Noninvasive Monitoring of Blood Glucose Using Color-Coded Photoplethysmographic Images of the Illuminated Fingertip Within the Visible and Near-Infrared Range: Opportunities and Questions
Open accessThe DOI matches its Crossref record; 37 of the 37 references it cites were found in a registry or an index.
Abstract
Noninvasive blood glucose assays have been promised for many years and various molecular spectroscopy-based methods of skin are candidates for achieving this goal. Due to the small spectral signatures of the glucose used for direct physical detection, moreover hidden among a largely variable background, broad spectral intervals are usually required to provide the mandatory analytical selectivity, but no such device has so far reached the accuracy that is required for self-monitoring of blood glucose (SMBG). A recently presented device as described in this journal, based on photoplethysmographic fingertip images for measuring glucose in a nonspecific indirect manner, is especially evaluated for providing reliable blood glucose concentration predictions.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“However, Heise and coworkers provide a complete analysis of these measurements and conclude that direct measurement of glucose is not possible at the measured wavelength bands and that the system, as described, lacks the ability to produce stable calibration functions required for practical clinical operation (240).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [240]
Citations
Cites
CRC Handbook of Chemistry and Physics
(2014) · doi:10.1201/b17118 · Cited by 57,706
Self-Organization and Associative Memory
Kohonen, T. (1989) · Springer Series in Information Sciences · doi:10.1007/978-3-642-88163-3 · Cited by 8,740
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206 · cited 1×
“Various physiological and pharmacological factors exist, which influence the systolic amplitude owing, for example, to poor finger perfusion, dysrhythmia, and others.(26) To round up the discussion of physiological parameters that are detectable in general by SW-NIR measurements, it is evident that hemoglobin chromoph…”
BackgroundEl Gamal, A., & Eltoukhy, H. (2005) · IEEE Circuits and Devices Magazine · doi:10.1109/mcd.2005.1438751 · Cited by 801
Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations
Chan, E. D., Chan, M. M., & Chan, M. M. (2013) · Respiratory Medicine · doi:10.1016/j.rmed.2013.02.004 · Cited by 710
Wearable non-invasive epidermal glucose sensors: A review
Kim, J., Campbell, A. S., & Wang, J. (2018) · Talanta · doi:10.1016/j.talanta.2017.08.077 · Cited by 586
Monte-Moreno, E. (2011) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2011.05.001 · Cited by 311 · cited 2×
“As physiological factors that are relevant for the method, blood viscosity and vessel compliance have been listed and claimed, while blood viscosity is mentioned to rely also on blood pressure and glucose level, altering the flux of blood in the corresponding vessels and the shape of the PPG pulse (see discussion on t…”
BackgroundWearable Continuous Glucose Monitoring Sensors: A Revolution in Diabetes Treatment
Cappon, G., Acciaroli, G., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2017) · Electronics · doi:10.3390/electronics6030065 · Cited by 237
New insights into the origin of remote PPG signals in visible light and infrared
Moço, A. V., Stuijk, S., & de Haan, G. (2018) · Scientific Reports · doi:10.1038/s41598-018-26068-2 · Cited by 168 · cited 2×
“However, there is also evidence that the PPG waveforms are also influenced from the dermal capillary density due to pressure variations from the larger blood vessels.(27,28) Lower frequency components are usually superimposed, which can be attributed to respiration, vasomotion, sympathetic nervous system activity, and…”
BackgroundMethodThe Technology Behind Glucose Meters: Test Strips
Hönes, J., Müller, P., & Surridge, N. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0005 · Cited by 156
Non-Invasive Glucose Monitoring: A Review of Challenges and Recent Advances
Lin, T. (2017) · Current Trends in Biomedical Engineering & Biosciences · doi:10.19080/ctbeb.2017.06.555696 · Cited by 142
The influence of glucose concentration upon the transport of light in tissue-simulating phantoms
Kohl, M., Essenpreis, M., & Cope, M. (1995) · Physics in Medicine and Biology · doi:10.1088/0031-9155/40/7/009 · Cited by 140 · cited 2×
“The values of the glucose absorption coefficients, as measured in D2O can be favorably compared with spectral data published by Kohl et al.14 Figure 1 also shows the magnitude of intrinsic glucose features and those signatures, which can be also induced by other solutes found in body fluids or temperature changes.”
