Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements
Open accessThe DOI matches its Crossref record; 15 of the 15 references it cites were found in a registry or an index.
Abstract
AIMS: Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error. We studied the severity of uncorrected velocity offset errors across sites and CMR systems. METHODS AND RESULTS: In a multi-centre, multi-vendor study, breath-hold through-plane retrospectively ECG-gated phase contrast acquisitions, as are used clinically for aortic and pulmonary flow measurement, were applied to static gelatin phantoms in twelve 1.5 T CMR systems, using a velocity encoding range of 150 cm/s. No post-processing corrections of offsets were implemented. The greatest uncorrected velocity offset, taken as an average over a 'great vessel' region (30 mm diameter) located up to 70 mm in-plane distance from the magnet isocenter, ranged from 0.4 cm/s to 4.9 cm/s. It averaged 2.7 cm/s over all the planes and systems. By theoretical calculation, a velocity offset error of 0.6 cm/s (representing just 0.4% of a 150 cm/s velocity encoding range) is barely acceptable, potentially causing about 5% miscalculation of cardiac output and up to 10% error in shunt measurement. CONCLUSION: In the absence of hardware or software upgrades able to reduce phase offset errors, all the systems tested appeared to require post-acquisition correction to achieve consistently reliable breath-hold measurements of flow. The effectiveness of offset correction software will still need testing with respect to clinical flow acquisitions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Cine or PC sequences can be used to quantify the stenosis, blockage, or regurgitation of these conduits.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, the quantification may be incorrect due to conduit artefacts [37,44].”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Subsequent CMR studies are usually performed if new clinical symptoms arise or every 3–5 years, depending on the previous clinical findings [34,36,37].”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“In paediatric centres with high patient volume, the first CMR is typically conducted around 8–9 years of age, when children can undergo the examination without anaesthesia.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Usually valved, these conduits—which may be synthetic or biological—can develop pulmonary regurgitation later on or function as a tiny conduit as people age.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Handbook of MRI Pulse Sequences
(2004) · Elsevier eBooks · doi:10.1016/b978-0-12-092861-3.x5000-6 · Cited by 1,610
Blood Flow Imaging by Cine Magnetic Resonance
Nayler, G. L., Firmin, D. N., & Longmore, D. B. (1986) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198609000-00001 · Cited by 634 · cited 1×
“Phase contrast cardiovascular magnetic resonance (CMR) [1] measurements of flow through planes transecting the great arteries are used clinically for calculations of cardiac output, shunt flow [2,3] or aortic or pulmonary regurgitation [4,5].”
Self-citationConcomitant gradient terms in phase contrast MR: Analysis and correction
Bernstein, M. A., Zhou, X. J., Polzin, J. A., King, K. F., Ganin, A., Pelc, N. J., & Glover, G. H. (1998) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910390218 · Cited by 547 · cited 2×
“Velocity offsets caused by Maxwell (concomitant) gradients in phase-contrast imaging are calculable allowing automatic correction during reconstruction [11].”
Rebergen, S. A., Chin, J. G., Ottenkamp, J., van der Wall, E. E., & de Roos, A. (1993) · Circulation · doi:10.1161/01.cir.88.5.2257 · Cited by 294
Gelfand, E., Hughes, S., Hauser, T., Yeon, S., Goepfert, L., Kissinger, K., Rofsky, N., & Manning, W. (2006) · Journal of Cardiovascular Magnetic Resonance · doi:10.1080/10976640600604856 · Cited by 265 · cited 2×
“In combination with measurements of left ventricular volume or mitral inflow, measurement of aortic outflow may also allow the indirect calculation of mitral regurgitation [5-7].”
Hundley, W. G., Li, H. F., Willard, J. E., Landau, C., Lange, R. A., Meshack, B. M., Hillis, L. D., & Peshock, R. M. (1995) · Circulation · doi:10.1161/01.cir.92.5.1151 · Cited by 229
Normal Thoracic Aorta Diameter on Cardiac Computed Tomography in Healthy Asymptomatic Adults
Mao, S. S., Ahmadi, N., Shah, B., Beckmann, D., Chen, A., Ngo, L., Flores, F. R., Gao, Y. L., & Budoff, M. J. (2008) · Academic Radiology · doi:10.1016/j.acra.2008.02.001 · Cited by 217 · cited 1×
“The software reported the most extreme mean velocity offset measured in a circular ‘great vessel’ region, 30 mm diameter [13], centered within specified distances (see below) from the centre of each velocity map.”
Quantification of mitral regurgitation by velocity-encoded cine nuclear magnetic resonance imaging
Fujita, N., Chazouilleres, A. F., Hartiala, J. J., O'sullivan, M., Heidenreich, P., Kaplan, J. D., Sakuma, H., Foster, E., Caputo, G. R., & Higgins, C. B. (1994) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(94)90642-4 · Cited by 198 · cited 1×
“In combination with measurements of left ventricular volume or mitral inflow, measurement of aortic outflow may also allow the indirect calculation of mitral regurgitation [ 5-7 ].”
