Accuracy of a 14-Day Factory-Calibrated Continuous Glucose Monitoring System With Advanced Algorithm in Pediatric and Adult Population With Diabetes
Open accessThe DOI matches its Crossref record; 21 of the 21 references it cites were found in a registry or an index.
Abstract
Background: In this study, we evaluated the analytical performance of the second-generation factory-calibrated FreeStyle Libre Flash Glucose Monitoring (FreeStyle Libre 2) System compared to plasma venous blood glucose reference, Yellow Springs Instrument 2300 (YSI). Methods: The study enrolled participants aged four and above with type 1 or type 2 diabetes at seven sites in the United States. Adult participants (18+ years) participated in three in-clinic sessions and pediatric participants (4-17 years) participated in up to two in-clinic sessions stratified to provide data for days 1, 2, 3, 7, 8, 9, 12, 13, or 14 of sensor wear. Participants aged 11+ underwent supervised glycemic manipulation during in-clinic sessions to achieve glucose levels across the measurement range of the System. Performance evaluation included accuracy measures such as the proportion of continuous glucose monitoring (CGM) values that were within ±20% or ±20 mg/dL of reference glucose values, and bias measures such as the mean absolute relative difference (MARD) between CGM and reference values. Results: Data from the 144 adults and 129 pediatric participants were analyzed. Percent of sensor results within ±20%/20 mg/dL of YSI reference were 93.2% and 92.1%, and MARD was 9.2% and 9.7% for the adults and pediatric participants, respectively. The System performed well in the hypoglycemic range, with 94.3% of the results for the adult population and 96.1% of the data for pediatric population being within 15 mg/dL of the YSI reference. The time lag was 2.4 ± 4.6 minutes for adults and 2.1 ± 5.0 minutes for pediatrics. Conclusions: The System demonstrated improved analytical accuracy performance across the dynamic range during the 14-day sensor wear period as compared to the previous-generation device. NCT#: NCT03607448 and NCT03820050
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Nonadjunctive CGM provides valuable information on glucose levels and is indicated to replace information obtained from standard BGM, without requiring verification, and to aid daily and long-term DM-related treatment decisions, such as insulin dosing (Table 2).7,15,17-19,24,29,35”
From Importance of FDA-integrated continuous glucose monitors to ensure accuracy of continuous glucose monitoring (2025) · cites this paper as [35]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Miller, K. M., Foster, N. C., Beck, R. W., Bergenstal, R. M., DuBose, S. N., DiMeglio, L. A., Maahs, D. M., & Tamborlane, W. V. (2015) · Diabetes Care · doi:10.2337/dc15-0078 · Cited by 1,254
Bolinder, J., Antuna, R., Geelhoed-Duijvestijn, P., Kröger, J., & Weitgasser, R. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31535-5 · Cited by 893
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685
Self-citationBeck, R. W., Riddlesworth, T. D., Ruedy, K., Ahmann, A., Haller, S., Kruger, D., McGill, J. B., Polonsky, W., Price, D., Aronoff, S., et al. (2017) · Annals of Internal Medicine · doi:10.7326/m16-2855 · Cited by 635
Haak, T., Hanaire, H., Ajjan, R., Hermanns, N., Riveline, J.-P., & Rayman, G. (2017) · Diabetes Therapy · doi:10.1007/s13300-016-0223-6 · Cited by 623
