Further Evidence of Severe Allergic Contact Dermatitis From Isobornyl Acrylate While Using a Continuous Glucose Monitoring System
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
In the past decade, new diabetes technologies, including continuous glucose monitoring (CGM) systems, support patients with diabetes in their daily struggle with achieving a good glucose control. However, shortly after the first CGM systems appeared on the market, also the first concerns about adverse skin reactions were raised. Most patients claimed to suffer from (sometimes severe) skin irritation, or even allergy, which they related to the (acrylate-based) adhesive part of the device. For a long time the actual substance that caused these skin reactions with, for example, the Flash Glucose Monitoring system (iscCGM; Freestyle® Libre) could not be identified; however, recently Belgian and Swedish dermatologists reported that the majority of their patients that have developed a contact-allergic while using iscCGM react sensitively to a specific acrylate, that is, isobornyl acrylate (IBOA). Subsequently they showed by means of gas chromatography-mass spectrometry that this substance is present in the case of the glucose sensor attached by an adhesive to the skin. We report three additional cases from Germany, including a 10-year-old boy, suffering from severe allergic contact dermatitis to IBOA.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In some cases, this has been linked to the presence of isobornyl acrylate, a skin sensitizer that can cause an additional spreading allergic reaction (119-121).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [119]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Herman, A., Aerts, O., Baeck, M., Bruze, M., De Block, C., Goossens, A., Hamnerius, N., Huygens, S., Maiter, D., Tennstedt, D., et al. (2017) · Contact Dermatitis · doi:10.1111/cod.12866 · Cited by 192
Self-citationAdhesives Used for Diabetes Medical Devices: A Neglected Risk With Serious Consequences?
Heinemann, L., & Kamann, S. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816662949 · Cited by 74 · cited 1×
“Until recently there were only a few reports on skin allergy caused by components of devices used for diabetes therapy.(2) In 1995, two young women, who showed a contact-allergic skin reaction to two different insulin pumps were described and the authors revealed that one of the culprit allergens was IBOA.”
Self-citationSchwensen, J. F. B., Friis, U. F., Zachariae, C. O. C., & Johansen, J. D. (2016) · Contact Dermatitis · doi:10.1111/cod.12503 · Cited by 62 · cited 1×
“A case of sensitization to cyanoacrylate, caused by exposure to a CGM system in another child with diabetes, was recently described.(4) In this particular case cyanoacrylate was shown to be a component of the thin layer glue that binds the sensor to the adhesive.”
Contact allergy to components of glue in insulin pump infusion sets
Busschots, A. M., Meuleman, V., Poesen, N., & Dooms‐Goossens, A. (1995) · Contact Dermatitis · doi:10.1111/j.1600-0536.1995.tb00554.x · Cited by 61
Two decades of occupational (meth)acrylate patch test results and focus on isobornyl acrylate
Christoffers, W. A., Coenraads, P., & Schuttelaar, M. A. (2013) · Contact Dermatitis · doi:10.1111/cod.12023 · Cited by 46 · cited 1×
“In a recent study Christoffers et al published an analysis of two decades of occupational patch testing with acrylates in 151 patients.(6) The authors also described one case of occupational ACD to IBOA caused by uncured UV ink and acrylate coating.”
Isobornyl acrylate: an impurity in alkyl glucosides
Foti, C., Romita, P., Rigano, L., Zimerson, E., Sicilia, M., Ballini, A., Ghizzoni, O., Antelmi, A., Angelini, G., Bonamonte, D., et al. (2015) · Cutaneous and Ocular Toxicology · doi:10.3109/15569527.2015.1055495 · Cited by 45 · cited 1×
“IBOA can also be easily released into materials flowing over surfaces made from it.(5) After recontacting the adhesive manufacturer itself, they once again confirmed that IBOA is not contained in the adhesive part, nor in the layer glue between the adhesive and the sensor.”
