Hematocrit Compensation in Electrochemical Blood Glucose Monitoring Systems
Open accessThe DOI matches its Crossref record; 38 of the 39 references it cites were found in a registry or an index.
Abstract
Background: Hematocrit (Hct) is a common interferent in test strips used by diabetes patients to self-monitor blood glucose (BG), resulting in measurement bias. Described is an electrochemical BG monitoring system (OneTouch® Verio™) that uses a cofacial sensor design, soluble enzyme chemistry, and multiphasic waveform to effectively correct for patient Hct, delivering an accurate reading for whole BG. Methods: The test strip comprises thin-film gold and palladium electrodes arranged cofacially and spatially separated with a thin spacer. Soluble glucose-sensing reagents are located on the lower palladium electrode and are hydrated on sample application. Blood glucose is oxidized by flavoprotein glucose dehydrogenase, with electron transfer via (reduced) potassium ferrocyanide mediator at the palladium electrode. Hematocrit levels are estimated by measuring oxidation of mediator diffusion to the upper gold electrode during the first portion of the assay. The Hct-corrected glucose levels are determined by an on-meter algorithm. Results: In performance testing of blood samples at five glucose levels (30–560 mg/dl) and five Hct levels (19–61%), using 12 test meters and 3 test strip lots, 100% of results ( N = 2700) met International Organization for Standardization accuracy criteria (within ± 15 mg/dl and ± 20% of reference results at glucose levels of <75 and ≥75 mg/dl, respectively). Furthermore, 99.9% (2698 of 2700) of results were within ±12 mg/dl and ± 15% of reference values at glucose levels <80 and ≥80 mg/dl, respectively. Conclusions: The technology used in this system provides accurate BG measurements that are insensitive to Hct levels across the range 20–60%.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
Dispersion and Absorption in Dielectrics I. Alternating Current Characteristics
Cole, K. S., & Cole, R. H. (1941) · The Journal of Chemical Physics · doi:10.1063/1.1750906 · Cited by 10,152
A treatise on electricity and magnetism
(1954) · Journal of the Franklin Institute · doi:10.1016/0016-0032(54)90053-8 · Cited by 9,077
Einstein, A. (1905) · Annalen der Physik · doi:10.1002/andp.19053220806 · Cited by 8,265
A Treatise on Electricity and Magnetism
(1873) · Nature · doi:10.1038/007478a0 · Cited by 4,537
(2000) · New England Journal of Medicine · doi:10.1056/nejm200002103420603 · Cited by 1,534
Standards of Medical Care in Diabetes—2008
Dora, J. M., Kramer, C. K., & Canani, L. H. (2008) · Diabetes Care · doi:10.2337/dc08-0109 · Cited by 1,409
Standards of Medical Care in Diabetes—2008
American Diabetes Association (2008) · Diabetes Care · doi:10.2337/dc08-s012 · Cited by 1,377
Cole, K. S. (1968) · University of California Press eBooks · doi:10.1525/9780520326514 · Cited by 992
Physical models of diffusion for polymer solutions, gels and solids
Masaro, L., & Zhu, X. X. (1999) · Progress in Polymer Science · doi:10.1016/s0079-6700(99)00016-7 · Cited by 913
Fricke, H. (1924) · Physical Review · doi:10.1103/physrev.24.575 · Cited by 908
Karter, A. J., Ackerson, L. M., Darbinian, J. A., D’Agostino, R. B., Ferrara, A., Liu, J., & Selby, J. V. (2001) · The American Journal of Medicine · doi:10.1016/s0002-9343(01)00742-2 · Cited by 595
The diffusion of electrolytes in a cation-exchange resin membrane I. Theoretical
Mackie, J. S., & Meares, P. (1955) · Proceedings of the Royal Society of London. Series A. Mathematical and Physical Sciences · doi:10.1098/rspa.1955.0234 · Cited by 588
Wu, W.-C., Schifftner, T. L., Henderson, W. G., Eaton, C. B., Poses, R. M., Uttley, G., Sharma, S. C., Vezeridis, M., Khuri, S. F., & Friedmann, P. D. (2007) · JAMA · doi:10.1001/jama.297.22.2481 · Cited by 586
Hematocrit Level and Associated Mortality in Hemodialysis Patients
MA, J. Z., EBBEN, J., XIA, H., & COLLINS, A. J. (1999) · Journal of the American Society of Nephrology · doi:10.1681/asn.v103610 · Cited by 577
THE BODY/VENOUS HEMATOCRIT RATIO: ITS CONSTANCY OVER A WIDE HEMATOCRIT RANGE
Chaplin, H., Mollison, P. L., & Vetter, H. (1953) · Journal of Clinical Investigation · doi:10.1172/jci102859 · Cited by 393
Assessing Glycemia in Diabetes Using Self-monitoring Blood Glucose and Hemoglobin A 1c
Saudek, C. D., Derr, R. L., & Kalyani, R. R. (2006) · JAMA · doi:10.1001/jama.295.14.1688 · Cited by 373
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
COLLINS, A. J., LI, S., PETER, W. S., EBBEN, J., ROBERTS, T., MA, J. Z., & MANNING, W. (2001) · Journal of the American Society of Nephrology · doi:10.1681/asn.v12112465 · Cited by 302
Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
Membranes, ions, and impulses : a chapter of classical biophysics
Cole, K. S. (1968) · University of California Press eBooks · Cited by 245
Jr., J. N. S., Jr., N. A. N., Tan, K. M., & Grava, M. B. (2005) · Current Medical Research and Opinion · doi:10.1185/030079904x20286 · Cited by 177
Fricke, H. (1925) · Physical Review · doi:10.1103/physrev.26.678 · Cited by 173 · cited 2×
“…including Hct. Compensatory methods are primarily based on measurement of the electrical properties impedance, resistivity, and conductivity, directly resulting from the classical work of Fricke25 and later Cole and Cole26,27 on the electric resistance and capacitance of cells and cell suspensions.”