Acciaroli, G., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2018) · Biosensors · doi:10.3390/bios8010024 · Cited by 122 · cited 1×
“For devices without direct access to capillary blood, as for needle-type CGM sensors that monitor the interstitial glucose level, effects from calibration to the inaccessible physiological skin site require additional attention.(4)”
MethodJungheim, K., & Koschinsky, T. (2002) · Diabetes Care · doi:10.2337/diacare.25.6.956 · Cited by 114
Wavelets and Their Applications
Byrnes, J. S., Byrnes, J. L., Hargreaves, K. A., & Berry, K. (1994) · Springer eBooks · doi:10.1007/978-94-011-1028-0 · Cited by 89
Noninvasive imaging of human skin hemodynamics using a digital red-green-blue camera
Nishidate, I., Тanaka, N., Kawase, T., Maeda, T., Yuasa, T., Aizu, Y., Yuasa, T., & Niizeki, K. (2011) · Journal of Biomedical Optics · doi:10.1117/1.3613929 · Cited by 82
Freckmann, G., Baumstark, A., Schmid, C., Pleus, S., Link, M., & Haug, C. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0208 · Cited by 77 · cited 1×
“Self-management plans for glycemic control usually require a measurement of (finger) capillary blood, as performed by widely used blood glucose test systems;(1) see report on exemplary devices that were tested for their analytical performance.(2) Alternate testing sites have also been suggested, but interferences from…”
MethodHuman tissue color as viewed in high dynamic range optical spectral transmission measurements
Petrov, G. I., Doronin, A. A., Whelan, H. T., Meglinski, I. V., & Yakovlev, V. V. (2012) · Biomedical Optics Express · doi:10.1364/boe.3.002154 · Cited by 72 · cited 1×
“Spectral data for eumelanin and pheomelanin are not given, as for the fingertips negligible absorption exists from these substances; for other skin areas these can influence the color of skin within the Vis/NIR spectral range significantly.(17) The inset in Figure 2A shows two exemplary reflection spectra of a fingert…”
BackgroundMicrovascularization of the human digit as studied by corrosion casting
Sangiorgi, S., Manelli, A., Congiu, T., Bini, A., Pilato, G., Reguzzoni, M., & Raspanti, M. (2004) · Journal of Anatomy · doi:10.1111/j.1469-7580.2004.00251.x · Cited by 71
Spatial frequency response of color image sensors: Bayer color filters and Foveon X3
Hubel, P. M., Liu, J., & Guttosch, R. J. (2004) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.561568 · Cited by 62 · cited 2×
“The pattern of colors in the CFA can be arranged in several ways, for example, as Bayer,21,22 multispectral,23 or sparse with panchromatic pixels.24 A less common technology is the vertically stacked sensor that does not require a separate CFA.22 In principle, also a completely custom pattern with custom filter dyes i…”
BackgroundDevice and method for noninvasive glucose assessment
Segman, Y. (. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818763457 · Cited by 55 · cited 4×
“Segman’s article describes an initial stage of development, as mentioned under his general notes.(10) For a more recent study,(11) improvement over the described stage is claimed.”
BackgroundInfluentialA HOLOGRAPHIC MODEL OF MEMORY, LEARNING AND EXPRESSION
Sutherland, J. G. (1990) · International Journal of Neural Systems · doi:10.1142/s0129065790000163 · Cited by 54 · cited 2×
“An associative memory implementation called “HNeT” (likely an abbreviation for holographic neural technology) that is cited as an emulation of the functions described,(35) has been presented in articles by Sutherland.(36,37) The latter article is a superset of the preceding one, as it is more extensive and also contai…”
BackgroundQu, J. Y., & Wilson, B. C. (1997) · Journal of Biomedical Optics · doi:10.1117/12.275453 · Cited by 46 · cited 1×
“Such estimates have been supported by simulations based on experimental tissue, water and glucose data, which have been carried out by Qu and Wilson for wavelengths of 800 and 960 nm.(25) They used a Monte Carlo simulation of the “photon random walk” based on wavelength-dependent absorption and scattering coefficients.”
MethodColor Restoration of RGBN Multispectral Filter Array Sensor Images Based on Spectral Decomposition
Park, C., & Kang, M. G. (2016) · Sensors · doi:10.3390/s16050719 · Cited by 44
Nutritional and Therapeutic Interventions for Diabetes and Metabolic Syndrome
Teran-Garcia, M. (2019) · Journal of Nutrition Education and Behavior · doi:10.1016/j.jneb.2019.03.018 · Cited by 42
Origin of Infrared Light Modulation in Reflectance-Mode Photoplethysmography
Sidorov, I. S., Владимирович, Р. Р., Koval, V. T., Giniatullin, R. A., & Kamshilin, A. A. (2016) · PLoS ONE · doi:10.1371/journal.pone.0165413 · Cited by 40 · cited 2×
“However, there is also evidence that the PPG waveforms are also influenced from the dermal capillary density due to pressure variations from the larger blood vessels.(27,28) Lower frequency components are usually superimposed, which can be attributed to respiration, vasomotion, sympathetic nervous system activity, and…”
BackgroundMethodEvaluation of a new noninvasive glucose monitoring device by means of standardized meal experiments