Chernobelsky, A., Shubayev, O., Comeau, C. R., & Wolff, S. D. (2007) · Journal of Cardiovascular Magnetic Resonance · doi:10.1080/10976640601187588 · Cited by 163 · cited 5×
“Opposite polarities from the same scanner have been reported previously [9].”
InfluentialCorrection of phase offset errors in main pulmonary artery flow quantification
Lankhaar, J., Hofman, M. B. M., Marcus, J. T., Zwanenburg, J. J. M., Faes, T. J. C., & Noordegraaf, A. V. (2005) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20361 · Cited by 101
Self-citation
Cited by
Items 101–200 of 234
Myocardial Blood Flow and Fibrosis in Hypertrophic Cardiomyopathy
Aquaro, G. D., Todiere, G., Barison, A., Strata, E., Marzilli, M., Pingitore, A., & Lombardi, M. (2011) · Journal of Cardiac Failure · doi:10.1016/j.cardfail.2011.01.006 · Cited by 21
“At the end of the examination, a phantom was placed in the MRI scanner and a velocity-encoded fast contrast gradient-echo cine sequence was acquired with the same acquisition plane and parameters used for the patient.(14)”
MethodSelf-citationNumerical simulation of time-resolved 3D phase-contrast magnetic resonance imaging
Puiseux, T., Sewonu, A., Moreno, R., Mendez, S., & Nicoud, F. (2021) · PLoS ONE · doi:10.1371/journal.pone.0248816 · Cited by 34
“: spatio-temporal resolution, encoding velocity, imaging artifacts) may limit the expected accuracy of the measurements and potentially lead to erroneous diagnosis [8, 9].”
BackgroundDeep Learning Automated Background Phase Error Correction for Abdominopelvic 4D Flow MRI
You, S., Masutani, E., Alley, M. T., Vasanawala, S. S., Taub, P. R., Liau, J., Roberts, A. C., & Hsiao, A. (2021) · Radiology · doi:10.1148/radiol.2021211270 · Cited by 28
Taylor, A. M., Dymarkowski, S., & Bogaert, J. G. (2000) · Medical radiology · doi:10.1007/174_2011_414 · Cited by 21
Spiral tissue phase velocity mapping in a breath‐hold with non‐cartesian SENSE
Simpson, R., Keegan, J., Gatehouse, P. D., Hansen, M. S., & Firmin, D. N. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24971 · Cited by 19
“correction based on static tissue has previously been used with TPVM (9) and would remove this inconvenience, although the accuracy of the correction is likely to be lower than with a phantom subtraction (23).”
MethodSelf-citationLin, H., Freed, D. H., Lee, T. W. R., Arora, R. C., Ali, A. A., Al-Moustadi, W., Xiang, B., Wang, F., Large, S., King, S. B., et al. (2011) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22587 · Cited by 21
“To address this issue, a multicenter, multivendor clinical trial of the accuracy of MRI flow measurements was recently conducted and reported (26).”
BackgroundShe, H. L., Roest, A. A. W., Calkoen, E. E., van den Boogaard, P. J., van der Geest, R. J., Hazekamp, M. G., Roos, A., & Westenberg, J. J. M. (2016) · Congenital Heart Disease · doi:10.1111/chd.12397 · Cited by 19
Cao, J. J., Wang, Y., McLaughlin, J., Rhee, P. D.-Y., Passick, M., Ngai, N., Cheng, J. Y., Gulotta, R. J., Berke, A. D., Petrossian, G., et al. (2012) · Circulation Cardiovascular Imaging · doi:10.1161/circimaging.112.973206 · Cited by 18
Bidhult, S. L., Töger, J., Heiberg, E., Carlsson, M., Arheden, H., Aletras, A. H., & Hedström, E. (2018) · Magnetic Resonance in Medicine · doi:10.1002/mrm.27392 · Cited by 18
Hofman, M. B. M., Rodenburg, M. J. A., Bloch, K. M., Werner, B., Westenberg, J. J. M., Buechel, E. R. V., Nijveldt, R., Spruijt, O. A., Kilner, P. J., van Rossum, A. C., et al. (2019) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-019-0538-3 · Cited by 18
“6 cm/s for reliable clinical CMR imaging of the volume flows, typically to obtain an error in cardiac output below 5% [4].”
BackgroundMethodInfluentialSelf-citationMinderhoud, S. C. S., van der Velde, N., Wentzel, J. J., van der Geest, R. J., Attrach, M., Wielopolski, P. A., Budde, R. P. J., Helbing, W. A., Roos‐Hesselink, J. W., & Hirsch, A. (2020) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-020-00659-3 · Cited by 22
“However, this drift is limited within a clinically acceptable range when scanning within the same imaging session on most systems [9].”
