Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements
Open accessThe DOI matches its Crossref record; 15 of the 15 references it cites were found in a registry or an index.
Abstract
AIMS: Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error. We studied the severity of uncorrected velocity offset errors across sites and CMR systems. METHODS AND RESULTS: In a multi-centre, multi-vendor study, breath-hold through-plane retrospectively ECG-gated phase contrast acquisitions, as are used clinically for aortic and pulmonary flow measurement, were applied to static gelatin phantoms in twelve 1.5 T CMR systems, using a velocity encoding range of 150 cm/s. No post-processing corrections of offsets were implemented. The greatest uncorrected velocity offset, taken as an average over a 'great vessel' region (30 mm diameter) located up to 70 mm in-plane distance from the magnet isocenter, ranged from 0.4 cm/s to 4.9 cm/s. It averaged 2.7 cm/s over all the planes and systems. By theoretical calculation, a velocity offset error of 0.6 cm/s (representing just 0.4% of a 150 cm/s velocity encoding range) is barely acceptable, potentially causing about 5% miscalculation of cardiac output and up to 10% error in shunt measurement. CONCLUSION: In the absence of hardware or software upgrades able to reduce phase offset errors, all the systems tested appeared to require post-acquisition correction to achieve consistently reliable breath-hold measurements of flow. The effectiveness of offset correction software will still need testing with respect to clinical flow acquisitions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Cine or PC sequences can be used to quantify the stenosis, blockage, or regurgitation of these conduits.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, the quantification may be incorrect due to conduit artefacts [37,44].”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Subsequent CMR studies are usually performed if new clinical symptoms arise or every 3–5 years, depending on the previous clinical findings [34,36,37].”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“In paediatric centres with high patient volume, the first CMR is typically conducted around 8–9 years of age, when children can undergo the examination without anaesthesia.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Usually valved, these conduits—which may be synthetic or biological—can develop pulmonary regurgitation later on or function as a tiny conduit as people age.”
From Multimodality imaging assessment of tetralogy of Fallot: From diagnosis to long-term follow-up (2023) · cites this paper as [37]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Handbook of MRI Pulse Sequences
(2004) · Elsevier eBooks · doi:10.1016/b978-0-12-092861-3.x5000-6 · Cited by 1,610
Blood Flow Imaging by Cine Magnetic Resonance
Nayler, G. L., Firmin, D. N., & Longmore, D. B. (1986) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198609000-00001 · Cited by 634 · cited 1×
“Phase contrast cardiovascular magnetic resonance (CMR) [1] measurements of flow through planes transecting the great arteries are used clinically for calculations of cardiac output, shunt flow [2,3] or aortic or pulmonary regurgitation [4,5].”
Self-citationConcomitant gradient terms in phase contrast MR: Analysis and correction
Bernstein, M. A., Zhou, X. J., Polzin, J. A., King, K. F., Ganin, A., Pelc, N. J., & Glover, G. H. (1998) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910390218 · Cited by 547 · cited 2×
“Velocity offsets caused by Maxwell (concomitant) gradients in phase-contrast imaging are calculable allowing automatic correction during reconstruction [11].”
Rebergen, S. A., Chin, J. G., Ottenkamp, J., van der Wall, E. E., & de Roos, A. (1993) · Circulation · doi:10.1161/01.cir.88.5.2257 · Cited by 294
Gelfand, E., Hughes, S., Hauser, T., Yeon, S., Goepfert, L., Kissinger, K., Rofsky, N., & Manning, W. (2006) · Journal of Cardiovascular Magnetic Resonance · doi:10.1080/10976640600604856 · Cited by 265 · cited 2×
“In combination with measurements of left ventricular volume or mitral inflow, measurement of aortic outflow may also allow the indirect calculation of mitral regurgitation [5-7].”
Hundley, W. G., Li, H. F., Willard, J. E., Landau, C., Lange, R. A., Meshack, B. M., Hillis, L. D., & Peshock, R. M. (1995) · Circulation · doi:10.1161/01.cir.92.5.1151 · Cited by 229
Normal Thoracic Aorta Diameter on Cardiac Computed Tomography in Healthy Asymptomatic Adults
Mao, S. S., Ahmadi, N., Shah, B., Beckmann, D., Chen, A., Ngo, L., Flores, F. R., Gao, Y. L., & Budoff, M. J. (2008) · Academic Radiology · doi:10.1016/j.acra.2008.02.001 · Cited by 217 · cited 1×
“The software reported the most extreme mean velocity offset measured in a circular ‘great vessel’ region, 30 mm diameter [13], centered within specified distances (see below) from the centre of each velocity map.”
