Adverse Events in Robotic Surgery: A Retrospective Study of 14 Years of FDA Data
Open accessThe DOI matches its Crossref record; 52 of the 53 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Use of robotic systems for minimally invasive surgery has rapidly increased during the last decade. Understanding the causes of adverse events and their impact on patients in robot-assisted surgery will help improve systems and operational practices to avoid incidents in the future. METHODS: By developing an automated natural language processing tool, we performed a comprehensive analysis of the adverse events reported to the publicly available MAUDE database (maintained by the U.S. Food and Drug Administration) from 2000 to 2013. We determined the number of events reported per procedure and per surgical specialty, the most common types of device malfunctions and their impact on patients, and the potential causes for catastrophic events such as patient injuries and deaths. RESULTS: During the study period, 144 deaths (1.4% of the 10,624 reports), 1,391 patient injuries (13.1%), and 8,061 device malfunctions (75.9%) were reported. The numbers of injury and death events per procedure have stayed relatively constant (mean = 83.4, 95% confidence interval (CI), 74.2-92.7 per 100,000 procedures) over the years. Surgical specialties for which robots are extensively used, such as gynecology and urology, had lower numbers of injuries, deaths, and conversions per procedure than more complex surgeries, such as cardiothoracic and head and neck (106.3 vs. 232.9 per 100,000 procedures, Risk Ratio = 2.2, 95% CI, 1.9-2.6). Device and instrument malfunctions, such as falling of burnt/broken pieces of instruments into the patient (14.7%), electrical arcing of instruments (10.5%), unintended operation of instruments (8.6%), system errors (5%), and video/imaging problems (2.6%), constituted a major part of the reports. Device malfunctions impacted patients in terms of injuries or procedure interruptions. In 1,104 (10.4%) of all the events, the procedure was interrupted to restart the system (3.1%), to convert the procedure to non-robotic techniques (7.3%), or to reschedule it (2.5%). CONCLUSIONS: Despite widespread adoption of robotic systems for minimally invasive surgery in the U.S., a non-negligible number of technical difficulties and complications are still being experienced during procedures. Adoption of advanced techniques in design and operation of robotic surgical systems and enhanced mechanisms for adverse event reporting may reduce these preventable incidents in the future.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures

![Fig 2 Data extraction and analysis flow from the FDA MAUDE database. The dictionaries of keywords for surgery types/specialties and surgical instruments were constructed based on the online information available on the da Vinci surgeries [32] and instruments catalog [33] from the manufacturer. The analysis tool was developed using open-source Python libraries for natural language processing, data analysis, and machine learning.](https://figures.cochat.ai/paper-figures/52d4daa60f2347c8938ae62f07d16891/adverse-events-in-robotic-surgery-a-retrospective-study-of-14-years-of-fda-data-figure-2.jpg)


Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“On health care despite the known advantages of using robots in surgeries, medical robots were reported to have a negative impact on patients’ lives due to inaccuracy mistakes and errors [210], or due to cyber-attacks (i.e. data exposure/leakage) such as the case of North Korea-Unit 180 (Lazarus) attack on UK’s National Health Service (NHS) in 2017 [211, 212].”
