Effect of the shapes of the oscillometric pulse amplitude envelopes and their characteristic ratios on the differences between auscultatory and oscillometric blood pressure measurements
Open accessThe DOI matches its Crossref record; 17 of the 17 references it cites were found in a registry or an index.
Abstract
Introduction Oscillometric noninvasive blood pressure (NIBP) devices determine pressure by analysing the oscillometric waveform using empirical algorithms. Many algorithms analyse the waveform by calculating the systolic and diastolic characteristic ratios, which are the amplitudes of the oscillometric pulses in the cuff at, respectively, the systolic and diastolic pressures, divided by the peak pulse amplitude. A database of oscillometric waveforms was used to study the influences of the characteristic ratios on the differences between auscultatory and oscillometric measurements. Methods Two hundred and forty-three oscillometric waveforms and simultaneous auscultatory blood pressures were recorded from 124 patients at cuff deflation rates of 2-3 mmHg/s. A simulator regenerated the waveforms, which were presented to two NIBP devices, the Omron HEM-907 [OMRON Europe B.V. (OMCE), Hoofddorp, The Netherlands] and the GE ProCare 400 (GE Healthcare, Tampa, Florida, USA). For each waveform, the paired systolic and paired diastolic pressure differences between device measurements and auscultatory reference pressures were calculated. The systolic and diastolic characteristic ratios, corresponding to the reference auscultatory pressures of each oscillometric waveform stored in the simulator, were calculated. The paired differences between NIBP measured and auscultatory reference pressures were compared with the characteristic ratios. Results The mean and standard deviations of the syst…
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“However, the range of optimal coefficients for accurate systolic and diastolic BP measurements varies greatly between devices [12, 45, 50, 54].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [54]
Citations
Cites
Blood pressure measuring devices: recommendations of the European Society of Hypertension
O'Brien, E., Waeber, B., Parati, G., Staessen, J., & Myers, M. G. (2001) · BMJ · doi:10.1136/bmj.322.7285.531 · Cited by 903
White, W. B., Berson, A. S., Robbins, C., Jamieson, M. J., Prisant, L. M., Roccella, E., & Sheps, S. G. (1993) · Hypertension · doi:10.1161/01.hyp.21.4.504 · Cited by 385
Characterization of the oscillometric method for measuring indirect blood pressure
Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F. (1982) · Annals of Biomedical Engineering · doi:10.1007/bf02367308 · Cited by 340
Noninvasive automatic determination of mean arterial pressure
Ramsey, M. (1979) · Medical and Biological Engineering and Computing · doi:10.1007/bf02440948 · Cited by 287
Validation of the Omron HEM-907 device for blood pressure measurement
El Assaad, M. A., Topouchian, J. A., Darné, B. M., & Asmar, R. G. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200208000-00006 · Cited by 279
White, W. B., & Anwar, Y. A. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200104000-00007 · Cited by 213
Ursino, M., & Cristalli, C. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.508540 · Cited by 179
Oscillometric blood pressure measurement: progress and problems
van Montfrans, G. A. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200112000-00004 · Cited by 175
Update on Methods & Simulators for Evaluation of Noninvasive Blood Pressure Monitors
NG, K.-G., & SMALL, C. F. (1994) · Journal of Clinical Engineering · doi:10.1097/00004669-199403000-00012 · Cited by 55
van Ittersum, F. J., Wijering, R. M. J., Lambert, J., Donker, A. J. M., & Stehouwer, C. D. A. (1998) · Journal of Hypertension · doi:10.1097/00004872-199816080-00007 · Cited by 52
Cited by
Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure
Sharman, J. E., Tan, I., Stergiou, G. S., Lombardi, C., Saladini, F., Butlin, M., Padwal, R., Asayama, K., Avolio, A., Brady, T. M., et al. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00693-x · Cited by 84
“However, the range of optimal coefficients for accurate systolic and diastolic BP measurements varies greatly between devices [12, 45, 50, 54].”
BackgroundSelf-citationAmoore, J. N., Lemesre, Y., Murray, I. C., Mieke, S., King, S. T., Smith, F. E., & Murray, A. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e3282f06ec7 · Cited by 56
Self-citationTaylor–Fourier Analysis of Blood Pressure Oscillometric Waveforms
Serna, J. A. D. L. O. (2013) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2013.2258245 · Cited by 37
“shapes [5] is one of the main causes of inaccuracy.”
BackgroundHow Important is the Recommended Slow Cuff Pressure Deflation Rate for Blood Pressure Measurement?
Zheng, D., Amoore, J. N., Mieke, S., & Murray, A. (2011) · Annals of Biomedical Engineering · doi:10.1007/s10439-011-0347-9 · Cited by 28
Self-citationMireles, S. A., Jaffe, R., Drover, D. R., & Brock‐Utne, J. G. (2009) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-009-9178-8 · Cited by 19
“Possible reasons for the discrepancy between noninvasive and invasive pressures are variations caused by differences in the oscillometric wave form produced by the device, beat to beat variability in the intra-arterial measurement and/or differences in the algorithms used by the monitor to generate blood pressure valu…”
Method
Metadata
- Authors
- Amoore, J. N., Vacher, E., Murray, I. C., Mieke, S., King, S. T., Smith, F. E., & Murray, A.
- Affiliations
- Edinburgh Royal Infirmary (GB)
- Université de la Méditerranée Aix-Marseille II (FR)
- University of Edinburgh (GB)
- Physikalisch-Technische Bundesanstalt (DE)
- Freeman Hospital (GB)
- Corresponding author
- Amoore, John N.
- Venue
- Blood Pressure Monitoring
- Published
- 2007-10
- Volume
- 12
- Issue
- 5
- Pages
- 297-305
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2060689584
- Semantic Scholar
- 206714b603b6a7fa93033741bcd2fe92bb306b7c
- Cited by
- 34
- References
- 17
- Citations by year
- 2025: 1 · 2024: 1 · 2023: 1 · 2022: 1 · 2021: 1 · 2019: 1 · 2018: 3 · 2017: 2 · 2016: 2 · 2015: 6 · 2013: 2 · 2012: 1
- FWCI
- 2.39
- Type
- journal article
- Language
- en
- Open access
- green
- doi.org
- Locations
- Blood Pressure Monitoring (publishedVersion)
- PubMed (publishedVersion)
- Newcastle University ePrints (Newcastle Univesity)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- oscillometric blood pressure measurement, auscultatory blood pressure, characteristic ratio, waveform simulation, cuff pressure
- MeSH terms
- Adolescent, Adult, Aged, Aged, 80 and over, Algorithms, Auscultation, Blood Pressure, Blood Pressure Determination, Child, Diastole, Humans, Middle Aged, Oscillometry, Systole, Blood Pressure Monitors
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Effect of the shapes of the oscillometric pulse amplitude envelopes and their characteristic ratios on the differences between auscultatory and oscillometric blood pressure measurements. https://app.cochat.ai/papers/10.1097/mbp.0b013e32826fb773
- The paper
- Amoore, J. N., Vacher, E., Murray, I. C., Mieke, S., King, S. T., Smith, F. E., & Murray, A. (2007). Effect of the shapes of the oscillometric pulse amplitude envelopes and their characteristic ratios on the differences between auscultatory and oscillometric blood pressure measurements. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0b013e32826fb773
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.