Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements
Open accessThe DOI matches its Crossref record; 15 of the 15 references it cites were found in a registry or an index.
Abstract
AIMS: Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error. We studied the severity of uncorrected velocity offset errors across sites and CMR systems. METHODS AND RESULTS: In a multi-centre, multi-vendor study, breath-hold through-plane retrospectively ECG-gated phase contrast acquisitions, as are used clinically for aortic and pulmonary flow measurement, were applied to static gelatin phantoms in twelve 1.5 T CMR systems, using a velocity encoding range of 150 cm/s. No post-processing corrections of offsets were implemented. The greatest uncorrected velocity offset, taken as an average over a 'great vessel' region (30 mm diameter) located up to 70 mm in-plane distance from the magnet isocenter, ranged from 0.4 cm/s to 4.9 cm/s. It averaged 2.7 cm/s over all the planes and systems. By theoretical calculation, a velocity offset error of 0.6 cm/s (representing just 0.4% of a 150 cm/s velocity encoding range) is barely acceptable, potentially causing about 5% miscalculation of cardiac output and up to 10% error in shunt measurement. CONCLUSION: In the absence of hardware or software upgrades able to reduce phase offset errors, all the systems tested appeared to require post-acquisition correction to achieve consistently reliable breath-hold measurements of flow. The effectiveness of offset correction software will still need testing with respect to clinical flow acquisitions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Cine or PC sequences can be used to quantify the stenosis, blockage, or regurgitation of these conduits.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, the quantification may be incorrect due to conduit artefacts [37,44].”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Subsequent CMR studies are usually performed if new clinical symptoms arise or every 3–5 years, depending on the previous clinical findings [34,36,37].”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“In paediatric centres with high patient volume, the first CMR is typically conducted around 8–9 years of age, when children can undergo the examination without anaesthesia.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Usually valved, these conduits—which may be synthetic or biological—can develop pulmonary regurgitation later on or function as a tiny conduit as people age.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Handbook of MRI Pulse Sequences
(2004) · Elsevier eBooks · doi:10.1016/b978-0-12-092861-3.x5000-6 · Cited by 1,610
Blood Flow Imaging by Cine Magnetic Resonance
Nayler, G. L., Firmin, D. N., & Longmore, D. B. (1986) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198609000-00001 · Cited by 634 · cited 1×
“Phase contrast cardiovascular magnetic resonance (CMR) [1] measurements of flow through planes transecting the great arteries are used clinically for calculations of cardiac output, shunt flow [2,3] or aortic or pulmonary regurgitation [4,5].”
Self-citationConcomitant gradient terms in phase contrast MR: Analysis and correction
Bernstein, M. A., Zhou, X. J., Polzin, J. A., King, K. F., Ganin, A., Pelc, N. J., & Glover, G. H. (1998) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910390218 · Cited by 547 · cited 2×
“Velocity offsets caused by Maxwell (concomitant) gradients in phase-contrast imaging are calculable allowing automatic correction during reconstruction [11].”
Rebergen, S. A., Chin, J. G., Ottenkamp, J., van der Wall, E. E., & de Roos, A. (1993) · Circulation · doi:10.1161/01.cir.88.5.2257 · Cited by 294
Gelfand, E., Hughes, S., Hauser, T., Yeon, S., Goepfert, L., Kissinger, K., Rofsky, N., & Manning, W. (2006) · Journal of Cardiovascular Magnetic Resonance · doi:10.1080/10976640600604856 · Cited by 265 · cited 2×
“In combination with measurements of left ventricular volume or mitral inflow, measurement of aortic outflow may also allow the indirect calculation of mitral regurgitation [5-7].”