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Aleppo, G., Ruedy, K. J., Riddlesworth, T. D., Kruger, D. F., Peters, A. L., Hirsch, I., Bergenstal, R. M., Toschi, E., Ahmann, A. J., Shah, V. N., et al. (2017) · Diabetes Care · doi:10.2337/dc16-2482 · Cited by 329
Fokkert, M. J., van Dijk, P. R., Edens, M. A., Abbes, S., de Jong, D., Slingerland, R. J., & Bilo, H. J. G. (2017) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2016-000320 · Cited by 209
Wadwa, R. P., Laffel, L. M., Shah, V. N., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0150 · Cited by 198
Shah, V. N., Laffel, L. M., Wadwa, R. P., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0143 · Cited by 197
Research works printed but not linked: 3
- 1. Abbott. Abbott Receives CE Mark for FreeStyle® Libre, A Revolutionary Glucose Monitoring System for People with Diabetes. Abbott Park, IL: Abbott; 2014. [Google Scholar]
- 18. Abbott. Abbotts’S FreeStyle® Libre 2, with Optional Real-Time Alarms Secures CE Mark for Use in Europe. Abbott Park, IL: Abbott; 2018. [Google Scholar]
- 20. Abbott. New Real-World Data Show that Abbott’s FreeStyle Libre System Delivers Positive Health Outcomes for People with Type 1 and Type 2 Diabetes. Abbott Park, IL: Abbott; 2020. [Google Scholar]
Cited by
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831
Continuous glucose monitors and automated insulin dosing systems in the hospital consensus guideline
Galindo, R. J., Umpierrez, G. E., Rushakoff, R. J., Basu, A., Lohnes, S., Nichols, J. H., Spanakis, E. K., Espinoza, J., Palermo, N. E., Awadjie, D. G., et al. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820954163 · Cited by 170
ISPAD Clinical Practice Consensus Guidelines 2022: Diabetes technologies: Glucose monitoring
Tauschmann, M., Forlenza, G., Hood, K., Cardona‐Hernandez, R., Giani, E., Hendrieckx, C., DeSalvo, D. J., Laffel, L. M., Saboo, B., Wheeler, B. J., et al. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13451 · Cited by 100
Self-citationHolzer, R., Bloch, W., & Brinkmann, C. K. (2022) · Sensors · doi:10.3390/s22052030 · Cited by 130
The Role of Wearable Devices in Chronic Disease Monitoring and Patient Care: A Comprehensive Review
Jafleh, E. A., Alnaqbi, F. A., Almaeeni, H. A., Faqeeh, S., Alzaabi, M. A., & Al Zaman, K. (2024) · Cureus · doi:10.7759/cureus.68921 · Cited by 125
Domingo‐Lopez, D. A., Lattanzi, G., Schreiber, L. H. J., Wallace, E. J., Wylie, R., O’Sullivan, J., Dolan, E. B., & Duffy, G. P. (2022) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2022.114280 · Cited by 126
Yoo, J. H., & Kim, J. H. (2023) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2022.0271 · Cited by 121
Expanding the Role of Continuous Glucose Monitoring in Modern Diabetes Care Beyond Type 1 Disease
Klupa, T., Czupryniak, L., Dzida, G. J., Fichna, P., Jarosz-Chobot, P. A., Gumprecht, J., Myśliwiec, M., Szadkowska, A., Bomba-Opon, D. A., Czajkowski, K., et al. (2023) · Diabetes Therapy · doi:10.1007/s13300-023-01431-3 · Cited by 97
Accuracy of the Third Generation of a 14-Day Continuous Glucose Monitoring System
Alva, S., Brazg, R., CASTORINO, K. N., Kipnes, M. S., Liljenquist, D. R., & Liu, H. (2023) · Diabetes Therapy · doi:10.1007/s13300-023-01385-6 · Cited by 90
Self-citationFreckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Tauschmann, M., Cardona‐Hernandez, R., DeSalvo, D. J., Hood, K. K., Лаптев, Д. Н., Olinder, A. L., Wheeler, B. J., & Smart, C. E. (2025) · Hormone Research in Paediatrics · doi:10.1159/000543156 · Cited by 83
Continuous Glucose Monitoring in the Hospital
