Metabolic Biofouling of Glucose Sensors in Vivo : Role of Tissue Microhemorrhages
Open accessAbstract
Objective: Based on our in vitro study that demonstrated the adverse effects of blood clots on glucose sensor function, we hypothesized that in vivo local tissue hemorrhages, induced as a consequence of sensor implantation or sensor movement post-implantation, are responsible for unreliable readings or an unexplained loss of functionality shortly after implantation. Research Design and Methods: To investigate this issue, we utilized real-time continuous monitoring of blood glucose levels in a mouse model. Direct injection of blood at the tissue site of sensor implantation was utilized to mimic sensor-induced local tissue hemorrhages. Results: It was found that blood injections, proximal to the sensor, consistently caused lowered sensor glucose readings, designated temporary signal reduction, in vivo in our mouse model, while injections of plasma or saline did not have this effect. Conclusion: These results support our hypothesis that tissue hemorrhage and resulting blood clots near the sensor can result in lowered local blood glucose concentrations due to metabolism of glucose by the clot. The lowered local blood glucose concentration led to low glucose readings from the still functioning sensor that did not reflect the systemic glucose level.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Temporary signal reductions may result from local tissue damage caused by forces exerted on the sensor and may be solved by using more flexible materials.
“Therefore, temporary signal reductions may actually be the result of local tissue damage caused by forces exerted on the sensor and may be solved by using more flexible materials.”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [133]
This reduction was explained by clot formation around the sensing portion of the CGM device, which formed a glucose sink until clearance.
“This reduction was explained by the clot formation around the sensing portion of the CGM device that formed a glucose sink until clearance.133”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [133]
Injection of blood, but not saline or plasma, led to a temporary signal reduction lasting several hours that eventually returned to normal.
“the injection of blood, but not saline or plasma, led to a temporary (i.e., several hours) signal reduction that eventually returned to normal.133”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [133]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Hem, A., Smith, A. J., & Solberg, P. (1998) · Laboratory Animals · doi:10.1258/002367798780599866 · Cited by 265 · see citing works
Continuous Subcutaneous Glucose Monitoring in Diabetic Patients
Maran, A., Crepaldi, C., Tiengo, A., Grassi, G., Vitali, E., Pagano, G., Bistoni, S., Calabrese, G., Santeusanio, F., Leonetti, F., et al. (2002) · Diabetes Care · doi:10.2337/diacare.25.2.347 · Cited by 198 · see citing works
Feldman, B., Brazg, R., Schwartz, S. L., & Weinstein, R. L. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322526978 · Cited by 184 · see citing works
Self-citationProtein interactions with subcutaneously implanted biosensors
Gifford, R., Kehoe, J. J., Barnes, S. L., Kornilayev, B. A., Alterman, M. A., & Wilson, G. S. (2006) · Biomaterials · doi:10.1016/j.biomaterials.2005.11.033 · Cited by 117 · see citing works
Influence of inflammatory cells and serum on the performance of implantable glucose sensors
Gerritsen, M., Jansen, J. A., Kros, A., Vriezema, D. M., Sommerdijk, N. A. J. M., Nolte, R. J. M., Lutterman, J. A., van Hövell, S. W. F. M., & van der Gaag, A. (2000) · Journal of Biomedical Materials Research · doi:10.1002/1097-4636(200101)54:1 · Cited by 89 · see citing works
Klueh, U., Dorsky, D. I., & Kreutzer, D. L. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.04.017 · Cited by 87 · see citing works
Self-citationThe Diabetes Research in Children Network (DirecNet) Study Group (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.109 · Cited by 82 · see citing works
Accuracy of the continuous glucose monitoring system in inpatient and outpatient conditions
Djakouré-Platonoff, C., Radermercker, R., Reach, G., Slama, G., & Jl, S. (2003) · Diabetes & Metabolism · doi:10.1016/s1262-3636(07)70023-x · Cited by 69 · see citing works
Continuous Glucose Monitoring in Normal Mice and Mice with Prediabetes and Diabetes
Klueh, U., Liu, Z., Cho, B., Ouyang, T., Feldman, B., Henning, T. P., Kaur, M., & Kreutzer, D. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.402 · Cited by 63 · see citing works
Self-citationBühling, K. J., Kurzidim, B., Wolf, C., Wohlfarth, K., Mahmoudi, M., Wäscher, C., Siebert, G., & Dudenhausen, J. W. (2004) · Experimental and Clinical Endocrinology & Diabetes · doi:10.1055/s-2004-830399 · Cited by 50 · see citing works
Critical role of tissue mast cells in controlling long-term glucose sensor function in vivo
Klueh, U., Kaur, M., Qiao, Y., & Kreutzer, D. L. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.02.023 · Cited by 48 · see citing works
Self-citationAssessment of Chronically Implanted Subcutaneous Glucose Sensors in Dogs
Ward, W. K., & Troupe, J. E. (1999) · ASAIO Journal · doi:10.1097/00002480-199911000-00009 · Cited by 43 · see citing works
Continuous Glucose Monitoring: Reliable Measurements for up to 4 Days with the SCGM1 System
Kapitza, C., Lodwig, V., Obermaier, K., et al. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322250622 · Cited by 30 · see citing works
Klueh, U., Liu, Z., Ouyang, T., Cho, B., Feldman, B., Henning, T. P., & Kreutzer, D. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100607 · Cited by 14 · see citing works
Self-citationKlueh, U., Liu, Z., Feldman, B., & Kreutzer, D. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400506 · Cited by 12 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831 · see citing works
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347 · see citing works
- Authors
- Klueh, U., Liu, Z., Feldman, B., Henning, T. P., Cho, B., Ouyang, T., & Kreutzer, D.
- Affiliations
- University of Connecticut (US)
- Abbott Diabetes Care (United States) (US)
- Corresponding author
- Klueh, Ulrike
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2011-05
- Volume
- 5
- Issue
- 3
- Pages
- 583-595
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3192625
- OpenAlex
- W2137240669
- Cited by
- 63
- References
- 15
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 2 · 2023: 3 · 2022: 1 · 2021: 3 · 2020: 4 · 2019: 2 · 2018: 7 · 2017: 5 · 2016: 7 · 2015: 6 · 2014: 8 · 2013: 10 · 2012: 1
- Type
- conference abstract
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Surgery › Pancreatic function and diabetes
- Keywords
- glucose sensor, microhemorrhages, blood clots, continuous glucose monitoring
- MeSH terms
- Animals, Blood Coagulation, Blood Glucose, Disease Models, Animal, Equipment Design, Female, Hemorrhage, Humans, Mice, Inbred ICR, Sensitivity and Specificity, Time Factors, Blood Glucose Self-Monitoring, Biosensing Techniques, Prostheses and Implants, Mice, Biofouling
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review