Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements
Open accessThe DOI matches its Crossref record; 15 of the 15 references it cites were found in a registry or an index.
Abstract
AIMS: Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error. We studied the severity of uncorrected velocity offset errors across sites and CMR systems. METHODS AND RESULTS: In a multi-centre, multi-vendor study, breath-hold through-plane retrospectively ECG-gated phase contrast acquisitions, as are used clinically for aortic and pulmonary flow measurement, were applied to static gelatin phantoms in twelve 1.5 T CMR systems, using a velocity encoding range of 150 cm/s. No post-processing corrections of offsets were implemented. The greatest uncorrected velocity offset, taken as an average over a 'great vessel' region (30 mm diameter) located up to 70 mm in-plane distance from the magnet isocenter, ranged from 0.4 cm/s to 4.9 cm/s. It averaged 2.7 cm/s over all the planes and systems. By theoretical calculation, a velocity offset error of 0.6 cm/s (representing just 0.4% of a 150 cm/s velocity encoding range) is barely acceptable, potentially causing about 5% miscalculation of cardiac output and up to 10% error in shunt measurement. CONCLUSION: In the absence of hardware or software upgrades able to reduce phase offset errors, all the systems tested appeared to require post-acquisition correction to achieve consistently reliable breath-hold measurements of flow. The effectiveness of offset correction software will still need testing with respect to clinical flow acquisitions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Cine or PC sequences can be used to quantify the stenosis, blockage, or regurgitation of these conduits.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, the quantification may be incorrect due to conduit artefacts [37,44].”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Subsequent CMR studies are usually performed if new clinical symptoms arise or every 3–5 years, depending on the previous clinical findings [34,36,37].”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“In paediatric centres with high patient volume, the first CMR is typically conducted around 8–9 years of age, when children can undergo the examination without anaesthesia.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Usually valved, these conduits—which may be synthetic or biological—can develop pulmonary regurgitation later on or function as a tiny conduit as people age.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Handbook of MRI Pulse Sequences
(2004) · Elsevier eBooks · doi:10.1016/b978-0-12-092861-3.x5000-6 · Cited by 1,610
Blood Flow Imaging by Cine Magnetic Resonance
Nayler, G. L., Firmin, D. N., & Longmore, D. B. (1986) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198609000-00001 · Cited by 634 · cited 1×
“Phase contrast cardiovascular magnetic resonance (CMR) [1] measurements of flow through planes transecting the great arteries are used clinically for calculations of cardiac output, shunt flow [2,3] or aortic or pulmonary regurgitation [4,5].”
Self-citationConcomitant gradient terms in phase contrast MR: Analysis and correction
Bernstein, M. A., Zhou, X. J., Polzin, J. A., King, K. F., Ganin, A., Pelc, N. J., & Glover, G. H. (1998) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910390218 · Cited by 547 · cited 2×
“Velocity offsets caused by Maxwell (concomitant) gradients in phase-contrast imaging are calculable allowing automatic correction during reconstruction [11].”
Rebergen, S. A., Chin, J. G., Ottenkamp, J., van der Wall, E. E., & de Roos, A. (1993) · Circulation · doi:10.1161/01.cir.88.5.2257 · Cited by 294
Gelfand, E., Hughes, S., Hauser, T., Yeon, S., Goepfert, L., Kissinger, K., Rofsky, N., & Manning, W. (2006) · Journal of Cardiovascular Magnetic Resonance · doi:10.1080/10976640600604856 · Cited by 265 · cited 2×
“In combination with measurements of left ventricular volume or mitral inflow, measurement of aortic outflow may also allow the indirect calculation of mitral regurgitation [5-7].”
Hundley, W. G., Li, H. F., Willard, J. E., Landau, C., Lange, R. A., Meshack, B. M., Hillis, L. D., & Peshock, R. M. (1995) · Circulation · doi:10.1161/01.cir.92.5.1151 · Cited by 229
Normal Thoracic Aorta Diameter on Cardiac Computed Tomography in Healthy Asymptomatic Adults
Mao, S. S., Ahmadi, N., Shah, B., Beckmann, D., Chen, A., Ngo, L., Flores, F. R., Gao, Y. L., & Budoff, M. J. (2008) · Academic Radiology · doi:10.1016/j.acra.2008.02.001 · Cited by 217 · cited 1×
“The software reported the most extreme mean velocity offset measured in a circular ‘great vessel’ region, 30 mm diameter [13], centered within specified distances (see below) from the centre of each velocity map.”
Quantification of mitral regurgitation by velocity-encoded cine nuclear magnetic resonance imaging
Fujita, N., Chazouilleres, A. F., Hartiala, J. J., O'sullivan, M., Heidenreich, P., Kaplan, J. D., Sakuma, H., Foster, E., Caputo, G. R., & Higgins, C. B. (1994) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(94)90642-4 · Cited by 198 · cited 1×
“In combination with measurements of left ventricular volume or mitral inflow, measurement of aortic outflow may also allow the indirect calculation of mitral regurgitation [ 5-7 ].”
Chernobelsky, A., Shubayev, O., Comeau, C. R., & Wolff, S. D. (2007) · Journal of Cardiovascular Magnetic Resonance · doi:10.1080/10976640601187588 · Cited by 163 · cited 5×
“Opposite polarities from the same scanner have been reported previously [9].”
