Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review
Open accessAbstract
Hemodynamic monitoring is essential in the postoperative management of neonates undergoing cardiac surgery, enabling early identification of circulatory failure and its underlying cause, optimization of oxygen delivery to tissues, and evaluation of treatment response. Despite its significant role, there is still no consensus and there remains substantial heterogeneity in bedside hemodynamic monitoring practices. Pediatric intensivists typically rely on macro- and microcirculatory indicators, including arterial blood pressure, urine output, capillary refill time, mixed venous oxygen saturation, lactate concentration, and serial echocardiographic evaluations. However, most of these are indirect hemodynamic indicators and provide only intermittent snapshots of the hemodynamic status, which can be very fluctuating following cardiac surgery. Technological advancements have driven a shift toward continuous, noninvasive monitoring techniques, such as near-infrared spectroscopy (NIRS), electrical biosensing technology, and microcirculatory assessment tools. Real-time, simultaneous tracking of multiple physiological variables through a multimodal hemodynamic monitoring protocol facilitates the understanding of systemic and regional perfusion and oxygenation. This narrative review aims to summarize current techniques and innovations in neonatal hemodynamic monitoring following cardiac surgery, combining clinical evaluation with both intermittent and continuous noninvasive techniques.
Overview
CoChat Summary
TL;DR
This narrative review summarizes current noninvasive and minimally invasive hemodynamic monitoring techniques for neonates after cardiac surgery, emphasizing that no single device meets all ideal requirements. It highlights NIRS, electrical biosensing, echocardiography, and microcirculatory tools, while noting limited validation, lack of standardization, and no consensus on thresholds or treatment targets.
Key findings
- Low cardiac output syndrome (LCOS) is the most common postoperative complication in children with congenital heart disease, with incidence 25–60% and typical onset within 6–18 h after surgery; definitions vary, with metabolic acidosis in 64%, inotropic support in 61%, and clinical manifestations in 59% of pediatric definitions.
- Indirect markers are limited: low cardiac output is clinically recognized in only 25% of pediatric intensive care cases; lactate is used by 99% of pediatric intensivists; postoperative acute kidney injury is reported in approximately 53.8% of neonates undergoing congenital heart surgery.
- Echocardiography remains the most widely used direct method, with neonatal cardiac output values generally 150–300 mL/kg/min, but it has a precision error around 30% versus dilution techniques and intra-/interobserver coefficients of variation from 2.1–22% and 3.1–22%, respectively; targeted neonatal echocardiography changes management in approximately 30–60% of hemodynamically unstable patients.
- NIRS average cerebral oxygen saturation is approximately 65–70% in term and preterm neonates versus 46–57% in cyanotic congenital heart disease; a cerebral oxygen saturation threshold <58% combined with lactate >7.4 mmol/L at 24 h post-neonatal cardiac surgery predicted survival and neurodevelopmental outcomes, but no universally accepted lower intervention threshold exists.
- The ESPR NICOM consensus states thoracic electrical biosensing technology is not currently reliable for precise cardiac output measurement in neonates and should not replace transthoracic echocardiography, though it may be useful for monitoring changes from an individual baseline.
Methods
- Narrative review with a systematic database search in PubMed and Embase for 2015–2025 using terms combining neonate/newborn/infant, congenital heart disease, cardiac surgery/procedure/intervention, and hemodynamic/hemodynamics; additional articles were found by manual search.
- One reviewer screened titles and abstracts for relevance and duplication; full texts in languages other than English were excluded. Evidence was synthesized narratively, with no formal meta-analysis or quantitative pooling described.
Limitations
- The paper states there is no gold standard or consensus in pediatric hemodynamic monitoring, leading to institutional variability; many studies of newer devices are small, single-center, and lack standardized reference values, limiting generalizability.
- The paper states newer devices are not routinely implemented, current expert recommendations support use primarily in research settings, and there are challenges in validation against reference methods, lack of measurement standardization, no consensus for guiding treatment, and limited data on patient outcomes.
- Evident from the text: the search and screening were performed by one reviewer, and no formal risk-of-bias assessment or quantitative synthesis is described, which may limit reproducibility and comprehensiveness.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
55 Valid · 15 Not valid · 48 Not checked · 77 Own results and interpretations
At least one cited source doesn’t support this claim or points to a different paper, and none confirmed it — see the verdicts below.
In the paper
“evaluating fluid responsiveness in children with shock [77]”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 4. Direct Hemodynamic Monitoring > 4.5. Thoracic Electrical Biosensing Technology
What the sources say
- Disputedcites [77] Awadhare, P., Patel, R., McCallin, T., Mainali, K., Jackson, K., Starke, H., and Bhalala, U. (2022). Non-invasive Cardiac Output Monitoring and Assessment of Fluid Responsiveness in Children With Shock in the Emergency Department. Front. Pediatr., 10. sourceExperimental — accuracy not yet measured
The Methods report measuring thoracic fluid content (TFC) with an electrical cardiometry device in children with shock. But the Results (Effects of Fluid Therapy and Fluid Responsiveness; page not available) found no significant difference in TFC change between fluid responders and non-responders. The paper identifies other measures—not TFC—as predictors of fluid responsiveness. (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Disputed
At least one cited source doesn’t support this claim or points to a different paper, and none confirmed it — see the verdicts below.
