Cardiac output monitoring: an integrative perspective
Open accessThe DOI matches its Crossref record; 41 of the 41 references it cites were found in a registry or an index.
Abstract
Cardiac output monitoring in the critically ill patient is standard practice in order to ensure tissue oxygenation [ 1 ] and has been traditionally accomplished using the pulmonary artery catheter (PAC). In recent years, however, the value of PAC has been questioned with some suggesting that its use might not only be unnecessary but also potentially harmful [ 1 ]. This notion, together with the availability of new less invasive cardiac output measuring devices, has markedly decreased the widespread use of the PAC [ 2 ]. Today, various devices are available to measure or estimate cardiac output using different methods. Some of these less invasive devices track stroke volume (SV) continuously and provide dynamic indices of fluid responsiveness, others allow assessment of volumetric preload variables, and some also provide continuous measurement of central venous saturation via the use of proprietary catheters that are attached to the same monitor. All these variables - together with cardiac output - may result in an improved hemodynamic assessment of the critically ill patient. However, it is important to appreciate that each device has its inherent limitations and that no cardiac output monitoring device can change patient outcome unless its use is coupled with an intervention that by itself has been associated with improved patient outcomes. Therefore, the concept of hemodynamic optimization is increasingly recognized as a cornerstone in the management of critically ill patients and has been shown to be associated with improved outcome in the perioperative [ 3 ] and in the intensive care unit (ICU) [ 4 ] setting.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
2 retracted
Boldt, J., Lenz, M., Kumle, B., et al. (1998) · Intensive Care Medicine · doi:10.1007/s001340050536 · Cited by 148 · cited 1×
“Although the majority of physicians use CVP in order to guide fl uid therapy [34], several studies have shown lack of correlation between CVP and SV [35,36].”
BackgroundRetractedMayer, J., Boldt, J., Poland, R., Peterson, A. J., & Manecke, G. R. (2009) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2009.03.003 · Cited by 142 · cited 1×
“This has resulted in an improved performance primarily in perioperative setting [15, 16].”
BackgroundRetractedEarly goal-directed therapy in the treatment of severe sepsis and septic shock
Rivers, E., Nguyen, B., Havstad, S., Ressler, J., Muzzin, A., Knoblich, B., Peterson, E., & Tomlanovich, M. (2001) · New England Journal of Medicine · doi:10.1056/nejmoa010307 · Cited by 10,743 · cited 2×
“As far as its clinical utility is concerned, ScvO2 has been used as a resuscitation endpoint in patients with severe sepsis and septic shock [43].”
BackgroundMethodThe Effectiveness of Right Heart Catheterization in the Initial Care of Critically III Patients
Connors, A. F. (1996) · JAMA: The Journal of the American Medical Association · doi:10.1001/jama.1996.03540110043030 · Cited by 2,210
Marik, P. E., Baram, M., & Vahid, B. (2008) · Chest · doi:10.1378/chest.07-2331 · Cited by 1,516 · cited 1×
“The different functional hemodynamic variables have been shown to be able to predict fluid responsiveness in various studies [39], whereas static preload variables have not [40].”
BackgroundMarik, P. E., Cavallazzi, R., Vasu, T., & Hirani, A. (2009) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181a590da · Cited by 1,272 · cited 1×
“The different functional hemodynamic variables have been shown to be able to predict fluid responsiveness in various studies [39], whereas static preload variables have not [40].”
BackgroundHarvey, S., Harrison, D. A., Singer, M., Ashcroft, J., Jones, C. M., Elbourne, D., Brampton, W., Williams, D., Young, D., & Rowan, K. (2005) · The Lancet · doi:10.1016/s0140-6736(05)67061-4 · Cited by 1,126
Kumar, A., Anel, R., Bunnell, E., Habet, K., Zanotti, S., Marshall, S., Neumann, A., Ali, A., Cheang, M., Kavinsky, C., et al. (2004) · Critical Care Medicine · doi:10.1097/01.ccm.0000114996.68110.c9 · Cited by 800 · cited 1×
“Although the majority of physicians use CVP in order to guide fluid therapy [34], several studies have shown lack of correlation between CVP and SV [35, 36].”
BackgroundPearse, R., Dawson, D., Fawcett, J., Rhodes, A., Grounds, R. M., & Bennett, E. D. (2005) · Critical Care · doi:10.1186/cc3887 · Cited by 698 · cited 1×
“Th e LiDCOTMplus system, in combination with a hemodynamic treatment protocol (targeting an oxygen delivery > 600 ml/min/m2, was shown to be associated with reduced complications and length of hospital stay in patients after major general surgery [14].”