BackgroundResultInfluentialClinical Applications of MRA 4D-Flow
Sierra-Galán, L. M., & François, C. J. (2019) · Current Treatment Options in Cardiovascular Medicine · doi:10.1007/s11936-019-0758-8 · Cited by 18
Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up
Moscatelli, S., Pergola, V., Motta, R., Fortuni, F., Borrelli, N., Sabatino, J., Leo, I., Avesani, M., Montanaro, C., Surkova, E., et al. (2023) · Children · doi:10.3390/children10111747 · Cited by 17
Flow quantification dependency on background phase correction techniques in 4D‐flow MRI
Callaghan, F. M., Burkhardt, B. E. U., Geiger, J., Buechel, E. R. V., & Kellenberger, C. J. (2019) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28085 · Cited by 17
Visual Exploration of Simulated and Measured Blood Flow
Vilanova, A., Preim, B., van Pelt, R., Gasteiger, R., Neugebauer, M., & Wischgoll, T. (2014) · Mathematics and visualization · doi:10.1007/978-1-4471-6497-5_25 · Cited by 17
Thavendiranathan, P., Guetter, C., da Silveira, J. S., Lu, X., SCANDLING, D. M., Xue, H., Jolly, M., Raman, S. V., & Simonetti, O. P. (2018) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2018.08.018 · Cited by 17
“Also, concerns about phase offset errors which continue to plague PC-CMR acquisitions are avoided with this method (18,19).”
BackgroundMethodThe muscle twitch profile assessed with motor unit magnetic resonance imaging
Heskamp, L., Birkbeck, M. G., Whittaker, R. G., Schofield, I. S., & Blamire, A. M. (2021) · NMR in Biomedicine · doi:10.1002/nbm.4466 · Cited by 17
“For the PC images, the velocity contained an offset, which can be attributed to the presence of B0 eddy currents, concomitant fields and gradient nonlinearity.(16) The offset was removed in two steps (see the supporting material: Methods – Removing offset in phase contrast images and Figure S1 for more information).”
BackgroundCardiac imaging in valvular heart disease
Choo, W. S., & Steeds, R. P. (2011) · British Journal of Radiology · doi:10.1259/bjr/54030257 · Cited by 18
“relative to the flow map and due to the problem of background phase-offset errors [43].”
BackgroundRigsby, C. K., Hilpipre, N., McNeal, G. R., Zhang, G., Boylan, E. E., Popescu, A. R., Choi, G. H., Greiser, A., & Deng, J. (2013) · Pediatric Radiology · doi:10.1007/s00247-013-2830-y · Cited by 17
“The cause of background phase errors is predominantly related to eddy current-induced errors, and these effects are aggravated by excessive slice offsets from isocenter [1].”
BackgroundMethodInfluentialVolpato, V., Badano, L. P., Figliozzi, S., Florescu, D. R., Parati, G., & Muraru, D. (2021) · Current Opinion in Cardiology · doi:10.1097/hco.0000000000000890 · Cited by 18
“CMR have the advantage to directly calculate TR volume and regurgitant fraction using standard phase-contrast sequences [31,71,72] (Fig.”
MethodCapron, T., Cautela, J., Scemama, U., Miola, C., Bartoli, A., Théron, A., Pinto, J., Porto, A., Collart, F., Lépidi, H., et al. (2020) · Diagnostic and Interventional Imaging · doi:10.1016/j.diii.2020.04.014 · Cited by 16
Bolen, M. A., Setser, R. M., Gabriel, R. S., Renapurkar, R. D., Tandon, Y. K., Lieber, M. L., Desai, M. Y., & Flamm, S. D. (2012) · International journal of cardiac imaging · doi:10.1007/s10554-012-0047-z · Cited by 15
Chemical shift‐induced phase errors in phase‐contrast MRI
Middione, M. J., & Ennis, D. B. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24262 · Cited by 16
Craiem, D. O., Pascaner, A. F., Casciaro, M. E., Gencer, U., Alcibar, J., Soulat, G., & Mousseaux, É. (2019) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-019-00765-z · Cited by 15
Aquaro, G. D., Barison, A., Todiere, G., Festa, P., Ait-Alì, L., Lombardi, M., & Bella, G. D. (2015) · Journal of Cardiovascular Medicine · doi:10.2459/jcm.0000000000000245 · Cited by 14
Self-citationComparison of two accelerated 4D-flow sequences for aortic flow quantification
Ebel, S., Dufke, J., Köhler, B., Preim, B., Rosemeier, S., Jung, B. A., Dähnert, I., Lurz, P. C., Borger, M. A., Grothoff, M., et al. (2019) · Scientific Reports · doi:10.1038/s41598-019-45196-x · Cited by 16
Flow evaluation software for four-dimensional flow MRI: a reliability and validation study
Burkhardt, B. E. U., Kellenberger, C. J., Callaghan, F. M., Buechel, E. R. V., & Geiger, J. (2023) · La radiologia medica · doi:10.1007/s11547-023-01697-4 · Cited by 13
“Therefore, several previous studies have investigated the impact of velocity offset errors on flow measurements [14–18].”
BackgroundValverde, I., Parish, V., Tzifa, A., Head, C., Sarikouch, S., Greil, G. F., Schaeffter, T., Razavi, R., & Beerbaum, P. (2011) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22573 · Cited by 15
“We appreciate there are disadvantages of free-breathing scanning, namely propensity for arrhythmia affecting retrospective phase reordering, background phase offset errors (36) and susceptibility to irregular breathing patterns during stress (34).”