InfluentialCorrection of phase offset errors in main pulmonary artery flow quantification
Lankhaar, J., Hofman, M. B. M., Marcus, J. T., Zwanenburg, J. J. M., Faes, T. J. C., & Noordegraaf, A. V. (2005) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20361 · Cited by 101
Self-citation
Cited by
Recommendations for Noninvasive Evaluation of Native Valvular Regurgitation
Zoghbi, W. A., Adams, D., Bonow, R. O., Enriquez-Sarano, M., Foster, E., Grayburn, P. A., Hahn, R. T., Han, Y., Hung, J., Lang, R. M., et al. (2017) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2017.01.007 · Cited by 3,556
“In addition, background phase, partial volume average from low spatial resolution and nonperpendicular plane selection, and complex jet flow pattern causing intravoxel dephasing and loss of phase coherence can all lead to inaccurate phase-contrast assessment.(82) Indirect quantification of TR can amplify errors and le…”
BackgroundSchulz-Menger, J., Bluemke, D. A., Bremerich, J., Flamm, S. D., Fogel, M. A., Friedrich, M. G., Kim, R. J., von Knobelsdorff-Brenkenhoff, F., Kramer, C. M., Pennell, D. J., et al. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-35 · Cited by 2,097
Self-citationTownsend, R. R., Wilkinson, I. B., Schiffrin, E. L., Avolio, A. P., Chirinos, J. A., Cockcroft, J. R., Heffernan, K. S., Lakatta, E. G., McEniery, C. M., Mitchell, G. F., et al. (2015) · Hypertension · doi:10.1161/hyp.0000000000000033 · Cited by 1,473
4D flow cardiovascular magnetic resonance consensus statement
Dyverfeldt, P., Bissell, M., Barker, A. J., Bolger, A. F., Carlhäll, C.-J., Ebbers, T., Francios, C. J., Frydrychowicz, A., Geiger, J., Giese, D., et al. (2015) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-015-0174-5 · Cited by 877
“Similar to 2D cine PC-CMR, the major sources of errors include eddy current effects [148], concomitant gradient field effects (Maxell terms) [149], gradient field nonlinearity [150, 151], and phase wraps resulting in velocity aliasing [152, 153].”
BackgroundMethodInfluentialSelf-citationFratz, S., Chung, T., Greil, G. F., Samyn, M. M., Taylor, A. M., Valsangiacomo Buechel, E. R., Yoo, S.-J., & Powell, A. J. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-51 · Cited by 507
“These include inappropriate setting of the venc, signal loss with complex turbulent flow, partial volume averaging, signal misregistration, and phase offset errors due to eddy currents or uncorrected concomitant gradients [70-72,110,120,121].”
4D Flow cardiovascular magnetic resonance consensus statement: 2023 update
Bissell, M. M., Raimondi, F., Ait Ali, L., Allen, B. D., Barker, A. J., Bolger, A., Burris, N., Carhäll, C.-J., Collins, J. D., Ebbers, T., et al. (2023) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-023-00942-z · Cited by 296
“Particular attention should be given to large fields-of-view where regions-of-interest may reside far away from the magnet isocenter as offset errors increase with distance from the magnet isocenter [59].”
BackgroundCardiovascular magnetic resonance phase contrast imaging
Nayak, K. S., Nielsen, J.-F., Bernstein, M. A., Markl, M., D. Gatehouse, P., M. Botnar, R., Saloner, D., Lorenz, C., Wen, H., S. Hu, B., et al. (2015) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-015-0172-7 · Cited by 264
“Its consequences and corrections are reported in [88–90].”
BackgroundSelf-citationCardiovascular magnetic resonance artefacts
Ferreira, P. F., Gatehouse, P. D., Mohiaddin, R. H., & Firmin, D. N. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-41 · Cited by 191
“generally available on scanners, but is available in post image-processing s sensitivity range (VENC)) can significantly affect measurements of shunt flow and cardiac output, and that these may occur easily on some systems [78].”
BackgroundSelf-citationAssessment of mitral valve regurgitation by cardiovascular magnetic resonance imaging
Garg, P., Swift, A. J., Zhong, L., Carlhäll, C.-J., Ebbers, T., Westenberg, J., Hope, M. D., Bucciarelli-Ducci, C., Bax, J. J., & Myerson, S. G. (2020) · Nature Reviews Cardiology · doi:10.1038/s41569-019-0305-z · Cited by 183
Heart valve disease: investigation by cardiovascular magnetic resonance
Myerson, S. G. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-7 · Cited by 171
Quantitative magnetic resonance imaging phantoms: A review and the need for a system phantom
Keenan, K. E., Ainslie, M., Barker, A. J., Boss, M. A., Cecil, K. M., Charles, C., Chenevert, T. L., Clarke, L., Evelhoch, J. L., Finn, P., et al. (2018) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26982 · Cited by 170
Penicka, M., Vecera, J., Mirica, D. C., Kotrc, M., Kockova, R., & Van Camp, G. (2018) · Circulation · doi:10.1161/circulationaha.117.029332 · Cited by 166
Clough, R. E., Waltham, M., Giese, D., Taylor, P. R., & Schaeffter, T. (2012) · Journal of Vascular Surgery · doi:10.1016/j.jvs.2011.11.005 · Cited by 159
Noninvasive Estimation of PA Pressure, Flow, and Resistance With CMR Imaging
Swift, A. J., Rajaram, S., Hurdman, J. A., Hill, C., Davies, C., Sproson, T., Morton, A. C., Capener, D. A., Elliot, C. A., Condliffe, R. A., et al. (2013) · JACC. Cardiovascular imaging · doi:10.1016/j.jcmg.2013.01.013 · Cited by 167
“Our study derives a prediction rule for mPAP based on physiological measurements, an estimate of PCWP based on LA volume and phase contrast CMR CO (39,40) to derive PVR.”