BackgroundPrognostic Value of Extravascular Lung Water in Critically Ill Patients
Sakka, S. G., Klein, M., Reinhart, K., & Meier‐Hellmann, A. (2002) · CHEST Journal · doi:10.1378/chest.122.6.2080 · Cited by 672 · cited 1×
“EVLW on the other hand can be used to diff erentiate between cardiac versus non-cardiac pulmonary edema, and has been identifi ed as an independent predictor of survival in critically ill patients [38].”
BackgroundNoninvasive cardiac output monitoring (NICOM): a clinical validation
Squara, P., Denjean, D., Estagnasie, P., Brusset, A., Dib, J. C., & Dubois, C. (2007) · Intensive Care Medicine · doi:10.1007/s00134-007-0640-0 · Cited by 320 · cited 2×
“In fact, initial validation studies reveal promising results [32,33].”
BackgroundMethodMayer, J., Boldt, J., Mengistu, A., Röhm, K. D., & Suttner, S. W. (2010) · Critical Care · doi:10.1186/cc8875 · Cited by 320 · cited 1×
“Nevertheless, a study using the FlotracTM/VigileoTM system for intraoperative hemodynamic optimization recently demonstrated a decreased complication rate and a reduced length of hospital stay [18].”
BackgroundMonnet, X., & Teboul, J.-L. (2008) · Intensive Care Medicine · doi:10.1007/s00134-008-0994-y · Cited by 317 · cited 1×
“Th is technique has been demonstrated to reliably determine fl uid responsiveness in critically ill patients [41].”
MethodErrors in the measurement of cardiac output by thermodilution
Nishikawa, T., & Dohi, S. (1993) · Canadian Journal of Anaesthesia · doi:10.1007/bf03011312 · Cited by 302 · cited 1×
“, mitral or tricuspid valve insuffi ciency, shunt) or misplacement of the PAC may infl uence reliable cardiac output assessment [6].”
BackgroundAbbas, S. M., & Hill, A. G. (2008) · Anaesthesia · doi:10.1111/j.1365-2044.2007.05233.x · Cited by 296 · cited 1×
“Despite some limitations of esophageal Doppler devices, their utility appears to be confirmed by several perioperative hemodynamic optimization studies that have consistently demonstrated a reduction in complication rates and hospital length of stay [24].”
BackgroundGoal-directed fluid management reduces vasopressor and catecholamine use in cardiac surgery patients
Goepfert, M. S. G., Reuter, D. A., Akyol, D., Lamm, P., Kilger, E., & Goetz, A. E. (2007) · Intensive Care Medicine · doi:10.1007/s00134-006-0404-2 · Cited by 254 · cited 1×
“GEDV could thus be used to better guide perioperative fluid therapy than pressure preload parameters [37].”
MethodNoninvasive Monitoring of Cardiac Output in Critically Ill Patients Using Transesophageal Doppler
Valtier, B., Cholley, B., Belot, J.-P., Coussaye, J.-E. D. L., Matéo, J., & Payen, D. M. (1998) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.158.1.9707031 · Cited by 235 · cited 1×
“Since probe position is crucial to obtaining an accurate measurement of aortic blood fl ow, this device is operator-dependent and studies have shown that 10– 12 insertions are required to obtain accurate measurements [22] with an intra- and inter-observer variability of 8–12% [23].”
BackgroundMulticenter Evaluation of Noninvasive Cardiac Output Measurement by Bioreactance Technique
Raval, N. Y., Squara, P., Cleman, M., Yalamanchili, K., Winklmaier, M., & Burkhoff, D. (2008) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-008-9112-5 · Cited by 229 · cited 1×
“In fact, initial validation studies reveal promising results [32,33].”
BackgroundDueck, M. H., Klimek, M., Appenrodt, S., Weigand, C., & Boerner, U. (2005) · Anesthesiology · doi:10.1097/00000542-200508000-00007 · Cited by 216 · cited 1×
“It is important to realize that absolute ScvO2 and SvO2 values may diff er considerably in diff erent clinical situations; however, a strong correlation of their trends over time has been demonstrated [44].”