From Robotics cyber security: vulnerabilities, attacks, countermeasures, and recommendations (2022) · cites this paper as [210]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Foundations of statistical natural language processing
Manning, C. D., & Schütze, H. (1999) · Cited by 9,816
Corke, P., Roberts, J., Cunningham, J., & Hainsworth, D. (2008) · Springer Handbook of Robotics · doi:10.1007/978-3-540-30301-5_50 · Cited by 3,543
Sample Size Tables for Clinical Studies
Machin, D., Campbell, M. J., Tan, S. B., & Tan, S. H. (2008) · Wiley eBooks · doi:10.1002/9781444300710 · Cited by 1,336
New Technology and Health Care Costs — The Case of Robot-Assisted Surgery
Barbash, G. I., & Glied, S. A. (2010) · New England Journal of Medicine · doi:10.1056/nejmp1006602 · Cited by 993
Minimally invasive mitral valve surgery: a systematic review and meta-analysis
Modi, P., Hassan, A., & Chitwood, W. R. (2008) · European Journal of Cardio-Thoracic Surgery · doi:10.1016/j.ejcts.2008.07.057 · Cited by 503
An open-source research kit for the da Vinci® Surgical System
Kazanzides, P., Chen, Z., Deguet, A., Fischer, G. S., Taylor, R. H., & DiMaio, S. P. (2014) · 2014 IEEE International Conference on Robotics and Automation (ICRA) · doi:10.1109/icra.2014.6907809 · Cited by 493
Raven-II: An Open Platform for Surgical Robotics Research
Hannaford, B., Rosen, J., Friedman, D. W., King, H., Roan, P., Cheng, L., Glozman, D., Ma, J., Kosari, S. N., & White, L. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2228858 · Cited by 372
Holzhey, D. M., Seeburger, J., Misfeld, M., Borger, M. A., & Mohr, F. W. (2013) · Circulation · doi:10.1161/circulationaha.112.001402 · Cited by 302
Hauser, R. G., Katsiyiannis, W. T., Gornick, C. C., Almquist, A. K., & Kallinen, L. M. (2010) · Europace · doi:10.1093/europace/eup375 · Cited by 264
The RAVEN: Design and Validation of a Telesurgery System
Lum, M. J. H., Friedman, D. C. W., Sankaranarayanan, G., King, H., Fodero, K., Leuschke, R., Hannaford, B., Rosen, J., & Sinanan, M. N. (2009) · The International Journal of Robotics Research · doi:10.1177/0278364909101795 · Cited by 264
Research works printed but not linked: 3
- L Eichel, 2005, Robotics in Urologic Surgery: Risks and Benefits, AUA Update Series, 24, 106
- Homa Alemzadeh, 2015, A Software Framework for Simulation of Safety Hazards in Robotic Surgical Systems, SIGBED Review (Special Issue on Medical Cyber Physical Systems Workshop), 12
- Russell H. Taylor, 2008, Springer handbook of robotics, 1199
Cited by
Items 401–464 of 464
Shenoy, R. S., Mederos, M. A., Ye, L., Mak, S. S., Begashaw, M. M., Booth, M. S., Shekelle, P. G., Wilson, M., Gunnar, W., Maggard-Gibbons, M. A., et al. (2021) · Figshare · doi:10.6084/m9.figshare.14478549
Shenoy, R. S., Mederos, M. A., Ye, L., Mak, S. S., Begashaw, M. M., Booth, M. S., Shekelle, P. G., Wilson, M., Gunnar, W., Maggard-Gibbons, M. A., et al. (2021) · Figshare · doi:10.6084/m9.figshare.14478555
Shenoy, R. S., Mederos, M. A., Ye, L., Mak, S. S., Begashaw, M. M., Booth, M. S., Shekelle, P. G., Wilson, M., Gunnar, W., Maggard-Gibbons, M. A., et al. (2021) · Figshare · doi:10.6084/m9.figshare.14478546
Shenoy, R. S., Mederos, M. A., Ye, L., Mak, S. S., Begashaw, M. M., Booth, M. S., Shekelle, P. G., Wilson, M., Gunnar, W., Maggard-Gibbons, M. A., et al. (2021) · Figshare · doi:10.6084/m9.figshare.14478561
Transparency, trust, and level of detail in user interface design for human autonomy teaming
Wang, T. (2023) · VTechWorks (Virginia Tech) · Cited by 3
Zwitterionic materials for biomedical applications
Cabanach Xifró, P. (2021) · Tesis Doctorals en Xarxa (Consorci de Serveis Universitaris de Catalunya) · Cited by 1
Assessing dependability of emergent large-scale autonomous systems in the wild
Jha, S. (2021) · Illinois Digital Environment for Access to Learning and Scholarship (University of Illinois at Urbana-Champaign)
Chung, S. A. (2023) · WakeSpace Scholarship (Wake Forest University)
Robotic Cholecystectomy versus Laparoscopic Cholecystectomy: A Comparative Study
Maloth, D. J., Vedire, N., Kumar, D. S., & Sachin, G. R. (2024) · International Journal of Advanced Robotic & Innovative Surgery · doi:10.4103/jris.jris_14_24
A Systematic Review of Surgical Robots and Controls for Teleoperations
Assan, K. N., Wayoe, P. T., Ayiku, C. N., Aku, A. D., & Quagraine, H. (2025) · Aviation Electronics Information Technology Telecommunications Electricals Controls (AVITEC) · doi:10.28989/avitec.v7i3.2941
“Communication barriers and resources also remains a challenge [74].”