MethodAdvanced flow MRI: emerging techniques and applications
Markl, M., Schnell, S., Wu, C., Bollache, É., Jarvis, K., Barker, A. J., Robinson, J. D., & Rigsby, C. K. (2016) · Clinical Radiology · doi:10.1016/j.crad.2016.01.011 · Cited by 173
Barker, A. J., Roldán‐Alzate, A., Entezari, P., Shah, S. J., Chesler, N. C., Wieben, O. E., Markl, M., & François, C. J. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25326 · Cited by 152
Swift, A. J., Rajaram, S., Condliffe, R. A., Capener, D. A., Hurdman, J. A., Elliot, C. A., Wild, J. M., & Kiely, D. G. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-40 · Cited by 149
The effect of supine positioning on maternal hemodynamics during late pregnancy
Humphries, A., Mirjalili, S. A., Tarr, G. P., Thompson, J. M. D., & Stone, P. (2019) · The Journal of Maternal-Fetal & Neonatal Medicine · doi:10.1080/14767058.2018.1478958 · Cited by 133
“flow measurements [35,36], a small effect in compari-”
BackgroundMultimodality Imaging of the Tricuspid Valve and Right Heart Anatomy
Khalique, O. K., Cavalcante, J. L., Shah, D. J., Guta, A. C., Zhan, Y., Piazza, N., & Muraru, D. (2019) · JACC. Cardiovascular imaging · doi:10.1016/j.jcmg.2019.01.006 · Cited by 127
Nordmeyer, S., Riesenkampff, E., Crelier, G. R., Khasheei, A., Schnackenburg, B., Berger, F. F., & Kühne, T. (2010) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22280 · Cited by 122
“Ideally, this should be performed in a larger group of patients and in a multicenter approach as performed by Gatehouse et al (22).”
BackgroundRecommendations towards standards for quantitative MRI (qMRI) and outstanding needs
Keenan, K. E., Biller, J. R., Delfino, J. G., Boss, M. A., Does, M. D., Evelhoch, J. L., Griswold, M. A., Gunter, J. L., Hinks, R. S., Hoffman, S. W., et al. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26598 · Cited by 137
Multi-modality image-based computational analysis of haemodynamics in aortic dissection
Dillon-Murphy, D., Noorani, A., Nordsletten, D., & Figueroa, C. A. (2015) · Biomechanics and Modeling in Mechanobiology · doi:10.1007/s10237-015-0729-2 · Cited by 120
“These methods, such as the Fictitious domain method (Glowinski et al. 1994), the mortar element method (Baai-123 jens 2001) and the immersed structural potential (ISPM) (Gil et al. 2010) allow for the imposition of the necessary kinematic (no-slip) and dynamic (balance of forces) compatibility conditions without resor…”
MethodLindstrøm, E. K., Ringstad, G. A., Mardal, K., & Eide, P. K. (2018) · NeuroImage Clinical · doi:10.1016/j.nicl.2018.09.006 · Cited by 121
Carlsson, M., Andersson, R., Bloch, K. M., Steding‐Ehrenborg, K., Mosén, H., Ståhlberg, F., Ekmehag, B., & Arheden, H. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-51 · Cited by 118
“Athletes had higher stroke volume index quantified on PC-CMR (66 ± 7 ml/m 2 for males and 60 ± 9 ml/m 2 for females) compared to age and gender matched healthy subjects (54 ± 6 ml/m 2 for males and 50 ± 7 ml/m 2 for females, p 0.001 for both).”
InfluentialCarlsson, M., Töger, J., Kanski, M., Bloch, K. M., Ståhlberg, F., Heiberg, E., & Arheden, H. (2011) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-13-55 · Cited by 105
“The accuracy is high when using a non-segmented approach [3,4] but several acquisitions are often required when assessing a patient and the newer segmented breath-hold sequences have showed lower accuracy [5,6].”
Hsiao, A., Lustig, M. S., Alley, M. T., Murphy, M., Chan, F. P., Herfkens, R. J., & Vasanawala, S. S. (2012) · American Journal of Roentgenology · doi:10.2214/ajr.11.6969 · Cited by 102
“A recent multi-institutional study concluded that systems from three major manufacturers have significant phase-offset errors requiring post-acquisition correction to achieve reliable flow measurements [31].”
BackgroundDiagnostic and Prognostic Utility of Cardiac Magnetic Resonance Imaging in Aortic Regurgitation
Kammerlander, A. A., Wiesinger, M., Duca, F., Aschauer, S., Binder, C., Zotter Tufaro, C., Nitsche, C., Badre-Eslam, R., Schönbauer, R., Bartko, P., et al. (2019) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2018.08.036 · Cited by 96
“For correct flow quantification, background correction is required to account for velocity offset (30).”
BackgroundAssessment of flow instabilities in the healthy aorta using flow‐sensitive MRI
Stalder, A. F., Frydrychowicz, A. P., Russe, M. F., Korvink, J. G., Hennig, J., Li, K., & Markl, M. (2011) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22512 · Cited by 119
Tariq, U., Hsiao, A., Alley, M. T., Zhang, T., Lustig, M. S., & Vasanawala, S. S. (2012) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.23936 · Cited by 95
“Several factors can influence the accuracy of flow quantification, including the presence of complex flow (6) and eddy-current-related phase offsets (7).”