Perez‐Guzman, M. C., Shang, T., Zhang, J. Y., Jornsay, D. L., & Klonoff, D. C. (2021) · Endocrinology and Metabolism · doi:10.3803/enm.2021.201 · Cited by 77
Shaw, J. L. V., Bannuru, R. R., Beach, L. A., ElSayed, N. A., Freckmann, G., Füzéry, A. K., Fung, A. W. S., Gilbert, J. D., Yun, H., Korpi-Steiner, N. L., et al. (2024) · Diabetes Care · doi:10.2337/dci24-0073 · Cited by 78
Benhalima, K., Beunen, K., Siegelaar, S. E., Painter, R., Murphy, H. R., Feig, D. S., Donovan, L. E., Polsky, S., Buschur, E., Levy, C. J., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(23)00116-x · Cited by 61
Shang, T., Zhang, J. Y., Thomas, A., Arnold, M. A., Vetter, B. N., Heinemann, L., & Klonoff, D. C. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211007212 · Cited by 71
Emerging Diabetes Technologies: Continuous Glucose Monitors/Artificial Pancreases
ALMURASHI, A. M., Rodriguez, E., & Garg, S. K. (2023) · Journal of the Indian Institute of Science · doi:10.1007/s41745-022-00348-3 · Cited by 67
Riveline, J.-P., Roussel, R., Vicaut, E., de Pouvourville, G., Detournay, B., Emery, C., Levrat-Guillen, F., & Guerci, B. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0085 · Cited by 64
Eichenlaub, M., Pleus, S., Rothenbühler, M., Bailey, T. S., Bally, L., Brazg, R., Bruttomesso, D., Diem, P., Boija, E. E., Fokkert, M. J., et al. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0465 · Cited by 55
Self-citationDivilly, P., Martine‐Edith, G., Zaremba, N., Søholm, U., MAHMOUDI, Z., Cigler, M., Ali, N., Abbink, E. J., Brøsen, J. M. B., de Galan, B. E., et al. (2024) · Diabetes Care · doi:10.2337/dc23-2332 · Cited by 51
Lindner, N., Kuwabara, A., & Holt, T. (2021) · Systematic Reviews · doi:10.1186/s13643-021-01644-2 · Cited by 47
Pemberton, J. S., Wilmot, E. G., Barnard‐Kelly, K., Leelarathna, L., Oliver, N., Randell, T., Taplin, C. E., Choudhary, P., & Adolfsson, P. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14962 · Cited by 44
Intralymphatic GAD-Alum (Diamyd®) Improves Glycemic Control in Type 1 Diabetes With HLA DR3-DQ2
Nowak, C., Lind, M., Šumník, Z., Pelikánová, T., Nattero‐Chávez, L., Lundberg, E., Rica, I., Martínez‐Brocca, M. A., de Adana, M. R., Wahlberg, J., et al. (2022) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgac343 · Cited by 39
Klonoff, D. C., Gabbay, M., Moon, S. J., & Wilmot, E. G. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241250357 · Cited by 36
Calibration algorithms for continuous glucose monitoring systems based on interstitial fluid sensing
Sun, T., Liu, J., & Chen, C. J. (2024) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2024.116450 · Cited by 34
Flash Glucose Monitoring and Patient Satisfaction: A Meta-Review of Systematic Reviews
Díez‐Fernández, A., Rodríguez‐Huerta, M. D., González, R. M., Laredo‐Aguilera, J. A., & Espinosa, N. M. M. (2021) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph18063123 · Cited by 33
Brummer, J., Glasbrenner, C., Figueroa, S. H., Koehler, K., & Höchsmann, C. (2024) · Frontiers in Nutrition · doi:10.3389/fnut.2023.1308348 · Cited by 29
Pleus, S., Eichenlaub, M., Dabla, P. K., Diem, P., Boija, E. E., Fokkert, M. J., Hinzmann, R., JENDLE, J. H., Klonoff, D. C., Lu, J., et al. (2025) · Clinica Chimica Acta · doi:10.1016/j.cca.2025.120728 · Cited by 27
Constrained IoT-Based Machine Learning for Accurate Glycemia Forecasting in Type 1 Diabetes Patients
Rodríguez‐Rodríguez, I., Campo-Valera, M., Rodríguez, J., & Frisa-Rubio, A. (2023) · Sensors · doi:10.3390/s23073665 · Cited by 25
