Novel methods of testing and calibration of oscillometric blood pressure monitors
Open accessThe DOI matches its Crossref record; 32 of the 32 references it cites were found in a registry or an index.
Abstract
We present a robust method for testing and calibrating the performance of oscillometric non-invasive blood pressure (NIBP) monitors, using an industry standard NIBP simulator to determine the characteristic ratios used, and to explore differences between different devices. Assuming that classical auscultatory sphygmomanometry provides the best approximation to intra-arterial pressure, the results obtained from oscillometric measurements for a range of characteristic ratios are compared against a modified auscultatory method to determine an optimum characteristic ratio, Rs for systolic blood pressure (SBP), which was found to be 0.565. We demonstrate that whilst three Chinese manufactured NIBP monitors we tested used the conventional maximum amplitude algorithm (MAA) with characteristic ratios Rs = 0.4624±0.0303 (Mean±SD) and Rd = 0.6275±0.0222, another three devices manufactured in Germany and Japan either do not implement this standard protocol or used different characteristic ratios. Using a reference database of 304 records from 102 patients, containing both the Korotkoff sounds and the oscillometric waveforms, we showed that none of the devices tested used the optimal value of 0.565 for the characteristic ratio Rs, and as a result, three of the devices tested would underestimate systolic pressure by an average of 4.8mmHg, and three would overestimate the systolic pressure by an average of 6.2 mmHg.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures




Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“For automated BPMDs, these processes are undertaken electronically and based on proprietary algorithms designed to estimate BP by analysis of cuff pressure waveform signals detected because they are transmitted from the cuff into the tubing system (and onwards to the pressure transducer), and ultimately processed by microcomputer. This approach has remained mostly unchanged for decades [13, 44, 45”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [45]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, the range of optimal coefficients for accurate systolic and diastolic BP measurements varies greatly between devices [12, 45, 50, 54].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [45]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
O??Brien, E., Petrie, J., Littler, W., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 772
Characterization of the oscillometric method for measuring indirect blood pressure
Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F. (1982) · Annals of Biomedical Engineering · doi:10.1007/bf02367308 · Cited by 340 · cited 2×
“Interestingly, Geddes, an early pioneer [14] reported that Rs decreased from 0.57 to 0.45 over an auscultatory systolic pressure ranging from 100 to 190 mm Hg, and was 0.55 at 120 mm Hg. Relative to the optimal value of Rs = 0.565, the Chinese manufactured NIBP devices overestimate the systolic…”
BackgroundInfluentialAccuracy of Cuff-Measured Blood Pressure
Picone, D. S., Schultz, M. G., Otahal, P., Aakhus, S., Al-Jumaily, A. M., Black, J. A., Bos, W. J., Chambers, J. B., Chen, C.-H., Cheng, H.-M., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.05.064 · Cited by 277
An arrhythmia classification system based on the RR-interval signal
Tsipouras, M. G., Fotiadis, D. I., & Sideris, D. (2005) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2004.03.007 · Cited by 231
Mauck, G. W., Smith, C. R., Geddes, L. A., & Bourland, J. D. (1980) · Journal of Biomechanical Engineering · doi:10.1115/1.3138195 · Cited by 222 · cited 2×
“In the conventional maximum amplitude algo-rithm (MAA), the maximum value of the OMWE is thought to correspond closely with the mean arterial pressure (MAP) [12].”
BackgroundOscillometric Blood Pressure Estimation: Past, Present, and Future
Forouzanfar, M., Dajani, H. R., Groza, V. Z., Bolic, M., Rajan, S., & Batkin, I. (2015) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2015.2434215 · Cited by 196
Ursino, M., & Cristalli, C. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.508540 · Cited by 179
Babbs, C. F. (2012) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-11-56 · Cited by 177
Baker, P. D., Westenskow, D. R., & Kück, K. (1997) · Medical & Biological Engineering & Computing · doi:10.1007/bf02530049 · Cited by 154
A review of common errors in the indirect measurement of blood pressure. Sphygmomanometry
Bailey, R. H. (1993) · Archives of Internal Medicine · doi:10.1001/archinte.153.24.2741 · Cited by 124
Cited by
Ozoemena, E. L., Iweama, C. N., Agbaje, O. S., Umoke, P. C. I., Ene, O. C., Ofili, P. C., Agu, B. N., Orisa, C. U., Agu, M. A., & Anthony, E. (2019) · Archives of Public Health · doi:10.1186/s13690-019-0349-x · Cited by 112
Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure
Sharman, J. E., Tan, I., Stergiou, G. S., Lombardi, C., Saladini, F., Butlin, M., Padwal, R., Asayama, K., Avolio, A., Brady, T. M., et al. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00693-x · Cited by 84
“Signal association with systolic and diastolic BP Korotkoff sounds denote systolic BP (phase I) and diastolic BP (phase V, or phase IV in absence of V): • Korotkoff phase I sound is the first appearance of two consecutive clear tapping sounds denoting the introduction of blood flow under the cuff • Korotkoff phase II–…”
BackgroundInfluentialPast, present and future of blood pressure measuring instruments and their calibration
Kumar, R., Dubey, P. K., Zafer, A., Kumar, A., & Yadav, S. (2020) · Measurement · doi:10.1016/j.measurement.2020.108845 · Cited by 50
Argha, A., Celler, B. G., & Lovell, N. H. (2020) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2020.3040715 · Cited by 44
Self-citationA Novel Automated Blood Pressure Estimation Algorithm Using Sequences of Korotkoff Sounds
Argha, A., Celler, B. G., & Lovell, N. H. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.3012567 · Cited by 29
Self-citationNoninvasive Blood Pressure Monitoring via a Flexible and Wearable Piezoresistive Sensor
Bijender, Kumar, S., Soni, A., Yadav, R., Singh, S. P., & Kumar, A. (2024) · ACS Omega · doi:10.1021/acsomega.3c04786 · Cited by 28
Oscillometric Waveform Evaluation for Blood Pressure Devices
Kumar, S., Yadav, S., & Kumar, A. (2022) · Biomedical Engineering Advances · doi:10.1016/j.bea.2022.100046 · Cited by 26
Wearable system for real-time monitoring of hemodynamic parameters: Implementation and evaluation
Girčys, R., Kazanavičius, E., Maskeliūnas, R., Damaševičius, R., & Woźniak, M. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.101873 · Cited by 18
Argha, A., Celler, B. G., Alinejad‐Rokny, H., & Lovell, N. H. (2022) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2022.3217865 · Cited by 20
“Oscillometric NIBP measurements are also generally considered less accurate than auscultatory measurements [18], [19], and the oscillogram can be disturbed by cardiac arrhythmia, frequently present in elderly patients aged > 65 with chronic conditions [20], [21].”
BackgroundSelf-citationKumar, S., Yadav, S., & Kumar, A. (2023) · Journal of Human Hypertension · doi:10.1038/s41371-023-00856-4 · Cited by 16
Technologies for Hemodynamic Measurements: Past, Present and Future
Corazza, I., Zecchi, M., Corsini, A., Marcelli, E., & Cercenelli, L. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-816861-5.00022-8 · Cited by 11
Accuracy of cuff blood pressure and systolic blood pressure amplification
Bui, T. V., Picone, D. S., Schultz, M. G., Peng, X., Black, J. A., Dwyer, N., Roberts‐Thomson, P., Adams, H. S. L., Chen, C., Cheng, H., et al. (2023) · Hypertension Research · doi:10.1038/s41440-023-01311-0 · Cited by 12
Principles of Construction, Structure and Features of the ECG and Blood Pressure Monitoring System
Demidov, A. V., Викторович, П. Д., & Krivonogov, L. Y. (2020) · Moscow Workshop on Electronic and Networking Technologies (MWENT) · doi:10.1109/mwent47943.2020.9067390 · Cited by 7
“• The oscillometric method used in the systems does not provide continuous measurement of BP, has low accuracy and reproducibility [13].”
MethodKumar, R., Zafer, A., Dubey, P. K., Kumar, A., Singh, M., Sharma, N. D., Jaiswal, S. K., Prakash, O., Kumar, H., Gupta, V. K., et al. (2023) · Review of Scientific Instruments · doi:10.1063/5.0100958 · Cited by 5
Celler, B. G., Yong, A. S. C., Rubenis, I., Butlin, M., Argha, A., Rehan, R., & Avolio, A. P. (2024) · Journal of Hypertension · doi:10.1097/hjh.0000000000003651 · Cited by 5
“However, theaccuracyoftheseNIBPmonitoringdevicesisindoubt [19– 24].”
Self-citationDeutsch, C., Bramlage, C. P., Botta, B., Krüger, R., Forstner, K., Bramlage, P., & Beime, B. (2021) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000528 · Cited by 3
Beat-to-Beat Oscillometric Blood Pressure Estimation: A Bayesian Approach With System Identification
Farzam, R., Azad, M. H., Moghaddam, H. A., & Forouzanfar, M. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3465663 · Cited by 3
“Some of these OMWE-based algorithms, presumably employed in proprietary commercial oscillometric devices, include the (MAA) [10], maximum/minimum slope algorithm (MSA) [11], and various empirically based methods [12], [13].”
New Perspectives on Non-invasive Blood Pressure Measurement
Argha, A., Celler, B. G., Yong, A. S. C., Rubenis, I., Butlin, M., & Avolio, A. P. (2023) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc40787.2023.10340757 · Cited by 2
“To develop automated NIBP measurement devices, other indirect methods have been proposed, e.g., oscillometric methods [10], [11], [11]–[13], in which oscillometric waveforms (cuff pressure oscillations or oscillogram) are processed for SBP and DBP estimation.”