Quantification of mitral regurgitation by velocity-encoded cine nuclear magnetic resonance imaging
Fujita, N., Chazouilleres, A. F., Hartiala, J. J., O'sullivan, M., Heidenreich, P., Kaplan, J. D., Sakuma, H., Foster, E., Caputo, G. R., & Higgins, C. B. (1994) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(94)90642-4 · Cited by 198 · cited 1×
“In combination with measurements of left ventricular volume or mitral inflow, measurement of aortic outflow may also allow the indirect calculation of mitral regurgitation [ 5-7 ].”
Chernobelsky, A., Shubayev, O., Comeau, C. R., & Wolff, S. D. (2007) · Journal of Cardiovascular Magnetic Resonance · doi:10.1080/10976640601187588 · Cited by 163 · cited 5×
“Opposite polarities from the same scanner have been reported previously [9].”
InfluentialCorrection of phase offset errors in main pulmonary artery flow quantification
Lankhaar, J., Hofman, M. B. M., Marcus, J. T., Zwanenburg, J. J. M., Faes, T. J. C., & Noordegraaf, A. V. (2005) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20361 · Cited by 101
Self-citation
Cited by
Items 201–234 of 234
Nadal, D., Meister, R., Rutishauser, C., Staubli, S., Stücheli, M., & Stücheli-Herlach, P. (2011)
Setser, R. M., Renapurkar, R. D., Desai, M. Y., Flamm, S. D., Bolen, M. A., Gabriel, R. S., Tandon, Y. K., & Lieber, M. L. (2013)
François, C. J. (2017) · Springer eBooks · doi:10.1007/978-3-319-53001-7_10
Advancements to Magnetic Resonance Flow Imaging in the Brain
Ragunathan, S. (2017)
Persistent ductus arteriosus in old patient with atrial fibrillation.
Rajewska-Tabor, J., Żabicki, B., Hrycow, M., Rozmiarek, S., & Pyda, M. (2018) · Europe PMC (PubMed Central) · doi:10.11909/j.issn.1671-5411.2018.05.007
MRI of the Cardiovascular System
von Knobelsdorff‐Brenkenhoff, F., & Schulz‐Menger, J. (2014) · Elsevier eBooks · doi:10.1016/b978-0-444-53632-7.00308-7
Self-citationSaadani‐Makki, F., Metanbou, S., Arbeaumont-Trocme, G., GYSEL, J. V., & Makki, M. (2020) · Diagnostics · doi:10.3390/diagnostics10060387
“When combined together and integrated over time, these inherent background errors produced inaccuracy in the assessment of flow measurements and characteristics and were more prominent in high-power gradient systems [7,8].”
BackgroundTechniques and Clinical Applications of Phase-Contrast MRI in CHD
Rajeshkannan, R. (2020) · Springer eBooks · doi:10.1007/978-981-15-6755-1_20
Is cardiac magnetic resonance ready for aortic regurgitation?
Markl, M., Lee, J., & Pradella, M. (2021) · Kardiologia Polska · doi:10.33963/kp.a2021.0094
“Third, accurate flow quantification using of 2D phase-contrast CMR requires careful 2D analysis plane orientation [10] along the aorta [11], correction for eddy-current induced background phase offset [12], and the use of sufficiently high velocity encoding sensitivity.”
MethodCurrent Awareness in NMR in Biomedicine
(2010) · NMR in Biomedicine · doi:10.1002/nbm.1457
(2013) · Elsevier eBooks · doi:10.1016/b978-0-12-415873-3.00035-3
Current Awareness in NMR in Biomedicine
John Wiley & Sons, L. (2010) · NMR in Biomedicine · doi:10.1002/nbm.1456
Oshinski, J. N., Sahu, A., & Hartlage, G. R. (2015) · Springer eBooks · doi:10.1007/978-3-319-22141-0_16
Schwitter, J. (2023) · Medical radiology · doi:10.1007/174_2023_404
Self-citationImaging aortic flows in 4D using MRI
van Ooij, P. (2024) · Elsevier eBooks · doi:10.1016/b978-0-323-95484-6.00015-4
Minderhoud, S. C. S., & Hirsch, A. (2024) · Medical radiology · doi:10.1007/174_2023_451
Berg, E. A. R., Fiorentini, S., Avdal, J., Grenne, B. L., Stensæth, K. H., While, P. T., Espeland, T., Wiseth, R., Torp, H. G., & Aakhus, S. (2024) · European Heart Journal - Imaging Methods and Practice · doi:10.1093/ehjimp/qyae138
“20 If gel phantom recordings were mispositioned or missing, a similar correction was applied using stationary patient tissue as the reference.”