BackgroundGoodacre, S., Sampson, F., Stevenson, M., Wailoo, A., Sutton, A., Thomas, S., Locker, T., & Ryan, A. (2006) · Health Technology Assessment · doi:10.3310/hta10150 · Cited by 180
Bloos, F., & Reinhart, K. (2005) · Intensive Care Medicine · doi:10.1007/s00134-005-2670-9 · Cited by 179
Hamzaoui, O., Monnet, X., Richard, C., Osman, D., Chemla, D., & Teboul, J.-L. (2008) · Critical Care Medicine · doi:10.1097/01.ccm.ob013e318161fec4 · Cited by 179 · cited 1×
“However, during situations of hemodynamic instability, calibration needs to be done more frequently (eventually every hour) [9].”
BackgroundButton, D., Weibel, L., Reuthebuch, O., Genoni, M., Zollinger, A., & Hofer, C. K. (2007) · British Journal of Anaesthesia · doi:10.1093/bja/aem188 · Cited by 176 · cited 1×
“Nevertheless, a variety of studies have successfully validated the PiCCOplusTM system in different patient populations [10, 11].”
BackgroundSelf-citationLefrant, J. Y., Bruelle, P., Aya, A. G. M., Saïssi, G., Dauzat, M., de La Coussaye, J. E., & Eledjam, J. J. (1998) · Intensive Care Medicine · doi:10.1007/s001340050578 · Cited by 176 · cited 1×
“Since probe position is crucial to obtaining an accurate measurement of aortic blood fl ow, this device is operator-dependent and studies have shown that 10– 12 insertions are required to obtain accurate measurements [22] with an intra- and inter-observer variability of 8–12% [23].”
BackgroundHofer, C. K., Furrer, L., Matter-Ensner, S., Maloigne, M., Klaghofer, R., Genoni, M., & Zollinger, A. (2005) · British Journal of Anaesthesia · doi:10.1093/bja/aei123 · Cited by 146 · cited 2×
“Although the majority of physicians use CVP in order to guide fl uid therapy [34], several studies have shown lack of correlation between CVP and SV [35,36].”
BackgroundSelf-citationRocca, G. D., Costa, M. G., Coccia, C., Pompei, L., Marco, P. D., Vilardi, V., & Pietropaoli, P. (2003) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03018714 · Cited by 140 · cited 1×
“Nevertheless, a variety of studies have successfully validated the PiCCOplusTM system in different patient populations [10, 11].”
BackgroundReinhart, K., & Bloos, F. (2005) · Current Opinion in Critical Care · doi:10.1097/01.ccx.0000158092.64795.cf · Cited by 132
Clinical review: Goal-directed therapy in high risk surgical patients
Lees, N., Hamilton, M. A., & Rhodes, A. (2009) · Critical Care · doi:10.1186/cc8039 · Cited by 129 · cited 1×
“Th erefore, the concept of hemodynamic optimization is increasingly recognized as a cornerstone in the management of critically ill patients and has been shown to be associated with improved outcome in the perioperative [3] and in the intensive care unit (ICU) [4] setting.”
BackgroundStover, J., Stocker, R. A., Lenherr, R., Neff, T. A., Cottini, S. R., Zoller, B., & Béchir, M. (2009) · BMC Anesthesiology · doi:10.1186/1471-2253-9-6 · Cited by 112 · cited 1×
“Regarding validation of the device, only limited published data are available [21].”
BackgroundScolletta, S., Romano, S. M., Biagioli, B., Capannini, G., & Giomarelli, P. (2005) · British Journal of Anaesthesia · doi:10.1093/bja/aei154 · Cited by 108 · cited 1×
“PRAM has been validated so far in a porcine model under various hemodynamic states [19] and in humans undergoing cardiac surgery [20].”
BackgroundCecconi, M., Dawson, D., Grounds, R. M., & Rhodes, A. (2008) · Intensive Care Medicine · doi:10.1007/s00134-008-1292-4 · Cited by 99 · cited 1×
“Clinical studies have demonstrated reliable estimation of cardiac output using PulseCO as long as no major hemodynamic changes are observed [13].”
MethodSelf-citationFunk, D., Sebat, F., & Kumar, A. (2009) · Current Opinion in Critical Care · doi:10.1097/mcc.0b013e32832e3825 · Cited by 92 · cited 2×
“Its use may be expanded in the perioperative and ICU setting.”