BackgroundNormal and Altered Three-dimensional Portal Venous Hemodynamics in Patients with Liver Cirrhosis
Stanković, Z. B., Csatari, Z., Deibert, P., Euringer, W., Blanke, P., Kreisel, W., Zadeh, Z. A., Kallfass, F., Langer, M., & Markl, M. (2012) · Radiology · doi:10.1148/radiol.11110127 · Cited by 91
Bollache, É., Redheuil, A. B., Clément-Guinaudeau, S., Defrance, C., Perdrix, L., Ladouceur, M., Lefort, M., Cesare, A. D., Herment, A., Diébold, B., et al. (2010) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-12-63 · Cited by 85
“Alternatively, these errors can be corrected using techniques presented in previous studies [33-35].”
MethodThe right ventricle and pulmonary hypertension
van de Veerdonk, M. C., Bogaard, H. J., & Voelkel, N. F. (2016) · Heart Failure Reviews · doi:10.1007/s10741-016-9526-y · Cited by 103
Quantitative Assessment of Mitral Regurgitation
Thavendiranathan, P., Phelan, D. M. J., Thomas, J. D., Flamm, S. D., & Marwick, T. H. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2012.05.048 · Cited by 80
4D Flow MRI Applications for Aortic Disease
Burris, N. S., & Hope, M. D. (2014) · Magnetic Resonance Imaging Clinics of North America · doi:10.1016/j.mric.2014.08.006 · Cited by 88
CT and MR Imaging of the Aortic Valve: Radiologic-Pathologic Correlation
Bennett, C. J., Maleszewski, J. J., & Araoz, P. A. (2012) · Radiographics · doi:10.1148/rg.325115727 · Cited by 82
Bollache, É., van Ooij, P., Powell, A., Carr, J., Markl, M., & Barker, A. J. (2016) · International journal of cardiac imaging · doi:10.1007/s10554-016-0938-5 · Cited by 73
Schwitter, J., Gold, M. R., Fagih, A. R. A., Lee, S., PETERSON, M. W., Ciuffo, A., Zhang, Y., Kristiansen, N., Kanal, E., & Sommer, T. (2016) · Circulation Cardiovascular Imaging · doi:10.1161/circimaging.115.004025 · Cited by 71
Self-citationBateman, G. A., & Brown, K. M. (2011) · Child s Nervous System · doi:10.1007/s00381-011-1617-4 · Cited by 73
“The most important consideration is to utilize background correction to remove the effect of phase offset errors and movement occurring during the cardiac cycle [13].”
BackgroundEide, P. K., Valnes, L. M., Lindstrøm, E. K., Mardal, K., & Ringstad, G. A. (2021) · Fluids and Barriers of the CNS · doi:10.1186/s12987-021-00251-6 · Cited by 78
“Moreover, we corrected for background velocity offset, which may be machine-dependent [32].”
BackgroundHsiao, A., Tariq, U., Alley, M. T., Lustig, M. S., & Vasanawala, S. S. (2014) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24578 · Cited by 67
“Several groups have now called upon MRI vendors to better optimize acquisition strategies and equipment to reduce the influence of several factors known to impair the accuracy of 2D-PC (28,29).”
BackgroundGiese, D., Haeberlin, M., Barmet, C., Pruessmann, K. P., Schaeffter, T., & Kozerke, S. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.23111 · Cited by 64
“Accordingly, the measurement of Dueð r !pÞ in N spatial positions allows solving the linear equation system (5) for Dkl.”
BackgroundMethodInfluentialReal‐time flow MRI of the aorta at a resolution of 40 msec
Joseph, A. A., Kowallick, J. T., Merboldt, K., Voit, D., Schaetz, S., Zhang, S., Sohns, J. M. S., Lotz, J., & Frahm, J. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24328 · Cited by 62
“These values were well below the recommended limit of 5% of the VENC value (25).”
BackgroundMathew, R. C., Loffler, A. I., & Salerno, M. (2018) · Current Cardiology Reports · doi:10.1007/s11886-018-1057-9 · Cited by 65
Rolf, M. P., Hofman, M. B. M., Gatehouse, P. D., Bloch, K. M., Heymans, M. W., Ebbers, T., Graves, M. J., Totman, J. J., Werner, B., van Rossum, A. C., et al. (2011) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-13-18 · Cited by 59
“5% error in an aortic regurgitation fraction [8].”
BackgroundResultInfluentialSelf-citationBock, J., Töger, J., Bidhult, S. L., Bloch, K. M., Arvidsson, P. M., Kanski, M., Arheden, H., Testud, F., Greiser, A., Heiberg, E., et al. (2018) · Acta Radiologica · doi:10.1177/0284185118784981 · Cited by 57
“A previous multi-center, multi-vendor 2D-flow validation showed that velocity offset varied between sites having the same vendor and sequence (18).”
BackgroundCardiovascular Magnetic Resonance Imaging for Structural and Valvular Heart Disease Interventions
Cavalcante, J. L., Lalude, O. O., Schoenhagen, P., & Lerakis, S. (2016) · JACC: Cardiovascular Interventions · doi:10.1016/j.jcin.2015.11.031 · Cited by 65
“Although commercially available software allows for the correction of these phase-offset errors, more studies are necessary to validate the accuracy of its algorithms (43).”
Reconstruction of divergence‐free velocity fields from cine 3D phase‐contrast flow measurements
Busch, J., Giese, D., Wissmann, L., & Kozerke, S. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24221 · Cited by 56
Improved cardiovascular flow quantification with time-resolved volumetric phase-contrast MRI
Hsiao, A., Alley, M. T., Massaband, P., Herfkens, R. J., Chan, F. P., & Vasanawala, S. S. (2011) · Pediatric Radiology · doi:10.1007/s00247-010-1932-z · Cited by 54
“Furthermore, in clinical practice, the results of 2-D PC-MRI are often inconsistent, even with correction techniques to address eddy-current-related phase-offsets [ 6 ].”