Urbano, F., Farella, I., Brunetti, G., & Faienza, M. F. (2023) · International Journal of Molecular Sciences · doi:10.3390/ijms241511980 · Cited by 26
Alva, S., Bhargava, A., Bode, B. W., Brazg, R., CASTORINO, K. N., Kipnes, M. S., Liljenquist, D. R., Tran, H., Liu, H., & Nada, M. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251329364 · Cited by 23
Self-citationAnalytical Performance of the FreeStyle Libre 2 Glucose Sensor in Healthy Male Adults
Fellinger, E., Brandt, T., Creutzburg, J., Rommerskirchen, T., & Schmidt, A. (2024) · Sensors · doi:10.3390/s24175769 · Cited by 22
Li, T., Wang, Q., An, Y., Guo, L. J., Ren, L., Lei, L., & Chen, X. (2025) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2025.107750 · Cited by 22
Jin, Z., Thackray, A. E., King, J. A., Deighton, K., Davies, M. J., & Stensel, D. J. (2023) · Sensors · doi:10.3390/s23177417 · Cited by 22
Dávila-Rúales, V., Córdoba, L. F. G., Medina, A. M. G., Múñoz, Ó., Serrano, M. N., & Henao, D. C. (2024) · Therapeutic Advances in Endocrinology and Metabolism · doi:10.1177/20420188241304459 · Cited by 22
Should continuous glucose monitoring be used to manage neonates at risk of hypoglycaemia?
Kalogeropoulou, M., Iglesias‐Platas, I., & Beardsall, K. (2023) · Frontiers in Pediatrics · doi:10.3389/fped.2023.1115228 · Cited by 21
March, C. A., Sherman, J., Bannuru, R. R., Fischer, A., Gabbay, R. A., Rodriguez, H., Rodriguez, J., Weissberg-Benchell, J. A., Woodward, C., & ElSayed, N. A. (2023) · Diabetes Care · doi:10.2337/dci23-0083 · Cited by 22
Edward, R., & Priefer, R. (2023) · Primary care diabetes · doi:10.1016/j.pcd.2023.10.009 · Cited by 20
Assessment of the FreeStyle Libre 2 interstitial glucose monitor in hypo- and euglycemic cats
Berg, A. S., Crews, C. D., Adin, C. A., Alfonso‐Castro, A., Hill, S. B., Mott, J., & Gilor, C. (2023) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.16820 · Cited by 19
Nakazawa, T., Sekine, R., Kitabayashi, M., Hashimoto, Y., Ienaka, A., Morishita, K., Fujii, T., Ito, M. A., & Matsushita, F. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.3.037001 · Cited by 17
Accuracy and Safety of the 15-Day CareSens Air Continuous Glucose Monitoring System
Kim, K., Lee, S., Yoo, W. S., & Park, C. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0468 · Cited by 16
Predicting risk for nocturnal hypoglycemia after physical activity in children with type 1 diabetes
Leutheuser, H., Bartholet, M., Marx, A., Pfister, M., Burckhardt, M., Bachmann, S., & Vogt, J. E. (2024) · Frontiers in Medicine · doi:10.3389/fmed.2024.1439218 · Cited by 16
Use of continuous glucose monitoring trend arrows in the younger population with type 1 diabetes
Elbarbary, N. S., Moser, O., Al-Yaarubi, S., Alsaffar, H., Shaikh, A. M. A., Ajjan, R., & Deeb, A. (2021) · Diabetes and Vascular Disease Research · doi:10.1177/14791641211062155 · Cited by 16
Hochfellner, D. A., Simic, A., Taucher, M. T., Sailer, L. S., Kopanz, J., Pöttler, T., & Mader, J. K. (2022) · Biosensors · doi:10.3390/bios12020106 · Cited by 15
Use of continuous glucose monitoring and hybrid closed‐loop therapy in pregnancy
Benhalima, K., & Yamamoto, J. M. (2024) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15999 · Cited by 15
Predicting glucose level with an adapted branch predictor
Koutný, T., & Mayo, M. (2022) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2022.105388 · Cited by 15
Flash glucose monitoring system in special situations