In the paper
“Scolletta et al. evaluated sublingual microcirculation in infants with cyanotic and acyanotic CHD using SDF, noting that while acyanotic infants maintained stable microvascular parameters, those with cyanotic CHD showed significant reductions in total vessel density (TVD) and proportion of perfused vessels (PPV) [86].”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 4. Direct Hemodynamic Monitoring > 4.6. Microcirculatory Monitoring Devices
What the sources say
- Disputedcites [86] Scolletta, 2016, Microcirculatory changes in children undergoing cardiac surgery: A prospective observational study, Br. J. Anaesth. sourceExperimental — accuracy not yet measured
The Table 2 results report a significant increase in PPV over time in cyanotic children and an increase, not a reduction, in TVD. The paper does describe comparatively stable PPV in acyanotic children, but its findings run contrary to the claimed reductions in the cyanotic group. (LLM, conf=0.90)
Checked by a reasoning model — may be wrong.
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Disputed
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Low cardiac output state is accurately clinically recognized in only 25% of cases in pediatric intensive care units [10].”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 3. Indirect Hemodynamic Monitoring
What the sources say
- Partly supportedcites [10] Tibby, 1997, Clinicians’ abilities to estimate cardiac index in ventilated children and infants, Arch. Dis. Child. sourceExperimental — accuracy not yet measured
Table 1 (Discussion; page unavailable) shows that clinicians correctly classified 12 of 47 measured low-cardiac-index estimates as low—about 25%. The Patients and methods section specifies that these estimates concerned ventilated, sedated children, so the result does not establish a 25% recognition rate for all pediatric intensive care patients. (LLM, conf=0.85)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“However, aggressive efforts to normalize lactate concentrations may lead to fluid overload, especially in the absence of other signs of tissue hypoperfusion [36,37].”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 3. Indirect Hemodynamic Monitoring > 3.6. Lactate
What the sources say
- Supportedcites [36] Kattan, 2020, A lactate-targeted resuscitation strategy may be associated with higher mortality in patients with septic shock and normal capillary refill time: A post hoc analysis of the ANDROMEDA-SHOCK study, Ann. Intensive Care source
The Discussion reports that patients with normal capillary refill time who received lactate-targeted resuscitation received more fluid boluses. It also describes an increased burden of over-resuscitation and cautions against pursuing lactate normalization without considering peripheral perfusion. These passages support the claim’s hedged warning in the septic-shock setting studied; page numbers were not available in our copy. (LLM, conf=0.88)
- Supportedcites [37] Bakker, 2016, Lactate-guided resuscitation saves lives: We are not sure, Intensive Care Med. source
In the paper’s discussion (page not provided), the authors warn that pursuing lactate normalization with further fluids or inotropes after other signs of tissue hypoperfusion have disappeared may expose patients to the toxicity of overresuscitation without clear benefit. This supports the citing sentence’s qualified warning about aggressive fluid administration. (LLM, conf=0.88)
hedgedstates a possible cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“TNE can lead to changes in clinical management in approximately 30–60% of patients with hemodynamic instability [50].”
§ Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review > 4. Direct Hemodynamic Monitoring > 4.1. Echocardiography
What the sources say
- Supportedcites [50] Singh, Y. (2017). Echocardiographic evaluation of hemodynamics in neonates and children. Front. Pediatr., 5. source
The Introduction reports that clinical management changed in 30–60% of cases after point-of-care echocardiography. The Abstract places its use in clinical decision-making for patients with hemodynamic instability, supporting the paraphrase. (LLM, conf=0.90)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
83 more claims · 26 with their cited sources checked
Create a free account to see all 195 claims and to verify the 2 references still open.
- It may state two separate claims as one. 77
- Probably not a substantial claim. 45
- The quote only partly states this claim. 2
Citations
Cites
Diagnosis, evaluation, and management of acute kidney injury: a KDIGO summary (Part 1)
Kellum, J. A., Lameire, N., & for the KDIGO AKI Guideline Work Group (2013) · Critical Care · doi:10.1186/cc11454 · Cited by 2,827 · cited 1×
“According to the KDIGO guidelines, acute kidney injury (AKI) is diagnosed based on elevated serum creatinine levels and/or diminished urine output [19].”
Gaies, M. G., Gurney, J. G., Yen, A. H., Napoli, M. L., Gajarski, R. J., Ohye, R. G., Charpie, J. R., & Hirsch, J. C. (2010) · Pediatric Critical Care Medicine · doi:10.1097/pcc.0b013e3181b806fc · Cited by 1,325 · cited 1×
“Gaies et al. applied VIS to neonates and small infants after cardiac surgery, defining high VIS as ≥ 20 during the first 24 h or ≥ 15 during the 24–48 h postoperative period [43].”
