The BpTRU automatic blood pressure monitor compared to 24 hour ambulatory blood pressure monitoring in the assessment of blood pressure in patients with hypertension
Open accessThe DOI matches its Crossref record; 21 of the 21 references it cites were found in a registry or an index.
Abstract
Background: Increasing evidence suggests that ABPM more closely predicts target organ damage than does clinic measurement. Future guidelines may suggest ABPM as routine in the diagnosis and monitoring of hypertension. This would create difficulties as this test is expensive and often difficult to obtain. The purpose of this study is to determine the degree to which the BpTRU automatic blood pressure monitor predicts results on 24 hour ambulatory blood pressure monitoring (ABPM). Methods: A quantitative analysis comparing blood pressure measured by the BpTRU device with the mean daytime blood pressure on 24 hour ABPM. The study was conducted by the Centre for Studies in Primary Care, Queen's University, Kingston, Ontario, Canada on adult primary care patients who are enrolled in two randomized controlled trials on hypertension. The main outcomes were the mean of the blood pressures measured at the three most recent office visits, the initial measurement on the BpTRU-100, the mean of the five measurements on the BpTRU monitor, and the daytime average on 24 hour ABPM. Results: The group mean of the three charted clinic measured blood pressures (150.8 (SD10.26) / 82.9 (SD 8.44)) was not statistically different from the group mean of the initial reading on BpTRU (150.0 (SD21.33) / 83.3 (SD12.00)). The group mean of the average of five BpTRU readings (140.0 (SD17.71) / 79.8 (SD 10.46)) was not statistically different from the 24 hour daytime mean on ABPM (141.5 (SD 13.25) / 79.7 (SD 7.79)). Within patients, BpTRU average correlated significantly better with daytime ambulatory pressure than did clinic averages (BpTRU r = 0.571, clinic r = 0.145). Based on assessment of sensitivity and specificity at different cut-points, it is suggested that the initial treatment target using the BpTRU be set at <135/85 mmHG, but achievement of target should be confirmed using 24 hour ABPM. Conclusion: The BpTRU average better predicts ABPM than does the average of the blood pressures recorded on the patient chart from the three most recent visits. The BpTRU automatic clinic blood pressure monitor should be used as an adjunct to ABPM to effectively diagnose and monitor hypertension.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(2003) · Journal of Hypertension · doi:10.1097/00004872-200311000-00002 · Cited by 2,672
Staessen, J. A. (1999) · JAMA · doi:10.1001/jama.282.6.539 · Cited by 1,592 · cited 3×
“The generally accepted blood pressure target on ABPM is a daytime average pressure of < 135/85 mmHG [2-4]; the target blood pressure when measured in the doctors office is < 140/90 mmHG.”
Parati, G., Pomidossi, G., Albini, F., Malaspina, D., & Mancia, G. (1987) · Journal of Hypertension · doi:10.1097/00004872-198702000-00013 · Cited by 952 · cited 1×
“ABPM also better predicted cardiovascular events such as MI, CHF, stroke and TIA [4,5] as well as other target organ damage such as ventricular hypertrophy, proteinuria, plasma creatinine, and stroke [5,7-14].”
Redon, J., Campos, C., Narciso, M. L., Rodicio, J. L., Pascual, J. M., & Ruilope, L. M. (1998) · Hypertension · doi:10.1161/01.hyp.31.2.712 · Cited by 420 · cited 2×
“ABPM also better predicted cardiovascular events such as MI, CHF, stroke and TIA [4,5] as well as other target organ damage such as ventricular hypertrophy, proteinuria, plasma creatinine, and stroke [5,7-14].”
Zanchetti, A., Bond, M. G., Hennig, M., Neiss, A., Mancia, G., Palù, C. D., Hansson, L., Magnani, B., Rahn, K.-H., Reid, J., et al. (1998) · Journal of Hypertension · doi:10.1097/00004872-199816070-00008 · Cited by 305 · cited 1×
“ABPM also better predicted cardiovascular events such as MI, CHF, stroke and TIA [4,5] as well as other target organ damage such as ventricular hypertrophy, proteinuria, plasma creatinine, and stroke [5,7-14].”
Ohkubo, T., Imai, Y., Tsuji, I., Nagai, K., Watanabe, N., Minami, N., Itoh, O., Bando, T., Sakuma, M., Fukao, A., et al. (1997) · Journal of Hypertension · doi:10.1097/00004872-199715040-00006 · Cited by 250 · cited 1×
“The Ohsama study found that cardiovascular mortality, but not all cause mortality, was more closely predicted by the daytime systolic blood pressure than clinic pressures [6].”
Little, P., Barnett, J., Barnsley, L., Marjoram, J., Fitzgerald-Barron, A., & Mant, D. (2002) · BMJ · doi:10.1136/bmj.325.7358.254 · Cited by 171 · cited 1×
“It has been found that the measurements taken at home, or while not in the presence of clinic staff, have a better correlation with daytime ABPM than do nurse or physician measurements [15-17].”
Little, P., Barnett, J., Barnsley, L., Marjoram, J., Fitzgerald-Barron, A., & Mant, D. (2002) · BMJ · doi:10.1136/bmj.325.7358.258 · Cited by 168 · cited 1×
“Home or clinic self measurement is also preferred by patients over 24 hour monitoring [18].”
