII) mit Insulinpumpen in der therapie des diabetes mellitus (Actual value of CSII in the therapy of diabetes mellitus). Swiss Med Weekly. 2000;130:1954-61.
  2. Renner R, Pfützner A, Trautmann M, Harzer O, Sauter K, Landgraf R. Use of insulin lispro in continuous subcutaneous insulin infusion treatment. Results of a multicenter trial. Diabetes Care. 1999;22(5):784-8.
  3. Poulsen C, Langkjaer L, Worsøe C. Precipitation of insulin products used forcontinuous subcutaneous insulin infusion. Diabetes Technol Ther. 2005;7(1):142-50.

Cited by

Sorted by citations

Authors
Schmid, V., Hohberg, C., Borchert, M., Forst, T., & Pfützner, A.
Affiliations
Institut für Klinische Forschung und Entwicklung (DE)
Corresponding author
Pfützner, Andreas
Venue
Journal of Diabetes Science and Technology
Published
2010-07
Volume
4
Issue
4
Pages
976-982
ISSN
1932-2968 (ISSN-L)
OpenAlex
W1966562602
Cited by
97
References
24
Citations by year
2026: 3 · 2025: 1 · 2024: 6 · 2023: 11 · 2022: 10 · 2021: 4 · 2020: 4 · 2019: 5 · 2018: 11 · 2017: 5 · 2016: 7 · 2015: 14 · 2014: 6 · 2013: 4 · 2012: 4
FWCI
2.42
Type
journal article
Language
en
Open access
Locations
  • Journal of Diabetes Science and Technology (publishedVersion)
  • Europe PMC (PubMed Central) (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
Keywords
insulin pump therapy, catheter dwell time, Teflon catheter, type 1 diabetes, glycemic control, adverse events, blood glucose variability, patient-reported outcomes, glycemic variability, hypoglycemia, hyperglycemia
MeSH terms
Adult, Blood Glucose, Catheterization, Diabetes Mellitus, Type 1, Female, Humans, Insulin Infusion Systems, Male, Middle Aged, Pilot Projects, Polytetrafluoroethylene, Prospective Studies, Surveys and Questionnaires, Skin, Blood Glucose Self-Monitoring, Kaplan-Meier Estimate, Young Adult

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review