Fabrication and in vitro deployment of a laser-activated shape memory polymer vascular stent

Open access

Abstract

Background: Vascular stents are small tubular scaffolds used in the treatment of arterial stenosis (narrowing of the vessel). Most vascular stents are metallic and are deployed either by balloon expansion or by self-expansion. A shape memory polymer (SMP) stent may enhance flexibility, compliance, and drug elution compared to its current metallic counterparts. The purpose of this study was to describe the fabrication of a laser-activated SMP stent and demonstrate photothermal expansion of the stent in an in vitro artery model. Methods: A novel SMP stent was fabricated from thermoplastic polyurethane. A solid SMP tube formed by dip coating a stainless steel pin was laser-etched to create the mesh pattern of the finished stent. The stent was crimped over a fiber-optic cylindrical light diffuser coupled to an infrared diode laser. Photothermal actuation of the stent was performed in a water-filled mock artery. Results: At a physiological flow rate, the stent did not fully expand at the maximum laser power (8.6 W) due to convective cooling. However, under zero flow, simulating the technique of endovascular flow occlusion, complete laser actuation was achieved in the mock artery at a laser power of ~8 W. Conclusion: We have shown the design and fabrication of an SMP stent and a means of light delivery for photothermal actuation. Though further studies are required to optimize the device and assess thermal tissue damage, photothermal actuation of the SMP stent was demonstrated.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1186/1475-925x-6-43

Cites

46 of 46 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1 SMP stent. Close-up of the SMP stent mounted over the SMP foam cylinder (black arrow) and the coaxial SMP diffuser (white arrow) prior to crimping (a) and after crimping (b). The distal end of the stainless steel hypotube containing the optical fiber is indicated by the asterisk in (a). An image of the diffuser emission, which decreases with distance from the optical fiber tip, is shown in (c). A cross-sectional diagram of the expanded stent with the diffuser and foam cylinder in place is shown in (d).
Figure 1 SMP stent. Close-up of the SMP stent mounted over the SMP foam cylinder (black arrow) and the coaxial SMP diffuser (white arrow) prior to crimping (a) and after crimping (b). The distal end of the stainless steel hypotube containing the optical fiber is indicated by the asterisk in (a). An image of the diffuser emission, which decreases with distance from the optical fiber tip, is shown in (c). A cross-sectional diagram of the expanded stent with the diffuser and foam cylinder in place is shown in (d).CC BY · PubMed Central
Figure 2 Artery model. Schematic of the artery model. WB: water bath, PP: peristaltic pump, TB: Touhy Borst valve, SS: stainless steel cylinder, OC: outer chamber, TC: thermocouple probe, MA: mock artery.
Figure 2 Artery model. Schematic of the artery model. WB: water bath, PP: peristaltic pump, TB: Touhy Borst valve, SS: stainless steel cylinder, OC: outer chamber, TC: thermocouple probe, MA: mock artery.CC BY · PubMed Central
Figure 3 Stent deployment. Timeline of SMP stent deployment in the mock artery (zero flow) as the laser power was gradually increased. Laser duration was approximately 6.3 min.
Figure 3 Stent deployment. Timeline of SMP stent deployment in the mock artery (zero flow) as the laser power was gradually increased. Laser duration was approximately 6.3 min.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 45 references

Research works printed but not linked: 2

  1. Donnelly RJ: Fluid dynamics. In A Physicist's Desk Reference: The Second Edition of Physics Vade Mecum. Edited by: Anderson HL. New York: American Institute of Physics; 1989:196–209.
  2. Anderson HL, Cohen ER: General section. In A Physicist's Desk Reference: The Second Edition of Physics Vade Mecum. Edited by: Anderson HL. New York: American Institute of Physics; 1989:37–39.

Cited by

See all 175

Metadata

Authors
Baer, G. M., Small, W., Wilson, T. S., Benett, W. J., Matthews, D. L., Hartman, J., & Maitland, D. J.
Affiliations
  • University of California, Davis (US)
  • Lawrence Livermore National Laboratory (US)
  • Kaiser Permanente (US)
Corresponding author
Small, Ward
Venue
BioMedical Engineering OnLine
Published
2007-11-27
Volume
6
Issue
1
Pages
43-43
ISSN
1475-925X (ISSN-L)
OpenAlex
W2161662294
Cited by
159
References
45
Citations by year
2026: 3 · 2025: 15 · 2024: 3 · 2023: 6 · 2022: 11 · 2021: 6 · 2020: 13 · 2019: 13 · 2018: 2 · 2017: 12 · 2016: 8 · 2015: 9 · 2014: 9 · 2013: 8 · 2012: 8
FWCI
15.26
Type
journal article
Language
en
Open access
Licence
  • cc-by
  • creativecommons.org/licenses/by/2.0 (tdm)
Locations
  • BioMedical Engineering OnLine (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) (submittedVersion)
  • University of North Texas Digital Library (University of North Texas) (submittedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
  • National Science Foundation (0120999)
  • U.S. Department of Energy (W-7405, -7405-ENG-48, W‐7405‐ENG‐48)
  • National Institutes of Health (W-7405-ENG-48)
  • University of California, Davis (PHY0120999)
  • National Institute of Biomedical Imaging and Bioengineering (R01EB000462)
  • Lawrence Livermore National Laboratory (W-7405-ENG 48, W-7405-ENG)
Topics
  • Physical Sciences › Engineering › Mechanical Engineering › Advanced Materials and Mechanics
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Coronary Interventions and Diagnostics
  • Physical Sciences › Materials Science › Polymers and Plastics › Polymer composites and self-healing
Keywords
photothermal actuation, thermoplastic polyurethane, vascular stents, arterial stenosis, laser etching, dip coating, convective cooling, light delivery
MeSH terms
Blood Vessel Prosthesis, Equipment Design, Lasers, Molecular Conformation, Polymers, Stainless Steel, Stents, Equipment Failure Analysis, Coated Materials, Biocompatible

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Fabrication and in vitro deployment of a laser-activated shape memory polymer vascular stent. https://app.cochat.ai/papers/10.1186/1475-925x-6-43
The paper
Baer, G. M., Small, W., Wilson, T. S., Benett, W. J., Matthews, D. L., Hartman, J., & Maitland, D. J. (2007). Fabrication and in vitro deployment of a laser-activated shape memory polymer vascular stent. BioMedical Engineering OnLine. https://doi.org/10.1186/1475-925x-6-43

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.