BackgroundCardiac output monitoring by pressure recording analytical method in cardiac surgery
GIOMARELLI, P. P., Biagioli, B., & Scolletta, S. (2004) · European Journal of Cardio-Thoracic Surgery · doi:10.1016/j.ejcts.2004.05.025 · Cited by 87 · cited 1×
“PRAM has been validated so far in a porcine model under various hemodynamic states [19] and in humans undergoing cardiac surgery [20].”
BackgroundHarvey, S. E., Stevens, K. J., Harrison, D. A., Young, D., Brampton, W., McCabe, C. J., Singer, M., & Rowan, K. (2006) · Health Technology Assessment · doi:10.3310/hta10290 · Cited by 81 · cited 1×
“This notion, together with the availability of new less invasive cardiac output measuring devices, has markedly decreased the widespread use of the PAC [2].”
BackgroundComparison of a supra-sternal cardiac output monitor (USCOM) with the pulmonary artery catheter
Thom, O., Taylor, D. M., Wolfe, R. E., Cade, J. F., Myles, P. S., Krum, H., & Wolfe, R. S. (2009) · British Journal of Anaesthesia · doi:10.1093/bja/aep296 · Cited by 73 · cited 1×
“Validation studies have revealed conflicting results, which could be explained primarily by the inherent problem of variable signal detection [25, 26].”
BackgroundGujjar, A. R., Muralidhar, K., Banakal, S. K., Gupta, R., Sathyaprabha, T. N., & Jairaj, P. S. (2008) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-008-9119-y · Cited by 68 · cited 1×
“Despite many adjustments of the mathematical algorithms, clinical validation studies continue to show conflicting results [30, 31].”
Pitfalls in haemodynamic monitoring based on the arterial pressure waveform
Camporota, L., & Beale, R. J. (2010) · Critical Care · doi:10.1186/cc8845 · Cited by 60 · cited 1×
“However, accuracy of the device during rapid hemodynamic changes remains a major concern [17].”
BackgroundCardiac output measurements in off-pump coronary surgery
Guéret, G., Kiss, G., Rossignol, B., Bezon, E., Wargnier, J. P., Miossec, A., Corre, O., & Arvieux, C. C. (2006) · European Journal of Anaesthesiology · doi:10.1017/s0265021506000573 · Cited by 32 · cited 1×
“There are several limitations to this device including the need for intubation and mechanical ventilation with fixed ventilator settings and minimal gas exchange abnormalities [27].”
BackgroundComparison of the Endotracheal Cardiac Output Monitor to Thermodilution in Cardiac Surgery Patients
Ball, T. R., Culp, B. C., Patel, V. B., et al. (2010) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2010.04.008 · Cited by 31 · cited 1×
“Despite many adjustments of the mathematical algorithms, clinical validation studies continue to show confl icting results [30,31].”
BackgroundChan, J. S., Segara, D., & Nair, P. (2006) · Critical Care and Resuscitation · doi:10.1016/s1441-2772(23)02064-1 · Cited by 30 · cited 1×
“Validation studies have revealed confl icting results, which could be explained primarily by the inherent problem of variable signal detection [25,26].”
BackgroundEvaluation of a new continuous cardiac output monitor in off‐pump coronary artery surgery
Leather, H. A., Vuylsteke, A., Bert, C., M'Fam, W., Segers, P., Sergeant, P., Vandermeersch, E., & Wouters, P. F. (2004) · Anaesthesia · doi:10.1111/j.1365-2044.2004.03668.x · Cited by 28 · cited 1×
“A so-called fast response continuous cardiac output catheter (truCCOMSTM, Omega Critical Care, East Klibride, GB) allows a more synchronized continuous cardiac output monitoring [8].”
Background
Cited by
Consensus guidelines for enhanced recovery after gastrectomy
Mortensen, K., Nilsson, M., Slim, K., Schäfer, M., Mariette, C., Braga, M., Carli, F., Demartines, N., Griffin, S. M., Lassen, K., et al. (2014) · British Journal of Surgery · doi:10.1002/bjs.9582 · Cited by 767
Metadata
- Authors
- Alhashemi, J. A., Cecconi, M., & Hofer, C. K.
- Venue
- Critical Care
- Published
- 2011
- Volume
- 15
- Issue
- 2
- Pages
- 214
- ISSN
- 1364-8535 (ISSN-L)
- OpenAlex
- W2031965614
- Cited by
- 202
- References
- 41
- FWCI
- 19.29
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.