Rigon, F., Ronsoni, M. F., Vianna, A. G. D., Schiavon, L. D. L., Hohl, A., & van de Sande‐Lee, S. (2022) · Archives of Endocrinology and Metabolism · doi:10.20945/2359-3997000000479 · Cited by 15
Three-wavelength PPG and support vector machine for non-invasive estimation of blood glucose
Castro-Pimentel, L. A., Téllez-Anguiano, A. D. C., Coronado-Reyes, O. I., & Díaz-Huerta, J. L. (2023) · Optical and Quantum Electronics · doi:10.1007/s11082-023-04927-1 · Cited by 15
Performance of a Novel Continuous Glucose Monitoring Device in People With Diabetes
Mader, J. K., Waldenmaier, D., Mueller-Hoffmann, W., Mueller, K., Angstmann, M., Vogt, G., Rieger, C. C., Eichenlaub, M., Forst, T. M., & Freckmann, G. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241267774 · Cited by 16
Wilmot, E. G., Evans, M. L., Barnard-Kelly, K. D., Burns, M., CRANSTON, I. C., Elliott, R. A., Gkountouras, G., Kanumilli, N., Krishan, A., Kotonya, C., et al. (2021) · BMJ Open · doi:10.1136/bmjopen-2021-050713 · Cited by 15
Molfetta, S. D., Rossi, A., Assaloni, R., Cherubini, V., Consoli, A., Bartolo, P. D., Guardasole, V., Laurenzi, A., Lombardo, F., Maffeis, C., et al. (2022) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2022.109867 · Cited by 14
Handy, C., Chaudhry, M. S., Qureshi, M. R. A., Love, B., Shillingford, J., Plum‐Mörschel, L., & Zijlstra, E. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231170242 · Cited by 13
Kaduk, K., Kaeber, M., Kühnel, A., Torrado, M. B., Grahlow, M., Derntl, B., & Kroemer, N. B. (2025) · EBioMedicine · doi:10.1016/j.ebiom.2025.106035 · Cited by 14
Use of Continuous Glucose Monitoring in Patients Following Bariatric Surgery: A Scoping Review
Yu, Y. H., & Groth, S. W. (2023) · Obesity Surgery · doi:10.1007/s11695-023-06704-1 · Cited by 12
Pleus, S., Link, M., Hinzmann, R., Öter, Ş., Eichenlaub, M., Baumstark, A., Haug, C., & Freckmann, G. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0489 · Cited by 12
Lameijer, A., Bakker, J. J., Kao, K., Xu, Y., Gans, R. O. B., Bilo, H. J. G., Dunn, T. C., & van Dijk, P. R. (2023) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2023.110735 · Cited by 12
Yoo, J. H., Moon, S. J., Park, C.-Y., & Kim, J. H. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241262106 · Cited by 11
Calibrating Glucose Sensors at the Edge: A Stress Generation Model for Tiny ML Drift Compensation
Sabatini, A., Cenerini, C., Vollero, L., & Pau, D. P. (2024) · BioMedInformatics · doi:10.3390/biomedinformatics4020083 · Cited by 11
Update on Patient Self-Testing with Portable and Wearable Devices: Advantages and Limitations
Lippi, G., Pighi, L., & Mattiuzzi, C. (2024) · Diagnostics · doi:10.3390/diagnostics14182037 · Cited by 11
Wang, R., Kyi, M., Krishnamoorthi, B., Tjahyadi, J., Connell, A., Chiang, C. Y., Renouf, D., Barmanray, R., & Fourlanos, S. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0604 · Cited by 11
Patient Adherence for Oral Combination Therapies in Diabetes Management: A Scoping Review
Zarrinkamar, M., Geran, M., & Geran, M. (2025) · Health Science Reports · doi:10.1002/hsr2.70780 · Cited by 11
Performance Assessment of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes
Kölle, J., Eichenlaub, M., Mende, J., Link, M., Vetter, B., Safary, E., Pleus, S., Haug, C., & Freckmann, G. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231159657 · Cited by 9
“Results of the established, market-approved FL2 largely agreed with the previous study 8 ; however, the importance of assessing individual sensor accuracy was highlighted.”