Pfützner, A., Strobl, S., Demircik, F., Redert, L., Pfützner, J., Pfützner, A. H., & Lier, A. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818758769 · Cited by 33
Yamakoshi, Y., Matsumura, K., Yamakoshi, T., Lee, J., Rolfe, P. J., Kato, Y., Shimizu, K., & Yamakoshi, K.-I. (2017) · Journal of Biomedical Optics · doi:10.1117/1.jbo.22.6.067001 · Cited by 31 · cited 2×
“The optical geometry is similar to the setup for side-scattered finger-photoplethysmography as published recently by Yamakoshi et al.16 For illustrating the basics of optical VIS/NIR technologies for noninvasive monitoring of tissue chromophores besides water and glucose, we compiled further relevant spectral intensit…”
MethodKhan, J. I. (1998) · IEEE Transactions on Neural Networks · doi:10.1109/72.668882 · Cited by 22
Noninvasive Blood Glucose Measurement
Uwadaira, Y., & Ikehata, A. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-812019-4.00036-2 · Cited by 21
The effect of varying LED intensity on pulse oximeter accuracy
De Kock, J. P., Reynolds, K. J., Tarassenko, L., & Moyles, J. T. B. (1991) · Journal of Medical Engineering & Technology · doi:10.3109/03091909109016208 · Cited by 16
Ramasahayam, S., Arora, L., & Chowdhury, S. R. (2016) · Smart sensors, measurement and instrumentation · doi:10.1007/978-3-319-47319-2_1 · Cited by 16
New Method for Computing Optical Hemodynamic Blood Pressure
Segman, Y. J. (2016) · Journal of Clinical & Experimental Cardiology · doi:10.4172/2155-9880.1000492 · Cited by 11
Evaluation of arterial oxygen saturation using RGB camera-based remote photoplethysmography
Haneishi, H., Nishidate, I., Nakano, K., Niizeki, K., Aizu, Y., & McDuff, D. (2018) · Optical Diagnostics and Sensing: Toward Point-of-Care Diagnostics · doi:10.1117/12.2290448 · Cited by 7
Schempp, W., & Segman, J. (1994) · Springer eBooks · doi:10.1007/978-94-011-1028-0_10 · Cited by 4
Kohonen, T. (1989) · Springer series in information sciences · doi:10.1007/978-3-642-88163-3_1 · Cited by 3 · cited 2×
“Implementation independent properties of such networks are summarized by Kohonen.(33) Principally, pairs of input and (desired) output vectors are learned.”
BackgroundIndexing of Image Databases Using Untrained 4D Holographic Memory Model
Gopalan, R. P., & Lee, G. (2002) · Lecture notes in computer science · doi:10.1007/3-540-36187-1_21 · Cited by 3 · cited 1×
“There is also an extended variant(38) that allows an interesting encoding of color images.(39) Our intention is to focus just on the parallels to and implication for Segman’s presentation.”
Background
Research works printed but not linked: 1
- Sutherland JG. The holographic neural method. In: Soucek B, ed. Fuzzy, Holographic, and Parallel Intelligence. 6th ed. New York, NY: John Wiley; 1992:7-92.
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Artificial Intelligence in Health Care: Current Applications and Issues
Park, C.-W., Seo, S. W., Kang, N., Ko, B., Choi, B. W., Park, C., Chang, D. K., Kim, H., Kim, H., Lee, H., et al. (2020) · Journal of Korean Medical Science · doi:10.3346/jkms.2020.35.e379 · Cited by 165
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 206
Delbeck, S., Vahlsing, T., Leonhardt, S. K., Steiner, G. R., & Heise, H. M. (2018) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-018-1395-x · Cited by 108
“Our focus was on the assessment of a strategy using color-coded photoplethysmographic imaging of the fingertip for non-invasive blood glucose testing [13].”
MethodSelf-citationReview of present method of glucose from human blood and body fluids assessment
Fiedorova, K., Augustýnek, M., Kubíček, J., Kudrna, P., & Bibbo, D. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2022.114348 · Cited by 103
Metadata
- Authors
- Vahlsing, T., Delbeck, S., Leonhardt, S., & Heise, H. M.
- Affiliations
- Federal Institute For Materials Research and Testing (DE)
- South Westphalia University of Applied Sciences (DE)
- RWTH Aachen University (DE)
- Corresponding author
- Heise, H. Michael
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2018-09-15
- Volume
- 12
- Issue
- 6
- Pages
- 1169-1177
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC6232738
- OpenAlex
- W2891502073
- Semantic Scholar
- 3dc505ae06d663369ddd04be882790c6f8b6bc07
- Cited by
- 28
- References
- 37
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 3 · 2023: 6 · 2022: 4 · 2021: 7 · 2020: 3 · 2019: 1 · 2018: 1
- FWCI
- 1.20
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion, cc-by)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- RWTH Publications (RWTH Aachen) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Spectroscopy Techniques in Biomedical and Chemical Research
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- non-invasive glucose monitoring, photoplethysmography, visible and near-infrared spectroscopy, self-monitoring of blood glucose, glucose concentration prediction
- MeSH terms
- Blood Glucose, Calibration, Color, Equipment and Supplies, Equipment Design, Fingers, Humans, Infrared Rays, Light, Lighting, Skin, Blood Glucose Self-Monitoring, Reproducibility of Results, Biosensing Techniques, Photoplethysmography, Spectroscopy, Near-Infrared
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.