Hundley, W. G., Li, H. F., Willard, J. E., Landau, C., Lange, R. A., Meshack, B. M., Hillis, L. D., & Peshock, R. M. (1995) · Circulation · doi:10.1161/01.cir.92.5.1151 · Cited by 229
Normal Thoracic Aorta Diameter on Cardiac Computed Tomography in Healthy Asymptomatic Adults
Mao, S. S., Ahmadi, N., Shah, B., Beckmann, D., Chen, A., Ngo, L., Flores, F. R., Gao, Y. L., & Budoff, M. J. (2008) · Academic Radiology · doi:10.1016/j.acra.2008.02.001 · Cited by 217 · cited 1×
“The software reported the most extreme mean velocity offset measured in a circular ‘great vessel’ region, 30 mm diameter [13], centered within specified distances (see below) from the centre of each velocity map.”
Quantification of mitral regurgitation by velocity-encoded cine nuclear magnetic resonance imaging
Fujita, N., Chazouilleres, A. F., Hartiala, J. J., O'sullivan, M., Heidenreich, P., Kaplan, J. D., Sakuma, H., Foster, E., Caputo, G. R., & Higgins, C. B. (1994) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(94)90642-4 · Cited by 198 · cited 1×
“In combination with measurements of left ventricular volume or mitral inflow, measurement of aortic outflow may also allow the indirect calculation of mitral regurgitation [ 5-7 ].”
Chernobelsky, A., Shubayev, O., Comeau, C. R., & Wolff, S. D. (2007) · Journal of Cardiovascular Magnetic Resonance · doi:10.1080/10976640601187588 · Cited by 163 · cited 5×
“Opposite polarities from the same scanner have been reported previously [9].”
InfluentialCorrection of phase offset errors in main pulmonary artery flow quantification
Lankhaar, J., Hofman, M. B. M., Marcus, J. T., Zwanenburg, J. J. M., Faes, T. J. C., & Noordegraaf, A. V. (2005) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20361 · Cited by 101
Self-citation
Cited by
Recommendations for Noninvasive Evaluation of Native Valvular Regurgitation
Zoghbi, W. A., Adams, D., Bonow, R. O., Enriquez-Sarano, M., Foster, E., Grayburn, P. A., Hahn, R. T., Han, Y., Hung, J., Lang, R. M., et al. (2017) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2017.01.007 · Cited by 3,556
“In addition, background phase, partial volume average from low spatial resolution and nonperpendicular plane selection, and complex jet flow pattern causing intravoxel dephasing and loss of phase coherence can all lead to inaccurate phase-contrast assessment.(82) Indirect quantification of TR can amplify errors and le…”
BackgroundSchulz-Menger, J., Bluemke, D. A., Bremerich, J., Flamm, S. D., Fogel, M. A., Friedrich, M. G., Kim, R. J., von Knobelsdorff-Brenkenhoff, F., Kramer, C. M., Pennell, D. J., et al. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-35 · Cited by 2,097
Self-citationTownsend, R. R., Wilkinson, I. B., Schiffrin, E. L., Avolio, A. P., Chirinos, J. A., Cockcroft, J. R., Heffernan, K. S., Lakatta, E. G., McEniery, C. M., Mitchell, G. F., et al. (2015) · Hypertension · doi:10.1161/hyp.0000000000000033 · Cited by 1,473
4D flow cardiovascular magnetic resonance consensus statement
Dyverfeldt, P., Bissell, M., Barker, A. J., Bolger, A. F., Carlhäll, C.-J., Ebbers, T., Francios, C. J., Frydrychowicz, A., Geiger, J., Giese, D., et al. (2015) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-015-0174-5 · Cited by 877
“Similar to 2D cine PC-CMR, the major sources of errors include eddy current effects [148], concomitant gradient field effects (Maxell terms) [149], gradient field nonlinearity [150, 151], and phase wraps resulting in velocity aliasing [152, 153].”
BackgroundMethodInfluentialSelf-citationFratz, S., Chung, T., Greil, G. F., Samyn, M. M., Taylor, A. M., Valsangiacomo Buechel, E. R., Yoo, S.-J., & Powell, A. J. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-51 · Cited by 507
“These include inappropriate setting of the venc, signal loss with complex turbulent flow, partial volume averaging, signal misregistration, and phase offset errors due to eddy currents or uncorrected concomitant gradients [70-72,110,120,121].”