Robotic health care is coming to a hospital near you
Milner, M. N., & Rice, S. (2019) · The Conversation · doi:10.64628/aai.xpuqm9vkd
Robótica e IA na prestação de cuidados médicos: desafios à responsabilidade civil
Couto, H. A. F. (2025) · Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT)
Tibial plateau plates - A national analysis of reported device and patient complications
Ahmad, M., Nassif, M., Strickler, I., Jensen, A., Al-Marrawi, Y., & Cavinatto, L. (2025) · Journal of Orthopaedic Reports · doi:10.1016/j.jorep.2025.100856
Debating artificial intelligence in patients’ care: A double-edged sword
Harth, T. A., Fannah, J. A., Dahri, R. A., Adawi, M. M. A., Araimi, R. A., Al‐Shehhi, W., Hashmi, A., Nabhani, M. S. A., Hinaai, S. A. A., & Harthi, H. A. (2025) · World Advances in Renal Medicine · doi:10.25259/warm_18_2025
A Stacking Ensemble Framework for Robust Detection of Data Poisoning Attacks
Alsaban, Y. B., & Fkih, F. (2026) · WSEAS Transactions on Signal Processing archive · doi:10.37394/232014.2026.22.1
Retention of a Detached Robotic Scissors Tip Cover in the Abdominal Wall: A Case Report
Kojima, M., Miyake, T., Tani, S., Muramoto, K., Nishina, Y., Kaida, S., Takebayashi, K., Maehira, H., Otake, R., Mori, H., et al. (2026) · Asian Journal of Endoscopic Surgery · doi:10.1111/ases.70242
When the action is “hybrid”–ethical challenges of the emerging technologies in the operating room
Valera, L., Irarrázaval, M. J., & Gabrielli, M. E. (2026) · Humanities and Social Sciences Communications · doi:10.1057/s41599-025-06455-7
“Despite the possibility of adverse events occurring during RAS due to multiple causes (e.g., Alemzadeh et al., 2016), it is of paramount importance to have emergency protocols for these situations.”
Verma, A. K., Pal, B., Sanfui, S., Mondal, A., Malek, R. R., Singh, R. S., & Reddy, G. H. V. (2026) · Cureus · doi:10.7759/cureus.104158
“In the same regard, predictive analytics of bleeding or anastomotic leak risks are also limited by data bias, poor generalization, regulatory restrictions, and the assimilation of the operating room setting [12].”
RetractedUzun, A. U., Demirok, H. M., Yıldız, Z. A., Coşkun, Ö. D., Erdeve, B., Serttürk, F., Cin, E. S., Beğer, B., Sarar, G. İ., Öcal, M. E., et al. (2026) · Akademisyen Kitabevi eBooks · doi:10.37609/akya.3878
Robot‐Assisted Ileocecal Resection Using the Saroa Surgical System
Maeda, H., Yamaguchi, S., Maeda, M., Marui, A., Kawanishi, Y., Fujisawa, K., Kitagawa, H., Namikawa, T., Oki, Y., Miyachi, H., et al. (2026) · Asian Journal of Endoscopic Surgery · doi:10.1111/ases.70270
“However, robot-specific intraoperative complications, including accidental organ injury and mechanical malfunction, have been re-ported [2], highlighting the need to improve the robot–human interface, instrument robustness, and surgeon training.”