MethodResultInfluentialParrilla, M. (2026) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2026.118789 · Cited by 9
Derks, T. G. J., Overduin, R. J., Grünert, S. C., Rossi, A., & CGM GSD Collaborators Group (2025) · Journal of Inherited Metabolic Disease · doi:10.1002/jimd.70040 · Cited by 8
Elabd, S. S., & Poole, R. (2023) · Practical Diabetes · doi:10.1002/pdi.2479 · Cited by 8
Continuous glucose monitoring in para cyclists: An observational study
Weijer, V., van der Werf, R., van der Haijden, M., Jeukendrup, A. E., van Loon, L. J. C., & van Dijk, J. (2024) · European Journal of Sport Science · doi:10.1002/ejsc.12220 · Cited by 8
Maeda, R., Onoue, T., Mizutani, K., Suzuki, K., Handa, T., Kobayashi, T., Iwama, S., Miyata, T., Sugiyama, M., Hagiwara, D., et al. (2025) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2025.112074 · Cited by 8
Advances in screening and diagnostic lab-on-chip tools for gynaecological cancers – a review
Nujhat, S., Leese, H. S., Lorenzo, M. D., Bowen, R. L., & Moise, S. (2023) · Artificial Cells Nanomedicine and Biotechnology · doi:10.1080/21691401.2023.2274047 · Cited by 8
Giza, S., Kotanidou, E. P., Tsinopoulou, V. R., Poukoulidou, T., Ntouma, S., & Galli-Tsinopoulou, A. (2025) · Children · doi:10.3390/children12121668 · Cited by 7
“Accuracy improvement and the optional alarm feature of new-generation CGM systems can be rather useful for the adoption of this technology in the daily life of the pediatric population with T1D [28].”
Pleus, S., Eichenlaub, M., Waldenmaier, D., & Freckmann, G. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241236504 · Cited by 7
Resi, V., Bianchi, C., Burlina, S., Grancini, V., Manicardi, E., Masulli, M., Scarpitta, A. M., Sorice, G. P., & Fresa, R. (2025) · Acta Diabetologica · doi:10.1007/s00592-025-02592-2 · Cited by 7
Noninvasive Glucose Monitor Using Dielectric Spectroscopy
Buehler, L. A., Balasubramanian, V., Baskerville, S., Bailey, R. J., McCarthy, K., Rippen, M., Bena, J., & Lansang, M. C. (2021) · Endocrine Practice · doi:10.1016/j.eprac.2021.09.010 · Cited by 7
Šimunović, M., Kumrić, M., Rušić, D., Šimunović, M. P., & Božić, J. (2024) · Diagnostics · doi:10.3390/diagnostics14242801 · Cited by 7
Accuracy of the Freestyle Libre 3 Continuous Glucose Monitoring System in Hypo- and Euglycemic Cats
Tardo, A. M., Crews, C. D., Mott, J., Porter, L. T., Adin, C. A., & Gilor, C. (2025) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.70048 · Cited by 6
de la Marca, M. D. L. O. N., Díaz-Soto, G., Ibáñez, M. S., Simón, S. D. A., Fernández-Velasco, P., Torres, B., Gómez, J. J. L., Hoyos, E. G., Buigues, A. O., García, E. D., et al. (2022) · Endocrinología Diabetes y Nutrición (English ed ) · doi:10.1016/j.endien.2022.01.009 · Cited by 6
Moving with technological advancements: blood glucose monitoring from a district nurse's perspective
Bailey, L. (2022) · British Journal of Community Nursing · doi:10.12968/bjcn.2022.27.10.480 · Cited by 6