BackgroundFischer, C., Wetzl, J., Schaeffter, T., & Giese, D. (2022) · Magnetic Resonance in Medicine · doi:10.1002/mrm.29363 · Cited by 13
Development and validation of a magneto-hydrodynamic solver for blood flow analysis
Kainz, W., Guag, J. W., Benkler, S., Szczerba, D., Neufeld, E., Krauthamer, V., Myklebust, J. B., Bassen, H., Chang, I. A., Chavannes, N., et al. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/23/005 · Cited by 14
Skaug, T. R., Amundsen, B. H., Hergum, T., Urheim, S., Torp, H. G., & Haugen, B. O. (2013) · European Heart Journal - Cardiovascular Imaging · doi:10.1093/ehjci/jet255 · Cited by 14
Four-Dimensional Flow MR Imaging
Wieben, O. E., Roberts, G. S., Corrado, P. A., Johnson, K. M., & Roldán‐Alzate, A. (2023) · Magnetic Resonance Imaging Clinics of North America · doi:10.1016/j.mric.2023.05.003 · Cited by 12
Zvaigzne, C. G., Howarth, A. G., & Patton, D. J. (2014) · Cardiology in the Young · doi:10.1017/s1047951113002321 · Cited by 13
“Turbulence, partial volume effects, and phase-offset or background errors can contribute to inaccuracies,(55) some of which can be corrected with the use of a phantom, or by measuring the phase-offset error in a region of no flow near the area of interest, typically the chest wall.(56) Aortic and main pulmonary artery…”
BackgroundPostigo, A., Viola, F., Chazo, C., Martínez‐Legazpi, P., González‐Mansilla, A., Rodríguez‐González, E., Fernández‐Avilés, F., del Álamo, J. C., Ebbers, T., & Bermejo, J. (2022) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2022.05.007 · Cited by 12
“4D flow acquisition entails inherent noise-related uncertainties and background phase offset, depending not only on the acquisition settings but also on the CMR scanner (Gatehouse et al. 2010).”
BackgroundMeierhofer, C., Lyko, C., Schneider, E. P., Stern, H. C., Martinoff, S. A., Hess, J., & Fratz, S. (2014) · Clinical Imaging · doi:10.1016/j.clinimag.2014.12.010 · Cited by 11
Quantitative Clinical Cardiac Magnetic Resonance Imaging
Reiter, U., Reiter, C., Kräuter, C., Nizhnikava, V., Fuchsjäger, M., & Reiter, G. (2019) · RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren · doi:10.1055/a-0999-5716 · Cited by 12
“Advantages of 2D over 4D flow measurement include short scan times (so that data can be acquired under breath holding) and the possibility of optimize velocity coding (VENC, as small as possible to maximize velocity-to-noise ratio, but greater than maximum velocities to avoid aliasing and thus potential errors in eval…”
BackgroundEspe, E. K. S., Zhang, L., & Sjaastad, I. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25023 · Cited by 11
Arvidsson, P. M., Nelsson, A., Edlund, J., Smith, J. G., Magnusson, M., Jin, N., Heiberg, E., Carlsson, M., Steding‐Ehrenborg, K., & Arheden, H. (2022) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00257.2022 · Cited by 14
Mokhtari, A., Corso, P., Jung, B. A., Ferrari, L., Zheng, S., & Obrist, D. (2025) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2025.110405 · Cited by 10
MacDonald, M. E., Forkert, N. D., Pike, G. B., & Frayne, R. (2016) · PLoS ONE · doi:10.1371/journal.pone.0149930 · Cited by 10
“For some regions, like the heart, this phase error has been reported to be as large as 10% to 25% [2, 15].”
BackgroundMethodResultInfluentialNacif, M. S., de Almeida, A. L. C., Young, A. A., Cowan, B. R., Armstrong, A. D. C., Yang, E., Sibley, C. T., Hundley, W. G., Liu, S., Lima, J. A. C., et al. (2017) · Arquivos Brasileiros de Cardiologia · doi:10.5935/abc.20170063 · Cited by 10
Middione, M. J., Wu, H. H., & Ennis, D. B. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25059 · Cited by 10
Enhancing the Velocity Data From 4D Flow MR Images by Reducing its Divergence
Mura, J., Pino, A. M., Sotelo, J. A., Valverde, I., Cristián Tejos, N., Andía, M. E., Pablo Irarrázaval, M., & Uribe, S. A. (2016) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2016.2570010 · Cited by 12
“volume effect and noise [11]–[17].”
Viola, F., Dyverfeldt, P., Carlhäll, C., & Ebbers, T. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26879 · Cited by 10
“A limit of acceptability of 5% error in stroke volume was considered acceptable by Gatehouse et al.31 Similarly, pathline analysis yielded better results for BC3-to-4 in both sequences, as the meand was lower than 5%, and the nonphysiological flow at its minimum, which was 18% and 17% for SGRE and EPI, respectively.”