ResultBusch, J., Vannesjo, S. J., Barmet, C., Pruessmann, K. P., & Kozerke, S. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-014-0097-6 · Cited by 54
“The study also showed that for such an accuracy to be achieved post-processing is required on all systems [16].”
BackgroundGiese, D., Wong, J., Greil, G. F., Buehrer, M., Schaeffter, T., & Kozerke, S. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-16-42 · Cited by 54
“Flow rates are known to be very sensitive to global phase offsets [31].”
BackgroundWestenberg, J. J. M., Roos, A., Grotenhuis, H. B., Paul Steendijk, A., Hendriksen, D., van den Boogaard, P. J., van der Geest, R. J., Bax, J. J., Jukema, J. W., & Reiber, J. H. C. (2010) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22359 · Cited by 56
“No correction for phase offset errors was made (31,32).”
BackgroundImage‐based background phase error correction in 4D flow MRI revisited
Busch, J., Giese, D., & Kozerke, S. (2017) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25668 · Cited by 53
Cardiovascular magnetic resonance in the evaluation of heart valve disease
Gulsin, G. S., Singh, A., & McCann, G. P. (2017) · BMC Medical Imaging · doi:10.1186/s12880-017-0238-0 · Cited by 57
“Even small velocity offset errors can lead to sizeable flow quantification errors, because flow volume is calculated by integrating velocities across the crosssectional area of the vessel over the entire cardiac cycle [9].”
BackgroundLin, H., Bender, J. A., Ding, Y., Chung, Y., Hinton, A. M., Pennell, M. L., Whitehead, K. K., Raman, S. V., & Simonetti, O. P. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.23273 · Cited by 52
Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance
Holland, B. J., Printz, B. F., & Lai, W. W. (2010) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-12-11 · Cited by 51
“A multicenter investigation of background phase offset errors using gelatin phantoms has highlighted the potential effect of phase offset errors in CMR on multiple scanning platforms [6].”
BackgroundNeuhaus, E., Weiss, K., Bastkowski, R., Koopmann, J., Maintz, D. C., & Giese, D. (2019) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-019-0573-0 · Cited by 51
Frydrychowicz, A. P., Wieben, O. E., Niespodzany, E., Reeder, S. B., Johnson, K. M., & François, C. J. (2013) · Investigative Radiology · doi:10.1097/rli.0b013e31829a4f2f · Cited by 47
“Phantom correction provides certainty with respect to unwanted phase offset contribution.(21) Although a 3-directionally velocity-encoded sequence is available, the vendor does not offer a 3-dimensional phase-contrast product sequence.”
BackgroundMethodInfluentialDefrance, C., Bollache, É., Kachenoura, N., Perdrix, L., Hrynchyshyn, N., Bruguière, É., Redheuil, A. B., Diébold, B., & Mousseaux, É. (2012) · Circulation Cardiovascular Imaging · doi:10.1161/circimaging.111.971218 · Cited by 59
Imaging Methods for Evaluation of Chronic Aortic Regurgitation in Adults
Ranard, L. S., Bonow, R. O., Nishimura, R. A., Mack, M. J., Thourani, V. H., Bavaria, J. E., O’Gara, P. T., Bax, J. J., Blanke, P., Delgado, V., et al. (2023) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2023.08.051 · Cited by 51
Giese, D., Schaeffter, T., & Kozerke, S. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24273 · Cited by 47
“Accordingly, errors in cardiac output remained below 5% and hence are clinically acceptable (27).”
BackgroundAdjudicating paravalvular leaks of transcatheter aortic valves: a critical appraisal
Abdelghani, M., Soliman, O. I. I., Schultz, C. J., Vahanian, A. S., & Serruys, P. W. J. C. (2016) · European Heart Journal · doi:10.1093/eurheartj/ehw115 · Cited by 52
“is generated with the systolic flow represented by area under the forward part of the curve, and area above the backward diastolic part, when present, indicates the retrograde-regurgitant flow volume (Figure 10).(112)”
Varga‐Szemes, Á., Muscogiuri, G., Schoepf, U. J., Wichmann, J. L., Suranyi, P. S., Cecco, C. N. D., Cannaò, P. M., Renker, M., Mangold, S., Fox, M. A., et al. (2015) · European Radiology · doi:10.1007/s00330-015-3952-4 · Cited by 46
Reconciling PC‐MRI and CFD: An in‐vitro study
Puiseux, T., Sewonu, A., Meyrignac, O., Rousseau, H., Nicoud, F., Mendez, S., & Moreno, R. (2019) · NMR in Biomedicine · doi:10.1002/nbm.4063 · Cited by 46
Töger, J., Bidhult, S. L., Revstedt, J., Carlsson, M., Arheden, H., & Heiberg, E. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25683 · Cited by 43
Hădăreanu, D.-R., Muraru, D., Volpato, V., Gavazzoni, M., Caravita, S., Tomaselli, M., Ciampi, P., Florescu, C., Bălşeanu, T. A., Parati, G., et al. (2022) · Journal of Clinical Medicine · doi:10.3390/jcm11020382 · Cited by 43
“Moreover, CMR can provide accurate TR severity grading based on the measurement of the RegVol and the regurgitant fraction, indirectly calculated by subtracting the pulmonic forward volume from the RV stroke volume [68], or directly measured by the use of either standard phase-contrast sequences [89,90], or by innovat…”
MethodCalkoen, E. E., Roest, A. A. W., Kroft, L. J. M., van der Geest, R. J., Jongbloed, M. R. M., van den Boogaard, P. J., Blom, N. A., Hazekamp, M. G., Roos, A., & Westenberg, J. J. M. (2014) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24735 · Cited by 40
“Local phase offset errors, which may lead to small systematic errors, should be taken into account (18).”