Styles, S. E., Wheeler, B. J., Boucsein, A., Crocket, H. R., de Lange, M., Signal, D., Wiltshire, E. J., Cunningham, V. J., Lala, A., Cutfield, W. S., et al. (2021) · Journal of Diabetes & Metabolic Disorders · doi:10.1007/s40200-021-00907-y · Cited by 6
Krakauer, M., Alves, S. T. D. F., Rodacki, M., Gabbay, M. A. L., Vianna, A. G. D., Scharf, M., Lamounier, R. N., Franco, D. R., Araujo, L. R., & Calliari, L. E. (2025) · Diabetology & Metabolic Syndrome · doi:10.1186/s13098-025-01797-3 · Cited by 5
Yafi, M., Shah, A., & Vélez, K. (2023) · Translational Pediatrics · doi:10.21037/tp-23-145 · Cited by 5
Magliozzo, M., Tumminia, A., Arpi, M. L., Deiana, E., Guglielmo, M., Giannone, G., Frasca, F., & Gullo, D. (2024) · Journal of Endocrinological Investigation · doi:10.1007/s40618-024-02472-6 · Cited by 5
Azuma, D., Okuda, H., & Saeger, B. (2025) · Frontiers in Clinical Diabetes and Healthcare · doi:10.3389/fcdhc.2025.1465732 · Cited by 4
Kalandyk-Osinko, K., Baran, J., Baran, R., Gierek, K., Juza, A., Kraśnianin, E., Januszek, S., Blajer‐Olszewska, B., Kluz, T., & Darmochwał-Kolarz, D. (2024) · Medical Science Monitor · doi:10.12659/msm.943644 · Cited by 4
Herman, J. L., Lidbury, J. A., Atiee, G. F., Washburn, S. E., & Patterson, C. A. (2025) · American Journal of Veterinary Research · doi:10.2460/ajvr.25.02.0055 · Cited by 4
Pros and cons of continous glucose monitoring
Ciechański, M., Witkowska, E., Ostańska, A., Szafran, A., Wiśniewska, K., Piasek, L., Godek, G., Więcław, K., Stańko, K., & Terelak, W. (2023) · Journal of Medical Science · doi:10.20883/medical.e873 · Cited by 4
Murata, T., Sakane, N., Hirota, Y., Toyoda, M., Matsuhisa, M., Kuroda, A., Itoh, A., Meguro, S., Miura, J., Matoba, Y., et al. (2024) · The Journal of Medical Investigation · doi:10.2152/jmi.71.225 · Cited by 4
An overview of the FreeStyle Libre monitoring system for individuals with type 1 and type 2 diabetes
Molfetta, S. D., Cutruzzolà, A., Gnasso, A., & Irace, C. (2025) · Expert Review of Medical Devices · doi:10.1080/17434440.2025.2489494 · Cited by 4
Preechasuk, L., Avari, P., Oliver, N. S., & Reddy, M. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0434 · Cited by 4
Babaya, N., Noso, S., Hiromine, Y., Taketomo, Y., Niwano, F., Yoshida, S., Yasutake, S., Kawabata, Y., Maeda, N., & Ikegami, H. (2024) · Scientific Reports · doi:10.1038/s41598-024-55124-3 · Cited by 4
Shimizu, N., Hayashi, A., Ito, S., Suzuki, A., Fujishima, R., Matoba, K., Wada, T., Takano, K., Katagiri, M., & Shichiri, M. (2021) · Endocrine Journal · doi:10.1507/endocrj.ej21-0540 · Cited by 4
Chen, A. X., Radhakutty, A., Zimmermann, A., Stranks, S. N., Thompson, C. H., & Burt, M. G. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0360 · Cited by 4
Liu, Y., Zhang, J., Chun, X., Gao, Y., Yao, R., Liang, Y., Zhu, L., He, Y., & Huang, W. (2024) · Endocrine Practice · doi:10.1016/j.eprac.2024.06.004 · Cited by 3