Versius from CMR—The next generation robotic surgical system
Collins, J. W., Morton, J. R., & Slack, M. C. (2026) · Elsevier eBooks · doi:10.1016/b978-0-443-21733-3.00016-x
Patel, Y., Shyu, E., Shahi, N., Kaplan, B. J., Taylor, J. S., & Damani, T. (2026) · Surgical Endoscopy · doi:10.1007/s00464-026-12653-4
“Despite these trends, operative time was approximately 18 minutes longer for robotic PEH repair before and after matching, aligning with previous studies, which attribute the increase to robotic setup and docking time [33–36].”
Ebihara, Y., Kyogoku, N., Takano, H., Wada, H., Nitta, T., Saikawa, D., Yamamura, Y., Takada, M., Shichinohe, T., & Hirano, S. (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-03295-8
When seconds matter – introducing a surgical salvage strategy by emergency robotic surgery: a review
Büdeyri, I., Ramspott, J. P., Juratli, M. A., Pascher, A., & Hölzen, J. P. (2026) · Patient Safety in Surgery · doi:10.1186/s13037-026-00483-1
“…problems, non-recoverable failures of the robotic system itself, such as persistent system errors, loss of console–cart communication, or arm malfunction that cannot be resolved intraoperatively, may require conversion to an alternative approach or termination of the procedure [21, 22].”
InfluentialVithya, T., Bollineni, S., Tasleem, F., Ganachari, P., Khan, S. A., & Prasad, S. (2026) · The Indian Journal of Medical Research · doi:10.25259/ijmr_1988_2025
“13 Even devices generally regarded as low risk, like Class B, were occasionally linked to serious issues, as shown by Ferner et al 14 and Deshwal et al .”
Transfer learning for abusive language detection
Bose, T. (2023) · theses.fr (ABES) · doi:10.70675/6b663c61zb592z4422zbc5fza68cbc285104 · Cited by 1
Ota, K., Ota, Y., Tasaka, K., Shiota, M., & Shimoya, K. (2026) · Asian Journal of Endoscopic Surgery · doi:10.1111/ases.70291
da Vinci robotic-assisted surgery: implications for modern anesthetic practice
Suria, S., Caballero, M.-J., Ouvrard, J., Lecanu, N., & Elmawieh, J. (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-03431-4
Parelkar, S., Malviya, S. J., Sanghvi, B., Gupta, R. K., Mudkhedkar, K. P., Shah, R., & Makhija, D. P. (2026) · Journal of Indian Association of Pediatric Surgeons · doi:10.4103/jiaps.jiaps_360_25
“[4] Other studies have proposed algorithms for lost needles; however, these protocols are designed for MAS in nonurologic, adult cases.”
The Criminal Liability Of Hinotori: The Japanese Surgical Robotic Doctor
Bal, E. N., & Abualrob, M. (2026) · Journal of Penal Law and Criminology / Ceza Hukuku ve Kriminoloji Dergisi · doi:10.26650/jplc2026-1711852
Tanaka, K., Nakajima, K., Masuda, S., Yamasaki, M., Momose, K., Yamashita, K., Makino, T., Saitoh, T., Takahashi, T., Kurokawa, Y., et al. (2026) · Innovative Surgical Trends · doi:10.1007/s44414-026-00021-7
Nechay, T. V., Игоревич, П. С., Yuldashev, A. G., Tolstopyatov, S. E., Петрович, П. М., & Sazhin, A. V. (2026) · Pirogov Russian Journal of Surgery · doi:10.17116/hirurgia202605191
Obana, R., & TAMAGAWA, M. (2026) · Journal of Advanced Research in Fluid Mechanics and Thermal Sciences · doi:10.37934/arfmts.139.1.111
Amarasinghe, R., Schott, C. A., Inuwa, A., Khalili, M., & Ng, M. (2026) · Cureus · doi:10.7759/cureus.110242
“One minor, yet well-known, complication with robotic port placement includes burning of the tissue surrounding the site [11].”
Blavette, L., Dacunha, S., Bondarenko, A., Piccoli, M., Rigaud, A., & Pino, M. (2026) · Frontiers in Digital Health · doi:10.3389/fdgth.2026.1802396
“From a safety and accountability perspective, automated systems introduce new modes of failure and responsibility that challenge existing mechanisms for monitoring system behavior, managing risk, and responding to adverse events, particularly when AI is involved (10, 23).”