BackgroundComputational fluid dynamics in Fontan patients to evaluate power loss during simulated exercise
Bossers, S. S. M., Cibiş, M., Gijsen, F. J. H., Schokking, M., Strengers, J. L. M., Verhaart, R. F., Moelker, A. D., Wentzel, J. J., & Helbing, W. A. (2014) · Heart · doi:10.1136/heartjnl-2013-304969 · Cited by 40
Muzzarelli, S., Monney, P. A., O’Brien, K., Faletra, F. F., Moccetti, T., Vogt, P. R., & Schwitter, J. (2013) · European Heart Journal - Cardiovascular Imaging · doi:10.1093/ehjci/jet129 · Cited by 39
“The most relevant include: (i) complex flow patterns with consequent intravoxel dephasing and underestimation in velocity measurement,(7,13) (ii) longitudinal excursion of the aortic annulus with a consequent systolic change in the aortic luminal volume between the fixed plane of velocity acquisition and the moving pl…”
BackgroundMethodSelf-citationMontalba, C., Jesús Urbina, R., Sotelo, J. A., Andía, M. E., Cristián Tejos, N., Pablo Irarrázaval, M., Hurtado, D. E., Valverde, I., & Uribe, S. A. (2017) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26834 · Cited by 38
“Phantoms have been used previously to assess the accuracy of 2D flow (38,39).”
MethodOechtering, T. H., Nowak, A., Sieren, M. M., Stroth, A. M., Kirschke, N., Wegner, F., Balks, M. F., König, I. R., Jin, N., Graessner, J., et al. (2023) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-023-00921-4 · Cited by 37
“For 2D PC measurements, the velocity offset between measurements with the same scanner and sequence at different sites is well acknowledged [24].”
BackgroundGatehouse, P. D., Rolf, M. P., Bloch, K. M., Graves, M. J., Kilner, P. J., Firmin, D. N., & Hofman, M. B. M. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-72 · Cited by 37
“Each centre acquired the same single plane at 45° oblique between transverse and sagittal planes, with anterior-posterior phase-encoding, at isocentre (the aortic plane shown in reference [11]).”
BackgroundMethodInfluentialSelf-citationDrake, W. M., Stiles, C. E., Bevan, J. S., Karavitaki, N., Trainer, P. J., Rees, A., Richardson, T., Baldeweg, S. E., Stojanovic, N. D., Murray, R. D., et al. (2016) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2016-2224 · Cited by 36
“It is not clear whether newer imaging modalities such as cardiac magnetic resonance imaging (MRI) will provide more accurate or reproducible assessment of mild degrees of regurgitation (29).”
BackgroundAn Echocardiographic Model Predicting Severity of Aortic Regurgitation in Congenital Heart Disease
Beroukhim, R. S., Graham, D. A., Margossian, R. E., Brown, D. W., Geva, T., & Colan, S. D. (2010) · Circulation Cardiovascular Imaging · doi:10.1161/circimaging.110.957175 · Cited by 35
Dandekar, V. K., Bauml, M. A., Ertel, A. W., Dickens, C., Gonzalez, R. C., & Farzaneh‐Far, A. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-16-24 · Cited by 35
Sieren, M. M., Berlin, C., Oechtering, T. H., Hunold, P., Drömann, D., Barkhausen, J., & Frydrychowicz, A. P. (2019) · PLoS ONE · doi:10.1371/journal.pone.0224121 · Cited by 35
Geiger, J., Callaghan, F. M., Burkhardt, B. E. U., Buechel, E. R. V., & Kellenberger, C. J. (2020) · Pediatric Radiology · doi:10.1007/s00247-020-04885-w · Cited by 38
“For flow quantification, background phase correction is necessary [42]; standard automated/semi-automated procedures facilitate the static tissue identification; however, in small children accuracy of delineation should be verified.”
MethodPhase contrast MR imaging measurements of blood flow in healthy human cerebral vessel segments
MacDonald, M. E., & Frayne, R. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/7/1517 · Cited by 32
Unravelling cardiovascular disease using four dimensional flow cardiovascular magnetic resonance
Kamphuis, V. P., Westenberg, J. J. M., van der Palen, R. L. F., Blom, N. A., Roos, A., van der Geest, R. J., Elbaz, M. S. M., & Roest, A. A. W. (2016) · International journal of cardiac imaging · doi:10.1007/s10554-016-1031-9 · Cited by 33
“Background phase offset errors are usually corrected by either performing a phantom velocity-encoded scan simulating static tissue and using this data set as a reference for background subtraction, or by the approach of fitting a multi-order polynomial through areas identified as static tissue, for correcting the loca…”
Demir, A., Wiesemann, S., Erley, J., Schmitter, S., Trauzeddel, R. F., Pieske, B. M., Hansmann, J., Kelle, S., & Schulz‐Menger, J. (2021) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.27804 · Cited by 32
“Plane positions that are more distant from the isocenter of the magnet rather lead to inaccurate flow results mainly caused by eddy currents yielding phase offsets.(29) Therefore, background correction has been applied by a linear fit to the static tissue.”