MethodInfluentialStudy of the right ventricle using magnetic resonance imaging
Alber, A. C., Astigarraga, E., de Paz, M. A., Larena, J. A., & Yeregui, T. S. (2012) · Radiología (English Edition) · doi:10.1016/j.rxeng.2012.07.004 · Cited by 12
“Yet, there is a margin of error due to limitations inherent to the technique.(4) Cardiac perfusion MR imaging requires high temporal resolution, compromising spatial resolution and anatomic coverage.”
BackgroundKocaoğlu, M., Pednekar, A. S., Tkach, J. A., & Taylor, M. D. (2021) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-021-00811-7 · Cited by 9
MRI Assessment of Diastolic and Systolic Intraventricular Pressure Gradients in Heart Failure
Jain, S., Londoño, F. J., Segers, P., Gillebert, T. C., Buyzere, M. D., & Chirinos, J. A. (2016) · Current Heart Failure Reports · doi:10.1007/s11897-016-0281-0 · Cited by 9
Increased biventricular hemodynamic forces in precapillary pulmonary hypertension
Pola, K., Bergström, E., Töger, J., Rådegran, G., Arvidsson, P. M., Carlsson, M., Arheden, H., & Ostenfeld, E. (2022) · Scientific Reports · doi:10.1038/s41598-022-24267-6 · Cited by 9
Design, construction, and evaluation of a dynamic MR compatible cardiac left ventricle model
Dieringer, M. A., Hentschel, J., de Quadros, T., von Knobelsdorff‐Brenkenhoff, F., Hoffmann, W., Niendorf, T., & Schulz‐Menger, J. (2012) · Medical Physics · doi:10.1118/1.4736954 · Cited by 10
Self-citationHedström, E., Bloch, K. M., Bergvall, E., Ståhlberg, F., & Arheden, H. (2010) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-12-70 · Cited by 9
“The velocity offset error was below 0.6 cm/s in all evaluated patients[ 13 ].”
BackgroundInfluentialPruitt, A. A., Jin, N., Liu, Y., Simonetti, O. P., & Ahmad, R. (2018) · Magnetic Resonance in Medicine · doi:10.1002/mrm.27572 · Cited by 9
Self-citationTruedsson, F., Polte, C. L., Gao, S. A., Johnsson, Å. A., Bech‐Hanssen, O., & Lagerstrand, K. M. (2021) · International journal of cardiac imaging · doi:10.1007/s10554-021-02341-w · Cited by 8
Keegan, J., Patel, H. C., Simpson, R., Mohiaddin, R. H., & Firmin, D. N. (2015) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-014-0105-x · Cited by 8
“in MR phase velocity mapping and we have minimised this by careful positioning of the region of interest close to isocentre [35].”
MethodSelf-citationPhase contrast MRI with minimized background phase errors
Loecher, M., & Ennis, D. B. (2024) · Magnetic Resonance in Medicine · doi:10.1002/mrm.30336 · Cited by 7
Self‐gated cine phase‐contrast balanced SSFP flow quantification at 0.55 T
McGrath, C., Bieri, O., Kozerke, S., & Bauman, G. (2023) · Magnetic Resonance in Medicine · doi:10.1002/mrm.29837 · Cited by 7
NO-HYPE: a novel hydrodynamic phantom for the evaluation of MRI flow measurements
Gadda, G., Cocozza, S., Gambaccini, M., Taibi, A., Tedeschi, E., Zamboni, P., & Palma, G. (2021) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-021-02390-2 · Cited by 7
Rengier, F., Delles, M., Unterhinninghofen, R., Ley, S., Müller–Eschner, M., Partovi, S., Geisbüsch, P., Dillmann, R., Kauczor, H., & von Tengg‐Kobligk, H. (2012) · International journal of cardiac imaging · doi:10.1007/s10554-012-0027-3 · Cited by 7
“offset errors remaining after correction also might have been a contributing factor [27], especially with regards to the impact of FOV position on the accuracy.”
Method for calculating confidence intervals for phase contrast flow measurements
Hansen, M. S., Olivieri, L., O’Brien, K., Cross, R. J., Inati, S., & Kellman, P. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-16-46 · Cited by 7
“One example is a study comparing methods for correction of biases caused by experimental imperfections [8-10].”
