Noninvasive Monitoring of Blood Glucose Using Color-Coded Photoplethysmographic Images of the Illuminated Fingertip Within the Visible and Near-Infrared Range: Opportunities and Questions
Open accessThe DOI matches its Crossref record; 37 of the 37 references it cites were found in a registry or an index.
Abstract
Noninvasive blood glucose assays have been promised for many years and various molecular spectroscopy-based methods of skin are candidates for achieving this goal. Due to the small spectral signatures of the glucose used for direct physical detection, moreover hidden among a largely variable background, broad spectral intervals are usually required to provide the mandatory analytical selectivity, but no such device has so far reached the accuracy that is required for self-monitoring of blood glucose (SMBG). A recently presented device as described in this journal, based on photoplethysmographic fingertip images for measuring glucose in a nonspecific indirect manner, is especially evaluated for providing reliable blood glucose concentration predictions.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“However, Heise and coworkers provide a complete analysis of these measurements and conclude that direct measurement of glucose is not possible at the measured wavelength bands and that the system, as described, lacks the ability to produce stable calibration functions required for practical clinical operation (240).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [240]
Citations
Cites
CRC Handbook of Chemistry and Physics
(2014) · doi:10.1201/b17118 · Cited by 57,706
Self-Organization and Associative Memory
Kohonen, T. (1989) · Springer Series in Information Sciences · doi:10.1007/978-3-642-88163-3 · Cited by 8,740
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206 · cited 1×
“Various physiological and pharmacological factors exist, which influence the systolic amplitude owing, for example, to poor finger perfusion, dysrhythmia, and others.(26) To round up the discussion of physiological parameters that are detectable in general by SW-NIR measurements, it is evident that hemoglobin chromoph…”
BackgroundEl Gamal, A., & Eltoukhy, H. (2005) · IEEE Circuits and Devices Magazine · doi:10.1109/mcd.2005.1438751 · Cited by 801
Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations
Chan, E. D., Chan, M. M., & Chan, M. M. (2013) · Respiratory Medicine · doi:10.1016/j.rmed.2013.02.004 · Cited by 710
Wearable non-invasive epidermal glucose sensors: A review
Kim, J., Campbell, A. S., & Wang, J. (2018) · Talanta · doi:10.1016/j.talanta.2017.08.077 · Cited by 586
Monte-Moreno, E. (2011) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2011.05.001 · Cited by 311 · cited 2×
“As physiological factors that are relevant for the method, blood viscosity and vessel compliance have been listed and claimed, while blood viscosity is mentioned to rely also on blood pressure and glucose level, altering the flux of blood in the corresponding vessels and the shape of the PPG pulse (see discussion on t…”
BackgroundWearable Continuous Glucose Monitoring Sensors: A Revolution in Diabetes Treatment
Cappon, G., Acciaroli, G., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2017) · Electronics · doi:10.3390/electronics6030065 · Cited by 237
New insights into the origin of remote PPG signals in visible light and infrared
Moço, A. V., Stuijk, S., & de Haan, G. (2018) · Scientific Reports · doi:10.1038/s41598-018-26068-2 · Cited by 168 · cited 2×
“However, there is also evidence that the PPG waveforms are also influenced from the dermal capillary density due to pressure variations from the larger blood vessels.(27,28) Lower frequency components are usually superimposed, which can be attributed to respiration, vasomotion, sympathetic nervous system activity, and…”
BackgroundMethodThe Technology Behind Glucose Meters: Test Strips
Hönes, J., Müller, P., & Surridge, N. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0005 · Cited by 156
Research works printed but not linked: 1
- Sutherland JG. The holographic neural method. In: Soucek B, ed. Fuzzy, Holographic, and Parallel Intelligence. 6th ed. New York, NY: John Wiley; 1992:7-92.
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Artificial Intelligence in Health Care: Current Applications and Issues
Park, C.-W., Seo, S. W., Kang, N., Ko, B., Choi, B. W., Park, C., Chang, D. K., Kim, H., Kim, H., Lee, H., et al. (2020) · Journal of Korean Medical Science · doi:10.3346/jkms.2020.35.e379 · Cited by 165
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 206
Delbeck, S., Vahlsing, T., Leonhardt, S. K., Steiner, G. R., & Heise, H. M. (2018) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-018-1395-x · Cited by 108
“Our focus was on the assessment of a strategy using color-coded photoplethysmographic imaging of the fingertip for non-invasive blood glucose testing [13].”
MethodSelf-citationReview of present method of glucose from human blood and body fluids assessment
Fiedorova, K., Augustýnek, M., Kubíček, J., Kudrna, P., & Bibbo, D. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2022.114348 · Cited by 103
Metadata
- Authors
- Vahlsing, T., Delbeck, S., Leonhardt, S., & Heise, H. M.
- Affiliations
- Federal Institute For Materials Research and Testing (DE)
- South Westphalia University of Applied Sciences (DE)
- RWTH Aachen University (DE)
- Corresponding author
- Heise, H. Michael
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2018-09-15
- Volume
- 12
- Issue
- 6
- Pages
- 1169-1177
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC6232738
- OpenAlex
- W2891502073
- Semantic Scholar
- 3dc505ae06d663369ddd04be882790c6f8b6bc07
- Cited by
- 28
- References
- 37
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 3 · 2023: 6 · 2022: 4 · 2021: 7 · 2020: 3 · 2019: 1 · 2018: 1
- FWCI
- 1.20
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion, cc-by)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- RWTH Publications (RWTH Aachen) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Spectroscopy Techniques in Biomedical and Chemical Research
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- non-invasive glucose monitoring, photoplethysmography, visible and near-infrared spectroscopy, self-monitoring of blood glucose, glucose concentration prediction
- MeSH terms
- Blood Glucose, Calibration, Color, Equipment and Supplies, Equipment Design, Fingers, Humans, Infrared Rays, Light, Lighting, Skin, Blood Glucose Self-Monitoring, Reproducibility of Results, Biosensing Techniques, Photoplethysmography, Spectroscopy, Near-Infrared
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Noninvasive Monitoring of Blood Glucose Using Color-Coded Photoplethysmographic Images of the Illuminated Fingertip Within the Visible and Near-Infrared Range: Opportunities and Questions. https://app.cochat.ai/papers/10.1177/1932296818798347
- The paper
- Vahlsing, T., Delbeck, S., Leonhardt, S., & Heise, H. M. (2018). Noninvasive Monitoring of Blood Glucose Using Color-Coded Photoplethysmographic Images of the Illuminated Fingertip Within the Visible and Near-Infrared Range: Opportunities and Questions. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296818798347
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.