Cited by
7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
7. Diabetes Technology: Standards of Medical Care in Diabetes—2020
American Diabetes Association (2020) · Diabetes Care · doi:10.2337/dc20-s007 · Cited by 250
7. Diabetes Technology: Standards of Care in Diabetes—2026
Bajaj, M. S., McCoy, R. G., Balapattabi, K., Bannuru, R. R., Bellini, N. J., Bennett, A. K., Beverly, E. A., Early, K. B., ChallaSivaKanaka, S., Echouffo-Tcheugui, J. B., et al. (2025) · Diabetes Care · doi:10.2337/dc26-s007 · Cited by 293
7. Diabetes Technology: Standards of Medical Care in Diabetes—2021
American Diabetes Association (2021) · Diabetes Care · doi:10.2337/dc21-s007 · Cited by 202
7. Diabetes Technology: Standards of Medical Care in Diabetes—2019
American Diabetes Association (2019) · Diabetes Care · doi:10.2337/dc19-s007 · Cited by 179
Yogev, D., Goldberg, T., Arami, A., Tejman‐Yarden, S., Winkler, T. E., & Maoz, B. M. (2023) · APL Bioengineering · doi:10.1063/5.0152290 · Cited by 166
Sherr, J. L., Heinemann, L., Fleming, G. A., Bergenstal, R. M., Bruttomesso, D., Hanaire, H., Holl, R. W., Petrie, J. R., Peters, A. L., & Evans, M. (2023) · Diabetologia · doi:10.1007/s00125-022-05744-z · Cited by 123
Self-citationSherr, J. L., Heinemann, L., Fleming, G. A., Bergenstal, R. M., Bruttomesso, D., Hanaire, H., Holl, R. W., Petrie, J. R., Peters, A. L., & Evans, M. (2022) · Diabetes Care · doi:10.2337/dci22-0018 · Cited by 136
Self-citationAllergic contact dermatitis caused by medical devices for diabetes patients: A review
Herman, A., de Montjoye, L., Tromme, I., Goossens, A., & Baeck, M. (2018) · Contact Dermatitis · doi:10.1111/cod.13120 · Cited by 95
“This was suggested by a previous analysis performed by Kamann et al.(21) These authors patch tested, in patients presenting a skin reaction to Freestyle Libre, the original adhesive obtained from the manufacturer, consisting of 3 different adhesives superimposed upon each other.”
MethodAsarani, N. A. M., Reynolds, A. N., Boucher, S. E., de Bock, M., & Wheeler, B. J. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819870849 · Cited by 81
“In practical terms, options include adequate drying prior to insertion and the use of various supplemental products such as adhesive barriers, tackifiers (to improve adhesion), or possibly off-label steroid sprays prior to insertion for those with known prior reactions (eg, fluticasone).(16,17,20,65,68,69) To minimize…”
BackgroundN,N‐dimethylacrylamide—A new sensitizer in the FreeStyle Libre glucose sensor
Mowitz, M., Herman, A., Baeck, M., Isaksson, M., Antelmi, A., Hamnerius, N., Pontén, A., & Bruze, M. (2019) · Contact Dermatitis · doi:10.1111/cod.13243 · Cited by 80
“According to the manufacturer of the adhesive patch, no IBOA is used in the adhesive in contact with the skin or in the adhesive used to fix the patch to the sensor.(12) However, in our analyses, we found that IBOA was also present in the extract of the adhesive patch that had been removed from the sensor, suggesting…”
MethodShang, T., Zhang, J. Y., Thomas, A., Arnold, M. A., Vetter, B. N., Heinemann, L., & Klonoff, D. C. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211007212 · Cited by 71
Self-citationContact Allergy—Emerging Allergens and Public Health Impact
Uter, W. J. C., Werfel, T., Lepoittevin, J., & White, I. R. (2020) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17072404 · Cited by 76
The Dexcom glucose monitoring system—An isobornyl acrylate‐free alternative for diabetic patients
Oppel, E. M., Kamann, S., Reichl, F., & Högg, C. (2019) · Contact Dermatitis · doi:10.1111/cod.13248 · Cited by 69
“Positive reactions after 48 and 72 hours were seen.(10) In the meantime, IBOA has also been detected by the use of gas chromatography in the patch pump Omnipod, which is a pump that sticks directly onto the skin and has no catheter system such as those in conventional insulin pumps.”