BackgroundSelf-citationLee, K., Yoo, S., Holtby, H. M., Grant, B., Mroczek, D., Wong, D., Grosse‐Wortmann, L., Benson, L. N., & Chaturvedi, R. R. (2011) · Heart · doi:10.1136/hrt.2011.225656 · Cited by 31
Vincenti, G. M., Masci, P. G., Rutz, T., Blois, J. D., Prša, M., Jeanrenaud, X., Schwitter, J., & Monney, P. A. (2016) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-017-0362-6 · Cited by 32
“By accurately measuring volumes and flows [9], CMR is an attractive method to quantify mitral regurgitant volume (RegVol) and regurgitant fraction (RF), two parameters that may be able to identify patients suitable for early surgery [10].”
MethodSelf-citationSeemann, F., Baldassarre, L. A., Llanos‐Chea, F., Gonzales, R. A., Grunseich, K., Hu, C., Sugeng, L., Meadows, J. L., Heiberg, E., & Peters, D. C. (2018) · Physiological Reports · doi:10.14814/phy2.13828 · Cited by 30
“Phase offset errors were corrected based on estimation in stationary tissue (Gatehouse et al. 2010).”
MethodChaturvedi, A., Hamilton-Craig, C. R., Cawley, P. J., Mitsumori, L. M., Otto, C. M., & Maki, J. H. (2015) · European Radiology · doi:10.1007/s00330-015-4120-6 · Cited by 30
Gorter, T. M., van Melle, J. P., Freling, H. G., Ebels, T., Bartelds, B., Pieper, P. G., Berger, R. M. F., van Veldhuisen, D. J., & Willems, T. P. (2015) · International journal of cardiac imaging · doi:10.1007/s10554-015-0670-6 · Cited by 30
Phase-contrast MRI volume flow – a comparison of breath held and navigator based acquisitions
Andersson, C., Kihlberg, J., Ebbers, T., Lindström, L., Carlhäll, C., & Engvall, J. E. (2016) · BMC Medical Imaging · doi:10.1186/s12880-016-0128-x · Cited by 29
“Gatehouse suggested that an error of 5 % could be acceptable in clinical practice, which would be equivalent to 4 ml when the stroke volume is 80 ml and 250 ml when cardiac output is 5000 ml/min [7].”
Becker, A. S., Boss, A., Klarhoefer, M., Finkenstaedt, T., Wurnig, M. C., & Rossi, C. (2017) · NeuroImage · doi:10.1016/j.neuroimage.2017.12.017 · Cited by 26
“Inaccuracy in the pre-emphasis of the applied gradient pulses (i.e., eddy current errors) and concomitant gradients with non-linear spatial dependence result in velocity offsets (Gatehouse et al., 2010), which lead to inaccuracy rates of up to 10% over the calculated flow (Barkhof et al., 1994).”
James, S. H., Wald, R. M., Wintersperger, B. J., Jiménez‐Juan, L., Deva, D. P., Crean, A. M., Nguyen, E. T., Paul, N., & Ley, S. (2013) · Canadian Association of Radiologists Journal · doi:10.1016/j.carj.2011.12.016 · Cited by 28
Simpson, R., Keegan, J., & Firmin, D. N. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-34 · Cited by 28
“The velocity maps from these phantom acquisitions were smoothed using a median filtering technique to reduce noise and then subtracted from the volunteer data on a pixel-by-pixel basis [30].”
MethodSelf-citationSkaug, T. R., Hergum, T., Amundsen, B. H., Skjærpe, T., Torp, H. G., & Haugen, B. O. (2009) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2009.10.005 · Cited by 28
Bracamonte, J., Saunders, S. K., Wilson, J. S., Truong, U. T., & Soares, J. S. (2022) · Applied Sciences · doi:10.3390/app12083954 · Cited by 35
de Beaufort, H. W., Shah, D. J., Patel, A. P., Jackson, M., Spinelli, D., Yang, E. Y., Ghosn, M. G., Autry, K., Igo, S. R., Lumsden, A. B., et al. (2018) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2018.06.022 · Cited by 28
“A further limitation is that eddy current corrections are not part of the analysis in the software that was used for 4D-flow dataset postprocessing, even though eddy current corrections can lead to differences in flow volume of up to 10 mL/beat.(21) Finally, the lack of follow-up data and small sample size limit the p…”
MethodBidhult, S. L., Hedström, E., Carlsson, M., Töger, J., Steding‐Ehrenborg, K., Arheden, H., Aletras, A. H., & Heiberg, E. (2019) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12582 · Cited by 27
“Comparison with earlier studies and future work The obtained limits of agreement for flow measurements in both the aorta and pulmonary artery were small in relation to reported flow volume errors caused by potential background velocity offset errors (Gatehouse et al., 2010).”
ResultScenario Adaptive Cuffless Blood Pressure Estimation by Integrating Cardiovascular Coupling Effects
Qiu, S., Zhang, Y., Lau, S.-K., & Zhao, N. (2022) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2022.3227235 · Cited by 28
“(a) Measurement of aortic blood flow by cardiovascular magnetic resonance (CMR) [20]; (b) Model of the blood flow velocity based on (2); (c) Recording of the ECG and PPG signals and feature extraction based on the proposed PF method.”