Palatini, P., Frigo, Bertolo, O., Roman, E., Da Cortà, R., & Winnicki, M. (1998) · PubMed · Cited by 134
Verdecchia, P., Reboldi, G., Porcellati, C., Schillaci, G., Pede, S., Bentivoglio, M., Angeli, F., Norgiolini, S., & Ambrosio, G. (2002) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(01)01827-7 · Cited by 129
Cited by
Measurement of Blood Pressure in Humans: A Scientific Statement From the American Heart Association
Muntner, P., Shimbo, D., Carey, R. M., Charleston, J. B., Gaillard, T., Misra, S., Myers, M. G., Ogedegbe, G., Schwartz, J. E., Townsend, R. R., et al. (2019) · Hypertension · doi:10.1161/hyp.0000000000000087 · Cited by 1,512
Daskalopoulou, S. S., Rabi, D. M., Zarnke, K. B., Dasgupta, K., Nerenberg, K., Cloutier, L., Gelfer, M., Lamarre-Cliche, M., Milot, A., Bolli, P., et al. (2015) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2015.02.016 · Cited by 684
Pickering, T. G., Miller, N. H., Ogedegbe, G., Krakoff, L. R., Artinian, N. T., & Goff, D. (2008) · Hypertension · doi:10.1161/hypertensionaha.107.189010 · Cited by 668
Baroreflex Activation Therapy Lowers Blood Pressure in Patients With Resistant Hypertension
Bisognano, J. D., Bakris, G., Nadim, M. K., Sanchez, L., Kroon, A. A., Schafer, J., de Leeuw, P. W., & Sica, D. A. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2011.06.008 · Cited by 600
Quinn, R. R., Hemmelgarn, B. R., Padwal, R. S., Myers, M. G., Cloutier, L., Bolli, P., McKay, D. W., Khan, N. A., Hill, M. D., Mahon, J., et al. (2010) · Canadian Journal of Cardiology · doi:10.1016/s0828-282x(10)70378-0 · Cited by 346
Hoppe, U. C., Brandt, M.-C., Wachter, R., Beige, J., Rump, L. C., Kroon, A. A., Cates, A. W., Lovett, E. G., & Haller, H. (2012) · Journal of the American Society of Hypertension · doi:10.1016/j.jash.2012.04.004 · Cited by 289
Myers, M. G., Godwin, M., Dawes, M., Kiss, A., Tobe, S. W., Grant, F. C., & Kaczorowski, J. (2011) · BMJ · doi:10.1136/bmj.d286 · Cited by 285
Self-citationDaskalopoulou, S. S., Khan, N. A., Quinn, R. R., Ruzicka, M., McKay, D. W., Hackam, D. G., Rabkin, S. W., Rabi, D. M., Gilbert, R. E., Padwal, R. S., et al. (2012) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2012.02.018 · Cited by 266
Self-citationMeasurement of Blood Pressure in the Office: Recognizing the Problem and Proposing the Solution
Myers, M. G., Godwin, M., Dawes, M., Kiss, A., Tobe, S. W., & Kaczorowski, J. (2010) · Hypertension · doi:10.1161/hypertensionaha.109.141879 · Cited by 260
Self-citationBakris, G. L., Nadim, M. K., Haller, H., Lovett, E. G., Schafer, J. E., & Bisognano, J. D. (2012) · Journal of the American Society of Hypertension · doi:10.1016/j.jash.2012.01.003 · Cited by 245
Herrera, J., Ferrebuz, A., MacGregor, E. G., & Rodriguez-Iturbe, B. (2006) · Journal of the American Society of Nephrology · doi:10.1681/asn.2006080918 · Cited by 233
A history of preeclampsia identifies women who have underlying cardiovascular risk factors
Smith, G. N., Walker, M. C., Liu, A., Wen, S. W., Swansburg, M., Ramshaw, H., White, R. R., Roddy, M., & Hladunewich, M. (2009) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2008.06.035 · Cited by 225
Roerecke, M., Kaczorowski, J., & Myers, M. G. (2019) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2018.6551 · Cited by 216
Use of automated office blood pressure measurement to reduce the white coat response
Myers, M. G., Valdivieso, M., & Kiss, A. (2009) · Journal of Hypertension · doi:10.1097/hjh.0b013e32831b9e6b · Cited by 209
Oscillometric Blood Pressure Estimation: Past, Present, and Future
Forouzanfar, M., Dajani, H. R., Groza, V. Z., Bolic, M., Rajan, S., & Batkin, I. (2015) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2015.2434215 · Cited by 196
Prevalence of pseudoresistant hypertension due to inaccurate blood pressure measurement
Bhatt, H., Siddiqui, M., Judd, E., Oparil, S., & Calhoun, D. (2016) · Journal of the American Society of Hypertension · doi:10.1016/j.jash.2016.03.186 · Cited by 77
Office blood pressure measurement in the 21st century
Myers, M. G., Asmar, R., & Staessen, J. A. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13276 · Cited by 34
Management of Hypertension in Diabetic Patients With Chronic Kidney Disease
Steigerwalt, S. (2008) · Diabetes Spectrum · doi:10.2337/diaspect.21.1.30 · Cited by 19
Metadata
- Authors
- Beckett, L., & Godwin, M.
- Venue
- BMC Cardiovascular Disorders
- Published
- 2005-06-28
- Volume
- 5
- Issue
- 1
- Pages
- 18-18
- ISSN
- 1471-2261 (ISSN-L)
- OpenAlex
- W1871270394
- Cited by
- 175
- References
- 21
- FWCI
- 3.39
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.