Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
BACKGROUND: This paper describes the National Electrical Manufacturers Association (NEMA) system performance of the Discovery MI 3-ring PET/CT (GE Healthcare) installed in Bruges, Belgium. This time-of-flight (TOF) PET camera is based on silicon photomultipliers instead of photomultiplier tubes. METHODS: The NEMA NU2-2012 standard was used to evaluate spatial resolution, sensitivity, image quality (IQ) and count rate curves of the system. Timing and energy resolution were determined. RESULTS: Full width at half maximum (FWHM) of spatial resolution in radial, tangential and axial direction was 4.69, 4.08 and 4.68 mm at 1 cm; 5.58, 4.64 and 5.83 mm at 10 cm; and 7.53, 5.08 and 5.47 mm at 20 cm from the centre of the field of view (FOV) for the filtered backprojection reconstruction. For non-TOF ordered subset expectation maximization (OSEM) reconstruction without point spread function (PSF) correction, FWHM was 3.87, 3.69 and 4.15 mm at 1 cm; 4.80, 3.81 and 4.87 mm at 10 cm; and 7.38, 4.16 and 3.98 mm at 20 cm. Sensitivity was 7.258 cps/kBq at the centre of the FOV and 7.117 cps/kBq at 10-cm radial offset. Contrast recovery (CR) using the IQ phantom for the TOF OSEM reconstruction without PSF correction was 47.4, 59.3, 67.0 and 77.0% for the 10-, 13-, 17- and 22-mm radioactive spheres and 82.5 and 85.1% for the 28- and 37-mm non-radioactive spheres. Background variability (BV) was 16.4, 12.1, 9.1, 6.6, 5.1 and 3.8% for the 10-, 13-, 17-, 22-, 28- and 37-mm spheres. Lung error was 8.5%. Peak noise equivalent count rate (NECR) was 102.3 kcps at 23.0 kBq/ml with a scatter fraction of 41.2%. Maximum accuracy error was 3.88%. Coincidence timing resolution was 375.6 ps FWHM. Energy resolution was 9.3% FWHM. Q.Clear reconstruction significantly improved CR and reduced BV compared with OSEM. CONCLUSION: System sensitivity and NECR are lower and IQ phantom's BV is higher compared with larger axial FOV (AFOV) scanners like the 4-ring discovery MI, as expected from the smaller solid angle of the 3-ring system. The other NEMA performance parameters are all comparable with those of the larger AFOV scanners.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Teoh, E. J., McGowan, D. R., Macpherson, R. E., Bradley, K. M., & Gleeson, F. V. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.159301 · Cited by 265
Studies of a Next-Generation Silicon-Photomultiplier–Based Time-of-Flight PET/CT System
Hsu, D. F. C., Ilan, E., Peterson, W. T., Uribe, J., Lubberink, M., & Levin, C. S. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.117.189514 · Cited by 241
Self-citationRausch, I., Cal-González, J., Dapra, D., Gallowitsch, H. J., Lind, P., Beyer, T., & Minear, G. (2015) · EJNMMI Physics · doi:10.1186/s40658-015-0132-1 · Cited by 133
Teoh, E. J., McGowan, D. R., Bradley, K. M., Belcher, E., Black, E., & Gleeson, F. V. (2016) · European Radiology · doi:10.1007/s00330-015-3832-y · Cited by 109
Initial experience with a SiPM-based PET/CT scanner: influence of acquisition time on image quality
Sonni, I., Baratto, L., Park, S., Hatami, N., Srinivas, S., Davidzon, G., Gambhir, S. S., & Iagaru, A. (2018) · EJNMMI Physics · doi:10.1186/s40658-018-0207-x · Cited by 74
Messerli, M., Stolzmann, P., Egger-Sigg, M., Trinckauf, J., D’Aguanno, S., Burger, I. A., von Schulthess, G. K., Kaufmann, P. A., & Huellner, M. W. (2018) · EJNMMI Physics · doi:10.1186/s40658-018-0223-x · Cited by 73
Cited by
Rogasch, J. M., Suleiman, S., Hofheinz, F., Bluemel, S., Lukas, M., Amthauer, H., & Furth, C. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-0270-y · Cited by 63
te Riet, J., Rijnsdorp, S., Roef, M. J., & Arends, A. J. (2019) · EJNMMI Physics · doi:10.1186/s40658-019-0262-y · Cited by 46
“Scans were performed on a TOF 3-ring Discovery 710 PET/CT scanner with lutetium yttrium orthosilicate (LYSO) crystals (GE Healthcare), see [21] for an overview of the performance of the system.”
MethodMoving the goalposts while scoring―the dilemma posed by new PET technologies
Rogasch, J. M. M., Boellaard, R., Pike, L., Borchmann, P., Johnson, P., Wolf, J., Barrington, S. F., & Kobe, C. (2021) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05403-2 · Cited by 20
Oddstig, J., Brolin, G., Trägårdh, E., & Minarik, D. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00366-7 · Cited by 10
“9 cps/MBq) performed on the three-ring system with analogue PMtubes used in the present study [12].”
Method
Metadata
- Authors
- Vandendriessche, D., Uribe, J., Bertin, H., & De Geeter, F.
- Venue
- EJNMMI Physics
- Published
- 2019-05-10
- Volume
- 6
- Issue
- 1
- Pages
- 8-8
- ISSN
- 2197-7364 (ISSN-L)
- OpenAlex
- W2947137705
- Cited by
- 38
- References
- 6
- FWCI
- 4.35
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.