BackgroundSelf-citationNeale, H., Garza‐Mayers, A. C., Tam, I., & Yu, J. (2020) · Journal of the American Academy of Dermatology · doi:10.1016/j.jaad.2020.11.002 · Cited by 66
Aerts, O., Herman, A., Mowitz, M., Bruze, M., & Goossens, A. E. (2020) · Dermatitis · doi:10.1097/der.0000000000000549 · Cited by 66
“French and German dermatologists confirmed, also in their patients, that IBOA was often the culprit sensitizer provoking ACD from the same glucose sensor.(18) Interestingly, and similar to earlier observations with insulin pumps,(10) the GC-MS investigations at the Swedish laboratory showed that the adhesive patch its…”
InfluentialSelf-citationRigo, R. S., Levin, L. E., Belsito, D. V., Garzon, M. C., Gandica, R., & Williams, K. M. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820918894 · Cited by 58
Herman, A., Baeck, M., de Montjoye, L., Bruze, M., Giertz, E., Goossens, A. E., & Mowitz, M. (2019) · Contact Dermatitis · doi:10.1111/cod.13374 · Cited by 57
“The patch test reactions were classified according to ESCD criteria.(1,2)”
BackgroundMethodDeiss, D., Szadkowska, A., Gordon, D., Mallipedhi, A., Schütz-Fuhrmann, I., Aguilera, E., Ringsell, C., Block, C. E. M. D., & Irace, C. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0397 · Cited by 62
“People with isobornyl acrylate oversensitivity: Recent studies highlight that exposure to isobornyl acrylate over an extended period of time (more than 4 days) can cause severe skin reactions such as redness, itching, or pain and in severe cases, to those who are susceptible, can cause an allergic immune response.(35,…”
BackgroundPrevalence and Prevention of Contact Dermatitis Caused by FreeStyle Libre: A Monocentric Experience
Pyl, J., Dendooven, E., van Eekelen, I., den Brinker, M., Dotremont, H., France, A., Foubert, K., Pieters, L. A. C., Lambert, J. R., Block, C. E. M. D., et al. (2020) · Diabetes Care · doi:10.2337/dc19-1354 · Cited by 53
Self-citationBaldo, F. D., Canton, C., Testa, S., Swales, H., Drudi, I., Golinelli, S., & Fracassi, F. (2020) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.15930 · Cited by 55
“the sensor itself and that can migrate into the adhesive part of the device and come into contact with the skin.(22) The mild erythema does not seem to create discomfort for the animal and can be considered a mild and acceptable side effect in the use of the FGMS.”
BackgroundComparative assessment of blood glucose monitoring techniques: a review
Ahmadian, N., Manickavasagan, A., & Ali, A. (2022) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2022.2100496 · Cited by 51
Modelling the influence of UV curing strategies for optimisation of inkjet based 3D printing
Zhao, P., He, Y., Trindade, G. F., Baumers, M., Irvine, D. J., Hague, R., Ashcroft, I., & Wildman, R. D. (2021) · Materials & Design · doi:10.1016/j.matdes.2021.109889 · Cited by 54
“These ink residuls can cause allergic effects, irritation, or inflammatory response of the surrounding tissue in both the short and long term [69].”
BackgroundPleus, S., Ulbrich, S., Zschornack, E., Kamann, S., Haug, C., & Freckmann, G. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0171 · Cited by 47
“7–9 Due to the prolonged sensor wear time the components spread and may get in contact with the skin.”
InfluentialSelf-citationReal-Time Continuous Glucose Monitoring in Gestational Diabetes: A Randomized Controlled Trial
Lane, A. S., Młynarczyk, M., de Veciana, M., Green, L. M., Baraki, D. I., & Abuhamad, A. (2019) · American Journal of Perinatology · doi:10.1055/s-0039-1678733 · Cited by 48
Herman, A., Uter, W. J. C., Rustemeyer, T., Matura, M., Aalto‐Korte, K., Johansen, J. D., Gonçalo, M., White, I. R., Balato, A., Giménez‐Arnau, A. M., et al. (2021) · Journal of the European Academy of Dermatology and Venereology · doi:10.1111/jdv.17238 · Cited by 45
Evaluation of Isobornyl Acrylate Content in Medical Devices for Diabetes Treatment
Kamann, S., Oppel, E. M., Liu, F., Reichl, F., Heinemann, L., & Högg, C. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0163 · Cited by 44
Self-citationContact dermatitis caused by glucose sensors, insulin pumps, and tapes: Results from a 5‐year period
Friis, U. F., Simonsen, A. B., Zachariae, C. O. C., Thyssen, J. P., & Johansen, J. D. (2020) · Contact Dermatitis · doi:10.1111/cod.13664 · Cited by 41
Cutaneous Adverse Effects of Diabetes Mellitus Medications and Medical Devices: A Review
Jedlowski, P. M., Te, C., Segal, R. J., & Fazel, M. T. (2018) · American Journal of Clinical Dermatology · doi:10.1007/s40257-018-0400-7 · Cited by 42
Oppel, E., Kamann, S., Heinemann, L., Reichl, F., & Högg, C. (2020) · Contact Dermatitis · doi:10.1111/cod.13392 · Cited by 37
“The transmitter, which is fixed with one of the two adhesives on top of the skin site where the sensor is implanted, has no direct skin contact, but allergens may be able to diffuse through adhesives.(6,8) Thus, as a precautionary measure, a daily change of the adhesive is recommended.”