Tateiwa, T., Shikuma, J., Takahashi, Y., Nakamura, I., Matsumura, H., Suzuki, R., & Yamamoto, K. (2025) · Life · doi:10.3390/life15101594 · Cited by 3
Bresci, A., Kim, Y., Jue, M., Foglia, L. P., Moores, E., So, P. T. C., & Kang, J. W. (2026) · Journal of Diabetes Science and Technology · doi:10.1177/19322968261445056 · Cited by 3
Aberer, F., Haberl, H. C., Elsayed, H., Pöttler, T., Hochfellner, D. A., & Mader, J. K. (2024) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15719 · Cited by 3
Nakazawa, T., Morishita, K., Ienaka, A., Fujii, T., Ito, M. A., & Matsushita, F. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.10.107001 · Cited by 3
Physiological reconstruction of blood glucose level using CGMS-signals only
Koutný, T. (2022) · Scientific Reports · doi:10.1038/s41598-022-09884-5 · Cited by 3
Jakubczyk, N., Leszczyńska-Knaga, E., Kurowski, M., Kuczapska, K., Gliwa, A., Ryglewicz, M., Fabian, D., Kwaśniewska, P., Moczydłowski, P., & Michalczuk, K. (2025) · Journal of Education Health and Sport · doi:10.12775/jehs.2025.79.58310 · Cited by 2
O'Donnell, D. J. P., Burns, S., Drever, S., Quesnelle, L., & Yuen, B. (2025) · JMIR Diabetes · doi:10.2196/73485 · Cited by 2
Santana, D., Mosteiro, A., Pedrosa, L., Pujol-Fontrodona, G., Zattera, L., Werner, M., Ortega, E., Llull, L., Torné, R., & Amaro, S. (2025) · Scientific Reports · doi:10.1038/s41598-025-29495-0 · Cited by 2
Letissier, N. B., Rault, C. C. S., Faucher, C., Rabois, E., Durocher, L., Bouleti, C., Ragot, S., Hadjadj, S., Mosbah, H., Piguel, X., et al. (2026) · Diabetes Technology & Therapeutics · doi:10.1177/15209156261423954 · Cited by 2
Metadata
- Authors
- Alva, S., Bailey, T., Brazg, R., Budiman, E. S., Castorino, K., Christiansen, M. P., Forlenza, G., Kipnes, M., Liljenquist, D. R., & Liu, H.
- Affiliations
- Abbott Diabetes Care (United States) (US)
- AMCR Institute (US)
- Rainier Clinical Research Center (US)
- Sansum Diabetes Research Institute (US)
- Diablo Clinical Research (US)
- University of Colorado Denver (US)
- DGD Clinic, San Antonio, TX, USA
- Rocky Mountain Diabetes and Osteoporosis Center (US)
- Corresponding author
- Alva, Shridhara
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2020-09-19
- Volume
- 16
- Issue
- 1
- Pages
- 70-77
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC8875061
- OpenAlex
- W3087175985
- Cited by
- 164
- References
- 21
- Citations by year
- 2026: 28 · 2025: 32 · 2024: 37 · 2023: 32 · 2022: 21 · 2021: 13 · 2020: 1
- FWCI
- 12.63
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- abbott diabetes care
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- continuous glucose monitoring, FreeStyle Libre 2, factory calibration, analytical accuracy, type 1 diabetes, type 2 diabetes, pediatric diabetes, hypoglycemia, mean absolute relative difference
- MeSH terms
- Adult, Aged, Algorithms, Blood Glucose, Child, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Glucose, Humans, Pediatrics, Blood Glucose Self-Monitoring, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.