Real-Time Multimodal Activity-Aware Error Detection in Robot-Assisted Surgery
Roodabeh, S. H. R., Li, Z., & Alemzadeh, H. (2026) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2026.3726373
“However, significant challenges remain in surgeon training and skill evaluation to ensure patient safety [1], [2].”
Self-citationPossard, M., & Bleichert, S. (2026) · Journal of Digital Health and Ethics · doi:10.66128/jdhe202602.2
Robotic and Laser-Assisted Techniques in Urologic Surgery: Current Applications
Rabcewicz, C. (2025) · MedCaseReports · doi:10.67417/medcasereports.129
Kamimura, G., Aoki, M., Tokunaga, T., Takeda, A., Maeda, K., & Ueda, K. (2026) · General Thoracic and Cardiovascular Surgery · doi:10.1007/s11748-026-02362-z
Wu, B., Chen, Q., Pan, W., & Li, E. (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-03727-5
“Adverse-event analyses based on robotic surgery reports further highlight the importance of system reliability, device malfunction management and team pre-paredness [15].”
Med-SAFE: Postmarket Safety-Actionability Screening for Robotic Medical DevicesPreprint
Bahaddad, A. A. S. (2026) · Research Square · doi:10.21203/rs.3.rs-9913804/v1
Advancing paediatric robotic surgery: building the foundation with databases
Marangoly, C. G., Blanc, T., Gundeti, M. S., Peters, C. A., Kimber, C., & Taghavi, K. (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-03774-y
“Robot-related complications are generally classified into two categories: issues with the robot or console (e.g. software issues, video loss) and robotic arms (e.g. collisions, component malfunctions) [27].”
Failure modes in soft-tissue surgical robots: an adjudicated FDA recall cohort, 2005–2026
Drobot, R. B. (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-03818-3
Robot-assisted appendectomy in children: need, feasibility, domestic experience, and prospects
Lozytskyi, O. O., Gonchar, V. V., Dvorakevych, A. O., & Kalinchuk, О. О. (2026) · Modern pediatrics Ukraine · doi:10.15574/sp.2026.3(155).121126
Risk Scrutiny and Future Opportunities of Robot-Assisted Implant Surgery: A Narrative Review
Wang, G., Lin, Y., Zheng, X., Shi, J., Huang, F., Zhao, Y., Wen, S., Gan, Y., Wang, Q., Kong, J., et al. (2026) · International Dental Journal · doi:10.1016/j.identj.2026.111105
“50,51 Another study involving 3935 robotic surgeries reported a 6.2% instrument failure rate, predominantly due to damage to instrument tips or joints (46.9%), cable failures (30.0%), control module malfunctions (5.3%), and abnormal drive shafts (3.2%).”
InfluentialDiamond, D. A., Moschovas, M. C., Dohler, D., Diamond, A., Patel, E., Gonzalez, J., & Casey, B. (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-03871-y
“…identified 10,624 adverse-event reports associated with the use of robotic systems, including 1,391 patient injuries and 144 deaths, with device malfunctions implicated in roughly three-quarters of these reports [5]; health complications are, moreover, widely understood to be under-reported [6].”