BackgroundMethodResultSelf-citationAllergic Contact Dermatitis to Components of Wearable Adhesive Health Devices
Khatsenko, K., Khin, Y., & Maibach, H. I. (2020) · Dermatitis · doi:10.1097/der.0000000000000575 · Cited by 39
“One case of ACD in a pediatric patient reports that Stomahesive (Convatec, Deeside, Flintshire, UK) appeared to be an adequate plaster for prevention of allergens in Freestyle Libre from penetrating into the skin.(11)We hope that awareness of the presence of these chemicals in wearable devices may aid with diagnostic…”
BackgroundResultPatch Pumps: Are They All the Same?
Heinemann, L., Waldenmaier, D., Kulzer, B., Ziegler, R., Ginsberg, B., & Freckmann, G. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818795150 · Cited by 36
Self-citationHerman, A., de Montjoye, L., & Baeck, M. (2020) · Contact Dermatitis · doi:10.1111/cod.13529 · Cited by 36
“sent in the original adhesives provided by the manufacturer (Adhesive Research, Glen Rock, Pennsylvania) but derived from the plastic part and migrated through the adhesive part of the FreeStyle Libre sensor.(15) Among our series of 35 patients sensitized to IBOA, tested with the adhesive part of the sensor (and not t…”
MethodDendooven, E., Foubert, K., Goossens, A. E., Gilles, P., Borggraeve, W. M. D., Pieters, L. A. C., Lambert, J. R., & Aerts, O. (2020) · Contact Dermatitis · doi:10.1111/cod.13706 · Cited by 35
Self-citationKamann, S., & Oppel, E. M. (2019) · Contact Dermatitis · doi:10.1111/cod.13329 · Cited by 34
“acryate (IBOA), be identified; this was followed by further reports and reviews on patients with Freestyle Libre and IBOA allergy.(1-3) Interestingly, in the case of Freestyle Libre, IBOA seems to be part of the sensor itself, whereas the adhesive and the thin glue layer that attaches the sensor to the adhesive are fr…”
BackgroundSelf-citationEvaluation of flash glucose monitoring after long-term use: a pediatric survey
Vergier, J., Samper, M., Dalla-Vale, F., Ventura, V., Baucher, F., Joubert, F., Pons, C., De Oliveira, M., Salet, R., Faure-Galon, N., et al. (2019) · Primary Care Diabetes · doi:10.1016/j.pcd.2018.08.004 · Cited by 32
Cameli, N., Silvestri, M., Mariano, M., Messina, C., Nisticò, S. P., & Cristaudo, A. (2022) · Dermatitis · doi:10.1097/der.0000000000000861 · Cited by 34
“In the FreeStyle Libre system, IBOA can be released from the glue used to attach the upper and lower part of the sensor and then migrate to other parts of the sensor.(15,16) A case of acquired leukoderma after ACD to IBOA in FreeStyle Libre has been also reported in a 35year-old woman by Herman et al(42): depigmentati…”
MethodContact dermatitis caused by glucose sensors in diabetic children
Herman, A., Darrigade, A. ‐., de Montjoye, L., & Baeck, M. (2019) · Contact Dermatitis · doi:10.1111/cod.13429 · Cited by 33
“It has been reported previously that patch testing using the piece of the sensor can provide false-negative reactions.(10) In this patient (Patient 11) displaying a negative reaction to IBOA, DMAA, and a piece of the sensor, the clinical reaction under the sensor was an eczematous annular reaction with a progressive w…”
BackgroundAllergic contact dermatitis caused by glucose sensors and insulin pumps: A full review
de Groot, A. C., van Oers, E. M., Ipenburg, N. A., & Rustemeyer, T. (2024) · Contact Dermatitis · doi:10.1111/cod.14698 · Cited by 28
“57,69 The investigators also convincingly showed that it was IBOA itself that caused the positive patch test reactions and not an impurity in the IBOA test material.”