Metadata
- Authors
- Gatehouse, P. D., Rolf, M. P., Graves, M. J., Hofman, M. B., Totman, J., Werner, B., Quest, R. A., Liu, Y., von Spiczak, J., Dieringer, M., Firmin, D. N., van Rossum, A., Lombardi, M., Schwitter, J., Schulz-Menger, J., & Kilner, P. J.
- Affiliations
- Royal Brompton Hospital (GB)
- Amsterdam UMC Location Vrije Universiteit Amsterdam (NL)
- University of Cambridge (GB)
- King's College London (GB)
- University Children's Hospital Zurich (CH)
- Hammersmith Hospital (GB)
- University of Auckland (NZ)
- Charité – Universitätsmedizin Berlin (DE)
- National Research Council (IT)
- University Hospital Zurich (CH)
- Corresponding author
- Kilner, Philip J
- ORCID
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2010-10
- Volume
- 12
- Issue
- 1
- Pages
- 5
- ISSN
- 1097-6647 (ISSN-L)
- PMC
- PMC2818657
- OpenAlex
- W2153286609
- Cited by
- 234
- References
- 15
- Citations by year
- 2026: 4 · 2025: 5 · 2024: 8 · 2023: 11 · 2022: 11 · 2021: 13 · 2020: 18 · 2019: 17 · 2018: 12 · 2017: 13 · 2016: 17 · 2015: 15 · 2014: 16 · 2013: 20 · 2012: 24
- FWCI
- 18.70
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- www.elsevier.com/legal/tdmrep-license (tdm)
- creativecommons.org/licenses/by/2/ (vor)
- Locations
- Journal of Cardiovascular Magnetic Resonance (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Pure Amsterdam UMC (publishedVersion, other-oa)
- Amsterdam UMC (VU Amsterdam) - Institutional Repository (publishedVersion, other-oa)
- MDC Repository (Max-Delbrueck-Center for Molecular Medicine) (acceptedVersion)
- Bournemouth University Research Online (Bournemouth University) (acceptedVersion)
- Zurich Open Repository and Archive (University of Zurich) (acceptedVersion, cc-by)
- National University of Singapore (submittedVersion, cc-by)
- Digital Academic REpository of VU University Amsterdam (Vrije Universiteit Amsterdam) (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Spiral (Imperial College London) (submittedVersion, cc-by)
- Repository for Publications and Research Data (ETH Zurich) (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- ETH Zürich Research Collection (cc-by)
- Zurich Open Repository and Archive (University of Zurich)
- Funders
- British Heart Foundation
- Vrije Universiteit Amsterdam
- Eidgenössische Technische Hochschule Zürich
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac Valve Diseases and Treatments
- Keywords
- phase-contrast MRI, flow measurement accuracy, regurgitant flow, shunt flow, aortic blood flow, pulmonary blood flow, cardiac output, gelatin phantom, offset correction
- MeSH terms
- Aorta, Blood Flow Velocity, Cardiac Output, Gelatin, Humans, Image Interpretation, Computer-Assisted, Materials Testing, Mitral Valve Insufficiency, Predictive Value of Tests, Pulmonary Artery, Reproducibility of Results, Respiratory Mechanics, Artifacts, Magnetic Resonance Imaging, Cine, Phantoms, Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. https://app.cochat.ai/papers/10.1186/1532-429x-12-5
- The paper
- Gatehouse, P. D., Rolf, M. P., Graves, M. J., Hofman, M. B., Totman, J., Werner, B., Quest, R. A., Liu, Y., von Spiczak, J., Dieringer, M., Firmin, D. N., van Rossum, A., Lombardi, M., Schwitter, J., Schulz-Menger, J., & Kilner, P. J. (2010). Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. Journal of Cardiovascular Magnetic Resonance. https://doi.org/10.1186/1532-429x-12-5
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.