Wernovsky, G., Wypij, D., Jonas, R. A., Mayer, J. E., Hanley, F. L., Hickey, P. R., Walsh, A. Z., Chang, A. C., Castañeda, A. R., Newburger, J. W., et al. (1995) · Circulation · doi:10.1161/01.cir.92.8.2226 · Cited by 1,010
Jansen, T. C., van Bommel, J., Schoonderbeek, F. J., Sleeswijk Visser, S. J., van der Klooster, J. M., Lima, A. P., Willemsen, S. P., & Bakker, J. (2010) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200912-1918oc · Cited by 894 · cited 1×
“Strategies aimed at reducing lactate levels have been linked to improved clinical outcomes, reduced organ dysfunction and lower mortality rates in various shock states [35].”
Marik, P. E., & Cavallazzi, R. (2013) · Critical Care Medicine · doi:10.1097/ccm.0b013e31828a25fd · Cited by 866 · cited 1×
“Central venous pressure (CVP), measured at the junction of the superior vena cava and right atrium, remains a vital part of the hemodynamic monitoring post-cardiopulmonary bypass, despite its limited utility in assessing fluid responsiveness [15,16].”
Cecconi, M., Rhodes, A., Poloniecki, J., Della Rocca, G., & Grounds, R. M. (2009) · Critical Care · doi:10.1186/cc7129 · Cited by 351 · cited 1×
“For instance, echocardiography shows a precision error of around 30% when compared to gold standard dilution techniques, which falls within the clinically accepted range [54,55].”
Vasoactive-Inotropic Score: Evolution, Clinical Utility, and Pitfalls
Belletti, A., Lerose, C. C., Zangrillo, A., & Landoni, G. (2021) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2020.09.117 · Cited by 295
Clinicians’ abilities to estimate cardiac index in ventilated children and infants
Tibby, S. M., Hatherill, M., Marsh, M. J., & Murdoch, I. A. (1997) · Archives of Disease in Childhood · doi:10.1136/adc.77.6.516 · Cited by 224 · cited 1×
“Low cardiac output state is accurately clinically recognized in only 25% of cases in pediatric intensive care units [10].”
Alderliesten, T., Dix, L., Baerts, W., Caicedo, A., van Huffel, S., Naulaers, G., Groenendaal, F., van Bel, F., & Lemmers, P. (2016) · Pediatric Research · doi:10.1038/pr.2015.186 · Cited by 199 · cited 1×
“Average ScO 2 values range between approximately 65 and 70% in both term and preterm neonates [68,69], in comparison to 46–57% in patients with cyanotic congenital heart disease [12].”
Charpie, J. R., Dekeon, M. K., Goldberg, C. S., Mosca, R. S., Bove, E. L., & Kulik, T. J. (2000) · The Journal of Thoracic and Cardiovascular Surgery · doi:10.1067/mtc.2000.106838 · Cited by 198 · cited 1×
“Elevated lactate concentrations have been associated with increased mortality and a higher risk for requiring ECMO post-cardiac surgery [34].”
Cited by
Li, D., & Yao, D. (2026) · Frontiers in Immunology · doi:10.3389/fimmu.2026.1776897 · Cited by 1
“Based on available pediatric data and expert consensus, we propose following provisional thresholds (7, 93, 132, 137, 146, 147): (cid:129) NT-proBNP/BNP: A change ≥ 30% from baseline (or ≥ 2-fold increase if no baseline available) is considered signi fi cant.”
Nedelcu, C., Cirstoveanu, C. G., Filip, C. I., et al. (2026) · Life · doi:10.3390/life16061001
Self-citation
Metadata
- Authors
- Nedelcu, C., Ionescu, N. S., Bizubac, A. M., Filip, C., & Cirstoveanu, C.
- Affiliations
- Carol Davila University of Medicine and Pharmacy (RO)
- Department of Neonatal Intensive Care, “Carol Davila” University of Medicine and Pharmacy, 020021 Bucharest, Romania
- Corresponding author
- Nedelcu, Carmina
- Ionescu, Nicolae Sebastian
- Venue
- Life
- Publisher
- MDPI AG
- Published
- 2025-10-17
- Volume
- 15
- Issue
- 10
- Pages
- 1621
- ISSN
- 2075-1729 (ISSN-L)
- PMC
- PMC12565398
- OpenAlex
- W4415299593
- Cited by
- 2
- References
- 85
- Citations by year
- 2026: 2
- FWCI
- 0.48
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Life (publishedVersion, cc-by)
- Preprints.org (acceptedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Congenital Heart Disease Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- noninvasive hemodynamic monitoring, neonates, neonatal cardiac surgery, postoperative care, circulatory failure, near-infrared spectroscopy, electrical biosensing, multimodal monitoring, regional perfusion, oxygen delivery, lactate, mixed venous oxygen saturation, capillary refill time, urine output, echocardiography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review. https://app.cochat.ai/papers/10.3390/life15101621
- The paper
- Nedelcu, C., Ionescu, N. S., Bizubac, A. M., Filip, C., & Cirstoveanu, C. (2025). Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review. Life. https://doi.org/10.3390/life15101621
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.