Pediatric Allergic Contact Dermatitis
Brown, C., & Yu, J. (2021) · Immunology and Allergy Clinics of North America · doi:10.1016/j.iac.2021.04.004 · Cited by 26
Corazza, M., Scuderi, V., Musmeci, D., Foti, C., Romita, P., & Borghi, A. (2018) · Contact Dermatitis · doi:10.1111/cod.13075 · Cited by 24
Lombardo, F., Passanisi, S., Tinti, D., Messina, M. F., Salzano, G., & Rabbone, I. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820947072 · Cited by 24
Allergic contact dermatitis caused by glucose sensors and insulin pumps: A full review
de Groot, A. C., van Oers, E. M., Ipenburg, N. A., & Rustemeyer, T. (2024) · Contact Dermatitis · doi:10.1111/cod.14697 · Cited by 22
Massa, G., Gys, I., Bevilacqua, E., Wijnands, A., & Zeevaert, R. (2019) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2019.05.015 · Cited by 24
Kamann, S., Heinemann, L., & Oppel, E. M. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819876964 · Cited by 22
“Further publications confirmed this report.(3,4) A key aspect of these publications is that IBOA is not only part of the adhesive with which the iscCGM system works but is also part of the plastic housing of the bottom part of the glucose sensor.”
BackgroundSelf-citationOppel, E. M., Högg, C., Oschmann, A., Summer, B., & Kamann, S. (2022) · Contact Dermatitis · doi:10.1111/cod.14141 · Cited by 21
“The sleigh only covers a small part of the adhesive patch and the skin reactions corresponded to the whole area of the adhesive patch, so there is no comparison to the housings of the Freestyle Libre.”
MethodSelf-citationSkin Involvement in Paediatric Patients with Type 1 Diabetes
Passanisi, S., Salzano, G., & Lombardo, F. (2021) · Current Diabetes Reviews · doi:10.2174/1573399817666210903153837 · Cited by 18
Passanisi, S., Salzano, G., Galletta, F., Aramnejad, S., Caminiti, L., Pajno, G. B., & Lombardo, F. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2022.846137 · Cited by 19
Preventing medical device-related skin damage
Liversedge, H. L. (2019) · Nursing Standard · doi:10.7748/ns.2019.e11375 · Cited by 18
Baldo, F. D., Diana, A., Canton, C., Linta, N., Chiocchetti, R., & Fracassi, F. (2021) · Animals · doi:10.3390/ani11020408 · Cited by 18
“It is thought to be due to the isobornyl acrylate contained in the sensor itself, which can migrate into the adhesive part of the device and as such comes into contact with the skin, leading to an allergic type 4 reaction [42].”
BackgroundFreckmann, G., Buck, S., Waldenmaier, D., Zschornack, E., Link, M., Jendrike, N., Obstfelder, I., Vetrugno, S., Kamann, S., & Haug, C. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820911105 · Cited by 17
“802 Journal of Diabetes Science and Technology 15(4)”
BackgroundSelf-citationThe need to disclose the composition of medical devices at the European level
Herman, A., & Goossens, A. E. (2019) · Contact Dermatitis · doi:10.1111/cod.13354 · Cited by 15
Modern diabetes devices for continuous blood sugar measuring: Limitations due to contact allergies
Kamann, S., Wagner, N., & Oppel, E. M. (2021) · JDDG Journal der Deutschen Dermatologischen Gesellschaft · doi:10.1111/ddg.14621 · Cited by 15
Self-citationContact dermatitis in children caused by diabetes devices
da Silva, C. A., Bregnhøj, A., Mowitz, M., Bruze, M., Andersen, K. E., & Sommerlund, M. (2022) · Contact Dermatitis · doi:10.1111/cod.14166 · Cited by 15
“ACD may be due to allergens in the devices.(21) ICD usually has a more abrupt presentation of symptoms, but since all our”
BackgroundWearable Woes: Allergens in Diabetic Devices
Hartsough, E. M., & Hylwa, S. A. (2020) · Dermatitis · doi:10.1097/der.0000000000000673 · Cited by 15
“It is highly utilized because of its ability to imbue hardness and flexibility into materials.(28) However, it has also been shown to play a role as a hidden allergen, because it can be easily released into materials that flow over surfaces made from IBOA.”