BackgroundKlug, G., Reinstadler, S. J., Feistritzer, H., Kremser, C., Schwaiger, J. P., Reindl, M., Mair, J. M., Müller, S., Mayr, A., Franz, W., et al. (2015) · European Radiology · doi:10.1007/s00330-015-4022-7 · Cited by 7
Estudio del ventrículo derecho mediante resonancia magnética
Alber, A. C., Astigarraga, E., de Paz, M. A., Larena, J. A., & Yeregui, T. S. (2011) · Radiología · doi:10.1016/j.rx.2011.05.019 · Cited by 8
Pola, K., Ashkir, Z., Myerson, S. G., Arheden, H., Watkins, H. C., Neubauer, S., Arvidsson, P. M., & Raman, B. (2024) · European Heart Journal - Imaging Methods and Practice · doi:10.1093/ehjimp/qyae074 · Cited by 5
“End-systolic volume, mL # 74 [21] 68 [23] 67 [21] 0.3 0.2 0.9 End-systolic volume/BSA, mL/m 2# 37 [11] 38 [11] 38 [11] Flow inefficiencies in non-obstructive HCM on 4D-flow CMR Peak KE and root mean square HDF and HDF ratio for systole and diastole are presented in Table 2 and Supplementary data…”
BackgroundMethodInfluentialMüller, M. L., Daud, E., Langer, G., Gröschel, J., Viezzer, D., Hadler, T., Jin, N., Giese, D., Schmitter, S., Schulz‐Menger, J., et al. (2024) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2024.1456814 · Cited by 5
Self-citationBlood-threshold CMR volume analysis of functional univentricular heart
Secchi, F., Alì, M., Petrini, M., Pluchinotta, F. R., Cozzi, A., Carminati, M., & Sardanelli, F. (2018) · La radiologia medica · doi:10.1007/s11547-017-0851-6 · Cited by 5
Assessment of the tricuspid valve using cardiovascular magnetic resonance
Zhan, Y., & Shah, D. J. (2021) · Current Opinion in Cardiology · doi:10.1097/hco.0000000000000887 · Cited by 5
“Additionally, phase contrast imaging should be corrected using phantoms or background soft tissue [16].”
BackgroundAccelerated aortic 4D flow MRI with wave‐CAIPI
Richter, J. A. J., Wech, T., Weng, A. M., Stich, M., Jin, N., Kosmala, A., Bley, T. A., & Köstler, H. (2020) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28605 · Cited by 5
Middione, M. J., Thompson, R. B., & Ennis, D. B. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24861 · Cited by 5
“The Qp/Qs ratios for SSRG with HBW þ TE IN,MIN were lower and significantly different than FCFE with LBW þ TE MID (1.00 6 0.02 vs. 1.05 6 0.04, P < 0.005).”
InfluentialLagerstrand, K. M., Svensson, F. R., Polte, C. L., Bech‐Hanssen, O., Starck, G., Chadorowski, A., & Johnsson, Å. A. (2020) · Journal of medical imaging · doi:10.1117/1.jmi.7.6.063502 · Cited by 4
Thut, T. L. Z., Buechel, E. R. V., Geiger, J., Kellenberger, C. J., Rücker, B., & Burkhardt, B. E. U. (2023) · Diagnostics · doi:10.3390/diagnostics13020180 · Cited by 3
“Offset errors were corrected using phantom acquisitions [21].”
MethodRutz, T., Meierhofer, C., Naumann, S., Martinoff, S. A., Ewert, P., Stern, H. C., & Fratz, S. (2017) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25701 · Cited by 4
“Offset errors have been discussed to induce a significant error.(18) There was no correction for offset errors; however, it has recently been shown that offset errors do not play a major role in clinical routine for the MR scanner used.”
BackgroundBlood Flow Velocity Assessment
Firmin, D. N. (2019) · Elsevier eBooks · doi:10.1016/b978-0-323-41561-3.00007-0 · Cited by 4
Self-citationDeep learning phase error correction for cerebrovascular 4D flow MRI
Srinivas, S., Masutani, E., Norbash, A., & Hsiao, A. (2023) · Scientific Reports · doi:10.1038/s41598-023-36061-z · Cited by 4
The roles of CMR and MSCT in adult congenital heart disease
Kilner, P. J., Nicol, E. D., & Rubens, M. B. (2015) · Oxford University Press eBooks · doi:10.1093/med/9780198703341.003.0042 · Cited by 4
Self-citationMeasurements of arterial pressure and flow in vivo
Sharman, J. E., Segers, P., & Chirinos, J. A. (2022) · Textbook of Arterial Stiffness and Pulsatile Hemodynamics in Health and Disease · doi:10.1016/b978-0-323-91391-1.00002-9 · Cited by 3
Increased contractility affects left ventricular kinetic energy in pulmonary hypertension
Bergström, E., Pola, K., Kjellström, B., Töger, J., Arvidsson, P. M., Carlsson, M., Rådegran, G., Arheden, H., & Ostenfeld, E. (2025) · Physiological Reports · doi:10.14814/phy2.70563 · Cited by 3
“Background phase errors were corrected with fitting to stationary tissue, using either first or second order correction based on visual assessment (Busch et al., 2017; Gatehouse et al., 2010).”