MethodSelf-citationHemodynamic Parameters Non-Invasive Hydro-Cuff Monitoring System
Markuleva, M. V., Сергеевич, Г. М., Викторович, П. Д., Gerashchenko, S. I., & Krivonogov, L. Y. (2019) · International Conference of Young Specialists on Micro/Nanotechnologies and Electron Devices · doi:10.1109/edm.2019.8823164 · Cited by 3
“In [17] the error rate of SBP and DBP measurement, the oscillometric method was defined as ±27.”
MethodAlgorithm for Estimating Arterial Pressure Based on Spectral Analysis of the Pulse Wave Form
Volkova, N. A., Gerashchenko, S. M., & Vasiliev, V. S. (2020) · Moscow Workshop on Electronic and Networking Technologies (MWENT) · doi:10.1109/mwent47943.2020.9067322 · Cited by 2
Pálhalmi, J. (2020) · Computing in cardiology · doi:10.22489/cinc.2020.359 · Cited by 1
“In the next section of the study, the systolic BP values measured by a reference sphygmomanometer and by a device under test (n=304) from a shared database [13] were modeled by Monte Carlo simulation and were resampled (m=1000) with and without expanding the sampled values with the measurement uncertainty under 95% co…”
MethodUpdates about TC-25: The Technical Committee on Biological and Medical Measurements
Balestrieri, E., Grassini, S., Groza, V. F., & Rapuano, S. (2021) · IEEE Instrumentation & Measurement Magazine · doi:10.1109/mim.2021.9580799 · Cited by 1
Development of an Innovative Algorithm for Accurate Measurement of Human Blood Pressure
Soni, A., Kumar, S., & Kumar, A. (2023) · MAPAN · doi:10.1007/s12647-023-00686-2 · Cited by 1
“After extracting the OMW, an upper envelope is created by fit-ting the positive peaks of oscillometric pulses using the Lorentzian function or cubic spline interpolation [9, 10].”
A FPGA-based high efficiency and low-power system for blood pressure measurement
Zou, F., Yuan, Y., Ma, S., & Cheng, Y. (2023) · International Conference on Integrated Circuits and Microsystems (ICICM) · doi:10.1109/icicm59499.2023.10365923 · Cited by 1
Návrh a realizace zařízení pro snímání hemodynamických parametrů ve veterinární medicíně
Vitáček, J. (2020) · Czech Technical University Digital Library (Czech Technical University in Prague)
DEVELOPMENT OF LOW-COST CLINICAL BLOOD PRESSURE DATA RECORDING UNIT WITH HIGH SAMPLING RATE
Ega, A. V., Ginanjar, G., & Tantokusumo, T. C. (2023) · Instrumentasi · doi:10.31153/instrumentasi.v47i1.331
“Up to now, there are some of noninvasive blood pressure (NIBP) calibrators that are used to calibrate the automated sphygmomanometer such as fluke biomedical with NIBP analyzer BP-Pump 2 and Cufflink NIBP Simulator (Han Wook Song et al., 2008; Doh et al., 2015; Celler et al., 2018).”
MethodCeller, B. G., Yong, A. S. C., Rubenis, I., Butlin, M., Argha, A., Rehan, R., & Avolio, A. P. (2024) · Journal of Hypertension · doi:10.1097/hjh.0000000000003698 · Cited by 1
“The ratio method requires the selection of values for RS and RD with the approximate range 0.43 RS 0.73 and 0.69 RD 0.86 recommended in the literature [12,26] However, despite many theoretical modelling [27–30] and experimental RD decrease as SBP and DBP increase.”
MethodInfluentialSelf-citationPereira, J. M. D., Ribeiro, G., & Postolache, O. A. (2025) · Applied Sciences · doi:10.3390/app15063198
Development of verification framework for a component to be integrated in medical devices
Monga, H. (2024) · Aaltodoc (Aalto University)
Metadata
- Authors
- Celler, B. G., Argha, A., Le, P. N., & Ambikairajah, E.
- Affiliations
- UNSW Sydney (AU)
- Corresponding author
- Celler, Branko G.
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2018-08-06
- Volume
- 13
- Issue
- 8
- Pages
- e0201123
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC6078288
- OpenAlex
- W2887864701
- Cited by
- 28
- References
- 31
- Citations by year
- 2025: 1 · 2024: 4 · 2023: 7 · 2022: 4 · 2021: 2 · 2020: 7 · 2019: 3
- FWCI
- 1.20
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- UNSWorks (University of New South Wales, Sydney, Australia) (submittedVersion, cc-by-nc-nd)
- RePEc: Research Papers in Economics (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Funders
- Australian Research Council (LP160101202)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- oscillometric blood pressure measurement, characteristic ratio, systolic blood pressure estimation, Korotkoff sounds, intra-arterial blood pressure, device performance comparison
- MeSH terms
- Adult, Aged, Aged, 80 and over, Algorithms, Blood Pressure, Blood Pressure Determination, Calibration, Female, Humans, Male, Middle Aged, Oscillometry, Blood Pressure Monitors, Databases, Factual, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.