BackgroundProceedings of the 14th Australasian database conference - Volume 17
Schewe, K., & Zhou, X. (2003) · Cited by 13
Ridder, F. D., Braspenning, R., Ordóñez, J., Klarenbeek, G., Lauwers, P., Ledeganck, K. J., Delbeke, D., & Block, C. E. M. D. (2024) · PLoS ONE · doi:10.1371/journal.pone.0301041 · Cited by 13
Allergic contact dermatitis caused by a new insulin pump system containing isobornyl acrylate
Enberg, J., Hamnerius, N., & Mowitz, M. (2023) · Contact Dermatitis · doi:10.1111/cod.14274 · Cited by 12
Diabetes Technology Meeting 2021
Xu, N. Y., Nguyen, K. T., DuBord, A. Y., Pickup, J. C., Sherr, J. L., Teymourian, H., Cengiz, E., Ginsberg, B. H., Cobelli, C., Ahn, D. T., et al. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221090279 · Cited by 11
Dorati, R., Cosco, D., Battaglia, L., Cirri, M., Conti, B., Borghi, T., Gabriele, M., Bruno, G. M., Rocco, P., Selmin, F., et al. (2025) · Drug Discovery Today · doi:10.1016/j.drudis.2025.104441 · Cited by 9
Teufel‐Schäfer, U., Huhn, C. K., Müller, S., Müller, C., & Grünert, S. C. (2021) · Pediatric Dermatology · doi:10.1111/pde.14767 · Cited by 8
Queirós, C. S., Alexandre, M. I., Garrido, P., Correia, T. E., & Filipe, P. (2020) · Contact Dermatitis · doi:10.1111/cod.13578 · Cited by 7
“reactions to FreeStyle Libre were described some months after its appearance, but the major allergen responsible for ACD to this device, IBOA, was identified in only 2017.(10,11) IBOA is widely used in coatings, inks, and paints.”
BackgroundDecisions in the Psychology of Glucose Monitoring
Reach, G. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819854109 · Cited by 8
Fink, H., Fernandes, N. M. D. B., Weißmann, J., & Frey, M. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820908656 · Cited by 6
Marsters, B. L., Styles, S. E., Galland, B. C., de Lange, M., Wiltshire, E. J., de Bock, M. I., Elbalshy, M. M., Tomlinson, P. A., Rayns, J. A., MacKenzie, K. E., et al. (2020) · Acta Diabetologica · doi:10.1007/s00592-020-01556-y · Cited by 7
Flash Continuous Glucose Monitoring: A Practical Guide and Call to Action for Pharmacists
Cowart, K., & Zgibor, J. C. (2021) · Journal of Pharmacy Practice · doi:10.1177/08971900211000273 · Cited by 5
“This is thought to originate from isobornyl acrylate, a compound found in the sensor.(68) However, the overall incidence of dermatologic complications from flash CGM use appears to be low based on available evidence (estimated to be 1 event per 8 weeks of sensor wear-time).”
BackgroundGatica‐Ortega, M. E., & Pastor‐Nieto, M. A. (2020) · Piel · doi:10.1016/j.piel.2019.11.003 · Cited by 5
Medical Device Contact Allergy: Glucose Monitors and Insulin Pumps
Bembry, R., Brys, A. K., & Atwater, A. R. (2022) · Current Dermatology Reports · doi:10.1007/s13671-021-00352-3 · Cited by 4
Polyelectrolyte Complex Coacervate Adhesive for Wearable Medical Devices
Montemayor, M. D., Iverson, E. T., Dahle, L., Green, A., Tenorio, E., Moran, A. V., Rodriguez‐Melendez, D., Chang, E. F., Karim, M. J., Haridas, B., et al. (2025) · Macromolecular Rapid Communications · doi:10.1002/marc.202500117 · Cited by 3
Monitoreo Continuo de Glucosa. ¿Qué evidencia tenemos en pediatría?