MethodHemodynamic force analysis is not ready for clinical trials on HFpEF
Arvidsson, P. M., Nelsson, A., Magnusson, M., Smith, J. G., Carlsson, M., & Arheden, H. (2022) · Scientific Reports · doi:10.1038/s41598-022-08023-4 · Cited by 28
Espe, E. K. S., Aronsen, J. M., Skrbic, B., Skulberg, V. M., Schneider, J. E., Sejersted, O. M., Zhang, L., & Sjaastad, I. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24226 · Cited by 23
Oscanoa, J., Middione, M. J., Syed, A., Sandino, C. M., Vasanawala, S. S., & Ennis, D. B. (2022) · Magnetic Resonance in Medicine · doi:10.1002/mrm.29441 · Cited by 22
“6cm/s as proposed by Gatehouse et al. 42,43 Accuracy is consistent across all reconstruction techniques while CD-DL shows higher precision compared to PICS. of pixel background phase offsets (cm/s) from differences within (Figure 2A) static background tissue and (Figure 2B) the vessel ROIs for PICS…”
MethodAshrafpoor, G., Bollache, É., Redheuil, A. B., Cesare, A. D., Giron, A., Defrance, C., Azarine, A., Perdrix, L., Ladouceur, M., Diébold, B., et al. (2014) · European Radiology · doi:10.1007/s00330-014-3488-z · Cited by 31
“Another limitation, which is inherent in MRI acquisition, included potential phase offset errors [40], which were not corrected in the present study, although they were minimized through the use of a 50 % rectangular field of view centered on the mitral annulus.”
MethodHassan, M. S., Wagdy, K., Kharabish, A. E., Selwanos, P., Nabil, A., ElGuindy, A. M., ElFaramawy, A., Elmahdy, M. F., Mahmoud-Elsayed, H. M., & Yacoub, M. H. (2017) · Circulation Heart Failure · doi:10.1161/circheartfailure.116.003592 · Cited by 22
Zoghbi, W. A., Adams, D., Bonow, R. O., Enriquez-Sarano, M. L., Foster, E., Grayburn, P. A., Hahn, R. T., Han, Y., Hung, J., Lang, R. M., et al. (2020) · Journal of The Indian Academy of Echocardiography & Cardiovascular Imaging · doi:10.4103/2543-1463.282191 · Cited by 36
“In addition, background phase, partial volume average from low spatial resolution and nonperpendicular plane selection, and complex jet flow pattern causing intravoxel dephasing and loss of phase coherence can all lead to inaccurate phase-contrast assessment.(82) Indirect quantification of TR can amplify errors and le…”
BackgroundThe Role of Cardiac Magnetic Resonance in Aortic Stenosis and Regurgitation
Guglielmo, M., Rovera, C., Rabbat, M., & Pontone, G. (2022) · Journal of Cardiovascular Development and Disease · doi:10.3390/jcdd9040108 · Cited by 25
“Another limitation of flow analysis is the existence of positive or negative phase offset errors, due to local turbulent currents [15].”
Background
Metadata
- Authors
- Gatehouse, P. D., Rolf, M. P., Graves, M. J., Hofman, M. B., Totman, J., Werner, B., Quest, R. A., Liu, Y., von Spiczak, J., Dieringer, M., Firmin, D. N., van Rossum, A., Lombardi, M., Schwitter, J., Schulz-Menger, J., & Kilner, P. J.
- Affiliations
- Royal Brompton Hospital (GB)
- Amsterdam UMC Location Vrije Universiteit Amsterdam (NL)
- University of Cambridge (GB)
- King's College London (GB)
- University Children's Hospital Zurich (CH)
- Hammersmith Hospital (GB)
- University of Auckland (NZ)
- Charité – Universitätsmedizin Berlin (DE)
- National Research Council (IT)
- University Hospital Zurich (CH)
- Corresponding author
- Kilner, Philip J
- ORCID
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2010-10
- Volume
- 12
- Issue
- 1
- Pages
- 5
- ISSN
- 1097-6647 (ISSN-L)
- PMC
- PMC2818657
- OpenAlex
- W2153286609
- Cited by
- 234
- References
- 15
- Citations by year
- 2026: 4 · 2025: 5 · 2024: 8 · 2023: 11 · 2022: 11 · 2021: 13 · 2020: 18 · 2019: 17 · 2018: 12 · 2017: 13 · 2016: 17 · 2015: 15 · 2014: 16 · 2013: 20 · 2012: 24
- FWCI
- 18.70
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- www.elsevier.com/legal/tdmrep-license (tdm)
- creativecommons.org/licenses/by/2/ (vor)
- Locations
- Journal of Cardiovascular Magnetic Resonance (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Pure Amsterdam UMC (publishedVersion, other-oa)
- Amsterdam UMC (VU Amsterdam) - Institutional Repository (publishedVersion, other-oa)
- MDC Repository (Max-Delbrueck-Center for Molecular Medicine) (acceptedVersion)
- Bournemouth University Research Online (Bournemouth University) (acceptedVersion)
- Zurich Open Repository and Archive (University of Zurich) (acceptedVersion, cc-by)
- National University of Singapore (submittedVersion, cc-by)
- Digital Academic REpository of VU University Amsterdam (Vrije Universiteit Amsterdam) (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Spiral (Imperial College London) (submittedVersion, cc-by)
- Repository for Publications and Research Data (ETH Zurich) (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- ETH Zürich Research Collection (cc-by)
- Zurich Open Repository and Archive (University of Zurich)
- Funders
- British Heart Foundation
- Vrije Universiteit Amsterdam
- Eidgenössische Technische Hochschule Zürich
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac Valve Diseases and Treatments
- Keywords
- phase-contrast MRI, flow measurement accuracy, regurgitant flow, shunt flow, aortic blood flow, pulmonary blood flow, cardiac output, gelatin phantom, offset correction
- MeSH terms
- Aorta, Blood Flow Velocity, Cardiac Output, Gelatin, Humans, Image Interpretation, Computer-Assisted, Materials Testing, Mitral Valve Insufficiency, Predictive Value of Tests, Pulmonary Artery, Reproducibility of Results, Respiratory Mechanics, Artifacts, Magnetic Resonance Imaging, Cine, Phantoms, Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.