Solana, A. B., Minderhoud, S. C. S., Wielopolski, P. A., Hernandez-Tamames, J. A., Budde, R. P. J., Helbing, W. A., Janich, M., & Hirsch, A. (2025) · Journal of Cardiovascular Magnetic Resonance · doi:10.1016/j.jocmr.2025.101895
“If the perpendicular vector between the linear fit and the scan plane is greater than 0.6 cm/s [5], the linear fit surface is subtracted from each velocity image, and the static tissue identification is restarted using these initial linear fit corrected velocity images.”
InfluentialEide, P. K., Valnes, L. M., Lindstrøm, E. K., Mardal, K.-A., & Ringstad, G. A. (2021) · Figshare · doi:10.6084/m9.figshare.14361566
Minderhoud, S. C. S., van der Velde, N., Wentzel, J. J., Geest, R. J. V. D., Attrach, M., Wielopolski, P. A., Budde, R. P. J., Helbing, W. A., Roos-Hesselink, J. W., & Hirsch, A. (2020) · Figshare · doi:10.6084/m9.figshare.12965480
Minderhoud, S. C. S., van der Velde, N., Wentzel, J. J., Geest, R. J. V. D., Attrach, M., Wielopolski, P. A., Budde, R. P. J., Helbing, W. A., Roos-Hesselink, J. W., & Hirsch, A. (2020) · Figshare · doi:10.6084/m9.figshare.12965477
Minderhoud, S. C. S., van der Velde, N., Wentzel, J. J., Geest, R. J. V. D., Attrach, M., Wielopolski, P. A., Budde, R. P. J., Helbing, W. A., Roos-Hesselink, J. W., & Hirsch, A. (2020) · Figshare · doi:10.6084/m9.figshare.12965483
Minderhoud, S. C. S., van der Velde, N., Wentzel, J. J., Geest, R. J. V. D., Attrach, M., Wielopolski, P. A., Budde, R. P. J., Helbing, W. A., Roos-Hesselink, J. W., & Hirsch, A. (2020) · Figshare · doi:10.6084/m9.figshare.12965471
Minderhoud, S. C. S., van der Velde, N., Wentzel, J. J., Geest, R. J. V. D., Attrach, M., Wielopolski, P. A., Budde, R. P. J., Helbing, W. A., Roos‐Hesselink, J. W., & Hirsch, A. (2020) · Figshare · doi:10.6084/m9.figshare.12965474
Phase contrast CMR in the descending aorta as a supportive reference for severe aortic regurgitation
Truedsson, F., Gao, S. A., Polte, C. L., Bech‐Hanssen, O., Johnsson, Å. A., & Lagerstrand, K. M. (2025) · Scientific Reports · doi:10.1038/s41598-025-31268-8
“Background offset compensation was applied to reduce eddy current-induced velocity offsets using post-acquisition adaptive filtering, resulting in a velocity offset below the acceptance limit (0.6 cm/s) 18 .”
InfluentialAortic Regurgitation as the Next Frontier in the TAVR Space
Nienaber, S., Curio, J., Tarantini, G., Wienemann, H., & Adam, M. (2025) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2025.11.029 · Cited by 1
Trauzeddel, R. F., Grandy, T. H., Daud, E., Müller, M., Gröschel, J., Viezzer, D., Hadler, T., Błaszczyk, E., Jin, N., Giese, D., et al. (2026) · Radiology Cardiothoracic Imaging · doi:10.1148/ryct.240247
Self-citationNumerical simulations for phase-contrast magnetic resonance imaging
Puiseux, T. (2019) · doi:10.70675/2df1ad5ezf6acz4f7cz8b17z6abda9a02b5b
“Moreover, machine-specific variability could affect the measurements depending on both hardware and software settings [60].”