Zhai, Z., Shi, M., Gao, D., & Cao, L. (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-03956-8
KIDA: Kinematic-Intent Dual-Path Alignment for Surgical Error Detection
Luo, Y., Yang, K., Li, Y., Fang, W., Li, Q., & Wang, Z. (2026) · Lecture notes in computer science · doi:10.1007/978-3-032-38233-7_22
Lyu, Z., Koyanagi, K., Masuta, H., Almassri, A. M. M., Li, F., & Sawai, K. (2026) · Robotics · doi:10.3390/robotics15100188
Medical Cyber-Physical Systems
Lee, I., Alemzadeh, H., Mangharam, R., Stankovic, J. A., & Weimer, J. E. (2026) · ACM Transactions on Cyber-Physical Systems · doi:10.1145/3842676
“…such as LLM-assisted hazard analysis [36, 87] grounded in regulatory guidelines, operational protocols, and real-world incident databases [10, 11]; extend formal frameworks to ML-enabled MCPS with end-to-end specifications and dynamic property adjustment under distribution shifts [17, 119];…”
Self-citationLaparoscopic and Robotic Device Malfunction: A Scoping Review
Tsutsumi, A., Bhandarkar, S., Darby, F., Schneider, E. B., Murthy, S. S., Blumberg, D., Brackett, A. L., & Ahuja, V. (2026) · Journal of Patient Safety · doi:10.1097/pts.0000000000001569
SEGURANÇA DO PACIENTE NA CIRURGIA ROBÓTICA: ATUAÇÃO DA ENFERMAGEM PERIOPERATÓRIA
Sousa, E. G., Campos, D. F., & dos Santos, I. A. R. (2026) · RECIMA21 - Revista Científica Multidisciplinar · doi:10.47820/recima21.v7i10.9277
Merlet, A. (2026) · Research Square · doi:10.21203/rs.3.rs-11394165/v1
Glosser, L., Tran, Q., Bettis, T., Maegawa, F. B., & Patel, S. (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-04156-0
Vidyasagar, K. E. C., Karadula, R., Suraram, M., & Saikia, M. J. (2026) · Bioengineering · doi:10.3390/bioengineering13111279
Schwartz, J. H., Almoussa, M. M., Blattman, N., & Makary, M. S. (2026) · Cureus · doi:10.7759/cureus.118938
Martinez, M. (2023) · SAP Gamification and Augmented Reality · doi:10.62486/gr20232
Singh, R., & Kanungo, P. (2026) · International Journal of Law Management & Humanities · doi:10.63108/ijlmh.13157
Chen, Y., Hu, Y., & 林春忠 (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-04159-x
Li, J. Z., Lao, Y., & Liu, X. (2026) · Current Problems in Surgery · doi:10.1016/j.cpsurg.2026.102225
Abushal, M. H. (2026) · Journal of Robotic Surgery · doi:10.1007/s11701-026-04162-2
Olsen, R. G., Stadager, M. W., Togsverd-Bo, M., Konge, L., Røder, M. A., & Bjerrum, F. (2026) · Surgical Endoscopy · doi:10.1007/s00464-026-13486-x
Metadata
- Authors
- Alemzadeh, H., Raman, J., Leveson, N., Kalbarczyk, Z., & Iyer, R. K.
- Affiliations
- University of Illinois Urbana-Champaign (US)
- Massachusetts Institute of Technology (US)
- Corresponding author
- Alemzadeh, Homa
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2016-04-20
- Volume
- 11
- Issue
- 4
- Pages
- e0151470
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4838256
- OpenAlex
- W1720752806
- Cited by
- 441
- References
- 53
- Citations by year
- 2026: 51 · 2025: 51 · 2024: 51 · 2023: 31 · 2022: 36 · 2021: 69 · 2020: 34 · 2019: 50 · 2018: 33 · 2017: 22 · 2016: 10
- FWCI
- 66.22
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- PLoS ONE (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- arXiv (Cornell University) (submittedVersion)
- RePEc: Research Papers in Economics (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- DSpace@MIT (Massachusetts Institute of Technology) (submittedVersion, cc-by-nc-sa)
- Figshare (submittedVersion, cc-by)
- arXiv (Cornell University)
- Funders
- National Science Foundation (1314891, 1545069)
- Society of Thoracic Surgeons
- University of Illinois at Urbana-Champaign
- Topics
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Life Sciences › Pharmacology, Toxicology and Pharmaceutics › Toxicology › Pharmacovigilance and Adverse Drug Reactions
- Social Sciences › Law › Medical Malpractice and Liability Issues
- Keywords
- robotic surgery, adverse events, equipment failure, patient harm, MAUDE database, natural language processing, conversion to open surgery, surgical specialties, minimally invasive surgery, patient safety, FDA adverse event reports, cardiothoracic surgery, head and neck surgery
- MeSH terms
- Equipment Failure, Humans, Laparoscopy, Retrospective Studies, Robotics, United States, United States Food and Drug Administration, Databases, Factual
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.