MethodBergenstal, R. M., & Gavin, J. R. (2005) · The American Journal of Medicine · doi:10.1016/j.amjmed.2005.07.055 · Cited by 166
The Technology Behind Glucose Meters: Test Strips
Hönes, J., Müller, P., & Surridge, N. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0005 · Cited by 156
Hirsch, I. B., Bode, B. W., Childs, B. P., Close, K. L., Fisher, W. A., Gavin, J. R., Ginsberg, B. H., Raine, C. H., & Verderese, C. A. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0104 · Cited by 120
Current evidence regarding the value of self-monitored blood glucose testing
Blonde, L., & Karter, A. J. (2005) · The American Journal of Medicine · doi:10.1016/j.amjmed.2005.07.053 · Cited by 95
The Electric Capacity of Suspensions of Red Corpuscles of a Dog
Fricke, H. (1925) · Physical Review · doi:10.1103/physrev.26.682 · Cited by 82 · cited 2×
“…including Hct. Compensatory methods are primarily based on measurement of the electrical properties impedance, resistivity, and conductivity, directly resulting from the classical work of Fricke25 and later Cole and Cole26,27 on the electric resistance and capacitance of cells and cell suspensions.”
MethodHeinemann, L. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0076 · Cited by 62
Effect of High Altitude on Blood Glucose Meter Performance
Fink, K. S., Christensen, D. B., & Ellsworth, A. J. (2002) · Diabetes Technology & Therapeutics · doi:10.1089/152091502320798259 · Cited by 57
Kilpatrick, E. S., Rumley, A. G., Myint, H., Dominiczak, M. H., & SMALL, M. P. (1993) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456329303000513 · Cited by 50
Musholt, P. B., Schipper, C., Thomé, N., Ramljak, S., Schmidt, M., Forst, T., & Pfützner, A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500520 · Cited by 46
The Determination of Surface Conductance from Measurements on Suspensions of Spherical Particles
Fricke, H., & Curtis, H. J. (1936) · The Journal of Physical Chemistry · doi:10.1021/j150375a003 · Cited by 42
Performance of a new interference-resistant glucose meter
Vanavanan, S., Santanirand, P., Chaichanajarernkul, U., Chittamma, A., DuBois, J. A., Shirey, T., & Heinz, M. (2010) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2009.09.010 · Cited by 39
Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer II
Barreau, P. B., & Buttery, J. E. (1988) · Diabetes Care · doi:10.2337/diacare.11.2.116 · Cited by 38
Lyon, M. E., & Lyon, A. W. (2011) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2011.01.005 · Cited by 28
Electrical impedance and erythrocyte sedimentation rate (ESR) of blood
Zhao, T., & LOCKNER, D. (1993) · Biochimica et Biophysica Acta (BBA) - Biomembranes · doi:10.1016/0005-2736(93)90411-r · Cited by 24
Research works printed but not linked: 1
- Lyon ME, 2011, Biochem., 44, 412
Cited by
Enzyme‐Based Glucose Sensor: From Invasive to Wearable Device
Lee, H., Hong, Y. J., Baik, S., Hyeon, T., & Kim, D. (2018) · Advanced Healthcare Materials · doi:10.1002/adhm.201701150 · Cited by 778
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 208
Lasserre, P., Balansethupathy, B., Vezza, V. J., Butterworth, A., Macdonald, A. R., Blair, E. O., McAteer, L., Hannah, S., Ward, A. C., Hoskisson, P. A., et al. (2022) · Analytical Chemistry · doi:10.1021/acs.analchem.1c04456 · Cited by 69
“SEP1 BIOTIP multichannel electrode printed circuit boards (PCB) (Bath, UK) were used for clinical sample testing and prepared according to the procedure described by Vezza et al. (2021)(29).”
MethodSelf-citationWu, H., Shi, C., Zhu, Q. Y., Li, Y., Xu, Z., Wei, C., Chen, D., & Huang, X. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112722 · Cited by 40
Metadata
- Authors
- Teodorczyk, M., Cardosi, M., & Setford, S.
- Affiliations
- Research and Development, LifeScan Inc., Milpitas, California
- LifeScan Scotland Ltd., Inverness, United Kingdom
- Corresponding author
- Teodorczyk, Maria
- ORCID
- Setford, Steven
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2012-05
- Volume
- 6
- Issue
- 3
- Pages
- 648-655
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3440048
- OpenAlex
- W2141602487
- Cited by
- 33
- References
- 39
- Citations by year
- 2026: 1 · 2025: 3 · 2023: 4 · 2022: 2 · 2021: 3 · 2020: 2 · 2019: 2 · 2018: 4 · 2017: 1 · 2016: 2 · 2015: 2 · 2014: 2 · 2013: 2 · 2012: 3
- FWCI
- 1.84
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Sensors
- Keywords
- blood glucose monitoring, diabetes self-monitoring, electrochemical glucose sensor, palladium electrodes, gold electrode, glucose dehydrogenase, measurement bias
- MeSH terms
- Algorithms, Blood Glucose, Electrodes, Equipment Design, Gold, Hematocrit, Humans, Materials Testing, Oxidation-Reduction, Palladium, Predictive Value of Tests, Reagent Strips, Signal Processing, Computer-Assisted, Blood Glucose Self-Monitoring, Reproducibility of Results, Biomarkers, Glucose 1-Dehydrogenase, Electrochemical Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.