BackgroundApport de la mécanique des fluides dans l'étude des flux sanguins aortiques
Bouaou, K. (2020) · theses.fr (ABES) · doi:10.70675/a8635e8az82efz494bz857az87d181f006fb
Pedersen, A. U., Pedersen, A. L. D., Mogensen, N. S. B., Nielsen, L. H., Nissen, H., Freeman, P., Hansson, N., Ellert‐Gregersen, J., Zaremba, T., Nørgaard, B. L., et al. (2026) · Journal of Cardiovascular Magnetic Resonance · doi:10.1016/j.jocmr.2026.102763 · Cited by 1
Kawaguchi, A., Kawaguchi, W., Hayashi, M., Yasuda, K., & Ohashi, Y. (2026) · Radiological Physics and Technology · doi:10.1007/s12194-026-01084-4
“RV volumetric analysis using LV short-axis images is susceptible to measurement error due to complex RV geometry [ 6], whereas PC flow measurements are subject to technical variability and reported Ao–PA flow differences of approximately 5% [7, 8 ].”
Velocity and flow imaging—cerebrospinal fluid, large vessels, and angiography
Bloch, K. M., & Töger, J. (2026) · Advances in magnetic resonance technology and applications · doi:10.1016/b978-0-443-21425-7.00003-8
Full 3D blood velocity mapping and flow quantification from Doppler echocardiographic images
Gómez, A. (2013) · UPT. Syiah Kuala University Library (Syiah Kuala University)
Metadata
- Authors
- Gatehouse, P. D., Rolf, M. P., Graves, M. J., Hofman, M. B., Totman, J., Werner, B., Quest, R. A., Liu, Y., von Spiczak, J., Dieringer, M., Firmin, D. N., van Rossum, A., Lombardi, M., Schwitter, J., Schulz-Menger, J., & Kilner, P. J.
- Affiliations
- Royal Brompton Hospital (GB)
- Amsterdam UMC Location Vrije Universiteit Amsterdam (NL)
- University of Cambridge (GB)
- King's College London (GB)
- University Children's Hospital Zurich (CH)
- Hammersmith Hospital (GB)
- University of Auckland (NZ)
- Charité – Universitätsmedizin Berlin (DE)
- National Research Council (IT)
- University Hospital Zurich (CH)
- Corresponding author
- Kilner, Philip J
- ORCID
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2010-10
- Volume
- 12
- Issue
- 1
- Pages
- 5
- ISSN
- 1097-6647 (ISSN-L)
- PMC
- PMC2818657
- OpenAlex
- W2153286609
- Cited by
- 234
- References
- 15
- Citations by year
- 2026: 4 · 2025: 5 · 2024: 8 · 2023: 11 · 2022: 11 · 2021: 13 · 2020: 18 · 2019: 17 · 2018: 12 · 2017: 13 · 2016: 17 · 2015: 15 · 2014: 16 · 2013: 20 · 2012: 24
- FWCI
- 18.70
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- www.elsevier.com/legal/tdmrep-license (tdm)
- creativecommons.org/licenses/by/2/ (vor)
- Locations
- Journal of Cardiovascular Magnetic Resonance (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Pure Amsterdam UMC (publishedVersion, other-oa)
- Amsterdam UMC (VU Amsterdam) - Institutional Repository (publishedVersion, other-oa)
- MDC Repository (Max-Delbrueck-Center for Molecular Medicine) (acceptedVersion)
- Bournemouth University Research Online (Bournemouth University) (acceptedVersion)
- Zurich Open Repository and Archive (University of Zurich) (acceptedVersion, cc-by)
- National University of Singapore (submittedVersion, cc-by)
- Digital Academic REpository of VU University Amsterdam (Vrije Universiteit Amsterdam) (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Spiral (Imperial College London) (submittedVersion, cc-by)
- Repository for Publications and Research Data (ETH Zurich) (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- ETH Zürich Research Collection (cc-by)
- Zurich Open Repository and Archive (University of Zurich)
- Funders
- British Heart Foundation
- Vrije Universiteit Amsterdam
- Eidgenössische Technische Hochschule Zürich
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac Valve Diseases and Treatments
- Keywords
- phase-contrast MRI, flow measurement accuracy, regurgitant flow, shunt flow, aortic blood flow, pulmonary blood flow, cardiac output, gelatin phantom, offset correction
- MeSH terms
- Aorta, Blood Flow Velocity, Cardiac Output, Gelatin, Humans, Image Interpretation, Computer-Assisted, Materials Testing, Mitral Valve Insufficiency, Predictive Value of Tests, Pulmonary Artery, Reproducibility of Results, Respiratory Mechanics, Artifacts, Magnetic Resonance Imaging, Cine, Phantoms, Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.