MRT zur Therapieplanung bei Patienten mit Vorhofseptumdefekt
Huber, A. M. A., Prompona, M., Kozlik-Feldmann, R. G., Mühling, O. M., Rummeny, E. J., Reiser, M. F., & Theisen, D. (2011) · Der Radiologe · doi:10.1007/s00117-010-1998-5 · Cited by 3
The Role of Cardiovascular Magnetic Resonance Imaging in the Assessment of Mitral Regurgitation
Botis, I., Bazmpani, M.-A., Daios, S., Ziakas, A., Kamperidis, V., & Karamitsos, T. D. (2024) · Diagnostics · doi:10.3390/diagnostics14060644 · Cited by 3
Background phase induced steady‐state effects in velocity quantification using phase‐contrast MRI
Fischer, C., Speier, P., Schaeffter, T., & Giese, D. (2024) · Magnetic Resonance in Medicine · doi:10.1002/mrm.30358 · Cited by 3
Rüegger, C. M., Makki, M., Capel, C., Gondry‐Jouet, C., & Balédent, O. (2014) · European Journal of Radiology Open · doi:10.1016/j.ejro.2014.09.005 · Cited by 3
Review of Journal of Cardiovascular Magnetic Resonance 2011
Pennell, D. J., Carpenter, J. P., Firmin, D. N., Kilner, P. J., Mohiaddin, R. H., & Prasad, S. K. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-78 · Cited by 3
“This manuscript describes a follow-up study to the multi-vendor, multi-centre study by Gatehouse in which they measured offsets in phase contrast velocity mapping studies [33].”
Self-citationReview of journal of cardiovascular magnetic resonance 2010
Pennell, D. J., Firmin, D. N., Kilner, P. J., Manning, W. J., Mohiaddin, R. H., & Prasad, S. K. (2011) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-13-48 · Cited by 3
“This multi-centre, multi-vendor study used static phantoms to test for such errors in the absence of any post-acquisition correction [23].”
MethodSelf-citationDunn, T. S., Patel, P., Abazid, B., Nagaraj, H. M., Desai, R. V., Gupta, H., & Lloyd, S. G. (2019) · Echocardiography · doi:10.1111/echo.14358 · Cited by 3
Niclauss, L., Masci, P. G., Pavon, A. G., Rodrigues, D., & Schwitter, J. (2020) · The Journal of Cardiovascular Surgery · doi:10.23736/s0021-9509.20.11150-9 · Cited by 3
Self-citationNickander, J., Lundin, M., Abdula, G., Jenner, J., Maret, E., Sörensson, P., Heiberg, E., Sigfridsson, A., & Ugander, M. (2020) · Scientific Reports · doi:10.1038/s41598-020-61812-7 · Cited by 3
Lagerstrand, K. M., Truedsson, F., Gao, S. A., Johansson, A., Bech‐Hanssen, O., Polte, C. L., & Johnsson, Å. A. (2021) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2021.09.010 · Cited by 2
Hughes, M. L., Muthurangu, V., & Taylor, A. M. (2011) · Medical radiology · doi:10.1007/174_2011_415 · Cited by 2
Effects of contour propagation and background corrections in different MRI flow software packages
Boye, D., Springer, O., Wassmer, F., Scheidegger, S., Remonda, L., & Berberat, J. (2015) · Acta Radiologica Open · doi:10.1177/2058460115589124 · Cited by 2
“(20) showed that a velocity offset correction of 2.”
BackgroundPulmonary Hypertension and Thromboembolic Disease
Ley, S., & Kreitner, K. (2017) · Medical radiology · doi:10.1007/174_2017_78 · Cited by 2
Thomas, B., Martins, J. D., Tavares, N. J., Lopes, A., Pinto, F. F., & Fragata, J. (2013) · Annals of Pediatric Cardiology · doi:10.4103/0974-2069.107230 · Cited by 2
“We have also used a phantom correction method to avoid analytical errors as recommended by recent studies.[11,12] All flow studies in our cohort were performed with the patient breathing freely, which is more physiological than apnea.”
CT and MRI of Aortic Valve Disease: Clinical Update
Hallett, R. L., Moainie, S. L., Hermiller, J. B., & Fleischmann, D. (2016) · Current Radiology Reports · doi:10.1007/s40134-016-0176-6 · Cited by 3
Paul, E. A., Solana, A. B., Duong, J., Shah, A. M., Lai, W. W., Tan, E. T., Hardy, C. J., & Chelliah, A. (2020) · Pediatric Radiology · doi:10.1007/s00247-020-04623-2 · Cited by 2
“vendors and can generate up to 10% error in shunt calculations [3].”
BackgroundAssessment of the Severity of Aortic Regurgitation by Noninvasive Imaging: Non-invasive MMI for AR
Ferrer‐Sistach, E., Teis, A., Escabia, C., & Delgado, V. (2023) · Current Cardiology Reports · doi:10.1007/s11886-023-02011-4 · Cited by 2
“Moreover, the technique is prone to background flow offset errors that can be compensated with correction methods to provide accurate measurements [38, 39].”
Bateman, G. A. (2011) · Child s Nervous System · doi:10.1007/s00381-011-1610-y · Cited by 2
“The phase offset error is due to eddy currents in the gradient coils and is an inherent error in each machine which must be removed from all measurements, by estimating its magnitude in an adjacent stationary region and subtracting it from the region of interest [2].”
BackgroundVisual Exploration of Simulated and Measured Blood FlowPreprint
Vilanova, A., Preim, B., van Pelt, R., Gasteiger, R., Neugebauer, M., & Wischgoll, T. (2012) · arXiv (Cornell University) · doi:10.48550/arxiv.1209.0999 · Cited by 2
“These effects are difficult to predict and therefore challenging to correct [14, 38].”