Ayala, D. V. S., Forero, C., Mauricio Coll, B., Durán, P. A., & Céspedes, C. (2021) · Andes pediatrica · doi:10.32641/andespediatr.v92i4.2973 · Cited by 5
Impact of trends in new and emerging contact allergens
Rodriguez, O., Brod, B. A., & James, W. D. (2022) · International Journal of Women’s Dermatology · doi:10.1097/jw9.0000000000000006 · Cited by 3
Automatisierte Insulindosierung: Ideale Diabeteskontrolle mit vielen offenen Fragen
Freckmann, G., Deiss, D., Ziegler, R., & Schlüter, S. (2023) · Deutsches Ärzteblatt · doi:10.3238/persdia.2023.03.24.01 · Cited by 1
Scoping review of subcutaneous glucose monitoring techniques
Hrubá, E., Kubíček, J., & Augustýnek, M. (2025) · Measurement · doi:10.1016/j.measurement.2025.119940 · Cited by 1
Kamann, S., Wagner, N., & Oppel, E. M. (2021) · JDDG Journal der Deutschen Dermatologischen Gesellschaft · doi:10.1111/ddg.14621_g · Cited by 1
Self-citationAcquired Leukoderma in a Patient with Allergic Contact Dermatitis to FreeStyle Libre®
Queirós, C. S., Alexandre, M. I., Garrido, P., de Almeida, L. S., Correia, T. E., & Filipe, P. (2020) · Journal of the Portuguese Society of Dermatology and Venereology · doi:10.29021/spdv.78.4.1272
Kontaktallergien – Neu aufkommende Allergene und Auswirkungen für das Gesundheitswesen
Uter, W. J. C., Werfel, T., Lepoittevin, J., & White, I. R. (2021) · Kompass Dermatologie · doi:10.1159/000515947
Self-monitoring in children with type 1 diabetes mellitus
Kašparová, P., & Kunovjánková, V. (2024) · Pediatrie pro praxi · doi:10.36290/ped.2024.051
Allergy in the Diabetic Patient
Bieck, C., & Skudlik, C. (2024) · Springer eBooks · doi:10.1007/978-3-031-65300-1_16
Adverse Skin and Systemic Reactions to Antidiabetic Treatments
Tétart, F., & Gáspár, K. (2024) · Springer eBooks · doi:10.1007/978-3-031-65300-1_29
London-Dawodu, L., Yin, X. J., & Yuvavanich, S. (2025) · Cosmetics · doi:10.3390/cosmetics12030089
Diabetes mellitus: nuevas técnicas para la monitorización de la glucemia en la especie canina.
Ropero, L. E., & Montoya, A. L. (2021) · Zaguan (Universidad de Zaragoza)
Batir, D. M., Boev, M., Gurău, G., Matei, M. N., Nechita, A., & Maftei, N.-M. (2026) · IntechOpen eBooks · doi:10.5772/intechopen.1014475
Olesen, B., Ehrmann, D., Kulzer, B., Hermanns, N., & Heinemann, L. (2026) · Journal of Diabetes Science and Technology · doi:10.1177/19322968261473290
Self-citation
Metadata
- Authors
- Kamann, S., Aerts, O., & Heinemann, L.
- Affiliations
- Pöcking, Germany
- University of Antwerp (BE)
- Science & Co, Düsseldorf, Deutschland, Germany
- Corresponding author
- Kamann, Stefanie
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2018-03-15
- Volume
- 12
- Issue
- 3
- Pages
- 630-633
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC6154227
- OpenAlex
- W2793234916
- Cited by
- 85
- References
- 6
- Citations by year
- 2026: 2 · 2025: 5 · 2024: 7 · 2023: 4 · 2022: 11 · 2021: 13 · 2020: 19 · 2019: 17 · 2018: 6
- FWCI
- 14.88
- Type
- journal article
- Language
- en
- Open access
- green
- europepmc.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Dermatology › Contact Dermatitis and Allergies
- Health Sciences › Medicine › Pharmacology (medical) › Drug-Induced Adverse Reactions
- Health Sciences › Medicine › Dermatology › Skin Diseases and Diabetes
- Keywords
- allergic contact dermatitis, isobornyl acrylate, continuous glucose monitoring, skin irritation, diabetes technology, pediatric patients
- MeSH terms
- Acrylates, Adhesives, Camphanes, Child, Diabetes Mellitus, Type 1, Humans, Male, Blood Glucose Self-Monitoring, Dermatitis, Allergic Contact
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.