BackgroundGómez, A., Giese, D., Pushparajah, K., Simpson, J. M., Schaeffter, T., & Penney, G. (2013) · Lecture notes in computer science · doi:10.1007/978-3-642-36961-2_28 · Cited by 1
“To analyse the significance of this number, we take into account that although 2D PC-MRI is the gold standard for non-invasive flow measurement, several parameters such as breathing, intercardiac cycle variability, remaining eddy currents or vessel wall segmentation introduce errors of up to 10% in flow rate computati…”
MethodSawhney, S., McLaughlin, J., Halloran, K., Pollack, S. Z., & Cao, J. J. (2016) · International Journal of Diagnostic Imaging · doi:10.5430/ijdi.v3n2p37 · Cited by 1
Le flux 4D : technique et principales applications pour l’étude de l’aorte thoracique
Gabiano, E., Silvestri, V., Pagniez, J., Simeone, A., Hennicaux, J., Longère, B., & Pontana, F. (2020) · Journal d imagerie diagnostique et interventionnelle · doi:10.1016/j.jidi.2020.09.006 · Cited by 1
PHASE CONTRAST MRI IN INTRACRANIAL ANEURYSMS
van Ooij, P. (2012) · UvA-DARE (University of Amsterdam) · Cited by 1
Advanced Imaging of the Right Ventricle
Kühne, T. (2014) · Respiratory medicine · doi:10.1007/978-1-4939-1065-6_4
Fairbairn, T. A. (2013) · White Rose eTheses Online (University of Leeds, The University of Sheffield, University of York)
Flow and Valvular Disease Studied by Cardiovascular Magnetic Resonance
Kilner, P. J. (2010) · Current Cardiovascular Imaging Reports · doi:10.1007/s12410-010-9008-x
Self-citation
Metadata
- Authors
- Gatehouse, P. D., Rolf, M. P., Graves, M. J., Hofman, M. B., Totman, J., Werner, B., Quest, R. A., Liu, Y., von Spiczak, J., Dieringer, M., Firmin, D. N., van Rossum, A., Lombardi, M., Schwitter, J., Schulz-Menger, J., & Kilner, P. J.
- Affiliations
- Royal Brompton Hospital (GB)
- Amsterdam UMC Location Vrije Universiteit Amsterdam (NL)
- University of Cambridge (GB)
- King's College London (GB)
- University Children's Hospital Zurich (CH)
- Hammersmith Hospital (GB)
- University of Auckland (NZ)
- Charité – Universitätsmedizin Berlin (DE)
- National Research Council (IT)
- University Hospital Zurich (CH)
- Corresponding author
- Kilner, Philip J
- ORCID
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2010-10
- Volume
- 12
- Issue
- 1
- Pages
- 5
- ISSN
- 1097-6647 (ISSN-L)
- PMC
- PMC2818657
- OpenAlex
- W2153286609
- Cited by
- 234
- References
- 15
- Citations by year
- 2026: 4 · 2025: 5 · 2024: 8 · 2023: 11 · 2022: 11 · 2021: 13 · 2020: 18 · 2019: 17 · 2018: 12 · 2017: 13 · 2016: 17 · 2015: 15 · 2014: 16 · 2013: 20 · 2012: 24
- FWCI
- 18.70
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- www.elsevier.com/legal/tdmrep-license (tdm)
- creativecommons.org/licenses/by/2/ (vor)
- Locations
- Journal of Cardiovascular Magnetic Resonance (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Pure Amsterdam UMC (publishedVersion, other-oa)
- Amsterdam UMC (VU Amsterdam) - Institutional Repository (publishedVersion, other-oa)
- MDC Repository (Max-Delbrueck-Center for Molecular Medicine) (acceptedVersion)
- Bournemouth University Research Online (Bournemouth University) (acceptedVersion)
- Zurich Open Repository and Archive (University of Zurich) (acceptedVersion, cc-by)
- National University of Singapore (submittedVersion, cc-by)
- Digital Academic REpository of VU University Amsterdam (Vrije Universiteit Amsterdam) (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Spiral (Imperial College London) (submittedVersion, cc-by)
- Repository for Publications and Research Data (ETH Zurich) (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- ETH Zürich Research Collection (cc-by)
- Zurich Open Repository and Archive (University of Zurich)
- Funders
- British Heart Foundation
- Vrije Universiteit Amsterdam
- Eidgenössische Technische Hochschule Zürich
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac Valve Diseases and Treatments
- Keywords
- phase-contrast MRI, flow measurement accuracy, regurgitant flow, shunt flow, aortic blood flow, pulmonary blood flow, cardiac output, gelatin phantom, offset correction
- MeSH terms
- Aorta, Blood Flow Velocity, Cardiac Output, Gelatin, Humans, Image Interpretation, Computer-Assisted, Materials Testing, Mitral Valve Insufficiency, Predictive Value of Tests, Pulmonary Artery, Reproducibility of Results, Respiratory Mechanics, Artifacts, Magnetic Resonance Imaging, Cine, Phantoms, Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.